The Nellcor Bedside SpO2 Patient Monitoring System is indicated for prescription use only for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, intra hospital transport, or home environments.
Device Story
Pulse oximeter for continuous, non-invasive monitoring of SpO2 and pulse rate; utilizes Nellcor OxiMax sensors; inputs red and infrared light signals interacting with tissue; processes signals via internal PCBA to calculate SpO2, pulse rate, and pulse amplitude; displays digital values and plethysmographic waveform/blip bar; includes SatSeconds alarm management system for desaturation/pulse rate limits; powered by AC or internal rechargeable lithium-ion battery; used in hospitals, clinical facilities, transport, and home environments by clinicians or patients; output informs clinical decision-making regarding patient oxygenation status; benefits include real-time monitoring and alarm notification for potential hypoxemia.
Clinical Evidence
No new clinical testing conducted. Substantial equivalence supported by bench testing and clinical data from a similar Nellcor device utilizing the same PCBA module. Bench testing confirmed SpO2 accuracy of ±2% (70-100%) without motion, ±3% (70-100%) with motion, and pulse rate accuracy of ±3% (20-250 BPM).
Technological Characteristics
Pulse oximeter; red/IR LED light source and photodetector; internal rechargeable 7.2V 83W/hr Li-ion battery or AC power (80-264V); PCBA-based processing; SatSeconds alarm management; connectivity for Nurse Call; compliant with ISO 9919, ISO 80601-2-61, IEC 60601-1 (3rd Ed), IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, and IEC 62304.
Indications for Use
Indicated for continuous non-invasive monitoring of functional oxygen saturation (SpO2) and pulse rate in neonatal, pediatric, and adult patients. Suitable for well or poorly perfused patients under motion or no-motion conditions in hospitals, hospital-type facilities, during intra-hospital transport, or in home environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Nellcor OxiMAX N-600x Pulse Oximetry System (K123581)
Nellcor Bedside SpO2 Patient Monitoring System (K120773)
Nellcor Bedside Respiratory Patient Monitoring System (K130320)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 26, 2015
Covidien, LLC Megan Fessenden Senior Regulatory Affairs Specialist 6135 Gunbarrel Avenue Boulder, CO 80301
Re: K142865
> Trade/Device Name: Nellcor Bedside SpO2 Patient Monitoring System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 24, 2015 Received: February 27, 2015
Dear Ms. Fessenden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### K142865 510(k) Number, if known: _
#### De vice Name : Nellcor Bedside SpO2 Patient Monitoring System
#### Indications for Use:
The Nellcor Bedside SpO2 Patient Monitoring System is indicated for prescription use only for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, intra hospital transport, or home environments.
#### Note:
- Hospital use typically covers such areas as general care floors (GCFs), operating rooms, ● special procedure areas, intensive and critical care areas within the hospital, and in hospital-type facilities.
- . Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
- . Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
- . Home environment includes any environment other than a professional healthcare facility or clinical laboratory where a device may be used.
Prescription Use 区 (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use □ (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evalutaion (ODE)
{3}------------------------------------------------
| Submitter | |
|---------------------|---------------------------------------------------------|
| Address | Covidien LLC<br>6135 Gunbarrel Ave<br>Boulder, CO 80301 |
| Phone | 303-305-2291 |
| Fax | 303-305-2212 |
| Company Contact | Megan Fessenden, Sr. Regulatory Affairs Specialist |
| Date Prepared | February 23, 2015 |
| Device | |
| Trade Name | Nellcor Bedside SpO2 Patient Monitoring System |
| Common Name | Pulse Oximeter |
| Classification Name | Oximeter (21 CFR 870.2700) |
| Classification | Class II |
| Product Code | DQA |
| Branch | Anesthesiology |
# Device Description
The Nellcor Bedside SpO2 Patient Monitoring System is a pulse oximeter designed for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended to be used with one of a range of Nellcor OxiMax oxygen transducers (sensors). The subject device displays digital values of SpO2 and pulse rate. Pulse amplitude is displayed by means of a 'blip bar' presentation or plethysmographic waveform. The Nellcor Bedside SpO2 Patient Monitoring System can be powered by an internal power supply operating on AC from a standard electrical utility receptacle (from 80V AC to 264V AC) or alternately by an integral sealed 7.2V, 83W/hr rechargeable lithium-ion battery.
The intended patient population includes neonatal, pediatric, and adult patients during both motion and no motion conditions. The patient may be well or poorly perfused and undergoing monitoring in hospitals, hospital-type facilities, intra-hospital transport, or home environments.
## Intended Use
The Nellcor Bedside SpO2 Patient Monitoring System is indicated for prescription use only for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended for use with neonatal, pediatric, and adult patients during both no motion conditions and for patients who are either well or poorly perfused and undergoing monitoring in hospitals, hospitaltype facilities, intra hospital transport, or home environments.
{4}------------------------------------------------
Note:
- Hospital use typically covers such areas as general care floors (GCFs), operating rooms, special . procedure areas, intensive and critical care areas within the hospital, and in hospital-type facilities.
- . Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
- . Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
- Home environment includes any environment other than a professional healthcare facility or clinical laboratory where a device may be used.
# Technological Characteristics
The subject and predicate devices measure functional oxygen saturation non-invasively via a light signal interacting with tissue, by utilizing the time-varying changes in tissue optical properties that occur with pulsatile blood flow.
At a high level, the subject and predicate devices are based on the following same technological characteristics:
- . Pulse Oximetry - Used to non-invasively measure functional oxygen saturation
- Regulation Number 870.2700 Oximeter ●
- . Product Code - DQA
- . Collect signal via light signal interacting with tissue - red & infrared light emitting diodes (LEDs) are used as light sources and a photodectector senses the signal strengths of the two wavelengths of light.
- . Use of software including the same oximetry algorithm to processes the information providing real-time values of SpO2, pulse rate, and pulse amplitude.
- . Use of an alarm management system – SatSeconds allows the caregiver to set the limits on desaturation and pulse rate including specifying acceptable duration and magnitude.
- Use of commonly used materials
- . Limited contact with skin, no contact with mucous membranes or non-intact skin designed cleaning and low-level disinfecting for multiple uses
- . Are electrically operated – come equipped with an internal battery supply and an external power cord
- . Operating and Storage Temperature and Humidity
- . Availability of Nurse Call connectivity
- . Compliance to Electrical Safety and EMC standards
- Intended patient population and sensor compatibility
- SpO2 and Pulse rate accuracy specifications
The following technological differences exist between the subject and predicate devices:
NOTE: When the difference does not apply to all predicates, the applicable predicate(s) are noted in parentheses.
- Updated compliance to current standards
- . Homecare use environment (Nellcor Bedside SpO2 Patient Monitoring System (Libra) and Nellcor Bedside Respiratory Patient Monitoring System only)
{5}------------------------------------------------
- Physical Dimensions and/or weight (All) ●
- . Operating Altitude (N-600x and Nellcor Bedside Respiratory Patient Monitoring System only)
- Storage Altitude (Nellcor Bedside Respiratory Patient Monitoring System only)
- Display size (N-600x and Nellcor Bedside Respiratory Patient Monitoring System only)
- . Printed Circuit Board Assembly (PCBA) (N-600x and Nellcor Bedside SpO> Patient Monitoring System only)
- . Ability to measure respiratory rate is only available on Nellcor Bedside Respiratory Patient Monitoring System
#### Substantial Equivalence
The Nellcor Bedside SpO2 Patient Monitoring System is considered to be substantially equivalent to the devices presented in Table 6-2.
| Device Manufacturer | Device Name | 510(k) Number | Date of Clearance |
|---------------------|-------------------------------------------------------------|---------------|-------------------|
| Covidien LLC | Nellcor OxiMAX N-600x<br>Pulse Oximetry System | K123581 | May 9, 2013 |
| Covidien LLC | Nellcor Bedside SpO2<br>Patient Monitoring<br>System | K120773 | July 12, 2012 |
| Covidien LLC | Nellcor Bedside<br>Respiratory Patient<br>Monitoring System | K130320 | February 4, 2014 |
#### Table 6-2 – Substantial Equivalence
## Performance Testing - Bench
Bench testing was conducted on the Nellcor Bedside SpO2 Patient Monitoring System to ensure the subject oximeter met design specifications. The device was tested to medical electrical equipment standards, environmental testing standards, medical device software standards, and usability engineering standards. These standards include:
| Standard | Title |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 9919 | Medical Electrical Equipment - Particular Requirements for the Basic Safety and Essential Performance of Pulse Oximeter Equipment for Medical Use |
| ISO 80601-2-61 | Medical Electrical Equipment - Part 2-61: Particular Requirements for the Basic Safety and Performance of Pulse Oximeter Equipment |
| IEC 60601-1: 3rd Edition | Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance |
| IEC 60601-1-2 | Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests |
| IEC 60601-1-8 | Medical Electrical Equipment - Part 1-8: General Requirements for Basic Safety and Essential Performance - Collateral Standard: General Requirements, Tests and Guidance for Alarm Systems in medical Electrical Equipment and Medical Electrical Systems |
| IEC 60601-1-11 | Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems used in the Home Healthcare Environment |
| IEC 60068-2-1 | Environmental Testing - Part 2-1: Tests - Test A: Cold |
| IEC 60068-2-2 | Environmental Testing - Part 2-2: Tests - Test B: Dry Heat |
{6}------------------------------------------------
#### Nellcor Bedside SpO2 Patient Monitoring System 510(k) Summary
| Standard | Title |
|----------------|-------------------------------------------------------------------------------------------------|
| IEC 60068-2-6 | Environmental Testing - Part 2-6: Tests - Fc: Vibration (sinusoidal) |
| IEC 60068-2-13 | Basic Environmental Testing Procedures - Part 2: Tests – Test M: Low Air Pressure |
| IEC 60068-2-27 | Environmental Testing - Part 2-27: Tests Test Ea and Guidance: Shock |
| IEC 60068-2-30 | Environmental Testing - Part 2-30: Tests – Test Db: Damp Heat, Cyclic |
| IEC 60068-2-64 | Environmental Testing - Part 2-64: Tests - Test Fh: Vibration, Broadband Random<br>and Guidance |
| IEC 62304 | Medical Device Software - Software Lifecycle Processes |
| IEC 62366 | Medical Devices - Application of Usability Engineering to Medical Devices |
The results of this testing show that the subject device meets the product requirements and specifications including: SpO2 accuracy without motion of ±2% from 70-100%, SpO2 accuracy with motion of ±3 from 70-100%, and pulse rate of ±3% from 20 to 250 BPM, and is substantially equivalent to the identified predicate devices.
## Performance Testing - Clinical
There was no new clinical testing conducted on the subject Nellcor Bedside SpO2 Patient Monitoring System. Clinical testing was conducted on a similar Nellcor pulse oximetry device utilizing the same PCBA module to demonstrate the accuracy of the oximetry performance during both motion and no motion conditions.
#### Summary
The Nellcor Bedside SpO2 Patient Monitoring System is a pulse oximeter that is substantially equivalent to the identified predicate pulse oximeters.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.