← Product Code [KWQ](/productcode/KWQ) · K143028

# Azure Anterior Cervical Plate System (K143028)

_Orthofix, Inc. · KWQ · May 29, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K143028

## Device Facts

- **Applicant:** Orthofix, Inc.
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** May 29, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Azure Anterior Cervical Plate System is a temporary implant intended for anterior fixation to the cervical spine from C2 to C7 and indicated for: a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies); b) Spondylolisthesis; c) Trauma (i.e., fracture or dislocation); d) Spinal stenosis; e) Deformities (i.e., scoliosis, kyphosis, and/or lordosis); f) Tumor: g) Pseudoarthrosis; h) Revision of previous surgery

## Device Story

Azure Anterior Cervical Plate System is a temporary spinal fixation implant; consists of titanium alloy plates and screws; used by surgeons in clinical settings to stabilize cervical spine (C2-C7). Device functions as mechanical support for vertebral bodies; surgeon attaches plates to anterior aspect of spine using screws. System modification involves removal of constrained screw options to eliminate risk of locking mechanism fracture; only semi-constrained screws remain. Output is physical stabilization of spine; aids in fusion/healing process. Benefits patient by providing structural support during recovery from trauma, deformity, or degenerative conditions.

## Clinical Evidence

Bench testing only. Modified Dynamic Axial Compression testing performed per ASTM F1717 to compare constrained vs. semi-constrained screw constructs. Results showed constrained screws caused locking mechanism fracture, while semi-constrained screws did not. Simulated testing validated the surgical technique for semi-constrained screws.

## Technological Characteristics

Materials: Titanium alloy (Ti6Al4V ELI per ASTM F136) and nitinol (per ASTM F2063). Components: Plates, screws, and instrumentation. Form factor: Anterior cervical plate system for spinal fixation. Sterilization: Non-sterile, single-use.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Azure Anterior Cervical Plate System ([K130825](/device/K130825.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected and appear to be a single, continuous line. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 29, 2015

Orthofix, Incorporated Ms. Natalia Volosen Senior Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056

Re: K143028

Trade/Device Name: Azure Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWO Dated: April 29, 2015 Received: April 30, 2015

Dear Ms. Volosen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Natalia Volosen

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

K143028

Page 1 of 1

510(k) Number (if known) K143028

Device Name

Azure Anterior Cervical Plate System

Indications for Use (Describe) The Azure Anterior Cervical Plate System is a temporary implant intended for anterior fixation to the cervical spine from C2 to C7 and indicated for: a) Degenerative disc disease (as defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies); b) Spondylolisthesis; c) Trauma (i.e., fracture or dislocation); d) Spinal stenosis; e) Deformities (i.e., scoliosis, kyphosis, and/or lordosis); f) Tumor: g) Pseudoarthrosis; h) Revision of previous surgery

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, bold letters. The "R" in "ORTHOFIX" has a registered trademark symbol next to it.

# 510(k) SUMMARY

# Azure Anterior Cervical Plate System

| 510(k) Owner Information<br>Name:<br>Address: | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056                                                         |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Telephone Number:<br>Fax Number:<br>Email:    | 214.937.2145<br>214-937-3322<br>nataliavolosen@orthofix.com                                                         |
| Registration Number:                          | 3008524126                                                                                                          |
| Contact Person:                               | Natalia Volosen<br>Senior Regulatory Affairs Specialist                                                             |
| Date Prepared:                                | May 14, 2015                                                                                                        |
| Name of Device                                |                                                                                                                     |
| Trade Name / Proprietary Name:                | Azure Anterior Cervical Plate System                                                                                |
| Common Name:                                  | Anterior Cervical Plate System                                                                                      |
| Product Code:                                 | KWQ - Appliance, Fixation, Spinal Intervertebral Body                                                               |
| Regulatory Classification:                    | Class II per 21 CFR § 888.3060                                                                                      |
| Review Panel:                                 | Orthopedic Device Panel                                                                                             |
| Predicate Devices:                            | K130825 - Azure Anterior Cervical Plate System, SE<br>5/7/2013<br>No reference devices were used in this submission |
| Reason for 510(k) Submission:                 | Removal of constrained screws from the Azure                                                                        |

system

## Device Description

The Azure Anterior Cervical Plate System is comprised of a variety of non-sterile, single use, titanium alloy (Ti6Al4V ELI per ASTM F136) with nitinol (per ASTM 2063) components that allow a surgeon to build a temporary anterior cervical implant construct. The system is attached to the anterior aspect of the vertebral body by means of screws to the cervical spine. The system consists of an assortment of screws, plates and associated instrumentation, which assists in the surgical implantation of the devices.

## Intended Use / Indications for Use

The AZURE Anterior Cervical Plate System is a temporary implant, intended for anterior fixation to the cervical spine from C2 to C7 and indicated for:

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, bold letters. There is a black line underneath the word "ORTHOFIX".

- a) Degenerative disc disease (as defined as back pain of discogenic origin with
- degenerative disc confirmed by patient history and radiographic studies);
- b) Spondylolisthesis;
- c) Trauma (i.e., fracture or dislocation);
- d) Spinal stenosis:
- e) Deformities (i.e., scoliosis, kyphosis, and/or lordosis);
- f) Tumor;
- q) Pseudoarthrosis:
- h) Revision of previous surgery

# Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices

There are no changes in the intended use, design, specifications and / or materials of Azure Cervical Plate system. The purpose of this 510(k) submission is to remove the constrained screw configuration from the Azure Anterior Cervical Plate system.

# PERFORMANCE DATA - Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence

A modified Dynamic Axial Compression test was conducted in accordance to ASTM F1717 standard for Azure constructs containing constrained and semi-constrained screws. The results of the modified ASTM F1717 testing show that constrained screw constructs were able to cause total fracture of the locking mechanism, similar to the failures experienced in the field. The semi-constrained screw constructs did not experience the same failure modes and therefore removing the constrained screws from the system will eliminate the risk of locking mechanism fracture.

In addition, Orthofix performed simulated testing with semi-constrained screws to validate the new recommended surgical technique (fully lagged screws or screws left above the locking mechanism).

These tests demonstrate that the Azure system containing semi-constrained screws is safe and effective for use and is substantially equivalent or better that its predicate device Azure Anterior Cervical Plate system K130825.

## Basis of Substantial Equivalence

There are no changes in the intended use, design, specifications and / or materials between the subject Azure system and the Azure Cervical Plate system K130825.

The subject Azure Anterior Cervical Plate System will be as safe and effective as its predicate device Azure Anterior Cervical Plate System (K130825). Usage of constrained or semi-constrained screws is a surgeon preference based on their training and experience.

---

**Source:** [https://fda-staging.innolitics.com/device/K143028](https://fda-staging.innolitics.com/device/K143028)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
