K142989 · Applied Medical Technology, Inc. · PIF · Jan 29, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K142989
Device Name
Mini ONE Enteral Extension Set
Applicant
Applied Medical Technology, Inc.
Product Code
PIF · Gastroenterology, Urology
Decision Date
Jan 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Mini ONE® Enteral Extension Set is intended for use as an extension set for the Mini ONE® Gastrostomy Button, G-Jet Gastrojejunal Feeding Tube, Mic-Key® Secure-Lok® feeding devices, or other compatible enteral feeding devices. The extension set is for administering feeding, medications, and decompression to compatible devices.
Device Story
Extension set for enteral feeding devices; connects feeding tubes (e.g., Mini ONE Gastrostomy Button, G-Jet) to external feeding bags or syringes. Components include: Mini ONE connector (straight or right-angle), flexible tubing (various lengths/diameters), pinch clamp, and ENFit enteral connectors (single or Y-port). Used in clinical or home settings; operated by clinicians or patients/caregivers. Facilitates delivery of nutrition/medication and gastric decompression. ENFit connectors designed per ISO/IEC 80369 series to reduce misconnection risk. Device is non-sterile, single-user, DEHP/latex-free.
Clinical Evidence
Bench testing only. Tests included tensile strength, fluid leakage, stress cracking, resistance to separation (axial load, unscrewing, overriding), and flow testing. Human factors study conducted to evaluate usability of ENFit connectors; results indicated no negative impact on device function or intended use. Misconnection testing performed per ISO 80369-1.
Technological Characteristics
Components: Mini ONE connector (straight/right-angle), flexible tubing, pinch clamp, ENFit connectors. Materials: DEHP-free, latex-free. Design: ISO/IEC 80369 series compliant connectors. Configurations: 2-24" lengths, 0.218" or 0.165" diameters, single or Y-port ENFit configurations. Non-sterile, single-user. Mechanical connection principle.
Indications for Use
Indicated for patients requiring enteral feeding, medication administration, or gastric decompression via compatible gastrostomy or gastrojejunal feeding tubes.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, stacked on top of each other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 30, 2015
Applied Medical Technology, Inc. Joshua Meinke OA/Regulatory Supervisor 8000 Katherine Boulevard Brecksville, OH 44141
Re: K142989
> Trade/Device Name: AMT Mini ONE® Enteral Extension Set Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIF Dated: October 30, 2014 Received: November 3, 2014
Dear Joshua Meinke,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### SECTION - 4
## c. - INDICATIONS FOR USE STATEMENT
510(k) Number: K142989
Trade Name: AMT Mini ONE® Enteral Extension Set Common Name: Enteral Extension Set
Indications for Use:
The Mini ONE® Enteral Extension Set is intended for use as an extension set for the Mini ONE® Gastrostomy Button, G-Jet Gastrojejunal Feeding Tube, Mic-Key® Secure-Lok® feeding devices, or other compatible enteral feeding devices. The extension set is for administering feeding, medications, and decompression to compatible devices.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## SECTION – 5
#### 510(K) Summary
AMT Mini ONE® Enteral Extension Set
| Date Prepared: | January 29, 2015 |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Joshua D. Meinke<br>QA/Regulatory Supervisor<br>Applied Medical Technology, Inc.<br>8006 Katherine Boulevard<br>Brecksville, OH 44141<br>Phone: 440-717-4252<br>Fax: 440-717-4200<br>Email: Josh.Meinke@appliedmedical.net<br>Contact Person: Joshua Meinke |
| Device<br>Information: | Trade Name:<br>AMT Mini ONE® Enteral Extension Set<br>Common Name:<br>Enteral Extension Set<br>Classification Name:<br>Gastroenterology and Urology (21 CFR 876.5980)<br>Regulatory Class:<br>II<br>Product Code:<br>PIF - Gastrointestinal Tubes with Enteral Specific Connectors |
| Predicate<br>Device: | AMT Low Profile Feeding Set Accessory (Applied Medical Technology, Inc.)<br>(cleared under K971757) |
| Intended Use: | The Mini ONE® Enteral Extension Set is intended for use with compatible enteral<br>feeding devices for administration of feeding, medications, and decompression. |
| Indications for<br>Use: | The Mini ONE® Enteral Extension Set is intended for use as an extension set for the<br>Mini ONE® Gastrostomy Button, G-Jet Gastrojejunal Feeding Tube, Mic-Key®<br>Secure-Lok® feeding devices, or other compatible enteral feeding devices. The<br>extension set is for administering feeding, medications, and decompression to<br>compatible devices. |
| Device<br>Description: | The AMT Mini ONE® Enteral Extension Set is intended for use with the AMT Mini<br>ONE® Gastrostomy Button, the AMT G-Jet Gastric-Jejunal Feeding Tube, Mic-<br>Key® Secure-Lok® feeding devices, or other compatible feeding devices. The<br>overall functionality and manufacturing methods remain similar to our original |
| Technological<br>Characteristics: | clearance of the Mini ONE Feeding Set in 510(K) K971757. The device materials and manufacturing methods were also recently cleared for use in our G-Jet Gastric-<br>Jejunal Feeding Tube 510(K) Submissions K110804 and K123716. The main difference between our original Mini ONE Feeding Set Clearance (K971757) and this submission is the inclusion of new end connectors that are designed to the ISO/IEC 80369 series.<br>The extension set consists of several different assembled components and materials that make up the device. Each extension set consists of several common components: A Mini ONE® connector that is able to connect to a compatible enteral feeding tube, flexible tubing, a pinch clamp to block the tubing, and ENfit enteral connector(s) to connect to a compatible feeding bag or syringe.<br>There are two types of Mini ONE® connectors that are used for the extension set. The first type is a straight through connector that is used for bolus or gravity feedings/administration of medication. The second type of connector is a right angle connector that better directs the tubing around the patient for longer feedings/administration of medication. Both connectors are identical in material and in design along the connecting portion that fits with a compatible enteral device. The Mini ONE® connectors are also compatible with Kimberly Clark Mic-Key® Gastrostomy and enteral devices. Mic-Key® compatibility was cleared through the 510(K) process in K971757, K110804, and K123716.<br>The flexible tubing used with each device comes in two different diameters and several different lengths. The two diameter options are provided to accommodate different types of feeds and medications to the patient. The tubing may be provided as straight tubing or in a coiled configuration. The different lengths are supplied to provide different options for patient orientation and maneuvering and include lengths of 2-24" lengths. Each tubing length also has a pinch clamp included to allow the user to close off the tubing while not in use.<br>The enteral connectors in this submission are provided in two different configurations with the extension sets. The first configuration is a single ENFit connector assembled directly onto the tubing. The single port connection is typically used for bolus or gravity use. The second configuration is a Y-Port connection with two ENFit connections, each with a separate plug. The two port connection allows the user to administer medication and feed through the device at the same time.<br>The Mini ONE® Enteral Extension Set is provided non-sterile for single user use only. It is made from DEHP and Latex free materials. The Mini ONE® Enteral Extension Set is provided in a number of configurations providing different Mini ONE® adapter designs, lengths of tubing, diameters of tubing, and number of ENFit connectors. Description of each difference in design configurations is listed below:<br>Feeding Set Configurations: Type of Mini ONE® Connector: Straight (bolus/gravity) or right angle (continuous/pump) Number of ENFit connectors: One or Two |
| | Tubing Diameter: 0.218" or 0.165" Tubing Length: 2-24" Part Number Scheme: 8-XXXX (where XXXX signifies tubing length, number of ENFit connectors, MiniONE® Connector Type, and tubing diameter) |
| Biocompatibility<br>Testing: | No additional Biocompatibility testing was performed for this submission. The device and materials in finished form for the Mini ONE® Enteral Extension Set were previously tested and cleared under 510(K) K123716 for permanent (greater than 30 days) skin contact. |
| Performance<br>Testing: | AMT conducted various performance tests on all components contained within the Mini ONE® Enteral Extension Set. A human factors study was also conducted to ensure that the new enteral connectors do not affect usability and intended use of the device. The study found that usability and intended use of our enteral extension set is not affected by the new ENFit connectors.<br><br>Bench tests have been carried out on samples of the AMT Mini ONE® Enteral Extension Sets with ENFit connectors. The tests carried out included: Tensile strength Fluid leakage Stress cracking Resistance to separation from axial load Resistance to separation from unscrewing Resistance to overriding Disconnection from unscrewing Flow testing Misconnection testing was conducted as recommended in the ISO 80369-1 standard. Connectors are made of rigid/semi rigid materials to reduce the risk of misconnection. Although the subject connector will not misconnect to other connectors in the 80369 series, misconnections with non ISO/IEC 80369 series connectors may still occur. |
| Conclusions: | The Mini ONE® Enteral Extension Set is substantially equivalent to the predicate device cleared under K971757 in intended use, patient population, design, biocompatibility and testing criteria, and method of operation. |
Applied Medical Technology, Inc. -510(k) Submission AMT Mini ONE® Enteral Extension Set
Section 5 — Page 5.1
{4}------------------------------------------------
Applied Medical Technology, Inc. -510(k) Submission AMT Mini ONE® Enteral Extension Set
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.