Bovie Ultimate High Frequency Electrosurgical Generator
K142975 · Bovie Medical Corporation · GEI · Dec 12, 2014 · General, Plastic Surgery
Device Facts
Record ID
K142975
Device Name
Bovie Ultimate High Frequency Electrosurgical Generator
Applicant
Bovie Medical Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 12, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bovie Ultimate™ High Frequency Electrosurgical Generator is indicated for delivery of RF energy and/or helium gas plasma to cut, coagulate, and ablate soft tissue during open and laparascopic sucrival procedures. The J-Plasma portion of the generator can be used only with dedicated Bovie J-Plasma handpieces
Device Story
Multipurpose electrosurgical generator (BVX-100H) delivers RF energy and/or helium gas plasma for soft tissue cutting, coagulation, and ablation. Used in hospital operating rooms by surgeons. Device features multiple cut, blend, coagulation, and bipolar modes; includes monopolar/bipolar outputs and helium gas outlet. Integrates return electrode contact quality monitoring (NEM) to prevent patient burns. User interface includes wattage/gas flow selection, LED displays, and alarm indicators. Device combines features of two predicate generators. Output waveforms (frequency/duration) determine clinical effect. Healthcare providers use handpieces or footswitches to activate energy delivery, enabling precise tissue management during surgery.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and IEC 60601-2-2 (high-frequency surgical equipment) was demonstrated to establish substantial equivalence.
Technological Characteristics
High-frequency electrosurgical generator; RF energy and helium gas plasma output. Features: 2 cut modes (up to 300W), 3 blend modes, 3 coagulation modes, 3 bipolar modes. Includes return electrode monitoring (NEM). Compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, RoHS (2011/65/EU). Connectivity: Monopolar/bipolar ports, helium gas outlet, footswitch ports. Software: Single component for user interface.
Indications for Use
Indicated for cutting, coagulating, and ablating soft tissue during open and laparoscopic surgical procedures using RF energy and/or helium gas plasma. J-Plasma function restricted to dedicated Bovie J-Plasma handpieces.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Bovie IDS-310 High Frequency Electrosurgical Generator (K134054)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, facing right. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 12, 2014
Bovie Medical Corporation Mr. Moshe Citronowicz Senior Vice-President 5115 Ulmerton Road Clearwater, Florida 33760
Re: K142975
Trade/Device Name: Bovie Ultimate " High Frequency Electrosurgical Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: October 2, 2014 Received: October 14, 2014
Dear Mr. Citronowicz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K142975
Device Name
Bovie Ultimate™ High Frequency Electrosurgical Generator
Indications for Use (Describe)
The Bovie Ultimate™ High Frequency Electrosurgical Generator is indicated for delivery of RF energy and/or helium gas plasma to cut, coagulate, and ablate soft tissue during open and laparascopic sucrival procedures. The J-Plasma portion of the generator can be used only with dedicated Bovie J-Plasma handpieces
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete the and review the collection of information. Send comments regarding this burden estimate or any other spect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K142975 Page 1 of 5
# 510(k) SUMMARY
# (Submitted As Required per 21 CFR 807.92)
# GENERAL INFORMATION:
| Submitter Name: | Bovie Medical Corporation |
|---------------------------------------|----------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3007593903 |
| Submitter Address: | 5115 Ulmerton Road<br>Clearwater, Florida 33760-4004<br>United States of America |
| Submitter Telephone Number: | (727) 803-8521 |
| Submitter FAX Number: | (727) 322-4465 |
| Contact Person: | Moshe Citronowicz<br>Senior Vice-President |
| Date Prepared: | October 2, 2014 |
# DEVICE IDENTIFICATION:
| Proprietary Name: | Bovie Ultimate™ High Frequency Electrosurgical<br>Generator |
|------------------------------------------|-------------------------------------------------------------------|
| Common Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| Model Number: | BVX-100H |
| Classification: | 21CFR 878.4400; Class II; Product Code GEI |
| Legally Marketed<br>Predicate Device(s): | 510(k) Number: K134054 |
| | Primary Predicate Device Name: |
| | Bovie IDS-310 High Frequency Electrosurgical<br>Generator |
| | Manufacturer: Bovie Medical Corporation |
| | 510(k) Number: K090586 |
| | Secondary Predicate Name: |
| | Bovie ICON GS Generator |
{4}------------------------------------------------
## K142975 Page 2 of 5
Manufacturer: Bovie Medical Corporation 510(k) SUMMARY
#### (Submitted As Required per 21 CFR 807.92)
#### Indications
The Bovie Ultimate™ High Frequency Electrosurgical Generator is used for delivery of RF energy and/or helium gas plasma to cut, coagulate, and ablate soft tissue during open and laparascopic surgical procedures. The J-Plasma portion of the generator can be used only with dedicated Bovie J-Plasma handpieces.
#### Device Description/Characteristics
The Bovie Ultimate™ High Frequency Electrosurgical Generator (BVX-100H) is a multipurpose electrosurgical device used for delivery of RF energy and/or helium gas plasma to cut, coagulate, and ablate soft tissue during open and laparascopic surgical procedures. The J-Plasma portion of the generator can be used only with dedicated Bovie J-Plasma handpieces. The generator combines the features of two cleared electrosurgical generators. The device is intended to be used in the Hospital Operating Room environment.
It features two (2) cut modes up to 300 watts, three (3) blend modes, three (3) coagulation modes, and three (3) bipolar modes. There is a characteristic electrical wave form associated with each mode. The electrical properties of the waveform (frequency and duration) produce the clinical effect (i.e. cut, coagulation). The shape and duration of waveforms are comparable between the generator and predicate devices. The generator offers one (1) monopolar handpiece output, one (1) monopolar foot controlled output, one (1) bipolar handpiece output, one (1) bipolar foot controlled output, and one J Plasma (helium) gas outlet. The generator has a return electrode contact and quality monitoring system (NEM) to reduce the risk of patient burns at the return electrode site. The padsensing feature allows the user to use either a split or solid return electrode. This feature is not required when using Helium gas mode for soft tissue coagulation. The Bovie Ultimate™ High Frequency Electrosurgical Generator contains a single software component that is solely used to support the user interface.
The generator is designed to comply with applicable Medical Electrical Equipment safety standards, including electromagnetic compatibility and other safety standards. It is also designed to meet the requirements of the European Restriction of Hazardous Substances Directive (2011/65/EU) directive.
The device consists of the generator, power cord, and a User's Guide. The main device components are a front panel containing the power button, wattage selection buttons, gas flow (rate) selection buttons, pulsing, LED numeric displays, alarm and return electrode indicator lights, and connector ports for accessories. The back panel consists of Helium
{5}------------------------------------------------
## K142975 Page 3 of 5
gas access ports, footswitch ports, volume controls, a power cord outlet, and a fuse. Internal components include printed circuit boards, speakers, and cabling.
Bench testing was performed to demonstrate that the generator met design specifications and to establish substantial equivalence with predicate devices. Device testing and evaluation demonstrated compliance with the following:
IEC 60601-1: 2005 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance
## 510(k) SUMMARY
## (Submitted As Required per 21 CFR 807.92)
IEC 60601-1-2: 2007
Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
IEC 60601-2-2: 2007
Medical Electrical Equipment - Part 2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
#### Substantial Equivalence:
| Characteristic | Bovie Ultimate™<br>High Frequency<br>Electrosurgical<br>Generator<br>(This Submission) | Bovie IDS-310<br>High Frequency<br>Electrosurgical<br>Generator<br>(K134054)<br>Primary Predicate | Bovie ICON GS<br>Generator<br>(K090586)<br>Secondary<br>Predicate |
|--------------------------------|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Regulation and<br>Product Code | 21 CFR 878.4400 /<br>GEI | 21 CFR 878.4400 /<br>GEI | 21 CFR 878.4400 /<br>GEI |
{6}------------------------------------------------
| | | | K142975 Page 4 of 5 |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Intended Use | Is used for delivery of RF energy and/or helium gas plasma to cut, coagulate, and ablate soft tissue during open and laparascopic surgical procedures. The J-Plasma portion of the generator can be used only with dedicated Bovie J-Plasma handpieces. | Is used to deliver RF energy via an assortment of surgical devices to cut and coagulate different kinds of tissue. | Utilizes helium gas for the coagulation of soft tissues during open soft tissue surgery. |
| Energy Type | High Frequency (RF) | High Frequency (RF) | High Frequency (RF) |
| Output | Monopolar and Bipolar | Monopolar and Bipolar | Monopolar and Bipolar |
{7}------------------------------------------------
# 510(k) SUMMARY
# (Submitted As Required per 21 CFR 807.92)
# Substantial Equivalence:
| Characteristic | Bovie Ultimate™<br>High Frequency<br>Electrosurgical<br>Generator<br>(BVX-100H)<br>(This Submission) | Bovie IDS-310<br>High Frequency<br>Electrosurgical<br>Generator<br>(K134054)<br>Primary Predicate | Bovie ICON GS<br>Generator<br>(GS100)<br>(K090586)<br>Secondary<br>Predicate |
|---------------------------------|------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Cut Modes | Has Two Cut Modes | Has Two Cut Modes | Not Applicable to<br>this Device |
| Blend Modes | Has Three Blend<br>Modes | Has Four Blend<br>Modes | Not Applicable to<br>this Device |
| Monopolar<br>Connectors | One | Dual | Not Applicable to<br>this Device |
| Plasma Connector | One | Not Applicable to<br>this Device | One |
| Auto Bipolar | Does Not Have<br>Auto Bipolar | Has Auto Bipolar | Not Applicable to<br>this Device |
| Bovie Bipolar | Does Not Have<br>Bovie Bipolar | Has Bovie Bipolar | Not Applicable to<br>this Device |
| Output Gas Flow | 1.0 - 5.0<br>Liters/minute | Not Applicable to<br>this Device | 1.0 - 5.0<br>Liters/minute |
| Bipolar Handpiece<br>Activation | Supported | Supported | Not Applicable to<br>this Device |
| Gas In Use Indicator | Present | Not Applicable to<br>this Device | Not Present |
## Conclusions
The Bovie Ultimate™ High Frequency Electrosurgical Generator has the same intended use, operating procedures, principles of operation, and utilizes the same technology as the predicate devices. The Bovie Ultimate™ High Frequency Electrosurgical Generator does not raise additional issues of safety or efficacy compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.