K142920 · K.M.S., Inc. · MQX · Jun 10, 2015 · General Hospital
Device Facts
Record ID
K142920
Device Name
KM Needles (Acupuncture Needle)
Applicant
K.M.S., Inc.
Product Code
MQX · General Hospital
Decision Date
Jun 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5580
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This KM Needles (Acupuncture Needle) is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the state regulations.
Device Story
KM Needles are simple stainless steel pins used by acupuncture practitioners to stimulate specific body points. Device consists of a needle body (0.16-0.70 mm diameter; 15-135 mm length) and a handle (20-30 mm length). Practitioners manually insert needles into the skin to treat various conditions. Device is single-use, sterile, and intended for professional clinical use. Benefits include facilitating traditional acupuncture therapy.
Clinical Evidence
Bench testing only. Testing included tensile strength verification and biocompatibility/sterility validation per ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 11137-2, AAMI/ISO/ANSI 11737-1, AAMI ANSI ISO 11607-1/2, and ASTM F1980-07. All results met established criteria.
Technological Characteristics
Material: STS 304 stainless steel. Design: Spring-type handle. Dimensions: 15-135 mm length, 0.16-0.70 mm diameter. Sterilization: Gamma radiation. Single-use, sterile. Biocompatibility: Conforms to ISO 10993-1.
Indications for Use
Indicated for use by qualified acupuncture practitioners to pierce the skin for acupuncture treatment.
Regulatory Classification
Identification
An acupuncture needle is a device intended to pierce the skin in the practice of acupuncture. The device consists of a solid, stainless steel needle. The device may have a handle attached to the needle to facilitate the delivery of acupuncture treatment.
Special Controls
*Classification.* Class II (special controls). The device, when it is an acupuncture point locator or a single use acupuncture needle, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9. Acupuncture needles must comply with the following special controls:(1) Labeling for single use only and conformance to the requirements for prescription devices set out in 21 CFR 801.109,
(2) Device material biocompatibility, and
(3) Device sterility.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 10, 2015
K.M.S. Inc. c/o Mr. Charlie Mack International Regulatory Consulting 12226 Washington Lane Parker, Arizona 85344
Re: K142920
Trade/Device Name: KM Needles (Acupuncture Needle) Regulation Number: 21 CFR 880.5580 Regulation Name: Acupuncture Needle Regulatory Class: II Product Code: MQX Dated: April 29, 2015 Received: May 06, 2015
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142920
Device Name KM Needles (Acupuncture Needle)
Indications for Use (Describe)
This KM Needles (Acupuncture Needle) is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the state regulations.
| Type of Use (Select one or both, as applicable) | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| <table border="0"><tr><td><div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div></td></tr><tr><td><div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | |
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
This summary of 510(k) substantial equivalence information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: May 29, 2015 510(k) Number: K142920
1. Company and Correspondent making the submission:
Name - K.M.S Inc. Address - F2, 81-1 Haengjeong-ri, Gwangdeok-myeon, Dongnam-gu, Cheonansi, Chungcheongnam-do, Korea 330922 Tel: 82-41-5587785 Fax: 82-41-5587786 Contact - Mr. Chul Hui.Kim Executive Director Email - charliemack@irc-us.com
2. Device :
Trade/proprietary name: KM Needles (Acupuncture Needle) Common Name : Acupuncture Needle Classification Name : needle, acupuncture, single use
- 3. Predicate Devices : Dong Bang Acupuncture Needles, (K972659)
- 4. Classifications Names & Citations : 21CFR 880.5580, MQX, Acupuncture Needle, Class 2
### 5. Description :
The KM Needles (Acupuncture Needle) are simple stainless steel pins that are used by an acupuncture practioner in the treatment of various conditions. Acupuncture is the timulation of specific acupuncture points along the skin of the body involving various methods such as penetration by thin needles or the application of heat, pressure, or laser light. Clinical practice varies depending on the country. Traditional acupuncture involves needle insertion, moxibustion, and cupping therapy. It is a form of alternative medicine and a key component of traditional Chinese medicine (TCM).
The KM Needles (Acupuncture Needle) are constructed of stainless steel, with pin body diameters ranging from 0.16 to 0.70 mm and vary in length from 15 to 135 mm. Each acupuncture needle has a handle for the practioner to manipulate the needle. The handle is also made from stainless steel and range in length from 20-30 mm.
{4}------------------------------------------------
- 6. Indication for use :
This acupuncture needle is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the state regulations.
- 7. Comparison with predicate device :
K.M.S Inc. believes that the KM Needles (Acupuncture Needle) is substantially equivalent to the Dong Bang Acupuncture Needles (K972659).
The predicate device chosen is comprised of the same material and both are single use devices delivered in a sterile condition. The predicate uses EO as the sterilization method and the KM Needles (Acupuncture Needle) are sterilized using radiation. Please note the table below, which demonstrates the similarities of the K.M.S Inc. Acupuncture Needles and the predicate device.
| UV YIVV.<br>Element of comparison | Subject Device | Claimed SE Device |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | K.M.S Inc. | Dong Bang Acupuncture<br>Needle |
| FDA510(K) Number | N/A | K972659 |
| Device Name | KM Needles<br>(Acupuncture Needle) | ACUPUNCTURE NEEDLES |
| Intended Use | This acupuncture needle is<br>intended to pierce the skin in<br>the practice of acupuncture by<br>qualified practitioners of<br>acupuncture as determined by<br>the state regulations. | This acupuncture needle is<br>intended to pierce the skin in<br>the practice of acupuncture by<br>qualified practitioners of<br>acupuncture as determined by<br>the states. |
| Principle of operation | General and hand type | General and hand type |
| Material:<br>Needle body<br>-<br>Handle<br>- | STS 304 | STS 304 |
| Specification:<br>Length<br>-<br>Diameter<br>- | Length: 15~135mm<br>Diameter: 0.16~0.70mm | Length: 7~150mm<br>Diameter: 0.16~0.50mm |
| Tensile Strength of the<br>Acupuncture Needle | Diameter :<br>0.20 (723.3gf)<br>0.25 (726.6gf)<br>0.35 (733.3gf) | Diameter :<br>0.20 (709.6gf)<br>0.25 (673.6gf)<br>0.35 (686.3gf) |
| Biocompatibility | Conform to ISO10993-1 | Conform to ISO10993-1 |
| Design | Spring | Spring |
| Reuse | Disposable, single use | Disposable, single use |
| Sterility | Gamma radiation | Sterilized by EO; SAL=10-6 |
The table provides a comparison of the KM Needles (Acupuncture Needle) with the predicate device:
{5}------------------------------------------------
The submitted K.M.S Inc. device and the predicate device have undergone testing to validate the tensile strength of the materials and also testing to verify the sterilization process and the sterile properties of the device, as delivered. The testing performed demonstrated that the devices met all of the criteria in these standards and that the submitted devices are similar in construction and delivered in the same state as the predicate Dong Bang acupuncture needles.
# 8. Performance Data :
Bench testing was performed to ensure the tensile strength of the stainless steel needle and Sterility testing (ISO10993-1 Biological evaluation of medical devices --Part 1: Evaluation and testing within a risk management process; ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization, ISO 1137-2 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, AAMI/ISO/ANSI 11737-1 Sterilization of medical devices - Microbiological methods Part 1: Determination of the population of microorganisms on product, 2ed. AAMI ANSI ISO 11607-1 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems, AAMI ANSI ISO 11607-2 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes, ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.) was performed. All test results were satisfactory.
## 9. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification K.M.S Inc. concludes that the KM Needles (Acupuncture Needle) are as safe and effective and substantially equivalent to predicate devices as described herein.
END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.