K142932 · Deltex Medical Limited · DPT · Jan 22, 2015 · Cardiovascular
Device Facts
Record ID
K142932
Device Name
Deltex Medical KDP72 Doppler Probe
Applicant
Deltex Medical Limited
Product Code
DPT · Cardiovascular
Decision Date
Jan 22, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The probe is for use with the Deltex Medical CardioQ-EDM and CardioQ-EDM+ for Monitoring of cardiac output and fluid status. The probe is only approved for oral placement into the esophagus of a single anesthetized patient 15 years of age or younger, 50cm (20") to 170cm (67") in height.
Device Story
KDP72 Doppler Probe is an oral extravascular blood flow probe used with CardioQ-EDM/CardioQ-EDM+ systems. Device consists of a spring-reinforced silicone shaft with an ultrasonic transmitting/receiving tip. Used in clinical settings (e.g., OR) by clinicians to monitor cardiac output and fluid status in anesthetized pediatric patients. The probe transmits ultrasonic signals to measure blood flow; the monitor processes these signals to provide hemodynamic data. Output assists clinicians in fluid management and hemodynamic optimization, potentially improving patient outcomes during anesthesia. The KDP72 features a shorter length, increased depth markings, and an open-coil spring design compared to previous models to enhance flexibility for pediatric anatomy.
Clinical Evidence
Bench testing only. Performance data provided in accordance with FDA guidance 'Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers'. Comparative flexibility testing conducted against predicate devices.
Technological Characteristics
Extravascular blood flow probe; spring-reinforced silicone shaft; ultrasonic transducer tip. Biocompatibility per ISO 10993-5, -7, -10. Electrical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37. Sterilization via ISO 11135-1. Packaging per ISO 11607-1/2. Single-patient use.
Indications for Use
Indicated for monitoring cardiac output and fluid status in anesthetized pediatric patients 15 years of age or younger, with heights between 50cm and 170cm, via oral esophageal placement.
Regulatory Classification
Identification
An extravascular blood flow probe is an extravascular ultrasonic or electromagnetic probe used in conjunction with a blood flowmeter to measure blood flow in a chamber or vessel.
Predicate Devices
Deltex Medical Ltd I2n Series Doppler Probe (K073593)
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES - USA
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 22, 2015
Deltex Medical Limited % Neil Armstrong RA Advisor To Deltex Medical Meddiquest Limited Quest Science, Orton Malborne Peterborough, PE2 5XS GB
Re: K142932
> Trade/Device Name: Deltex Medical KDP72 Doppler Probe Regulation Number: 21 CFR 870.2120 Regulation Name: Extravascular Blood Flow Probe Regulatory Class: Class II Product Code: DPT Dated: October 10, 2014 Received: October 14, 2014
Dear Neil Armstrong,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Melissa A. Torres -S
Bram D. Zuckerman, M.D. For Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) Unknown - not yet assigned by FDA
Device Name KDP72, 72 Hour Doppler Probe
Indications for Use (Describe)
The probe is for use with the Deltex Medical CardioQ-EDM and CardioQ-EDM+ for Monitoring of cardiac output and fluid status. The probe is only approved for oral placement into the esophagus of a single anesthetized patient 15 years of age or younger, 50cm (20") to 170cm (67") in height.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
#### Contains Nonbinding Recommendations Appendix G
## Appendix G: Example Diagnostic Ultrasound Indications For Use Format
CardioQ-EDM and CardioQ-EDM+ cardiac output and fluid status monitor System: KDP72 , 72 Hour Doppler Probe Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows;
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | N | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "Deltex medical". The word "Deltex" is in a large, bold, blue font. Below "Deltex" is the word "medical" in a smaller, light gray font. The logo is simple and clean, with a focus on the company name.
Deltex Medical Limited Terminus Road Chichester United Kingdom, PO19 8TX 011 44 1243 774837 Tel: Fax: 011 44 1243 532534
## 510(k) Summary
as required by 21 CFR 807.92(c)
#### Submitter and Owner:
This Premarket Notification, 510(k), is submitted by: Graham Lowe Marketing and Operations Director For and on behalf of the Owner: Deltex Medical Ltd Terminus Road Chichester West Sussex PO19 8TX United Kingdom Tel: 011 44 1243 523174 Fax: 011 44 1243 532534
#### Date
This Summary was prepared on September 22, 2014
## Classification name:
Extravascular blood flow probe
The FDA has classified: "Extravascular blood flow probe" in 21 CFR 870.2120 as a Class II medical device with Product Code DPT
## Common/Usual Name:
probe, blood-flow, extravascular
{5}------------------------------------------------
K142932 page 2 of 4
#### Proprietary Name:
Deltex Medical Ltd KDP72 Doppler Probe
#### Establishment Reqistration Number:
- The device will be manufactured by: Deltex Medical Ltd Terminus Road Chichester West Sussex PO19 8TX United Kingdom Establishment Registration Number 9680933
and sterilized by: Sterigenics UK, Ltd Cotes Park Estate Somercotes Derbyshire DE55 4NJ
## Establishment Registration Number 3002807091
United Kingdom
### Substantial Equivalence:
The Deltex Medical Ltd KDP72 Doppler Probe is substantially equivalent in design, use and materials to the:
Deltex Medical Ltd I2n Series Doppler Probe - K073593 Deltex Medical Ltd DP240 Doppler Probe - K052989 and the Sometic Dynemo 3000 - K972798, now Arrow Hemosonic probe
Like the Inn Series Probe is for use with the Deltex Medical CardioQ-EDM and CardioQ-EDM+ systems, K111542 and K132139 respectively.
The Deltex Medical Ltd KDP72 Pediatric Doppler Probe is designed to be a shorter version of the current legally marketed Dp240 product with more depth markings. It has an open-coil spring, as used on the 12 series products instead of the closed coil used on the Dp240, to increase flexibility.
The Sometic Dynemo 3000 K972798 has been superseded by the Arrow Hemosonic probe (still K972798) which is indicated for pediatric use for patients above 15kg. The Deltex Medical Ltd KDP72 will be shorter and more flexible enabling it to be indicated for patients above 3kg.
{6}------------------------------------------------
K142932 page 3 of 4
#### Description:
The Deltex Medical Ltd KDP72 Pediatric Doppler Probe is an oral extravascular blood flow probe designed to work with the CardioQ-EDM and CardioQ-EDM+ Systems (K111542 and K132139 respectively). It consists of a shaft, which is a spring reinforced silicone tube, with an electrical connector on the machine end and an ultrasonic transmitting and receiving tip on the patient end. The tip is fully covered and sealed to the shaft with a silicone rubber boot and by wires running through the shaft to the connector. Visual product identification is provided at the machine end and the device is provided single packed sterile for single patient use.
Standard ultrasonic diagnostic extravascular blood flow probes have been successfully used with pediatric patients; however the KDP72 has been specially designed for pediatric patients with a shorter length, more depth markings and an open coil spring to increase the flexibility.
#### Intended Use:
The probe is for use with Deltex Medical CardioQ-EDM and CardioQ-EDM+ for monitoring of cardiac output and fluid status. The probe is only approved for oral placement into the esophagus of a single anesthetized patient 15 years of age or younger.
#### Performance Data
The performance data recommended in "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers", issued on September 9, 2008, has been included.
Additionally a flexibility test has been conducted on the subject and predicate devices which demonstrates the comparative flexibility.
#### Conclusion
Based on the information above, Deltex Medical Ltd concludes that the Deltex Medical Ltd KDP72 Doppler Probe in substantially equivalent to:
- the Deltex Medical Ltd I2 Series and Dp240 Doppler Probes (K073593 and . K052989), and
- the Arrow International, Inc, Hemosonic 100 Esophageal Probe (K972798 . Sometic Dynemo 3000).
{7}------------------------------------------------
| Feature | Deltex Medical<br>'KDP72' | Deltex Medical<br>'I₂n-series' | Deltex Medical<br>'DP240' | Arrow<br>'Hemosonic 100 Esophageal Probe' |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| | Subject Device | Predicate Device<br>#1 | Predicate Device<br>#2 | Predicate Device<br>#3 |
| FDA 510k number | This Submission | K073593 | K052989 | K972798 |
| product name | KDP72 | I₂n-series | DP240 Probe | Transesophageal<br>Probe |
| common name | 72 Hour Pediatric<br>Doppler probe | I₂n-series Doppler<br>probe | 240 Hour Doppler<br>Probe | Trans-esophageal<br>Probe & Sterile<br>Jacket |
| product number | 9081-7002 | 9090 - 7015<br>9090 - 7016<br>9090 - 7017 | 9070- 7006 | HSP - 02150 &<br>HSS-02150 |
| Device Description | | | | |
| Sterilization | Sterilized in<br>accordance with<br>ISO11135-1<br>(version of<br>standard current<br>at time of<br>submission) | Sterilized in<br>accordance with<br>ISO11135-1<br>(version of<br>standard current<br>at time of<br>submission) | Sterilized in<br>accordance with<br>ISO11135-1<br>(version of<br>standard current<br>at time of<br>submission) | Unknown |
| Shelf Life | ISO11607-1<br>(version of<br>standard current<br>at time of<br>submission) | ISO11607-1<br>(version of<br>standard current<br>at time of<br>submission) | ISO11607-1<br>(version of<br>standard current<br>at time of<br>submission) | Unknown |
| Biocompatibility | Tested in<br>accordance with<br>ISO 10993-5, ISO<br>10993-10, ISO<br>10993-7 (version<br>of standard<br>current at time of<br>submission) | Tested in<br>accordance with<br>ISO 10993-5, ISO<br>10993-10, ISO<br>10993-7 (version<br>of standard<br>current at time of<br>submission) | Tested in<br>accordance with<br>ISO 10993-5, ISO<br>10993-10, ISO<br>10993-7 (version<br>of standard<br>current at time of<br>submission) | Unknown |
| EMC and<br>Electrical Safety | IEC60601-1,<br>IEC60601-1-2,<br>IEC606001-2-37<br>(version of<br>standard current<br>at time of<br>submission) | IEC60601-1,<br>IEC60601-1-2,<br>IEC606001-2-37<br>(version of<br>standard current<br>at time of<br>submission) | IEC60601-1,<br>IEC60601-1-2,<br>IEC606001-2-37<br>(version of<br>standard current<br>at time of<br>submission) | Unknown |
| Packaging | ISO 11607-1, ISO<br>11607-2 (version<br>of standard<br>current at time of<br>submission) | ISO 11607-1, ISO<br>11607-2 (version<br>of standard<br>current at time of<br>submission) | ISO 11607-1, ISO<br>11607-2 (version<br>of standard<br>current at time of<br>submission) | Unknown |
a
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.