A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, powder-free nitrile examination gloves; scented (grape, apple, peppermint, vanilla). Worn by healthcare personnel or patients on hands/fingers to provide a physical barrier against contamination. Manufactured from carboxylated butadiene acrylonitrile. Used in clinical or home settings. Device functions as a passive barrier; performance verified through physical property testing (tensile strength, elongation), barrier integrity (water leak test), and biocompatibility (skin irritation/sensitization). Output is the physical protection provided by the glove material during medical examinations or procedures.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ASTM D6319-10 (dimensions, physical properties, freedom from holes, powder-free status) and ISO 10993-10 (primary skin irritation and dermal sensitization). All tests met specifications.
Technological Characteristics
Material: Carboxylated butadiene acrylonitrile. Form: Powder-free, scented blue nitrile examination glove. Standards: ASTM D6319-10 (dimensions, physical properties, freedom from holes), ASTM D6124-10 (residual powder), ISO 10993-10 (biocompatibility). Non-sterile. Single-use.
Indications for Use
Indicated for use as a disposable medical device worn on the examiner's hand or fingers to prevent contamination between patient and examiner. For over-the-counter use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Non-Sterile, On Line Powder Free Nitrile Blue & White Color Examination Gloves (K052502)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract design of an eagle with three human profiles incorporated into its body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 12, 2015
YTY Industry (Manjung) SDN. BHD. Ms. Punitha Samy, Assistant Manager DC/RA Lot 1422-1424, Batu 10 Lekir Sitiawan, Perak Darul Ridzuan Malaysia 32020
Re: K143055
Trade/Device Name: Non-Sterile, Powder Free Scented Blue Nitrile Examination Gloves -Grape, Apple, Peppermint, Vanilla Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: January 2, 2015 Received: January 8, 2015
Dear Ms. Samy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Samy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D. Teiashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
# K143055
Device Name
NON-STERILE, POWDER FREE SCENTED BLUE NITRILE EXAMINATION GLOVES - GRAPE, APPLE, PEPPERMINT, VANILLA
Indications for Use (Describe)
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a green symbol with a gold outline. The symbol appears to be abstract, with a horizontal curved line at the top and three vertical rectangular shapes below it. The two outer rectangular shapes have additional horizontal extensions on their outer sides, creating a stylized, symmetrical design.
# YTY INDUSTRY (MANJUNG) SDN BHD
(Company No. 380830-P) Lot 1422-1424, Batu 10 Lekir, 32020 Sitiawan, Perak Darul Ridzuan, Malaysia Tel: 05-679 2288 (Hunting Line), 679 2443 & 679 2445 Fax: 05-6791188
K143055
| | 510 (K) SUMMARY SHEETS |
|-----|------------------------|
| 1.0 | |
| | 510 (K) SUMMARY | |
|-----|------------------------|--------------------------------------------------------------------------------------------------------|
| 2.0 | Submitter | YTY INDUSTRY (MANJUNG) SDN. BHD.,<br>Lot 1422-1424, Batu 10 Lekir<br>32020 Sitiawan, Perak<br>Malaysia |
| | Tel | 605-6792288 |
| | Fax | 605-6791188 |
| | Name of Contact Person | 1. MS. PUNITHA SAMY<br>E-mail: punitha@ytygroup.com.my |
| | Date Summary Prepared | February 12, 2015 |
| 3.0 | Name of Device | |
|-----|------------------------|------------------------------------------------------------------------------------------------------|
| | Trade Name: | Non-Sterile, Powder Free Scented Blue Nitrile Examination Gloves – Grape, Apple, Peppermint, Vanilla |
| | Common Name: | Nitrile Examination Gloves |
| | Classification Name: | Patient examination glove |
| | Device Classification: | I |
| | Regulation Number: | 21 CFR 880.6250 |
| | Panel: | General Hospital (80) |
| | Product Code: | LZA |
| 4.0 | Identification of The Legally Marketed Devices | |
|-----|------------------------------------------------|--------------------------------------------------------------------------------|
| | Predicate Device Name: | Non-Sterile, On Line Powder Free Nitrile Blue & White Color Examination Gloves |
| | Predicate 510(K) number: | K052502. |
Manufacturer's Name: YTY Industry (Manjung) Sdn Bhd Lot 1422-1424, Batu 10 Lekir, 32020 Sitiawan, Perak Darul Ridzuan
{4}------------------------------------------------
#### 5.0 Description of The Device
Non-Sterile, Powder Free Scented Blue Nitrile Examination Gloves - Grape, Apple, Peppermint, and Vanilla meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.
#### 6.0 Indications for Use
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. This device is for over-the counter use.
#### 7.0 Summary of the Technological Characteristic of the Device compared to the Predicate Device for substantial equivalent discussion
There is no difference in technology characteristic compared to the predicate device. Gloves are made from nitrile latex compound. Non-sterile, Powder Free Scented Blue Nitrile Examination Gloves have the below technological characteristic compared to ASTM or Equivalent standards.
| Characteristic | Standards | Performance of<br>Non-Sterile, Powder Free<br>Scented Blue Nitrile<br>Examination Gloves |
|---------------------|-----------------------------------------|------------------------------------------------------------------------------------------|
| Dimension | ASTM D6319-10 | Meets |
| Physical Properties | ASTM D6319-10 | Meets |
| Freedom from holes | ASTM D6319-10 | Meets |
| Powder-free | ASTM D6319-10 | Meets |
| Bio-compatibility | Primary skin irritation<br>ISO 10993-10 | Non-Irritant |
| | Dermal Sensitization<br>ISO 10993-10 | Non-sensitizer |
{5}------------------------------------------------
K143055
| Test | T OF THE FINAL OF COLLED OF CRUPA ART THE FOR THE T PLATE T PATT TAGALATE HOME TEMP ASSE<br>FDA<br>1000ml<br>Water Leak<br>Test | YTY Powder Free Nitrile Examination Gloves | Non-Sterile On<br>Line Powder<br>Free Nitrile<br>Blue & White<br>Color<br>Examination<br>Glove | Characteristics | Applicant | | | | Predicate K052502 | Medical Glove Manual (1661) | | | |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-----------------------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Watertight (1000ml) | Multiple<br>Normal | Blue-<br>Grape | Blue-<br>Apple | Blue-<br>Peppermint | Blue-<br>Vanilla | Predicate<br>K052502 | | Blue - Grape | Blue - Apple | Blue - Peppermint | Blue - Vanilla | | |
| ASTM D5151-06 (2011)<br>Test<br>ASTM D6319-10 | GII<br>AQL = 2.5<br>ASTM<br>D6319-10 | Holes<br>found: 0<br>(Accept 1,<br>Reject 7) | Holes<br>found: 0<br>(Accept 1,<br>Reject 7) | Holes found:<br>0<br>(Accept 1,<br>Reject 7) | Holes<br>found: 0<br>(Accept 1,<br>Reject 7) | Holes found: 0<br>(Accept 1,<br>Reject 7) | Company Name | YTY Industry (Manjung) Sdn Bhd | | | | YTY Industry (Manjung) Sdn Bhd | FDA |
| 2. Length (mm)<br>Size<br>M | Min 230 | 240-243 | 240-249 | 240-249 | 240-251 | 240-251 | Product Name | Non-Sterile, Powder Free Scented Blue Nitrile Examination Gloves | | | | Non-Sterile, On Line Powder Free Nitrile Blue & White Color Examination Gloves | Patient Examination Gloves |
| 3. Palm width (mm)<br>Size<br>M | 95 + 10 | તેર-તેવે | તેર-તેત્ર | તેર-તેત્ર | 94-96 | 94-96 | Product Code | LZA | | | | LZA | LZA - (Polymer - other than Vinyl (includes Nitrile, Polyurethane, etc.) |
| 4. Thickness (mm)<br>(Single Layer) | | | | | | | Indication for Use | A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. | | | | This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patient. | <b>Powder-Free Examination Gloves</b><br>A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. |
| Finger<br>Palm | Min 0.05<br>Min 0.05 | 0.10-0.14<br>0.07-0.08 | 0.10-0.14<br>0.07-0.08 | 0.11-0.14<br>0.07-0.08 | 0.11-0.14<br>0.08-0.09 | 015-0.19<br>0.12-0.16 | Device Description | Non-Sterile, Powder Free Scented Blue Nitrile Examination Gloves meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. This device is for over-the counter single use. | | | | Class 1 Nitrile Patient Examination Glove 80 LZA, powder free that meets all requirements of ASTM Standard D6319-00a3 and FDA water leak test. | This gloves meet all current specifications listed under ASTM specifications D6319-10 |
| 5. Physical Properties | | | | | | | Over the Counter Use | This device is for over-the counter single use. | | | | This device is for over-the counter single use. | Indication for use |
| Before Aging<br>Tensile Strength (MPa)<br>Ultimate Elongation<br>(%) | Min 14<br>Min 500 | 25.70-<br>29.83<br>520-580 | 25.99-<br>28.39<br>520-580 | 24.11-28.98<br>520-580 | 24.54-<br>30.14<br>540-580 | 26.00-30.00<br>750-800 | Use | Single Use | | | | Single Use | Directions for use |
| After Aging<br>Tensile Strength (MPa) | Min 14 | 27.76-<br>31.67 | 28.39-<br>31.61 | 26.53-30.77 | 29.25-<br>32.35 | 25.00-28.00 | Non Sterile or Sterile | Non Sterile | | | | Non Sterile | Sterilization |
| Ultimate Elongation<br>(%) | Min 400 | 460-500 | 460-500 | 440-500 | 460-480 | 670-730 | Powder Free | Powder Free | | | | Powder Free | Process & Attribute labeling |
| 6. Residual Powder<br>ASTM-D6124-10<br>(Reapproved 2011) | Max<br>2.0mg/glove | 0.16mg/<br>glove | 0.10mg/<br>glove | 0.12mg/<br>glove | 0.12mg/<br>glove | 0.20%/ glove | Compare materials | | | | | | |
| 7. Biocompatibility<br>Primary Skin Irritation<br>Dermal Sensitization | ISO 10993-<br>10<br>Non-irritant<br>Non-<br>sensitizer | Under the<br>conditions<br>of the<br>study, the<br>device is<br>non-irritant<br>or non-<br>sensitizer | Under the<br>conditions<br>of the<br>study, the<br>device is<br>non-irritant<br>or non-<br>sensitizer | Under the<br>conditions of<br>the study,<br>the device is<br>non-irritant<br>or non-<br>sensitizer | Under the<br>conditions<br>of the<br>study, the<br>device is<br>non-irritant<br>or non-<br>sensitizer | Non-irritant<br>Non-sensitizer | Materials | Carboxylated Butadiene Acrylonitrile | | | | Carboxylated Butadiene Acrylonitrile | LZA - (Polymer - other than Vinyl (includes Nitrile, Polyurethane, etc.) |
Performance data of gloves based on ASTM D6319-10 and FDA 1000ML water leak test.
{6}------------------------------------------------
K143055
# 8.0 Substantial Equivalence Comparison Table with Predicate Device, K052502
{7}------------------------------------------------
| | K143055 | | | | | | | |
|--------------------------------------------|--------------------------------|-------------------|------------|--------------------------------|-------------------------------------------------------|-------------------------|--------------------------------|--|
| | K143055 | | | | | | Medical Glove | |
| Characteristics | Blue -<br>Apple | Blue -<br>Grape | | Blue -<br>Peppermint | Blue - Vanilla | Predicate K052502 | Manual (1661) | |
| Company Name | | | | YTY Industry (Manjung) Sdn Bhd | YTY Industry<br>(Manjung) Sdn Bhd | | | |
| Specifications and | | | | | Dimension: Finger & Palm Thickness min: 0.05mm | | Current gloves | |
| Performance | | | | | Physical Properties: Min 14MPa Before and After Aging | | meet ASTM | |
| | | Barrier: AQL 2.5 | | | | | D6319-10 Nitrile | |
| Tensile Strength | 25.70- | 25.99- | | 24.11- | 24.54- | 26.00-30.00 | gloves | |
| before aging (MPa) | 29.83 | 29.83 | | 28.98 | 30.15 | | | |
| Tensile Strength after | 27.76- | 28.39- | | 26.53- | 29.25- | 25.00-28.00 | ASTM D6124-06 | |
| aging (MPa) | 31.67 | 31.61 | | 30.77 | 32.35 | | (Reapproved | |
| Ultimate Elongation | 520-580 | 520-580 | 520-580 | | 540-580 | 750-800 | 2011) | |
| before aging (%) | | | | | | | Residual Powder | |
| Ultimate Elongation | 460-500 | | 460-500 | 440-500 | 460-480 | 670-730 | | |
| after aging (%) | | | | | | | ASTM D5151-06 | |
| Dimensions Length | 240-243 | | 240-249 | 240-249 | 240-251 | 240-251 | (Reapproved | |
| (mm) | | | | | | | 2011) Detection<br>of Holes in | |
| Dimensions Width | तेर-तेत्र | | d5-99 | d3-dd | 96-99 | 94-96 | Medical Gloves | |
| (mm) | | | | | | | | |
| Thickness Finger | 0.10-0.14 | | 0.11-0.14 | 0.11-0.14 | 0.11-0.14 | 0.15-0.19 | | |
| (mm) | | | | | | | | |
| Thickness Cuff (mm) | 0.06 | | 0.06 | 0.06-0.07 | 0.07-0.08 | 0.09-0.10 | | |
| Thickness Palm<br>(mm) | 0.07-0.08 | | 0.07-0.08 | 0.07-0.08 | 0.08-0.09 | 0.12-0.16 | | |
| AQL | AQL 2.5 | | AQL 2.5 | AQL 2.5 | AQL 2.5 | AQL 2.5 | | |
| | Result: 0 | | Result: 0 | Result: 0 | Result: 0 | Result: 0 | | |
| Residual Powder | 0.16 | | 0.10 | 0.16 | 0.12 | 0.20 | | |
| (mg/glove) | | | | | | | | |
| Size | M | | M | M | M | M | | |
| Bio-compatibility | Under the | | Under the | Under the | Under the | Non-irritant | ISO 10993-10 | |
| | condition | | condition | condition | condition | Non-sensitizer | Test for Irritation | |
| | of the | | of the | of the | of the study | | and Skin | |
| | study the | | study the | study the | the device | | Sensitization | |
| | device is | | device is | device is | is non- | | | |
| | non- | | non- | non- | irritant and | | | |
| | irritant | | irritant | irritant | non- | | | |
| | and non- | | and non- | and non- | sensitizer | | | |
| | sensitizer | | sensitizer | sensitizer | | | | |
| Labeling for legally<br>marketed predicate | -Powder Free<br>-devices color | | | | -Powder Free<br>-devices color: | Chapter 4 -<br>Labeling | | |
| | -Scent | | | | Clear (Blue) | | | |
| -Patient Examination Glove | | | | -Patient | | | | |
| | -Non sterile | | | | Examination Glove | | | |
| | | -Single Use Only | | | -Non sterile | | | |
| | | -Manufactured for | | | -Single Use Only | | | |
| | -Lot | | | | | -Manufactured for: | | |
| | -Intended use | | | | | -Lot | | |
| | | | | | | | | |
{8}------------------------------------------------
#### 9.0 Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the device, Non-Sterile, Powder Free Scented Blue Nitrile Examination Gloves – Grape, Apple, Peppermint and Vanilla and the predicate device is substantially equivalent based on intended uses, physical properties, technological characteristics and non-clinical performance.
Sec 1 Sum Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.