← Product Code [LXJ](/productcode/LXJ) · K143034

# Jintronix (K143034)

_Jintronix, Inc. · LXJ · Apr 22, 2015 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143034

## Device Facts

- **Applicant:** Jintronix, Inc.
- **Product Code:** [LXJ](/productcode/LXJ.md)
- **Decision Date:** Apr 22, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5360
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Jintronix is a software system used with the Microsoft Kinect v2 intended to be used to support the physical rehabilitation of adults in the clinic/ at home. The system includes rehabilitation exercises for the upper extremity, trunk, and lower extremity with audio-visual feedback & graphic movement representations for patients as well as remotely accessible patient performance metrics for the medical professional. Patient assessment, exercise guidance and approval by the medical professional is required prior to use.

## Device Story

Jintronix is a software system used with Microsoft Kinect v2 hardware to support physical rehabilitation. The system tracks patient joint positions (upper extremity, trunk, lower extremity) via the Kinect sensor; transforms movement data into graphic representations and audio-visual feedback for the patient; and records performance metrics (speed, precision). Used in clinics or homes under the guidance of a medical professional. Clinicians access a remote web application to monitor patient progress and customize exercise programs (e.g., difficulty levels). The device provides no direct patient contact and delivers no energy. Benefits include remote monitoring and standardized exercise guidance, potentially improving rehabilitation adherence and outcomes.

## Clinical Evidence

Bench testing only. No clinical trials were conducted. An independent bench study compared Microsoft Kinect v2 performance against Optotrak (gold standard) and goniometry. Results demonstrated that Jintronix range of motion assessments are reproducible within an error tolerance comparable to intra-rater precision error of goniometry. System compatibility and data processing accuracy were validated through software testing.

## Technological Characteristics

Software system utilizing Microsoft Kinect v2 hardware for optical motion capture. Non-invasive, non-sterile, no energy delivery. Connectivity via networked web application for remote clinician access. Software developed per AAMI/ANSI/IEC 62304 and AAMI TIR45. Kinect v2 hardware conforms to IEC 60825-1:2007 Class 1 laser standards.

## Regulatory Identification

Measuring exercise equipment consist of manual devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. These devices also include instrumentation, such as the pulse rate monitor, that provide information used for physical evaluation and physical planning purposes., Examples include a therapeutic exercise bicycle with measuring instrumentation, a manually propelled treadmill with measuring instrumentation, and a rowing machine with measuring instrumentation.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a measuring exerciser or an interactive rehabilitation exercise device for prescription use only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- Jintronix Rehabilitation System ([K130847](/device/K130847.md))
- Peak Motus Motion Measurement System ([K030714](/device/K030714.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The caduceus is composed of three abstract human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 22, 2015

Jintronix Inc. c/o Navneet Sekhon AxSource Consulting Inc. 336 Bronte Street South, Suite 224-225 Milton, Ontario L9T 7W6

Re: K143034

Trade/Device Name: Jintronix Version 2.0 Regulatory Class: Unclassified Product Code: LXJ Dated: March 19, 2015 Received: March 20, 2015

Dear Ms. Sekhon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Felipe Aquel -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K143034

Device Name Jintronix Version 2.0

#### Indications for Use (Describe)

Jintronix is a software system used with the Microsoft Kinect v2 intended to be used to support the physical rehabilitation of adults in the clinic/ at home. The system includes rehabilitation exercises for the upper extremity, trunk, and lower extremity with audio-visual feedback & graphic movement representations for patients as well as remotely accessible patient performance metrics for the medical professional. Patient assessment, exercise guidance and approval by the medical professional is required prior to use.

| Type of Use (Select one or both, as applicable)                          |                                                                         |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### JINTRONIX TRADITIONAL 510(k)

# 7. 510(k) Summary

# 7.1 Owner Information

| Name:                | Jintronix Inc.                                     |
|----------------------|----------------------------------------------------|
| Device common name:  | System, optical position/movement recording        |
| Address:             | 999 3rd Avenue, Suite 3400, Seattle, WA 98104      |
| Phone:               | 1-514-754-6688                                     |
| Fax:                 | None                                               |
| Contact:             | Mark Evin                                          |
| Title:               | Regulatory Affairs & Quality Assurance Head (RAQA) |
| Email:               | mark@jintronix.com                                 |
| Date of Preparation: | March 13, 2015                                     |

# 7.2 Regulatory Correspondent Information

| Name:           | AxSource Consulting Inc.                                        |
|-----------------|-----------------------------------------------------------------|
| Address:        | 336 Bronte Street South, Suite 224-225 Milton, Ontario, L9T 7W6 |
| Office Phone:   | 905-854-6059                                                    |
| Contact Person: | Ms. Navneet Sekhon, President                                   |
| Email:          | nav.sekhon@axsource.ca                                          |

# 7.3 Device Information

| Trade Name                | Jintronix                                   |
|---------------------------|---------------------------------------------|
| Common Name               | System, optical position/movement recording |
| Classification name       | System, optical position/movement recording |
| Model Number              | Version 2.0.140905                          |
| 510(k) Submitter / Holder | Jintronix Inc.                              |
| Device Panel              | Physical Medicine                           |
| Product Code              | LXJ                                         |
| Classification Regulation | Unclassified                                |

# 7.4 Predicate(s) / Substantially Equivalent Device(s)

| General<br>information | 510(k)      | Predicate Device(s)<br>[510(k) summaries attached – Appendices 6 & 7] |
|------------------------|-------------|-----------------------------------------------------------------------|
| Trade Name             | Jintronix   | Peak Motus Motion<br>Measurement System                               |
| Model Number           | Version 1.0 | Unknown                                                               |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Jintronix. The logo features a stylized square shape with interconnected blue and green squares inside, resembling a network or circuit board. Below the graphic is the company name, "JINTRONIX," in a bold, sans-serif font, with "JINTRON" in blue and "IX" in green, mirroring the colors used in the square graphic above.

| General<br>information    | 510(k)            | Predicate Device(s)<br>[510(k) summaries attached – Appendices 6 & 7] |  |
|---------------------------|-------------------|-----------------------------------------------------------------------|--|
| 510(k) Submitter / Holder | Jintronix Inc.    | Peak<br>Technologies Inc.                                             |  |
| 510(k) Number             | K130847           | K030714                                                               |  |
| Device Panel              | Physical Medicine | Physical Medicine                                                     |  |
| Product Code              | LXJ               | LXJ                                                                   |  |

## 7.5 Device Description

Jintronix Version 2.0.140905, a software system comprising of a client application and a web application is to be used with the Microsoft Kinect version 2 to support the physical rehabilitation of adults in the clinic/ at home.

Through the client application, the patient engages in their rehabilitation exercises. Using the Microsoft Kinect for Windows, which tracks the patient's upper extremity, trunk, and lower body joint positions as they move, the software provides a graphic representation of the patient's movements, as well as activities that include visual and audio feedback to support the patient in performing therapeutic exercise.

As the patient engages in exercises, the system records performance metrics, such as the speed and precision of their movements. These metrics are then sent to a remote web application.

Clinicians, through a "clinician interface" of the Client application, securely access the remote web application to remotely monitor the progress of their patients, and modify the exercise program through program customizations (for example, difficulty level) based on patients' needs.

Patient assessment, exercise guidance and approval by the medical professional is required prior to use. There is no direct contact with the patient since the device is a software product. No energy is delivered to the subject.

### 7.6 Indications for Use

Jintronix is a software system used with the Microsoft Kinect v2 intended to be used to support the physical rehabilitation of adults in the clinic/ at home. The system includes rehabilitation exercises for the upper extremity, trunk, and lower extremity with audio-visual feedback & graphic movement representations for patients as well as remotely accessible patient performance metrics for the medical professional. Patient assessment, exercise guidance and approval by the medical professional is required prior to use.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains a logo for a company called Jintronix. The logo features a stylized square shape with rounded corners, filled with a pattern of blue and green squares. Below the square is the company name, "JINTRONIX," written in a bold, sans-serif font, with the "JIN" in blue and the "TRONIX" in green.

# 

|                        | Predicate Device(s)<br>[510(k) summaries attached]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                     | Jintronix<br>(v2.0.140905)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Jintronix and<br>Predicate(s)<br>Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                        | Jintronix<br>Rehabilitation<br>System (JRS)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Peak Motus                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Indications<br>for Use | A software system<br>used with the<br>Microsoft Kinect<br>intended to be<br>used to support<br>the physical<br>rehabilitation of<br>adults in the clinic/<br>at home. The<br>system includes<br>rehabilitation<br>exercises for the<br>upper extremity<br>and trunk with<br>audio-visual<br>feedback &<br>graphic movement<br>representations for<br>patients as well as<br>remotely<br>accessible patient<br>performance<br>metrics for the<br>medical<br>professional.<br>Patient<br>assessment,<br>exercise guidance<br>and approval by<br>the medical<br>professional is<br>required prior to<br>use. 1 | Computer and<br>video system used<br>to quantify and<br>graphically display<br>human movement<br>patterns<br>and techniques for<br>uses such as<br>assessment and<br>training of limb or<br>body motion in gait<br>analysis, prosthetic<br>design, pre/post<br>rehabilitation<br>evaluation, physical<br>therapy, and the<br>like.2 | A software<br>system used<br>with the<br>Microsoft Kinect<br>version 2<br>intended to be<br>used to support<br>the physical<br>rehabilitation of<br>adults in the<br>clinic/ at home.<br>The system<br>includes<br>rehabilitation<br>exercises for the<br>upper extremity,<br>trunk, and lower<br>extremity, with<br>audio-visual<br>feedback &<br>graphic<br>movement<br>representations<br>for patients as<br>well as remotely<br>accessible<br>patient<br>performance<br>metrics for the<br>medical<br>professional.<br>Patient<br>assessment,<br>exercise<br>guidance and<br>approval by the<br>medical<br>professional is<br>required prior to<br>use. | Jintronix shares<br>the same<br>indications for<br>use as the<br>predicates. All<br>devices are<br>intended for<br>physical<br>rehabilitation. All<br>devices include<br>graphic<br>movement<br>representations<br>and performance<br>metrics for<br>assessment.<br>Like the JRS,<br>Jintronix is for<br>clinic/at home<br>physical<br>rehabilitation of<br>the upper<br>extremity and<br>trunk.<br>Like predicate<br>Peak Motus,<br>Jintronix is<br>indicated for<br>limbs and full<br>body.<br>No difference<br>between<br>Jintronix and<br>predicates. |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains a logo for Jintronix. The logo features a blue and green square with a network of blue squares inside. Below the square is the word "JINTRONIX" in green, with the letters in a bold, sans-serif font. The logo appears to be for a technology or software company.

1 FDA 510(k) Database. Jintronix Rehabilitation System. 510(k) Summary (attached). Retrieved August 30, 2014, from http://www.accessdata.fda.gov/cdrh_docs/pdf13/K130847.pdf 2 FDA 510(k) Database. Peak Motus Motion Measurement System. 510(k) Summary (attached). Retrieved September 23, 2014, from http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?lD=11114

## 7.7 Comparison of Technological Characteristics with Predicate Device(s)

Principles of operation and fundamental design and technology considerations are shared by Jintronix and its predicates.

Specifically, Jintronix and its predicate(s)

- Are software / hardware systems or computer systems using optical capture motion technology 1.
- 2. Track movement and provide visual feedback or graphic movement representations
- 3. Report on kinematic parameters or provide performance metrics for assessment
- 4. Are non-invasive devices and so do not deliver energy to patients
- 5. Do not pose any issues in terms of electrical, chemical, thermal or radiation safety
- 6. Are non-sterile devices

Jintronix is validated for system compatibility and performance.

The following technological differences exist between Jintronix and its predicates.

- 1. Jintronix improves usability with its new design features which include enhanced software user interfaces and a simple range of motion assessment feature desired by clinicians.
- 2. Jintronix surpasses predicate JRS in performance. Jintronix uses the Kinect v2 for optical motion capture. The Kinect v2 is more precise over a greater tracking range than Kinect v1 used with predicate JRS.

### 7.8 Performance Data

### 7.8.1 Performance Data Summary / Conclusions

Jintronix system performance is validated as suitable for its indications for use surpassing the performance of predicate JRS. With improved performance accuracy, Jintronix has a better safety and effectiveness profile when compared to predicate JRS.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains a logo with the word "JINTRONIX" in green, with the top half of the logo containing a blue and green square design. The blue squares are connected to each other, and the green squares are on the top right of the design. The logo is surrounded by a green border.

# TRADITIONAL 510(k)

| Jintronix<br>System                                  | Supporting<br>Performance Data                                                                        | Performance Conclusions & SE Comparison<br>to Predicate(s)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Data Processing<br>Accuracy                          | 1. System Validation                                                                                  | 1. Jintronix data processing accuracy verified<br>by software testing.<br>2. Clinician feedback.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| System<br>Compatibility                              | System Validation<br>(software testing)                                                               | System compatibility of Jintronix software,<br>Microsoft Kinect v2 hardware and the<br>recommended computer operating system in<br>Jintronix labeling assured by software testing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Kinect<br>Performance<br>Accuracy /<br>Effectiveness | Bench Study                                                                                           | Jintronix surpasses predicate JRS in<br>performance. Jintronix uses the Kinect v2 (for<br>optical motion capture) which is more precise<br>over a greater tracking range than Kinect v1<br>used with predicate JRS.<br><br>Jintronix can be consistently reproduced within<br>an error tolerance that is comparable to intra-<br>rater precision error1 of goniometry. Shoulder<br>range of motion measurements can only be used<br>to describe a patient's relative change in joint<br>angle, not absolute joint angle. Shoulder range of<br>motion measurements have not been validated<br>in standardized postures associated with<br>goniometric measurements.<br><br>Bench study recommendations for optimal<br>performance have been incorporated or<br>accounted for. |
| Kinect Electrical<br>Safety                          | Refer to FDA<br>recognized standard<br>and test reports for<br>Class I laser (section<br>7.8.3 below) | Kinect deemed electrically safe.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

# 7.8.2 System Validation for System Compatibility & Performance Accuracy

Jintronix system accuracy in data processing was validated by software testing. Jintronix software is of "moderate" level of concern. All potential device hazards have been mitigated through appropriate controls. Jintronix software has been validated with Windows Defender, anti-virus software.

Jintronix conforms partially to the following FDA recognized standards and associated FDA guidance documents.

- 1. Medical device software Software life cycle processes AAMI / ANSI / IEC 62304:2006
<sup>1</sup> Defined as the degree of agreement among repeated administrations of a diagnostic test performed by a single rater

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Jintronix. The logo features a stylized graphic of interconnected squares in shades of blue and green, enclosed within a rounded rectangular border, also in blue and green. Below the graphic, the word "JINTRONIX" is written in a bold, sans-serif font, with "JINTRON" in blue and "IX" in green.

- 2. Guidance on the use of AGILE practices in the development of medical device software AAMI TIR45:2012
- 3. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff, May 2005
- 4. General Principles of Software Validation; Final Guidance for Industry and FDA Staff, January 2002

## 7.8.3 Electrical Safety

Jintronix is software for use with Microsoft Kinect v2 hardware. The Microsoft Kinect v2 used with Jintronix software is electrically safe. The Kinect for Windows v2 hardware conforms to the International Standard IEC 60825-1:2007:03 for a Class 1 laser product and with 21 CFR 1040.10 & 1040.11 except for deviations pursuant to Laser Notice No. 50, dated June 24, 2007.

### 7.8.4 Bench Study - Kinect v2 Performance Accuracy / Effectiveness

Jintronix Inc. commissioned an independent study to determine the performance accuracy of Microsoft Kinect v2, the optical motion capture technology for use with Jintronix. Kinect's accuracy was determined in terms of kinematics over a range of distances between the Kinect and the user. In the study, the Kinect v2 was compared to the Optotrak, a gold standard in motion sensing technology with a reported precision of less than 0.5mm. Kinect v2 performance was compared to that of Kinect v1 and goniometry, the clinical standard for range of motion measurements.

The bench study found that Jintronix range of motion (ROM) assessments using the Kinect v2 can be consistently reproduced within an error tolerance that is comparable to intra-rater precision error2 of goniometry.

### 7.9 Other Information

### 7.9.1 Jintronix Compliance to FDA Quality System Regulation (QSR) 21 CFR 820

Jintronix has been implemented with appropriate design and change controls per Jintronix Inc.'s quality system which is compliant with FDA's Quality System Regulation 21 CFR 820. Standard Operating Procedures relevant to this 510(k) have been referenced herein.

### 7.9.2 Jintronix Labeling Controls

Jintronix is a prescription use device and various labeling precaution statements further serve as potential device hazard mitigation controls. This includes requiring an in person patient-exercise suitability assessment prior to Jintronix prescription.

<sup>2</sup> Defined as the degree of agreement among repeated administrations of a diagnostic test performed by a single rater

{9}------------------------------------------------

### 7.10 This summary

- includes only information that is also covered in the body of the 510(k). ●
- . does not contain any puffery or unsubstantiated labeling claims.
- . does not contain any raw data, i.e., contains only summary data.
- . does not contain any trade secret or confidential commercial information.
- does not contain any patient identification information. .

---

**Source:** [https://fda-staging.innolitics.com/device/K143034](https://fda-staging.innolitics.com/device/K143034)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).