K142990 · Cardinal Health200, LLC · FXX · Mar 4, 2015 · General, Plastic Surgery
Device Facts
Record ID
K142990
Device Name
Insta-Gard Surgical and Procedure Face Mask
Applicant
Cardinal Health200, LLC
Product Code
FXX · General, Plastic Surgery
Decision Date
Mar 4, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Cardinal Health Insta-Gard® surgical and procedure mask is intended to be worn by operating room personnel and other general healthcare workers to protect both patients and healthcare workers of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health Insta-Gard® surgical and procedure masks are single use, disposable devices provided non-sterile.
Device Story
Insta-Gard surgical and procedure masks; three-layer nonwoven construction; sonically welded edges; includes polyolefin ties or polyester-spandex earloops; malleable nosepiece for fit; optional eye shield; anti-fog foam strips; vapor barriers. Used in operating rooms and clinical settings by healthcare personnel. Single-use, disposable, non-sterile. Provides physical barrier against microorganisms, blood, body fluids, and airborne particulates. Protects both patient and wearer.
Clinical Evidence
Bench testing only. Device tested in accordance with ASTM F2100-11 Standard Specification for Performance of Materials Used in Medical Face Masks. All models met Level 1 criteria. Testing performed at an AQL of 4% (32 samples tested).
Technological Characteristics
Three-layer construction; materials include nonwoven tissue, cellulose, and polyolefin; sonically welded; includes polyolefin ties or polyester-spandex earloops; malleable nosepiece; optional anti-fog foam strip, vapor barrier, and eye shield. Tested to ASTM F2100-11 Level 1. Non-sterile.
Indications for Use
Indicated for operating room personnel and general healthcare workers to protect patients and wearers from microorganisms, blood, body fluids, and airborne particulates.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2015
Cardinal Health 200, LLC Ms. Lavenia Ford Manager, Regulatory Affairs 1500 Waukegan Road Waukegan, Illinois 60085
Re: K142990
Trade/Device Name: Cardinal Health Insta-Gard® Surgical and Procedure Masks Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 4, 2015 Received: February 5, 2015
Dear Ms. Ford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142990
#### Device Name
Cardinal Health Insta-Gard® surgical and procedure masks
Indications for Use (Describe)
Cardinal Health Insta-Gard® surgical and procedure mask is intended to be worn by operating room personnel and other general healthcare workers to protect both patients and healthcare workers of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health Insta-Gard® surgical and procedure masks are single use, disposable devices provided non-sterile.
This submission covers 15 catalog numbers of Insta-Gard® surgical and procedure masks models, see Table I below. Each model is a three-laver mask of varying composition and has been tested according to ASTM F2100-11 Standard Specification for Performance of Materials Used in Medical Face Masks and meets Level 1 criteria.
Table 1: Product Description and Catalog Number
Catalog # Description, Color
AT771145 Insta-Gard® surgical mask Ties Dancing Bandages Print AT71235 Insta-Gard® surgical mask Anti-fog foam strip, and tiesBlue AT71035 Insta-Gard® surgical mask ties Blue AT71039 Insta-Gard@ surgical mask Horizontal ties Blue AT73035Insta-Gard® surgical maskTiesWhite AT752007 Insta-Gard® surgical mask Anti-fog foam strip, vapor barrier, and ties Green AT752005 Insta-Gard® surgical mask Anti-fog foam strip, vapor barrier, and ties Green AT73835 Insta-Gard® surgical mask Anti-fog foam strip, and ties Green AT72835 Insta-Gard® surgical mask Anti-fog foam strip, and ties Green AT7505P Insta-Gard® surgical mask Ties Blue AT73335 Insta-Gard® surgical mask Ties White AT71021 Insta-Gard® procedure mask Earloops Blue AT70021 Insta-Gard® procedure mask Earloops Yellow AT7511-WE Insta-Gard® procedure mask Eye shield, anti-fog foam strip, and earloops Blue AT771141 Insta-Gard® procedure mask Earloops Dancing Bandages Print
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
#### 1500 Waukegan Road Waukegan, IL 60085
www.cardinalhealth.com
### 510(k) SUMMARY Insta-Gard® Surgical and Procedure Masks
Manufacturer:
Cardinal Health 200, LLC 1500 Waukegan Road Waukegan, IL 60085
1500 Waukegan Road
Lavenia Ford
(847) 887-3323
December 31, 2014
Regulatory Affairs Contact:
Waukegan, IL 60085
Telephone Number:
Fax Number: (847) 785-2461
Date summary Prepared:
Trade Name:
Insta-Gard® Surgical Mask
Insta-Gard® Procedure Mask
Regulation Number/Device Class: Class II per 21 CFR § 878.4040
FXX
Surgical Apparel Regulation Name:
Common Name:
Surgical Mask Procedure Mask
Product Code:
K110455 -Kimberly-Clark, KC100 Face Mask(s)
Predicate Device:
{4}------------------------------------------------
# Description
The Cardinal Health Insta-Gard surgical and procedure masks are identified by Regulation 21 CFR 878.4040 with product code FXX.
The Cardinal Health Insta-Gard® surgical and procedure masks are constructed of varying materials (see Table 3). The three layers of the mask body are collated and sonically welded around the edges to enclose the filter media. The mask is provided with nonwoven polyolefin ties or polyester-spandex earloops (both attached by ultrasonic welding). A malleable nosepiece is placed within the binding for comfort and individualized fit around the wearer's nose. The surgical masks will be provided with or without an eye shield. Cardinal Health surgical and procedure masks are a single use, disposable device provided non-sterile.
### Indications for Use
Cardinal Health Insta-Gard® surgical and procedure masks are intended to be worn by operating room personnel and other general healthcare workers to protect both patients and healthcare workers against transfer of microorganisms, blood and body fluids, and airborne particulates. The Cardinal Health Insta-Gard® surgical and procedure masks are single use, disposable devices provided non-sterile.
This submission covers 15 catalog numbers of Insta-Gard® surgical and procedure masks models, see Table 1 below. Each model is a three-layer mask of varying composition and has been tested according to ASTM F2100-11 Standard Specification for Performance of Materials Used in Medical Face Masks and meets Level 1 criteria.
| Catalog # | Trade Name | Design Features | Color |
|-----------|----------------------------|--------------------------------------------------|---------------------------|
| AT771145 | Insta-Gard® surgical mask | Ties | Dancing<br>Bandages Print |
| AT71235 | Insta-Gard® surgical mask | Anti-fog foam strip, and ties | Blue |
| AT71035 | Insta-Gard® surgical mask | ties | Blue |
| AT71039 | Insta-Gard® surgical mask | Horizontal ties | Blue |
| AT73035 | Insta-Gard® surgical mask | Ties | White |
| AT752007 | Insta-Gard® surgical mask | Anti-fog foam strip, vapor<br>barrier, and ties | Green |
| AT752005 | Insta-Gard® surgical mask | Anti-fog foam strip, vapor<br>barrier, and ties | Green |
| AT73835 | Insta-Gard® surgical mask | Anti-fog foam strip, and ties | Green |
| AT72835 | Insta-Gard® surgical mask | Anti-fog foam strip, and ties | Green |
| AT7505P | Insta-Gard® surgical mask | Ties | Blue |
| AT73335 | Insta-Gard® surgical mask | Ties | White |
| AT71021 | Insta-Gard® procedure mask | Earloops | Blue |
| AT70021 | Insta-Gard® procedure mask | Earloops | Yellow |
| AT7511-WE | Insta-Gard® procedure mask | Eye shield, anti-fog foam strip,<br>and earloops | Blue |
| AT771141 | Insta-Gard® procedure mask | Earloops | Dancing<br>Bandages Print |
#### Table 1: Product Description and Catalog Number
{5}------------------------------------------------
# Device and Predicate Device Technical Characteristics
The Cardinal Health Insta-Gard® surgical and procedure masks are substantially equivalent to the Kimberly Clark surgical and procedure mask with regards to claims, safety and effectiveness, design, technology, and intended use. See Table 4 below.
| Table 4: Comparison of Predicate device and Cardinal Health Insta-Gard® Surgical and<br>Procedure Masks | | |
|---------------------------------------------------------------------------------------------------------|------------------------|-------------------------------------------------------------|
| Element of Comparison | Kimberly-Clark (KC100) | Cardinal Health Insta-Gard®<br>Surgical and Procedure Masks |
| Element of Comparison | Kimberly-Clark (KC100) | Cardinal Health Insta-Gard Surgical and Procedure Masks |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Kimberly Clark KC100 Face<br>Mask(s) is intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms, body<br>fluids, and particulate material.<br>These face masks are intended<br>for use in infection control<br>practices to reduce the potential<br>exposure of the wearer to blood<br>and body fluids. The Kimberly<br>Clark KC100 face mask(s) is a<br>single use, disposable device(s),<br>provided non-sterile. | Insta-Gard® Surgical Mask<br>Cardinal Health Insta-Gard®<br>surgical masks are intended to<br>be worn by operating room<br>personnel and other general<br>healthcare workers to protect<br>both patients and healthcare<br>workers against transfer of<br>microorganisms, blood and body<br>fluids, and airborne particulates.<br>The Cardinal Health Insta-Gard®<br>surgical masks are single use,<br>disposable devices provided non-<br>sterile.<br><br>Insta-Gard® Procedure Mask<br>Cardinal Health Insta-Gard®<br>procedure masks are intended to<br>be worn by operating room<br>personnel and other general<br>healthcare workers to protect<br>both patients and healthcare<br>workers against transfer of<br>microorganisms, blood and body<br>fluids, and airborne particulates.<br>The Cardinal Health Insta-Gard®<br>procedure masks are single use,<br>disposable devices provided non-<br>sterile |
| Material Composition | 3 layer mask made from<br>nonwoven polyester blends,<br>tissue, cellulose, and<br>polypropylene materials | Insta-Gard® Surgical Mask<br>3 layer mask made from<br>nonwoven tissue, cellulose, and<br>polyolefin materials<br>Insta-Gard® Procedure Mask<br>3 layer mask made from<br>nonwoven tissue, cellulose, and<br>polyolefin materials |
| Specifications and Dimensions | KC100 Surgical Mask<br>6.875"x3.75"<br><br>KC100 Procedure Mask<br>6.9"x4.5" | Insta-Gard® Surgical Mask<br>AT771145 – 7"x4"<br>AT71235 – 7"x4"<br>AT71035 – 7"x4"<br>AT71039 – 7.5"x7.5"<br>AT73035 |
{6}------------------------------------------------
| | | | AT752007 – 7"x4"<br>AT752005 – 7"x4"<br>AT73835 – 7"x4"<br>AT72835 – 7"x4"<br>AT7505P – 7"x4"<br>AT73335 – 7"x4" | |
|------------------|--|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | Insta-Gard® Procedure Mask<br>AT7511-WE – 7"x3.375"<br>AT771141 – 7"x3.375"<br>AT71021 – 7"x3.375"<br>AT70021 – 7"x3.375" | |
| | | | See Appendix I for mask<br>specifications | |
| Mask Style | | Pleated | Insta-Gard® Surgical Mask<br>Pleated<br>Insta-Gard® Procedure Mask<br>Pleated | |
| Design Features | | KC100 Surgical Mask<br>Fog-Free | Insta-Gard® Surgical Mask<br>AT771145 – Ties<br>AT71235 – Anti-fog foam strip<br>and Ties<br>AT71035 – Ties<br>AT71039 – Horizontal Ties<br>AT73035 – Ties<br>AT752007 – Anti-fog foam strip,<br>Vapor barrier, and Ties | |
| | | KC100 Procedure Mask<br>Fog-Free | AT752005 – Anti-fog foam strip,<br>Vapor barrier, and Ties<br>AT73835 – Anti-fog foam strip<br>and Ties<br>AT72835 – Anti-fog foam strip<br>and Ties<br>AT7505P – Ties<br>AT73335 – Ties | |
| | | | Insta-Gard® Procedure Mask<br>AT7511-WE – Eye shield, Anti-<br>foam fog strip, and Earloops<br>AT771141 – Earloops<br>AT71021 – Earloops<br>AT70021 – Earloops | |
| Physical Testing | | | Predicate device was tested<br>according to ASTM F2100-11<br>Level 1 in previous 510(k)<br>submission K110455. | Insta-Gard® Surgical Mask<br>Device was tested in accordance<br>with ASTM F2100-11, and meets<br>Level 1 requirements.<br>Insta-Gard® Procedure Mask<br>Device was tested in accordance<br>with ASTM F2100-11, and meets<br>Level 1 requirements. |
. . . .
{7}------------------------------------------------
All results of testing met ASTM F2100-11 Level 1 acceptance criteria. ASTM F2100-11 Level 1 was performed at an AQL of 4% (32 samples tested, accept on 3 failures and reject on 4 failures).
# Conclusion Statement
The Insta-Gard® surgical and procedure masks are substantially equivalent to the predicate device, in terms of general intended use performance testing, material composition, configurations/dimensions, and safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.