K143046 · Toshibamedical Systems Corporation · IYN · Nov 19, 2014 · Radiology
Device Facts
Record ID
K143046
Device Name
Xario 100, V3.0
Applicant
Toshibamedical Systems Corporation
Product Code
IYN · Radiology
Decision Date
Nov 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Diagnostic Ultrasound System Xario 100 Model TUS-X100 and Xario 100 Model TUS-X100S are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diaqnosis in the following clinical applications: fetal, abdominal, intra-operative(abdominal), laparoscopic, pediatric, small orqans, neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal (superficial), cardiac adult, cardiac pediatric, trans-esoph(cardiac) and peripheral vessel.
Device Story
Mobile diagnostic ultrasound system (TUS-X100/TUS-X100S V3.0) for clinical imaging. Inputs: ultrasound signals from wide array of probes (linear, convex, sector; 2-12 MHz). Processing: system performs signal/image processing to generate diagnostic images; includes BEAM feature for needle visualization enhancement during biopsy. Output: real-time ultrasound images displayed for physician/clinician. Used in clinical settings (OR, clinic) by trained professionals. Output aids in diagnosis and procedural guidance (e.g., needle placement). Benefits: improved visualization of anatomical structures and dynamic processes; enhanced needle visibility during interventional procedures.
Clinical Evidence
Bench testing only. Performance testing included a comparative study of an existing imaging mode versus the BEAM needle enhancement feature, confirming improved needle visualization. Software validation and acoustic output testing (UD3, 2004) were performed to verify system modifications.
Technological Characteristics
Mobile diagnostic ultrasound system using linear, convex, and sector array transducers (2-12 MHz). Features include B-mode, M-mode, PWD, CWD, Color Doppler, THI, Advanced Dynamic Flow, Power Doppler, CHI, 4D, TDI, ApliPure, Precision Imaging, and BEAM. Conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, and ISO 10993-1. Sterilization: N/A (transducers are reusable).
Indications for Use
Indicated for visualization of structures and dynamic processes in the human body using ultrasound for diagnostic imaging. Clinical applications include fetal, abdominal, intra-operative (abdominal), laparoscopic, pediatric, small organs (thyroid, breast, testicle), neonatal/adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional/superficial), cardiac (adult/pediatric/trans-esophageal), and peripheral vessel.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Xario100 Model TUS-X100, TUS-X100S, v1.0 (K133277)
MicroMaxx High Resolution Ultrasound System (C3 Series) (K053069)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 19, 2014
Toshiba Medical Systems Corporation % Mr. Orlando Tadeo. Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K143046
Trade/Device Name: Xario 100. V3.0 (Models: TUS-X100 and TUS-X100S) Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN Dated: October 21, 2014 Received: October 22, 2014
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This determination of substantial equivalence applies to the following transducers intended for use with the Xario 100 TUS-X100, TUS-X100S V3.0 Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
| PSU-25BT | PSU-30BT | PSU-50BT |
|------------|------------|------------|
| PVU-375BT | PVU-382BT | PVU-674BT |
| PVU-674MV | PVU-712BT | PVU-745BTF |
| PVU-745BTH | PVU-745BTV | PVU-681MVL |
| PVU-770ST | PVU-781VT | PVL-715RS |
| PLU-704BT | PLU-1005BT | PLU-1204BT |
| PET-512MC | PET-805LA | PC-20M |
| PC-50M | | |
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K143046
Device Name
Xario 100 TUS-X100 and TUS-X100S, V3.0
### Indications for Use (Describe)
The Diaqnostic Ultrasound System Xario 100 Model TUS-X100 and Xario 100 Model TUS-X100S are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diaqnosis in the following clinical applications: fetal, abdominal, intra-operative(abdominal), laparoscopic, pediatric, small orqans, neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal (superficial), cardiac adult, cardiac pediatric, trans-esoph(cardiac) and peripheral vessel.
Type of Use (Select one or both, as applicable)
ك Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Toshiba America Medical Systems, Inc.
### System: _Xario 100 TUS-X100, TUS-X100S V3.0
Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | | | | | |
|------------------------------------|---|---|-------------------|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Specific<br>(Tracks 3) | B | M | | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | P | P | | P | P | P | 2 | P | P | P | | P | 4,5,6,7,8 |
| Abdominal | P | P | | P | P | P | 2,3 | P | P | P | | P | 4,5,6,7,8 |
| Intra-operative (Abdominal) | N | N | | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Intra-operative (Neuro) | | | | | | | | | | | | | |
| Laparoscopic | N | N | | N | N | N | 2 | N | N | N | | | 5,7 |
| Pediatric | P | P | | P | P | P | 2,3 | P | P | P | | P | 4,5,6,7,8 |
| Small Organ (Note 1) | P | P | | P | P | P | 2 | P | P | P | | | 4,5,6,7,8,9,1 |
| Neonatal Cephalic | P | P | | P | P | P | 3 | P | P | P | | | 5,6,7,8 |
| Adult Cephalic | P | P | | P | P | P | 3 | P | P | P | | | 7 |
| Trans-rectal | P | P | | P | P | P | 2 | P | P | P | | N | 4,5,6,7,8 |
| Trans-vaginal | P | P | | P | P | P | 2 | P | P | P | | N | 4,5,6,7,8 |
| Trans-urethral | | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | P | P | | P | P | P | 2 | P | P | P | | | 4,5,6,7,8,9,1 |
| Musculo-skeletal<br>(Superficial) | P | P | | P | P | P | 2 | P | P | P | | | 4,5,6,7,8,9,1 |
| Intravascular | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | |
| Cardiac Adult | P | P | | P | P | P | 3 | P | P | P | | | 4, 7 |
| Cardiac Pediatric | P | P | | P | P | P | 3 | P | P | P | | | 4, 7 |
| Intravascular (Cardiac) | | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | N | N | | N | N | N | 3 | N | | | | | 4,7 |
| Intra-cardiac | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | |
| Peripheral vascular | P | P | | P | N | P | 2 | P | P | P | | | 4,5,6,7,8,9,1 |
| Other (Specify) | | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{4}------------------------------------------------
System: _ Xario 100 TUS-X100, TUS-X100S V3.0 Transducer: PSU-25BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Specific<br>(Tracks 3) | Mode of Operation | | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
|------------------------------------------------|-------------------|--|---|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Ophthalmic | | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | | |
| Abdominal | | | N | N | N | N | N | 3 | N | N | N | | | 7 |
| Intra-operative (Abdominal) | | | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | | |
| Pediatric | | | N | N | N | N | N | 3 | N | N | N | | | 7 |
| Small Organ (Specify) (1) | | | | | | | | | | | | | | |
| Neonatal Cephalic | | | N | N | N | N | N | 3 | N | N | N | | | 7 |
| Adult Cephalic | | | N | N | N | N | N | 3 | N | N | N | | | 7 |
| Trans-rectal | | | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | | |
| Cardiac Adult | | | N | N | N | N | N | 3 | N | N | N | N | | 4,7 |
| Cardiac Pediatric | | | N | N | N | N | N | 3 | N | N | N | N | | 4,7 |
| Intravascular (Cardiac) | | | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 ApliPure Plus Note 7 Precision Imaging Note 8 Differential THI Note 9 Elastography Note 10 BEAM
{5}------------------------------------------------
### Xario 100 TUS-X100,TUS-X100S V3.0 System: _
Transducer:_ PSU-30BT
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | P | P | P | P | P | 3 | P | P | P | | | 7 |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | P | P | P | P | P | 3 | P | P | P | | | 7 |
| Small Organ (Specify) (1) | | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | | | 7 |
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | | | 7 |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | | | 4, 7 |
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | | | 4, 7 |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K133277
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{6}------------------------------------------------
#### System: Xario 100 TUS-X100, TUS-X100S V3.0 Transducer:_ PSU-50BT
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | N | N | N | N | N | 3 | N | N | N | | | 7 |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | N | N | N | N | N | 3 | N | N | N | | | 7 |
| Small Organ (Specify) (1) | | | | | | | | | | | | |
| Neonatal Cephalic | N | N | N | N | N | 3 | N | N | N | | | 7 |
| Adult Cephalic | N | N | N | N | N | 3 | N | N | N | | | 7 |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | N | N | N | N | N | 3 | N | N | N | N | | 4,7 |
| Cardiac Pediatric | N | N | N | N | N | 3 | N | N | N | N | | 4,7 |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{7}------------------------------------------------
#### System: _Xario 100 TUS-X100,TUS-X100S V3.0 Transducer: PVU-375BT
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | P | P | P | P | P | 2 | P | P | P | | | 4,5,6,7 |
| Abdominal | P | P | P | P | P | 2 | P | P | P | | | 4,5,6,7 |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | P | P | P | P | P | 2 | P | P | P | | | 4,5,6,7 |
| Small Organ (Specify) (1) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K133277
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{8}------------------------------------------------
#### System: Xario 100 TUS-X100, TUS-X100S V3.0 Transducer: PVU-382BT
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Abdominal | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Small Organ (Specify) (1) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
Note 7 Precision Imaging
- Note 8 Differential THI
- Note 9 Elastography
Note 10 BEAM
{9}------------------------------------------------
#### System: Xario 100 TUS-X100, TUS-X100S V3.0 Transducer: PVU-674BT
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Abdominal | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Small Organ (Specify) (1) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{10}------------------------------------------------
#### System: _Xario 100 TUS-X100,TUS-X100S V3.0 Transducer: PVU-674MV
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | P | P | P | P | P | 2 | P | P | P | | P | 5,6,7 |
| Abdominal | P | P | P | P | P | 2 | P | P | P | | P | 5,6,7 |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | P | P | P | P | P | 2 | P | P | P | | P | 5,6,7 |
| Small Organ (Specify) (1) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducers: K133277
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{11}------------------------------------------------
#### System: Xario 100 TUS-X100, TUS-X100S V3.0 Transducer: PVU-712BT
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----------|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI<br>2D | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Intra-operative (Abdominal) | | | | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Small Organ (Specify) (1) | | | | | | | | | | | | |
| Neonatal Cephalic | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{12}------------------------------------------------
#### System: Xario 100 TUS-X100, TUS-X100S V3.0 Transducer: PVU-745BTF
| Clinical Application | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] |
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Intra-operative (Abdominal) | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Intra-operative (Neuro) | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ (Specify) (1) | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{13}------------------------------------------------
#### System: Xario 100 TUS-X100, TUS-X100S V3.0 Transducer: PVU-745BTH
| Clinical Application | Mode of Operation | | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|--|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 | |
| Intra-operative (Abdominal) | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 | |
| Intra-operative (Neuro) | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (Specify) (1) | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | | | | | |
| Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | | | | | |
| Intra-cardiac | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | | | |
| Other (Specify) | | | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
- Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 ApliPure Plus
Note 7 Precision Imaging
Note 8 Differential THI
Note 9 Elastography
Note 10 BEAM
{14}------------------------------------------------
#### System: Xario 100 TUS-X100, TUS-X100S V3.0 Transducer: PVU-745BTV
| Clinical Application | Mode of Operation | | | | | | | | | | | | |
|------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------------------|-------|-----|----|-----------------|--|
| Specific<br>(Tracks 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Advanced<br>Dynamic<br>Flow | Power | CHI | 4D | Other<br>[Note] | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 | |
| Intra-operative (Abdominal) | N | N | N | N | N | 2 | N | N | N | | | 5,6,7,8 | |
| Intra-operative (Neuro) | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (Specify) (1) | N…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.