← Product Code [FBN](/productcode/FBN) · K142922

# SpyGlass DS Direct Visualization System (K142922)

_Boston Scientific Corporation · FBN · Jan 16, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142922

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [FBN](/productcode/FBN.md)
- **Decision Date:** Jan 16, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The SpyGlass DS Direct Visualization System is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts. The SpyGlass DS Direct Visualization System comprises two components: the SpyScope DS Access and Delivery Catheter and the SpyGlass DS Digital Controller. The SpyScope DS Access and Delivery Catheter is intended to provide direct visualization and to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts. The SpyGlass DS Digital Controller is intended to provide illumination and receive, process, and output images from the SpyScope DS Access and Delivery Catheter for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

## Device Story

System consists of single-use SpyScope DS Access and Delivery Catheter and SpyGlass DS Digital Controller. Catheter, introduced via duodenoscope, features articulation control, working channel for accessories/aspiration, irrigation channels, and distal imaging sensor. Controller provides LED illumination and processes video signals from catheter for display on external monitor. Used by physicians in clinical settings for direct visualization of pancreatico-biliary anatomy. Output allows real-time guidance of diagnostic/therapeutic instruments, aiding clinical decision-making and potentially improving procedural outcomes.

## Clinical Evidence

Bench testing only. Evaluated diameter, working length, channel diameter, field of view, resolution, articulation reliability, leakage, compatibility (duodenoscope/pump/accessory), image noise, video latency, frame rate, and illumination intensity. Biocompatibility tested per ISO 10993-1 (cytotoxicity, irritation, sensitization, USP <661>, latex). Electrical safety and EMC evaluated per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18.

## Technological Characteristics

System includes flexible video cholangiopancreatoscope (single-use) and digital controller. Features CMOS image sensor, LED illumination, and working/irrigation channels. Connectivity via DVI, VGA, or S-Video. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601 series.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- ACMI DUR-Digital Ureteroscope and Choledochoscope System ([K060269](/device/K060269.md))
- SpyScope Access and Delivery Catheter ([K090170](/device/K090170.md))
- SpyGlass Direct Visualization Probe/Ocular ([K052194](/device/K052194.md))
- Olympus XCHF-B180Y1 Choledochoscope ([K080586](/device/K080586.md))
- Stryker LED Light Source ([K082813](/device/K082813.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol of a human figure, represented by three overlapping profiles facing to the right, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 16, 2015

Boston Scientific Corporation Carter Navarro Manager, Regulatory Affairs 100 Boston Scientific Way Mail Stop M-11 Marlborough, MA 01752

Re: K142922

> Trade/Device Name: SpyGlass DS Direct Visualization System Regulation Number: 21 CFR 876.1500 Regulation Name: Gastroenterology/Urology Regulatory Class: Class II Product Code: FBN, NTN, KQM Dated: December 19, 2014 Received: December 22, 2014

Dear Carter Navarro,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -A

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K142922

Device Name SpyGlass DS Direct Visualization System

Indications for Use (Describe)

The SpyGlass DS Direct Visualization System is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

The SpyGlass DS Direct Visualization System comprises two components: the SpyScope DS Access and Delivery Catheter and the SpyGlass DS Digital Controller.

The SpyScope DS Access and Delivery Catheter is intended to provide direct visualization and to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

The SpyGlass DS Digital Controller is intended to provide illumination and receive, process, and output images from the SpyScope DS Access and Delivery Catheter for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

| Type of Use (Select one or both, as applicable)                                       |                                                                                      |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Boston Scientific. The words "Boston Scientific" are written in a serif font. The color of the text is blue. 

Boston Scientific Corporation 300 Boston Scientific Way Marlborough, MA 01752 (508) 382-4000 www.bostonscientific.com

### 510(k) Summary for SpyGlass™ DS Direct Visualization System

1. Submitter Boston Scientific Corporation Endoscopy Division 100 Boston Scientific Way Marlborough, MA 01752

Contact: Carter Navarro Manager, Regulatory Affairs Phone: (508) 683-4793 Fax: (508) 683-5939 E-mail: carter.navarro@bsci.com

Date Prepared: October 7, 2014

#### 2. Device

| Trade Name:                | SpyGlass DS Direct Visualization System           |
|----------------------------|---------------------------------------------------|
| Common Name:               | Choledochoscope and accessories, flexible/rigid;  |
|                            | Surgical camera and accessories; LED light source |
| Product Code:              | FBN, KQM, NTN                                     |
| Device Class and Panel:    | Class II, Gastroenterology/Urology (FBN, NTN)     |
|                            | Class I, General and plastic surgery (KQM)        |
| Classification Regulation: | 21 CFR 876.1500 (FBN, NTN)                        |
|                            | 21 CFR 878.4160 (KQM)                             |

## 3. Predicate Devices

Clearance Number:

| Trade Name:                | ACMI DUR-Digital Ureteroscope and<br>Choledochoscope System |
|----------------------------|-------------------------------------------------------------|
| Manufacturer:              | ACMI Corporation                                            |
| Clearance Number:          | K060269                                                     |
| Common Name:               | Choledochoscope and accessories, flexible/rigid             |
| Product Code:              | FBN                                                         |
| Device Class and Panel:    | Class II, Gastroenterology/Urology                          |
| Classification Regulation: | 21 CFR 876.1500                                             |
| Trade Name:                | SpyScope Access and Delivery Catheter                       |
| Manufacturer:              | Boston Scientific Corporation                               |

K090170

SpyGlass DS Direct Visualization System Premarket Notification

{4}------------------------------------------------

| Common Name:<br>Product Code: | Mini endoscope, gastroenterology-urology<br>ODF |
|-------------------------------|-------------------------------------------------|
| Device Class and Panel:       | Class II, Gastroenterology/Urology              |
| Classification Regulation:    | 21 CFR 876.1500                                 |
| Trade Name:                   | SpyGlass Direct Visualization Probe/Ocular      |
| Manufacturer:                 | Boston Scientific Corporation                   |
| Clearance Number:             | K052194                                         |
| Common Name:                  | Endoscope and/or accessories                    |
| Product Code:                 | KOG                                             |
| Device Class and Panel:       | Class II, Gastroenterology/Urology              |
| Classification Regulation:    | 21 CFR 876.1500                                 |
| Trade Name:                   | Olympus XCHF-B180Y1 Choledochoscope             |
| Manufacturer:                 | Olympus Medical Systems Corporation             |
| Clearance Number:             | K080586                                         |
| Common Name:                  | Choledochoscope and accessories, flexible/rigid |
| Product Code:                 | FBN                                             |
| Device Class and Panel:       | Class II, Gastroenterology/Urology              |
| Classification Regulation:    | 21 CFR 876.1500                                 |
| Trade Name:                   | Stryker  LED Light Source                       |
| Manufacturer:                 | Stryker Endoscopy                               |
| Clearance Number:             | K082813                                         |
| Common Name:                  | Light source, fiberoptic, routine               |
| Product Code:                 | FCW                                             |
| Device Class and Panel:       | Class II, Gastroenterology/Urology              |
| Classification Regulation:    | 21 CFR 876.1500                                 |
|                               |                                                 |

#### 4. Device Description

The SpyGlass DS Direct Visualization System comprises two components: (1) the sterile, single-use SpyScope DS Access and Delivery Catheter ("Catheter") and (2) the SpyGlass DS Digital Controller (the "Controller").

The Catheter is a flexible video cholangiopancreatoscope, introduced into the pancreatico-biliary system via a duodenoscope. The Catheter comprises a handle, an insertion tube, and a connection cable. The handle includes two articulation control knobs, a lever to lock the control knobs in place, connectors for irrigation and aspiration, a working channel port, and a strap to attach the Catheter to a duodenoscope. The insertion tube contains one working channel for accessory devices and aspiration, two channels for irrigation, two optical fibers to transmit illumination from the Controller, and wiring to transmit video signals to the Controller. The bending section at the distal portion of the insertion tube is controlled by the user via the articulation control knobs on the handle. The distal end of the insertion tube contains an imaging sensor for capturing

{5}------------------------------------------------

video and transmitting it to the Controller, elements for transmitting illumination from the Controller, and the distal openings of the irrigation and working channels. The connection cable connects the Catheter handle to the Controller for transmitting illumination and video signals.

The Controller is an endoscopic video imaging system that combines the functionality of a camera and an LED light source. The Controller receives video signals from the Catheter, processes the video signals, and outputs video images to an attached monitor. The Controller also generates and controls the illumination transmitted to the distal end of the Catheter. The user interface of the Controller comprises a power button, a receptacle to connect the Catheter connection cable, buttons to turn illumination on or off and to control the illumination intensity, and an illumination intensity indicator. The Controller outputs video images to an attached monitor via DVI, VGA, or S-Video ports, and the user may select NTSC or PAL video formats according to the geographic region of use.

### 5. Indication for Use

The SpyGlass DS Direct Visualization System is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

The SpyGlass DS Direct Visualization System comprises two components: the SpyScope DS Access and Delivery Catheter and the SpyGlass DS Digital Controller.

The SpyScope DS Access and Delivery Catheter is intended to provide direct visualization and to guide both optical and accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

The SpyGlass DS Digital Controller is intended to provide illumination and receive, process, and output images from the SpyScope DS Access and Delivery Catheter for diagnostic and therapeutic applications during endoscopic procedures in the pancreaticobiliary system including the hepatic ducts.

#### 6. Technological Characteristics

The proposed SpyGlass DS Direct Visualization System shares similar design features and functions with the predicate devices.

The proposed device and predicate ACMI DUR-Digital Ureteroscope and Choledochoscope System share similar indications for use, both being intended for use for examination of the bile duct, and facilitating diagnostic and therapeutic procedures in the bile duct. The proposed device and predicate ACMI DUR-Digital Ureteroscope and

{6}------------------------------------------------

Choledochoscope System share similar technological characteristics, including CMOS image sensors for visualization, light-emitting diodes (LEDs) for illumination (the light source is located in the handheld component of the predicate device), and similar video output capabilities. The differences in respect to the dimensions of the Catheter and location of the LEDs do not alter the suitability of the proposed device for its intended use.

In overall appearance, dimensions, and mechanical performance, the proposed SpyScope DS Access and Delivery Catheter is very similar to the SpyScope Access and Delivery Catheter, while adding the optical functionality of the SpyGlass Probe/Ocular. The minor dimensional differences between the proposed SpyScope DS Access and Deliverv Catheter and the predicate SpyScope Access and Delivery Catheter do not raise new questions of safety or effectiveness for the proposed device's intended use in the pancreatico-biliary system. The applicable dimensions are within the range of the predicate Olympus XCHF-B180Y1 Choledochoscope, which has the same indications for use.

The proposed SpyGlass DS Digital Controller and the predicate Stryker LED Light Source share similar technological characteristics, both utilizing LEDs for illumination, and both offering brightness control.

### 7. Substantial Equivalence

A direct comparison of key characteristics demonstrates that the SpyGlass DS Direct Visualization System is substantially equivalent to the predicate devices in terms of intended use, technological characteristics, and performance characteristics. The SpyGlass DS Direct Visualization System is as safe and effective as the predicate devices.

## 8. Performance Data

Non-clinical testing was successfully performed on the proposed SpyGlass DS Direct Visualization System.

Performance testing (bench) was successfully completed to establish substantial equivalence between the proposed SpyGlass DS Direct Visualization System and the predicate devices. This testing included the following:

- Diameter ●
- . Working length
- Working channel diameter ●
- Field of view ●
- Direction of view ●
- Resolution
- Articulation direction ●
- Articulation angle ●
- . Surface and edges
- Irrigation and working channel system leakage
- Articulation reliability
- Attachment to duodenoscope reliability ●
- Duodenoscope compatibility ●

{7}------------------------------------------------

- Irrigation pump compatibility ●
- Accessory compatibility .
- Image noise ●
- Video latency ●
- System frame rate ●
- Illumination intensity ●
- Default automatic light control setpoint ●
- Automatic light control response time ●
- Simultaneous video output ●

Biocompatibility of the SpyScope DS Access and Delivery Catheter was evaluated in accordance with ISO 10993-1:2009 for the body contact category of "surface - mucosal membrane" with a contact duration of "limited (≤24 hours)," and the following tests were performed: Cytotoxicity, Irritation, Sensitization, USP Physiochemical <661>, and Latex. All evaluation acceptance criteria were met.

Electrical safety and electromagnetic compatibility of the SpyGlass DS Direct Visualization System were evaluated in accordance with IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012, IEC 60601-1-2:2007, and IEC 60601-2-18:2009. All evaluation acceptance criteria were met.

The results of non-clinical testing demonstrate that the SpyGlass DS Direct Visualization System is considered safe and effective for its intended use.

#### 9. Conclusion

Boston Scientific has demonstrated that the proposed SpyGlass DS Direct Visualization System is substantially equivalent to the currently marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K142922](https://fda-staging.innolitics.com/device/K142922)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).