VLP Mini-Mod Small Bone Plating System 1.5mm Plates and Screws
K143050 · Smith & Nephew, Inc. · HRS · Dec 18, 2014 · Orthopedic
Device Facts
Record ID
K143050
Device Name
VLP Mini-Mod Small Bone Plating System 1.5mm Plates and Screws
Applicant
Smith & Nephew, Inc.
Product Code
HRS · Orthopedic
Decision Date
Dec 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Bone plates and bone screws from the VLP Mini-Mod Small Bone Plating System are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
Device Story
The VLP Mini-Mod Small Bone Plating System is an internal fixation system consisting of titanium alloy (Ti-6Al-4V) bone plates and locking/non-locking bone screws. It is a size-extension (1.5mm) of the previously cleared Variable-Angle Locking Mini-Fragment Plating System. The device is used by surgeons in clinical settings for the stabilization and reduction of small bone fractures or fragments. The plates feature a radiused cross-section with 6 to 24 holes and are available in Y, straight, and column configurations. The system provides mechanical support to bone fragments to facilitate healing. The device is provided sterile and is intended for single use only.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including Finite Element Analysis (FEA) to identify worst-case designs, four-point bend fatigue testing, torque-to-failure testing per ASTM F543-07, and axial pull-out strength testing per ASTM F543. Results demonstrated non-inferiority to predicate devices.
Technological Characteristics
Implant-grade titanium alloy (Ti-6Al-4V). Plates feature radiused cross-sections with 6-24 holes; Y, straight, and column designs. Locking and non-locking screws. Sterilized via Gamma irradiation. Mechanical testing performed per ASTM F543-07.
Indications for Use
Indicated for adolescent (12-18 years), transitional adolescent (18-21 years), and adult populations, including patients with osteopenic bone. Used for fracture fixation, arthrodesis, reconstruction, replantation, or reduction of small bones and fragments, and non-load bearing stabilization/reduction of long bone fragments.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (K132886)
Synthes (USA) 1.5mm Mini Fragment LCP System (K090047)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure. The figure is composed of three overlapping profiles, suggesting a sense of community and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2014
Smith & Nephew, Inc. Ms. Samantha Staubach Regulatory Affair Specialist I 1450 Brooks Road Memphis. Tennessee 38116
Re: K143050
Trade/Device Name: VLP Mini-Mod Small Bone Plating System 1.5mm Plates and Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: October 22, 2014 Received: October 23, 2014
Dear Ms. Staubach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
Page 2 - Ms. Samantha Staubach
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Premarket Notification Indications for Use Statement
K143050 510(k) Number (if known):
Device Name: VLP Mini-Mod Small Bone Plating System 1.5mm Plates and Screws
Indications for Use:
The Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System is indicated for adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System is indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bones and fragments. This system is also indicated for nonload bearing stabilization and reduction of bone fragments in long bones.
Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System are for single use only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Smith & Nephew. The logo features an orange flower-like symbol to the left of the text. The text reads "We are smith&nephew", with "We are" in black and "smith&nephew" in orange.
| Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Date of Summary: | October 22, 2014 |
| Contact Person and Address: | Samantha Staubach<br>Regulatory Affairs Specialist I<br>T (901) 399-6132<br>F (901) 566-7596 |
| Name of Device: | Smith & Nephew, Inc. VLP Mini-Mod Small Bone Plating<br>System 1.5mm Plates and Screws |
| Common Name: | Bone Plates and Screws |
| Device Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories<br>Class II |
| | 21 CFR 888.3040 Smooth or threaded metallic bone fixation<br>fastener<br>Class II |
| Panel Code: | Orthopedics/87 |
#### Device Description
The subject premarket notification describes 1.5mm line additions to the VLP Mini-Mod Small Bone Plating System. The subject internal fixation plating system is comprised of assorted implantable, locking bone plates and compatible locking and non-locking bone screws to be used on various small bones and fragments. The VLP Mini-Mod Small Bone Plating System is a subset of the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System, which was previously cleared via Traditional 510(k) K132886. Subject plates consist of groups of devices with a radiused cross-section where the number of holes in the plates will range from 6 through 24. Specific plate designs include Y-plates, Straight plates, and Column plates. All described implant devices are manufactured from implant-grade titanium alloy material (Ti-6Al-4V) and designed for single-use. They will be provided in a sterile packaged option and will be sterilized via Gamma irradiation. Bone screws will be offered in both locking and non-locking options in a variety of lengths.
#### Intended Use
Bone plates and bone screws from the VLP Mini-Mod Small Bone Plating System are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
{4}------------------------------------------------
#### Indications for Use
The Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System is indicated for adolescent (12-18 years) and transitional adolescent (18-21 years) subpopulations and adults, as well as patients with osteopenic bone. The Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System is indicated for fracture fixation, arthrodesis, reconstruction. replantation or reduction of small bones and fragments. This system is also indicated for nonload bearing stabilization and reduction of bone fragments in long bones.
Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System are for single use only.
The indications for use for the subject VLP Mini-Mod Small Bone Plating System 1.5mm Plates and Screws are identical to those previously cleared for the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System.
#### Technological Characteristics
Device comparisons described in this premarket notification demonstrated that the proposed bone plates and bone screws are substantially equivalent to the legally marketed predicate devices listed below with regard to intended use, indications for use, materials and performance characteristics. The subject VLP Mini-Mod 1.5mm Bone Plates and Screws feature characteristics very similar to the VLP Mini-Mod Small Bone Plating System cleared via K132886, with the primary difference being the 1.5mm size offering. To accommodate for the size difference between the 1.5mm subject plates and the 2.0mm and 2.4mm offerings cleared via K132886, mechanical testing was conducted against the secondary predicates.
#### Substantial Equivalence Information
When compared to the predicate devices listed below, substantial equivalence is based upon similar function, intended use, indications for use and overall design to the devices listed in the table below.
| Manufacturer | Description | Submission<br>Number | Clearance<br>Date |
|----------------------|----------------------------------------------------------------------------------------------|----------------------|-------------------|
| Smith & Nephew, Inc. | Smith & Nephew Variable-Angle Locking<br>Mini-Fragment Plating System (Primary<br>Predicate) | K132886 | 2/4/14 |
| Synthes (USA) | Synthes (USA) 1.5mm Mini Fragment LCP<br>System | K090047 | 4/1/09 |
| Smith & Nephew, Inc. | Smith & Nephew Bone Plate System | K993106 | 12/9/99 |
Table 1: Substantially Equivalent Predicates to the VLP Mini-Mod Small Bone Plating System 1.5mm Plates and Screws
{5}------------------------------------------------
#### Summary of Pre-Clinical Testing
- Finite element analysis (FEA) was conducted on the proposed 1.5mm bone plate o designs to determine the worst case plate as well as the high stress region of the plate to be used for subsequent mechanical testing. The acceptance criteria was met in that the subject bone plates exhibited similar or superior structural strength compared to the existing 1.5mm predicates. Results of the FEA demonstrated that the plates identified for mechanical testing were the appropriate bone plates because they possessed the highest stress concentrations.
- o Four-point bend fatigue testing was conducted on worst-case examples of the proposed bone plates, as identified through FEA testing. Results of the testing concluded that the number of log cycles to failure achieved by the proposed bone plates met the acceptance criteria in that they were found to be non-inferior to the fatigue log cycles to failure of a previously cleared predicate.
- Torque-to-failure testing for the proposed 1.5mm bone screws was conducted per the o guidelines of ASTM F543-07. The acceptance criterion was met in that all tested 1.5mm screws met the minimum torsional strength requirement dictated by ASTM F543-07.
- o Axial pull-out testing was conducted on the subject 1.5mm cortex screws following the guidelines set forth in ASTM F543 with maximum force denoted as pull-out strength. The subject implants met the acceptance criteria in that the maximum pullout force of the subject screws was non-inferior to the maximum pull-out force of the predicate.
#### Conclusion
This 510(k) premarket notification is being submitted to request clearance for the VLP Mini-Mod Small Bone Plating System 1.5mm Plates and Screws. Based on the similarities to the predicate components and a review of the mechanical testing performed, the devices are substantially equivalent to the predicate plating systems listed in Table 1 on page 5-2.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.