HEPATIQ is a nuclear medicine software application used to display and process liver-spleen images. The results obtained may be used as a tool, by a nuclear physician, in quantifying liver-spleen images. The data processed may be derived from any nuclear medicine liver-spleen procedure. The HEPATIQ software should only be used by qualified nuclear medicine professionals.
Device Story
HEPATIQ is a Windows-based software application for nuclear medicine workstations. It interfaces with hospital PACS via DICOM to retrieve liver-spleen images. The device automatically identifies regions of interest (ROI) for the liver, spleen, and intervening marrow; users review, approve, or manually adjust these ROIs. The software calculates the Perfused Hepatic Mass (PHM) parameter using standard formulas and generates a report for PACS storage. Used by nuclear medicine physicians to assist in image quantification, the device streamlines workflow by automating calculations previously performed manually. It provides a tool for clinical assessment of liver-spleen images.
Clinical Evidence
No clinical testing performed. Substantial equivalence established via non-clinical validation comparing automated PHM calculations from HEPATIQ against manual calculations from the predicate device (PEGASYS) on a set of test images. Results showed a coefficient of determination (R2) of 97%, exceeding the 90% acceptance criterion.
Technological Characteristics
Software-based image processing application; runs on Windows XP SP3 or later; interfaces with PACS via DICOM protocols. Features automated ROI identification for liver, spleen, and marrow; manual override capability; automated calculation of Perfused Hepatic Mass (PHM). Developed under 21 CFR 820.30 design controls.
Indications for Use
Indicated for qualified nuclear medicine professionals to display, process, and quantify liver-spleen images derived from nuclear medicine procedures.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure in profile, with three overlapping heads or faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2014
HEPATIQ LLC % Mr. Dipu Ghosh CEO 1150 Main Street. Suite E IRVINE CA 92614
Re: K142891 Trade/Device Name: HEPATIO Regulation Number: 21CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 27, 2014 Received: Oct. 3. 2014
Dear Mr. Ghosh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert A Ochs
Robert A. Ochs, PhD Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142891
Device Name HEPATIQ
Indications for Use (Describe)
HEPATIQ is a nuclear medicine software application used to display and process liver-spleen images. The results obtained may be used as a tool, by a nuclear physician, in quantifying liver-spleen images. The data processed may be derived from any nuclear medicine liver-spleen procedure. The HEPATIQ software should only be used by qualified nuclear medicine professionals.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a document header for HEPATIQ software. The title is "HEPATIQ SOFTWARE" and the subtitle is "510(k) Premarket Notification". The document number is H-029, the version is 0, and the page number is 14 of 82. The company is HEPATIQ LLC and the document is marked as confidential.
#### 5 510(K) SUMMARY
In accordance with 21 CFR 807.92, the following summary of the information in the 510(k) submission is provided.
#### 5.1 807.92(A)(1)
| Summary Date: | 9/27/2014 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | HEPATIQ LLC<br>1150 Main Street, Suite E<br>Irvine, California 92614<br>USA. |
| Primary Contact: | Dipu Ghosh, MSEE, MBA.<br>Chief Executive Officer<br>Phone: (949) 250-1065 ext. 124<br>Fax: (949) 250-1075<br>E-mail: dghosh@hepatiq.com |
| Secondary Contact: | John Hoefs, MD.<br>Chief Operating Officer<br>Phone: (714) 883-2208<br>Fax: (949) 250-1075<br>E-mail: jhoefs@hepatiq.com |
#### 5.2 807.92(A)(2)
| Trade Name: | HEPATIQ |
|-------------------------|-------------------------------------------------|
| Common Name: | Radiology Image Processing Software |
| Device: | System, Image Processing, Radiological |
| Regulation Number: | 21 CFR 892.2050 |
| Regulation Description: | Picture Archiving & Communication System (PACS) |
| Review Panel: | Radiology |
| Device Class: | Class II |
| Product Code: | LLZ |
{4}------------------------------------------------
| HEPATIQTM<br>Hepatic Quantitation | TITLE: | HEPATIQ SOFTWARE | |
|-----------------------------------|--------------|-------------------------------|-------------|
| | SUBTITLE: | 510(k) Premarket Notification | |
| | NUMBER: | H-029 | VER: |
| | CONFIDENTIAL | COMPANY: | HEPATIQ LLC |
| Indication For Use: | HEPATIQ is a nuclear medicine software application<br>used to display and process liver-spleen images.<br>The results obtained may be used as a tool, by a<br>nuclear physician, in quantifying liver-spleen images.<br>The data processed may be derived from any<br>nuclear medicine liver-spleen procedure.<br>The HEPATIQ software should only be used by<br>qualified nuclear medicine professionals. |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### 5.3 807.92(A)(3) — Predicate Device
| Predicate Device: | PEGASYS Ultra |
|-------------------------|-------------------------------------------------|
| Manufacturer: | ADAC Laboratories |
| 510(k) Number: | K993946 |
| Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Regulation Description: | Picture Archiving & Communication System (PACS) |
#### 5.4 807.92(A)(4) — Device Description
| Device Description: | HEPATIQ is Microsoft Windows software that allows<br>the user to display and process nuclear medicine liver-<br>spleen images.<br><br>HEPATIQ software runs on any nuclear medicine<br>workstation running Windows XP SP3 or later.<br><br>HEPATIQ software provides the user a means for<br>quantification of nuclear medicine liver-spleen images. |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{5}------------------------------------------------
| HEPATIO" " " | HEPATIQ SOFTWARE | | |
|---------------------------------------|------------------------------------------|--------|------------------|
| | subtitle: 510(k) Premarket Notification | | |
| Hepatic Quantitation NUMBER: H-029 | | VER: 0 | |
| CONFIDENTIAL | COMPANY: HEPATIQ LLC | | PAGE: 16 of 82 |
#### 5.5 807.92(A)(5) — Intended Use
| Intended Use: | HEPATIQ is a nuclear medicine software application<br>used to display and process liver-spleen images.<br>The results obtained may be used as a tool, by a<br>nuclear physician, in quantifying liver-spleen images.<br>The data processed may be derived from any nuclear<br>medicine liver-spleen procedure. |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The HEPATIQ software should only be used by<br>qualified nuclear medicine professionals. |
#### 5.6 807.92(A)(6) — Technological Characteristics
HEPATIQ software has two major technologies:
Image Display: HEPATIQ interfaces with the hospital PACS to display nuclear medicine liver-spleen images using the standard Digital Imaging and Communications in Medicine (DICOM) protocols..
Image Quantification: Regions of interest (ROI) are identified automatically for the liver and spleen. An ROI is also automatically identified for the marrow between the liver and the spleen. The user is required to review and approve these ROI or re-draw them manually. ROI parameters are computed. Image quantification parameter Perfused Hepatic Mass (PHM) is calculated from the ROI parameters using standard formulas. A report is prepared showing serial PHM values and saved on PACS.
## 5.6.1 Technological Similarities
The following table summarizes the similarities in the technological characteristics of HEPATIQ and the predicate device PEGASYS.
| Technology | PEGASYS<br>(Predicate) | HEPATIQ | Comparison |
|-----------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Computer | General purpose<br>commercial<br>computer (Sun) | General purpose<br>commercial<br>computer (Intel) | Both products run on general<br>purpose commercial computers.<br>Neither product is relying on a<br>specialized computer. |
| Technology | PEGASYS<br>(Predicate) | HEPATIQ | Comparison |
| Operating<br>System | General purpose<br>commercial<br>operating<br>system (UNIX) | General purpose<br>commercial<br>operating<br>system<br>(Windows) | Both products use a general<br>purpose commercial operating<br>system. Neither product is relying<br>on a specialized operating system.<br>Both UNIX and Windows provide<br>standard operating system<br>functions for the display and<br>processing of images. |
| Launch | PEGASYS<br>Image<br>Manipulation<br>application<br>launched by<br>clicking on an<br>icon. | HEPATIQ<br>application<br>launched by<br>clicking on an<br>icon. | Both products use a graphical user<br>interface icon to launch the<br>application. |
| PACS<br>Interface | DICOM | DICOM | Both products use DICOM<br>protocols to interface to the<br>hospital PACS for downloading<br>images and uploading reports. |
| Configuration | PEGASYS<br>screens to set<br>network and<br>operational<br>parameters. | HEPATIQ<br>screens to set<br>network and<br>operational<br>parameters. | Both products use a graphical user<br>interface dialog screen to set<br>configuration parameters. |
| Manual<br>image<br>processing | General purpose<br>software tools<br>for manual<br>image<br>processing. | General purpose<br>software tools<br>for manual<br>image<br>processing. | Both products provide similar<br>manual software tools for:<br>·Image Display<br>·Frame Selection<br>·Spleen Length<br>·Summarized Transaxial Image<br>·ROI Identification<br>•ROI Calculations |
| User | Qualified nuclear<br>medicine<br>professionals | Qualified nuclear<br>medicine<br>professionals | Both products require users to be<br>qualified nuclear medicine<br>professionals. |
| Technology PEGASYS<br>(Predicate) | HEPATIQ | Comparison | |
| Automated<br>image<br>processing | No automation<br>(manual<br>calculation of<br>PHM only) | Automated<br>calculation of<br>PHM | The only difference in the image<br>processing by PEGASYS and HEPATIQ<br>is the automation feature. Both produce<br>results that are highly correlated with<br>each other. Automation of the image<br>processing is a feature of HEPATIQ that<br>saves the user a tremendous amount of<br>time.<br>For validation of the automation feature,<br>the PHM calculation was performed<br>manually using PEGASYS and<br>automatically using HEPATIQ for a set<br>of test images. The PHM value obtained<br>from the two devices were compared<br>and found to be highly correlated thus<br>validating the automation feature. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the word "HEPATIQ" in large, bold, red letters, with the letters slightly overlapping each other. Below the word is a green swoosh and the words "Hepatic Quantitation" in a smaller, green font. At the bottom of the image, there is a gray rectangle with the word "CONFIDENTIAL" in bold, black letters.
HEPATIQ SOFTWARE TITLE: subtitle: 510(k) Premarket Notification H-029 NUMBER: VER: 0
COMPANY: HEPATIQ LLC
PAGE: 17 of 82
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a document header with the HEPATIQ logo and the words "Hepatic Quantitation" below it. The title of the document is "HEPATIQ SOFTWARE" and the subtitle is "510(k) Premarket Notification". The document number is H-029, the version is 0, the company is HEPATIQ LLC, and the page number is 18 of 82.
## 5.6.2 Technological Differences
The following table summarizes the differences in the technological characteristics of HEPATIQ and the predicate device PEGASYS.
As the HEPATIQ and predicate device PEGASYS calculation of PHM are highly correlated, the automation feature of HEPATIQ does not raise any different questions of safety or efficacy.
#### 5.7 807.92(B)(1) - Non-Clinical Performance Data
# 5.7.1 Design Controls
HEPATIQ was developed under the Quality System Regulation using Design Controls 21 CFR 820.30. This included establishing and maintaining procedures to ensure design requirements are met. All elements of the process have been documented, including design and development plan, design input, design output, design review, design verification, design validation, design transfer, design changes, and design history file.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a document header for a company called HEPATIQ LLC. The title of the document is "HEPATIQ SOFTWARE" and the subtitle is "510(k) Premarket Notification". The document number is H-029, the version is 0, and the page number is 19 of 82.
### 5.7.2 Software Verification
HEPATIQ has successfully undergone extensive verification testing to ensure that all requirements for the software have been met.
A Software Verification Protocol was developed and executed. All verification tests passed and this was documented in a Software Verification Report. These results provide objective evidence that the outputs of the software design activity meet all of the specified requirements for that activity. For details on the verification testing performed, please see Section 12.9 - Verification Testing.
### 5.7.3 Software Validation
Substantial equivalence was established by conducting a (non-clinical) validation test that collected performance data on HEPATIQ and the predicate device PEGASYS. The validation test consisted of directly comparing the manual processing of the liver-spleen images using the PEGASYS predicate device with the automatic processing of the same images using HEPATIQ.
A correlation analysis was performed between the manual PHM calculations using PEGASYS and the automatic PHM calculations using HEPATIQ. The analysis produced a coefficient of determination R2 of 97% which exceeded the validation criterion of a minimum R2 of 90%. Thus the validation test produced performance data that proved that HEPATIQ was substantially equivalent to PEGASYS.
As the HEPATIQ and predicate device PEGASYS calculation of PHM are highly correlated, the automation feature of HEPATIQ does not raise anv different questions of safety or efficacy. For details on the validation testing performed, please see Section 12.10 - Validation Testing (Performance Data).
### 5.7.4 Site Testing
Site testing was performed at a beta test hospital to test the installation procedures. The site testing was successful.
For details on the site test protocol, please see Attachment 11: H-026 -Site Test Protocol.
For details on the site test results, please see Attachment 12: H-027 -Site Test Report.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a document header with the HEPATIQ logo and the words "Hepatic Quantitation" below it. The title of the document is "HEPATIQ SOFTWARE" and the subtitle is "510(k) Premarket Notification". The document number is H-029, the version is 0, the company is HEPATIQ LLC, and the page number is 20 of 82. The word "CONFIDENTIAL" is written in bold letters on the left side of the header.
#### 807.92(B)(2) — Clinical Performance Data 5.8
Based on the criteria listed under Section F of the guidance document: "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications", no clinical testing of HEPATIQ was required or performed. For a detailed rationale, please see Section 20: Performance Testing - Clinical.
#### 5.9 807.92(B)(3) - Conclusion
Substantial equivalence of HEPATIQ and PEGASYS has been positively established in accordance with the flowchart in Appendix A of the guidance document: "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications"
| Decision | Question | Answer |
|----------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Is the predicate device<br>legally marketed? | Yes - PEGASYS is a legally marketed device |
| 2 | Do the devices have<br>the same intended<br>use? | Yes - HEPATIQ has the same intended use as<br>PEGASYS. |
| 3 | Do the devices have<br>the same technological<br>characteristics? | No - There is one difference:- PEGASYS<br>processes liver-spleen images manually while<br>HEPATIQ processes them automatically (with<br>manual overrides available to the user). |
| 4 | Do the different<br>technological<br>characteristics of the<br>devices raise different<br>questions of safety<br>and effectiveness? | No - The automation feature of HEPATIQ<br>does not raise different questions of safety<br>and effectiveness. Validation testing proved<br>that the PEGASYS manual processing and<br>HEPATIQ automated processing are highly<br>correlated. |
| 5a | Are the methods<br>acceptable? | Yes - Direct comparison of the processing of<br>the same images on both PEGASYS and<br>HEPATIQ was performed and standard<br>statistical analysis used for comparison. |
| 5b | Do the data<br>demonstrate<br>substantial<br>equivalence? | Yes - Manual processing with PEGASYS and<br>automatic processing with HEPATIQ were<br>correlated with an R2 of 97%, exceeding the<br>validation requirement of a minimum of 90%. |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a document with the title "HEPATIQ SOFTWARE" and subtitle "510(k) Premarket Notification". The document number is H-029 and the version is 0. The company is listed as HEPATIQ LLC and the page number is 21 of 82. The document is marked as confidential.
Furthermore, no applicable mandatory performance standards or special controls exist for HEPATIQ. The product design, development and manufacturing conform to Quality System Regulations. Results of testing and standards conformance demonstrate safety and effectiveness of HEPATIQ.
We conclude that HEPATIQ and PEGASYS are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.