K142968 · Health & Life Co., Ltd. · DXN · Dec 19, 2014 · Cardiovascular
Device Facts
Record ID
K142968
Device Name
Full Automatic (NIBP) Blood Pressure Monitor
Applicant
Health & Life Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Dec 19, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
HL858CB automatically measuers human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use. HL858CB detects the appearance of irregular heartbeats during measurement, an indicated symbol will appear with measuring reading. And the device features a built-in "Bluetooth Data Transmission'' function, which enables the device automatically transmit measuring results to paired Bluetooth device.
Device Story
Oscillometric blood pressure monitor; measures systolic/diastolic pressure and heart rate at upper arm. Features: irregular heartbeat detection (rhythm >25% deviation from average); Risk Category Indicator; Triple Check mode (three consecutive measurements at 1-minute intervals); Rest Assure function (5-minute countdown timer); Bluetooth data transmission to paired devices. Used in home environment by adult patients. Output displayed on LCD panel; Bluetooth transmission allows logging/tracking. Benefits: enables self-monitoring of blood pressure levels and irregular heartbeats, facilitating user tracking and management.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2009. Study included 90 subjects with qualified distribution. Results confirmed mean difference and standard deviation met required limits for automated non-invasive sphygmomanometers.
Technological Characteristics
Oscillometric measurement; ABS housing/keys. Power: 6V 1A (4x AA batteries or AC adapter). Connectivity: Bluetooth. Standards: IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30, ISO 10993 (biocompatibility), IEC 60601-1-2 (EMC). Software lifecycle per IEC 62304. Usability per IEC 62366.
Indications for Use
Indicated for adults aged 18+ with arm circumference 9-17 inches (23-43 cm) for home-use blood pressure and heart rate monitoring via oscillometric method.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Full Automatic (NIBP) Blood Pressure Monitor, Model: HL858CA (K131121)
Full Automatic (NIBP) Blood Pressure Monitor, Model: HL888HD (K113238)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
August 10, 2021
Health & Life Co.,Ltd. Sarah Su Director 9F, No. 186, Jian Yi Road Zhonghe District , New Taipei City, Taiwan 23553 Taiwan
Re: K142968
Trade/Device Name: Full Automatic (NIBP) Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN
Dear Sarah Su:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 19, 2014. Specifically, FDA is updating this SE Letter as an administrative correction for an incorrect 510(k) Summary.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact LCDR Stephen Browning, OHT2: Office of Cardiovascular Devices, (240) 402-5241, stephen.browning@fda.hhs.gov.
Sincerely,
Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of a human figure. The figure is composed of three overlapping profiles, suggesting a sense of community and collaboration.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2014
Health & Life Co., Ltd. % Ms. Sarah Su Director Regulatory Affairs 9F, No. 186, Jian Yi Road Zhonghe District , New Taipei City, 23553 TW
> K142968 Trade/Device Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CB Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: October 15, 2014 Received: October 20, 2014
Dear Ms. Sarah Su,
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
Bram D. Zuckerman, M.D. For Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K142968
Page __ 1__ of__1__
510(k) Number (if known): K142968
Device Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CB
### Indications For Use:
HL858CB automatically measuers human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
HL858CB detects the appearance of irregular heartbeats during measurement, an indicated symbol will appear with measuring reading. And the device features a built-in "Bluetooth Data Transmission'' function, which enables the device automatically transmit measuring results to paired Bluetooth device.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Over-The-Counter Use
{4}------------------------------------------------
K142968 Page 1 of 7
# PREMARKET NOTIFICATION
# 510(k) SUMMARY
(As Required By 21 CFR 807.92)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K142968 Date:
### 1. Submitter:
Health & Life Co., Ltd. 9F. No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C TEL: +886-2-8227-1300 FAX: +886-2-8227-1301
Contact person: Sarah Su/ Regulatory Affairs Dept. E-mail: sarah.su@hlmt.com.tw Tel: 886-2-8227-1300 ext.1201 Fax: 886-2-8227-1301
#### 2. Name of the Device:
Trade Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858CB Common Name: Blood Pressure Monitor Classification Name: Non-invasive Blood Pressure Measurement System Classification: Class II, 21 CFR 870.1130 Classification Panel: 74 Cardiovascular Product Code: DXN
## 3. Information for the 510(k) Cleared Device (Predicate Device):
{5}------------------------------------------------
A. Full Automatic (NIBP) Blood Pressure Monitor, Model: HL858CA (K131121) B. Full Automatic (NIBP) Blood Pressure Monitor, Model: HL888HD (K113238)
#### 4. Device Description:
HL858CB automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
The device will display a symbol wor or ma, to indicate the detection of irregular heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.
Besides, When Triple Check mode is turned on by user, the symbol ( & TIME ) will display on the LCD. Then press Start/Stop button the device will take three consecutive measurements automatically at 1 minute intervals. After measurements are completed, LCD will display the average values of the three measurements. And this device is designed with Rest Assure function as a countdown timer to help user in relax state for 5 minutes before taking measurement. Furthermore, the user can use the Bluetooth Data Transmission function, which provides users an optional choice to log, track and store their measurement data.
## 5. Intended Use
HL858CB automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
{6}------------------------------------------------
HL858CB detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the device features a built-in "Bluetooth Data Transmission" function, which enables the device automatically transmit measuring results to paired Bluetooth device.
# 6. Comparison of device to predicate device: Product Specification Comparison Table of Subject Device HL858CB, and Predicate Device HL858CA(K131121)
| Item | Predicate Device<br>HL858CA (K131121) | Subject Device<br>HL858CB |
|-------------------------------------------|-----------------------------------------------------------------------------|---------------------------|
| Method of<br>measurement | Oscillimetric | Same as left |
| Range of<br>measurement | Pressure 0- 300mmHg,<br>Pulse 40-199 Beats/minute | Same as left |
| Accuracy | Pressure ± 3mmHg<br>Pulse ± 5% | Same as left |
| Inflation | Automatic inflation<br>(Air pump) | Same as left |
| Deflation | Automatic air release<br>control valve | Same as left |
| Exhaust | Automatic exhaust valve | Same as left |
| Display | Liquid Crystal Digital | Same as left |
| Power Supply | 6V 1A, 4 × AA/1.5V (LR6) Alkaline<br>batteries, or<br>AC adapter (optional) | Same as left |
| Storage/<br>Transportation<br>Environment | - 25°C<br>~ + 70℃ (- 13°F~ +158°F),<br>≤ 93% R.H. | Same as left |
| Operating | 5℃~40℃ (41°F~104°F), | Same as left |
{7}------------------------------------------------
| Environment | 15% ~ 93% R.H. | |
|------------------------------------|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Material | ABS housing and<br>ABS keys | Same as left |
| Sets of memory | 2*120, total 240 | Same as left |
| Number of<br>Push Button | 7 +<br>2 switch control<br>( Triple check, Rest assure) | Same as left |
| Storage pouch | Yes | Same as left |
| Cuff size | Arm circumference approx.<br>23~33 cm /9~13 inches(Normal cuff)<br>33~43cm/13~17 inches(Large cuff) | Arm circumference<br>approx.<br>23~43cm / 9~17 inch<br>(Universal cuff) |
| Unit Weight | Approx. 393 ± 10g<br>(Excluding cuff and Batteries ) | Same as left |
| Risk Category<br>Indicator | Yes | Same as left |
| Irregular<br>Heartbeat<br>Detector | Yes | Same as left |
| Data Link<br>function | Yes<br>(Via USB cable) | Yes<br>(Via Bluetooth) |
#### Changes from the predicate devices HL858CA (K131121):
- * Changing the cuff size from Normal cuff (9~13 inches/23-33cm) and Large cuff (13~17 inches/33-43cm) to Universal cuff (9~17 inches/23~43cm).
- * Modifying the Data Link Function from USB into Bluetooth Data Transmission Function.
These additional feature has been verified and validated and do not affect the safety and effectiveness of subject device HL858CB.
For the product feature of Bluetooth data transmission, was compared with the other our own predicate device HL888HD (K113238).Please refer to Section 12. Substantial Equivalence Discussion for detail information.
{8}------------------------------------------------
K142968 Page 5 of 7
#### 7. Discussion of Clinical Tests Performed:
HL858CB is compliant to the standard of ISO 81060-2: First Edition 2009-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device HL858CB in the group of 90 subjects with qualified distribution. Thus, all the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.
# 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The subject device was tested to evaluate its safety and effectiveness, including the followings:
- a. EMC Test: IEC 60601-1-2 Edition 3:2007-03, Medical Electrical Equipment -Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Tests
#### Radio Frequency Wireless Test: b.
-ETSI EN 300 328, Electromagnetic compatibility and Radio spectrum Matters (ERM); Wideband transmission systems; Data transmission equipment operating in the 2.4 GHz ISM band and using wide band modulation techniques; Harmonized EN covering the essential requirements of article 3.2 of the R&TTE Directive
-ETSI EN 301 489-1, Electromagnetic compatibility and Radio spectrum Matters (ERM); Electromagnetic Compatibility (EMC) standard for radio equipment and services; Part 1: Common technical requirements
-ETSI EN 301 489-17, Electromagnetic compatibility and Radio spectrum Matters (ERM); Electro Magnetic Compatibility (EMC) standard for radio equipment; Part 17: Specific conditions for Broadband Data Transmission Svstems
{9}------------------------------------------------
## c. Safety Test:
-IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
-IEC 60601-1-11:2010, Medical electrical equipment-Part 1-11:General Requirement for basic safety and essential performance- Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment
## d. FCC Test:
FCC 47 CFR Part 15, Subpart B & FCC 47 CFR Part 15, Subpart C
## e. Biocompatibility Test:
-ISO 10993-1:2009, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process
-ISO 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity
-ISO 10993-10:2010, Third Edition Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization
#### f. Reliability Test:
IEC 80601-2-30 Edition1.1 2013-07 Medical electricalequipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers.
- g. Risk Assessment: ISO 14971:2007 Second Edition, Medical devices Application of risk management to medical devices
## h. Software Verification and Validation:
-IEC 62304 Ed.1.0 (2006), Medical device software - Software life cycle processes, -IEC 60601-1-4 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1
# i. Usability Validation:
-IEC 62366:2007 Medical devices - Application of usability engineering to medical devices
-IEC 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
{10}------------------------------------------------
# K142968 Page 7 of 7
# 9. Conclusions:
The subject device was tested and fulfilled the requirements of those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.