K142997 · Heartvista, Inc. · LNH · Dec 17, 2014 · Radiology
Device Facts
Record ID
K142997
Device Name
RTHawk, HeartVista Workstation with RTHawk
Applicant
Heartvista, Inc.
Product Code
LNH · Radiology
Decision Date
Dec 17, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The HeartVista Workstation with the RTHawk application software is an accessory to 1.5T and 3.0T whole-body magnetic resonance diagnostic devices (MRDD or MR). The HeartVista Workstation with the RTHawk application software is intended to operate alongside, and in parallel with, the existing MR console to acquire real-time and accelerated images. The HeartVista Workstation with the RTHawk application software is indicated for Cardiovascular MR (CMR) applications. The Heart Vista Workstation with the RTHawk application software produces static and dynamic transverse, coronal, sagittal, and oblique cross-sectional images that display the internal structures and/or functions of the entire body. The images produced reflect the spatial distribution of nuclei exhibiting magnetic resonance. The magnetic resonance properties that determine image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that may assist in the determination of a diagnosis. The Heart Vista Workstation with the RTHawk application software is intended for use as an accessory to the following OEM, MRI system, and software release versions: · GE Healthcare (GEHC) Signa HDxt 1.5T, 3.0T. Software versions 15 and 16 - · GE Healthcare (GEHC) DVMR 1.5T, 3.0T. Software version 24
Device Story
RTHawk is a Linux-based workstation accessory for 1.5T/3.0T MRI scanners; operates in parallel with existing MR consoles via private Ethernet. Inputs raw MR data; transforms data via optimized acquisition apps, pipelined reconstruction engine, and pulse sequence control. Produces real-time/accelerated static and dynamic cross-sectional images (transverse, coronal, sagittal, oblique). Used in clinical settings by radiologists/technicians. Provides real-time interactive imaging, system tuning, and shimming. Output displayed on workstation monitor; assists physicians in diagnostic decision-making by providing high-quality cardiovascular MR images. Benefits include efficient data acquisition and real-time scan control.
Clinical Evidence
Bench testing only. Clinical images acquired with RTHawk were compared to images from the predicate device on the same patients during the same sessions. Images were evaluated by radiologists for diagnostic accuracy and image quality. Safety parameters (SAR, dB/dt, acoustic noise) were verified against consensus standards and predicate results.
Technological Characteristics
Linux-based workstation; color monitor; keyboard; mouse. Connects via private Ethernet to OEM MR scanner. Functional modules: Acquisition, Analysis, Application (pulse sequences), Information System, Reconstruction (pipelined), Scan Control, Sequencer, Storage (DICOM), Visualization. Complies with ANSI/AAMI ES60601-1, IEC 60601-2-33, NEMA PS3.1-3.20 (DICOM), and ISO 14971.
Indications for Use
Indicated for Cardiovascular MR (CMR) applications in patients undergoing 1.5T or 3.0T whole-body magnetic resonance imaging. Used as an accessory to existing MR systems to acquire real-time and accelerated images for diagnostic interpretation by a trained physician.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
RTHawk, or HeartVista Workstation with RTHawk (K133848)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol depicts a stylized eagle with three human profiles incorporated into its design. The profiles are facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2014
HeartVista, Inc. % Mr. James Rogers FDA Regulatory Affairs, Quality Assurance, and Clinical Studies 998 Hamilton Ave. MENLO PARK CA 94025
Re: K142997 Trade/Device Name: RTHawk, or HeartVista Workstation with RTHawk Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: November 5, 2014 Received: November 17, 2014
Dear Mr. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert A. Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K142997
Device Name
RTHawk, or HeartVista Workstation with RTHawk
### Indications for Use (Describe)
The HeartVista Workstation with the RTHawk application software is an accessory to 1.5T and 3.0T whole-body magnetic resonance diagnostic devices (MRDD or MR). The HeartVista Workstation with the RTHawk application software is intended to operate alongside, and in parallel with, the existing MR console to acquire real-time and accelerated images. The HeartVista Workstation with the RTHawk application software is indicated for Cardiovascular MR (CMR) applications.
The Heart Vista Workstation with the RTHawk application software produces static and dynamic transverse, coronal, sagittal, and oblique cross-sectional images that display the internal structures and/or functions of the entire body. The images produced reflect the spatial distribution of nuclei exhibiting magnetic resonance. The magnetic resonance properties that determine image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that may assist in the determination of a diagnosis.
The Heart Vista Workstation with the RTHawk application software is intended for use as an accessory to the following OEM, MRI system, and software release versions:
· GE Healthcare (GEHC) Signa HDxt 1.5T, 3.0T. Software versions 15 and 16
- · GE Healthcare (GEHC) DVMR 1.5T, 3.0T. Software version 24
| Type of Use (Select one or both, as applicable) | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|-----------------------------------------------|
| <table><tr><td> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </td></tr><tr><td> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </td></tr></table> | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | | |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "HEARTVISTA" in a stylized font. The word is written in black, except for the last letter "A", which is red. There is also a red dot above the "V".
# SPECIAL 510(k) Summary RTHawk 510(k) Number: K142997
Submitted in accordance with the requirements of SMDA 1990 and 21CFR807.92.
#### Medical Establishment Registration 1.0
Medical Establishment Registration No.: pending
### 2.0 Contact Information
James Jochen Rogers FDA Regulatory Affairs, Quality Assurance, and Clinical Studies T: 724 713.2298 E: jr@heartvista.com
#### 3.0 Establishment Name and Address
HeartVista, Inc. 998 Hamilton Ave Menlo Park, CA 94025
#### 4.0 Submission Date
October 8, 2014, December 16, 2014
#### 5.0 Device Information
Trade/Proprietary Name: RTHawk, or HeartVista Workstation with RTHawk Common Name: RTHawk, or HeartVista Workstation with RTHawk Model Number(s):
- O 9001 HeartVista Workstation with RTHawk
- 9002 RTHawk O
Regulation Number: 892.1000 Regulation Name: Magnetic resonance diagnostic device (MRDD) Regulatory Class: Class II Device Classification Name: System, Nuclear Magnetic Resonance Imaging Classification Panel: Radiology Classification Product Code(s): LNH
### 6.0 Predicate Device(s)
| 510(k) # | Device | 510(k) Sponsor | 510(k) Clearance<br>Date |
|----------|-----------------------------------------------|----------------|--------------------------|
| K133848 | RTHawk, or HeartVista Workstation with RTHawk | HeartVista | 6/25/2014 |
#### Device Description 7.0
The HeartVista Workstation with RTHawk application software are a software and dedicated hardware product intended to provide a platform for efficient real-time MRI data acquisition, data transfer, image reconstruction, and interactive scan control and display of static and dynamic MR imaging data.
minum
{4}------------------------------------------------
The HeartVista Workstation consists of a stand-alone linux-based computer workstation, color monitor, keyboard and mouse. It is designed to operate alongside, and in parallel with, the existing MR console with no hardware modifications required to be made to the MR system or console. A private ethernet network connects the HeartVista Workstation to the MR scanner computer. When not in use, the HeartVista Workstation may be detached from the MR scanner with no detrimental, residual impact upon MR scanner function, operation, or throughput.
RTHawk is a linux operating system-level software application that is intended to control the MR scanner, acquiring high quality, real-time MRI image data and performing post-processing. The RTHawk software includes optimized image acquisition applications, a pipelined raw data image reconstruction engine, a rich graphical user interactive scan control, real-time adjustment of pulse sequence parameters, and display of reconstructed images, and drivers and protocols for communications with, and control of, the OEM MR scanner console.
RTHawk applications ("apps") support real-time interactive imaging, high-resolution imaging, and system tuning and shimming calibration modules. RTHawk apps are currently optimized for cardiovascular MR (CMR) imaging and measurements.
| Reference # | Title |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI ES60601-1:2005/<br>(R)2012 +C1 +A2 | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance (IEC 60601-1:2005, MOD)<br>Section 14 Programmable Electrical Medical Systems (PEMS) |
| IEC 60601-2-33 Ed 3.0 (2010-03) | Medical electrical equipment - Part 2-33: Particular requirements<br>for the basic safety and essential performance of magnetic<br>resonance equipment for medical diagnostic (radiology). |
| MS1-2008 | Determination of Signal-to-Noise Ratio (SNR) in Diagnostic<br>Magnetic Resonance Imaging |
| MS3-2008 | Determination of Image Uniformity in Diagnostic Magnetic<br>Resonance Images |
| MS4-2010 | Acoustic Noise Measurement Procedure for Diagnostic Magnetic<br>Resonance Imaging Devices |
| MS8-2008 | Characterization of the Specific Absorption Rate (SAR) for<br>Magnetic Resonance Imaging Systems |
| NEMA PS3.1 - 3.20 (2011) | Digital Imaging And Communications In Medicine (DICOM) Set. |
| ISO 14971:2007 | Medical devices - Application of risk management to medical<br>devices. |
RTHawk has been designed to comply with the FDA Recognized Consensus Standards listed in the table below, as applicable to device features and components:
## 8.0 Indications for Use
The HeartVista Workstation with the RTHawk application software is an accessory to 1.5T and 3.0T whole-body magnetic resonance diagnostic devices (MRDD or MR). The HeartVista Workstation with the RTHawk application software is intended to operate alongside, and in parallel with, the existing MR console to acquire real-time and accelerated images. The HeartVista Workstation with the RTHawk application software is indicated for Cardiovascular MR (CMR) applications.
The HeartVista Workstation with the RTHawk application software produces static and dynamic transverse, coronal, sagittal, and oblique cross-sectional images that display the internal structures and/or functions of the entire body. The images produced reflect the spatial distribution of nuclei exhibiting magnetic resonance. The magnetic resonance properties that determine image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that may assist in the determination of a diagnosis.
Image /page/4/Figure/8 description: The image shows a red wave pattern that starts with a high frequency and gradually decreases. The wave eventually flattens out into a straight line. The number 3 is located in the bottom right corner of the image.
{5}------------------------------------------------
The HeartVista Workstation with the RTHawk application software is intended for use as an accessory to the following OEM, MRI system, and software release versions:
- O GE Healthcare (GEHC) Signa HDxt 1.5T, 3.0T. Software versions 15 and 16
- O GE Healthcare (GEHC) DV 1.5T, 3.0T. Software version 24.
### 9.0 Technological Characteristics Comparison to Predicate Device
Identical.
The RTHawk software is identical to comprised of the following functional modules:
- O Acquisition - responsible for the transfer of MR raw data from the MR scanner to the HeartVista Workstation
- O Analysis contains the image post-processing tools
- O Application - HeartVista APPs. Each APP is comprised of a pulse sequence, user parameters, a reconstruction pipeline, and a specific user interface
- Information System the central repository of all relevant MRI system configuration, patient, ○ study, scan, etc., parameters associated with the current patient study
- O Reconstruction - responsible for the efficient processing of raw data to generate MR images via a flexible, pipelined topology
- Scan Control responsible for the real-time network transfer of controlling orders for APPs. APPs o parameters modifications, and dynamic information from the MR host in response to user or program requests
- Sequencer creates and provides a specific set of pulse sequence waveforms to control the O MR scanner
- o Storage - obtains current patient and scan information, performs non-volatile local storage, exports images and data in DICOM format, and logs events.
- O Visualization - implements all aspects of the user iterface, including APP selection, controls to modify APP parameters, image display, graphical slice prescription, and image review, save, and export.
Instructions for use are included within the device labeling, and the information provided enables the user to operate the device in a safe and effective manner.
### 10.0 Performance Data - Discussion of Non-Clinical Tests
Design controls quality assurance measures during the development of RTHawk include:
- O Code reviews
- o Design reviews
- O Unit and integration level testing
- O Verification testing, including System and Manual testing
- Safety testing, including SAR, dB/dt, and acoustic noise O
- Performance testing, including SNR and uniformity
- O Validation testing
Risk management, compliant with ISO 14971:2007, identified hazards, sequences of events, and resultant harms; developed, implemented, and tested risk-controlling mitigations; and evaluated residual risks.
Image /page/5/Figure/26 description: The image shows a red wave pattern that starts with high frequency and gradually decreases. The wave eventually flattens out into a straight line. The number 4 is present in the bottom right corner of the image.
{6}------------------------------------------------
# 11.0 Safety Parameters
| | RTHawk | Predicate Device |
|---------------------------------------------|--------------------------|--------------------------|
| Magnet field strength | 1.5T, 3.0T | 1.5T, 3.0T |
| Operating Modes<br>IEC 60601-2-33 (2010-03) | 1st Level Operating Mode | 1st Level Operating Mode |
| Safety Parameter Display | SAR, dB/dt | SAR, dB/dt |
| Max SAR | <4W/kg whole-body | <4W/kg whole-body |
| Max dB/dt | 1st Level Operating Mode | 1st Level Operating Mode |
# 12.0 Performance Data - Bench Testing
Clinical images were acquired using RTHawk, and were compared to images acquired on the same patient during the same imaging session using, where possible, equivalent pulse sequences and postprocessing as necessary from the predicate device. Images from both devices were evaluated and rated on the basis of diagnostic accuracy and image quality. Where no directly comparable images were available from the predicate device, RTHawk images were evaluated directly based upon radiologist expertise.
Bench testing results for safety parameters of SAR, dB/dt, and acoustic noise were compared to results from K133848 and were consistent with previously reported results.
### 13.0 Conclusions
Based upon verification and validation testing, safety testing, and compliance with voluntary standards, the Company believes that the RTHawk, or the HeartVista Workstation with the RTHawk application software is substantially equivalent to the predicate device, and does not raise any new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.