Premium Surgiclip III

K142869 · Covidien · FZP · Oct 30, 2014 · General, Plastic Surgery

Device Facts

Record IDK142869
Device NamePremium Surgiclip III
ApplicantCovidien
Product CodeFZP · General, Plastic Surgery
Decision DateOct 30, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Premium Surgiclip™ III clip application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy, and radiographic markings.

Device Story

Premium Surgiclip™ III is a manual surgical instrument used to occlude vessels and tubular structures; also used for vagotomy, sympathectomy, and radiographic marking. Device consists of an applier shaft, handles, and an integrated cartridge containing 20 titanium clips. Operation: surgeon places jaw around vessel/structure; squeezing handles closes clip; internal ratchet mechanism ensures full handle squeeze before clip release; releasing handles automatically loads next clip. Device includes a lockout feature after the final clip is used. Used in clinical surgical settings. Benefits include secure vessel occlusion and procedural efficiency via automated clip loading.

Clinical Evidence

Bench testing only. In-vitro testing included clip over clip, jaw side load, jaw deflection, clip twist resistance, firing force, clip formation, and clip security. In-vivo testing included hemostasis, clip security, clip removal, and jaw twist deflection resistance.

Technological Characteristics

Manual surgical clip applier. Components constructed of materials compliant with ISO 10993-1. Features integrated cartridge with 20 titanium clips, ratchet mechanism for full-closure assurance, and automatic clip loading. Mechanical operation.

Indications for Use

Indicated for use in surgical procedures requiring occlusion of vessels and tubular structures, vagotomy, sympathectomy, and radiographic markings.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of three human profiles facing to the right, representing health and human services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 30, 2014 Covidien Mr. Michael Koczocik Product Specialist, Regulatory Affair 60 Middletown Avenue North Haven, Connecticut 06473 Re: K142869 Trade/Device Name: Premium Surgiclip™ III Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: September 30, 2014 Received: October 1, 2014 Dear Mr. Koczocik: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S Radiological Health Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Enclosure {2}------------------------------------------------ ### 5. Indication for Use Statement ### K142869 Page 1 of 1 DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: | OMB No. 0910-0120 | |------------------|--------------------------| | Expiration Date: | January 31, 2017 | | | See PRA Statement below. | | 510(k) Number ( <i>if known</i> ) | | |-----------------------------------|--| | Device Name | | Premium Surgiclip™ III Indications for Use (Describe) The Premium Surgiclip™ III clip application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy, and radiographic markings. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 {3}------------------------------------------------ ## 6. 510(k) Summary This 510(k) summary of data used to demonstrate substantial equivalence is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR క్తి807.98 | NAME: | Covidien | |----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ADDRESS: | 60 Middletown Avenue<br>North Haven, Connecticut 06473 USA | | CONTACT PERSON: | Michael Koczocik<br>Product Specialist, Regulatory Affairs | | PHONE NUMBER: | 203-492-6312 | | FAX NUMBER: | 203-492-5029 | | DATE PREPARRED: | September 30, 2014 | | TRADE/PROPRIETARY NAME: | Premium SurgiclipTM III | | COMMON/USUAL NAME: | Implantable Clip, Applier | | CLASSIFICATION NAME: | Implantable Clip | | PRODUCT CODE: | FZP | | CLASSIFICATION PANEL NAME: | General and Plastic Surgery | | FDA PANEL NUMBER: | 79 | | DEVICE CLASS: | Pursuant to 21 CFR § 878.44300, an Implantable<br>clip is a Class II device | | PREDICATE DEVICE(S): | Premium SurgiclipTM Small 9.0" Clip Applier<br>[K853650] (formerly AutoSutureTM Hemostatic<br>Clips) | | DEVICE DESCRIPTION: | The Premium SurgiclipTM III clip applier consists of<br>an applier shaft with attached handles and<br>integrated cartridge containing 20 titanium clips.<br>The clip applier jaw is placed around a vessel or<br>other tubular structure. As the handles of the | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square with a white square in the center and a light blue cross. To the right of the square is the word "COVIDIEN" in blue, sans-serif font. applier are brought together, the clip is closed around the vessel or structure. The device contains a ratchet mechanism to ensure that a clip is not allowed to release until a full handle squeeze is applied. As the handles are released, a new clip is automatically loaded into the clip applier jaw. After the last clip is used, the device is designed to lockout which prevents the hands from being closed. The Premium Surgiclip™ III clip applier has application in many types of surgical procedures to occlude vessels and other tubular structures and for vagotomy, sympathectomy and radiographic markings. SUMMARY COMPARING THE TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICE(S): INTENDED USE: The proposed device (Premium Surgiclip™ III clip applier) is a design modification to the predicate Surgiclip™ (formerly AutoSuture™ Premium Hemostatic Clips) cleared under [K853650] clip applier, while maintaining the same technological characteristics. Modifications have been made to the jaw to accommodate the clip feeding mechanism and jaw closing mechanism. Modifications to the clip geometry have been made to increase retention forces. The dimensional characteristics of the clips have changed slightly. The addition of a ratcheting mechanism has been introduced which prevents the jaws from opening until the handles have been squeezed and the clip has been fully formed. MATERIALS: All components of the Premium Surgiclip™ III clip applier are comprised of materials which are in accordance with ISO 10993-1. PERFORMANCE DATA: Bench performance evaluations were completed to show Premium Surgiclip™ III is substantially equivalent to the predicate device and performs as intended. > The tests performed to show substantial equivalence of the Premium Surgiclip™ III clip applier to the predicate device are as follows: CONFIDENTIAL {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the Covidien logo. The logo consists of a blue square with a white square inside, and a blue vertical line running through the center of the white square. To the right of the square is the word "COVIDIEN" in blue, sans-serif font. The logo is simple and modern, and the colors are clean and professional. ### In-vitro - Clip over clip ● - Jaw side load ● - Jaw deflection ● - Clip twist resistance . - . Firing force - Clip formation ● - Clip security . - In-vivo - Hemostasis ● - Clip security . - Clip Removal ● - Jaw twist deflection resistance . ### CONCLUSION: The results of the performance and comparative testing described above demonstrate that the Premium Surgiclip™ III clip applier is substantially equivalent to the predicate device, AutoSture™ Hemostatic Clips cleared under [K853650].
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...