K142910 · Venus Concept USA, Inc. · GEI · Dec 19, 2014 · General, Plastic Surgery
Device Facts
Record ID
K142910
Device Name
Legacy BX
Applicant
Venus Concept USA, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 19, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Venus Legacy BX Device is a non-invasive device intended for use in dermatologic and general surgical procedures for females for the non-invasive treatment of moderate to severe facial wrinkles and rhytides in Fitzpatrick skin types I- IV.
Device Story
Non-invasive, non-ablative aesthetic device; delivers 1MHz Radio Frequency (RF) energy and 15Hz Pulsed Magnetic Field (PMF) energy to skin. RF energy heats tissue to trigger cellular changes; PMF energy acts as adjunct to reduce healing downtime. System comprises RF power module, PMF module, two handpiece applicators, and console with LCD touch screen. Operated by clinician in clinical setting. Output provides thermal energy to skin; healthcare provider monitors treatment via console interface. Device intended to improve facial wrinkles and rhytides, potentially benefiting patient through non-surgical skin rejuvenation.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing comparing RF output parameters, PMF output parameters, and temperature stability to the predicate device.
Technological Characteristics
Non-ablative aesthetic device. RF output: 1MHz, 150W max. PMF output: 15Hz, 15 Gauss max. Materials: Biocompatible per ISO 10993. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-2 (high-frequency surgical equipment). Form factor: Console (40cm x 40cm x 100cm, 40kg) with two handpiece applicators. Connectivity: Standalone system with LCD touch screen control.
Indications for Use
Indicated for females with moderate to severe facial wrinkles and rhytides, Fitzpatrick skin types I-IV.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2014
Venus Concept USA Incorporated % Ms. Ahava Stein A. Stein - Regulatory Affairs Consulting LTD % MEDX Ventures Group LLC 175 Derby Street, Unit 27 Suite 1 Hingham, Massachusetts 02043
Re: K142910
Trade/Device Name: Venus Legacy BX Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 19, 2014 Received: November 24, 2014
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Binita S. Ashar -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K 142910
Device Name
Venus Legacy BX Device
Indications for Use (Describe)
The Venus Legacy BX Device is a non-invasive device intended for use in dermatologic and general surgical procedures for females for the non-invasive treatment of moderate to severe facial wrinkles and rhytides in Fitzpatrick skin types I-IV
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY VENUS LEGACY BX DEVICE
## 510(k) Number K_______________________________________________________________________________________________________________________________________________________________
## Applicant Name:
| Company Name: | Venus Concept USA Inc. |
|---------------|------------------------------------------------------------|
| Address: | 4556 N. Hiatus Road,<br>Sunrise, FL 33351<br>United States |
| | Tel: +1-416-907-0115 |
| | Fax: +954-572-5680 |
| | E-mail: lilia@venus-concept.com |
## Contact Person:
| Official Correspondent: | Ahava Stein |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
| Address: | C/O MEDX Ventures Group LLC<br>175 Derby St., Unit 27 Suite 1<br>Hingham, MA 02043<br>USA<br>Tel: 1-888-4433867<br>Fax: 1-718-374-6529<br>E-mail: ahava@asteinrac.com |
| Date Prepared: | September 30, 2014 |
| Trade Name: | Venus Legacy BX Device |
| Classification Name: | CFR Classification section 878.4400; (Product code GEI) |
| Classification: | Class II Medical Device |
## Predicate Device:
The Venus Legacy BX Device is substantially equivalent to the previously cleared, Venus Freeze (MP)2 device, also manufactured by Venus Concept Ltd.
| Device | Manufacturer | 510(k) No. |
|-----------------------|--------------------|------------|
| Venus Freeze<br>(MP)2 | Venus Concept Ltd. | K111670 |
{4}------------------------------------------------
### Device Description:
The Venus Legacy BX device is a non-invasive, non-ablative medical aesthetic device. It is designed to deliver RF (Radio Frequency, 1MHz) energy and PMF (Pulsed Magnetic Field, 15Hz) energy to the skin. RF energy heats the tissue to trigger cellular changes for the intended use. The RF treatment is supplemented with an adjunct magnetic energy for the reduction of the downtime healing process.
The device consists of an RF Power Module, a Pulsed Magnetic Field (PMF) module, two hand piece applicators, and a console with control electronics and an LCD touch screen.
#### Device Specifications:
Maximal RF Output Power: 150W RF Output Frequency: 1[MHz] Magnetic Frequency: Pulse, 15Hz Maximal Magnetic Field: 15 Gauss Dimension: 40cm W x 40cm D x 100cm H (15.7' W x 15.7'' D x 39.4'' H) Weight: 40 Kg (88 lbs) Main Line Frequency (nominal): 50-60 Hz Input Voltage (nominal): 100-240 VAC
#### Intended Use/Indication for Use:
The Venus Legacy BX device is a non-invasive device intended for use in dermatologic and general surgical procedures for females for the non-invasive treatment of moderate to severe facial wrinkles and rhytides in Fitzpatrick skin types I- IV.
### Performance Standards:
The Venus Legacy BX device has been tested and complies with the following voluntary recognized standards:
- IEC 60601-1, (Third Edition, 2005 / 2006), Medical Electrical Equipment Part 0 1: General requirements for basic safety and essential performance
- IEC 60601-1-2. (Third Edition, 2007). Medical Electrical Equipment Part 1-2: ● General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests
- IEC 60601-2-2 (Fifth Edition, 2009): Medical Electrical Equipment Part 2: . Particular requirements for the safety of high frequency surgical equipment.
#### Non-Clinical (Bench) Performance Data:
A bench test was performed to measure the RF output parameters; PMF output parameters and temperature stability in the Venus Legacy BX device and compare them to the RF output measurements, PMF output parameters and temperature stability in the Venus Freeze (MP) device. The results of the bench test demonstrated that the Venus
{5}------------------------------------------------
Legacy BX device has the same RF and PMF output specifications and temperature stability profile as those reported for the predicate device and therefore, is substantially equivalent to the predicate device.
#### Pre-Clinical (Animal Study) Performance Data:
Not Applicable
#### Clinical Performance Data:
Not Applicable
#### Substantial Equivalence:
The indications for use and technological characteristics of the Venus Legacy BX device are substantially equivalent to the indications for use and technological characteristics of the Venus Freeze (MP)2 device.
The Venus Legacy BX device is a modified device of the Venus Freeze (MP)2 device, previously cleared under 510k No. K111670. The modifications are exclusively design related with a new outer design yet with the same technology and components as in the Venus Freeze (MP)2 device.
The inner design and components in the Venus Legacy BX device, including the console (with power supply, RF generator, controller and display panel) and the hand piece applicators (with cable, connector to console) are similar to the design and components found in the Venus Freeze (MP) device. The safety features and compliance with safety standards in the Venus Legacy BX device are similar to the safety features and compliance with safety standards found in the predicate devices (provided in Section 7). The patient contact materials are biocompatible in compliance with ISO10993 Standard and similar to materials found in the Venus Freeze (MP)device. The Venus Legacy BX device underwent performance testing, including software validation testing (provided in Section 12) and electrical and mechanical safety testing according to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2 (provided in Section 12). The performance specifications (including RF electrical power output, PMF output and temperature stability) of the Venus Legacy BX device are substantially equivalent to those in the Venus Freeze (MP) device, as shown in the performance bench tests provided in section 12. The results of the tests demonstrate that the specifications are similar to those of the predicate device. These performance tests demonstrate that the device specifications meet the system requirements and that the minor differences in design do not raise new safety or effectiveness concerns.
The general method of treating the patient's skin, the levels of energies used in treatment of skin, the size of treatment area and treatment zone, the active electrode area, durations of treatment, total energies delivered, are all similar to the respective methods and parameters in the predicate device.
{6}------------------------------------------------
Consequently, it can be concluded that the Venus Legacy BX device is substantially equivalent to the predicate Venus Freeze (MP)2 device, cleared under 510(k) K111670; and therefore, may be legally marketed in the USA.
## Conclusions:
Based on the performance testing and comparison to predicate devices, the Venus Legacy BX device is substantially equivalent to the Venus Freeze (MP)2 predicate device for the mentioned intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.