K142850 · Mckesson Medical Imaging Company · LLZ · Jan 6, 2015 · Radiology
Device Facts
Record ID
K142850
Device Name
McKesson Radiology Mammography Plus
Applicant
Mckesson Medical Imaging Company
Product Code
LLZ · Radiology
Decision Date
Jan 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
McKesson Radiology Mammography Plus™ ("Mammography Plus"), is an accessory to McKesson Radiology Station™, a component of McKesson Radiology™. McKesson Radiology is medical image and information management software that is intended to receive, transmit, store, archive, retrieve, manage, display, print and process digital medical images, digital medical video and associated patient and medical information. McKesson Radiology™ includes a suite of standalone, web-enabled software components, and is intended for installation and use with off-the-shelf hardware that meets or exceeds minimum specifications. McKesson Radiology Station is the primary software component used for processing and presentation of medical images on display devices with network access to McKesson Radiology. McKesson Radiology Station is intended to process and display lossless and non-lossless compressed medical images provided from DICOM conformant modalities such as X-Ray Radiography (including digital mammography), X-Ray Computed Tomography, Magnetic Resonance Imaging, Ultrasound, and Nuclear Medicine, as well as medical images obtained from other DICOM-compliant modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using cleared monitors intended for mammography display. McKesson Radiology is indicated for use by qualified healthcare professionals including, but not restricted to, radiologists, non-radiology specialists, physicians and technologists.
Device Story
Mammography Plus is a software accessory to McKesson Radiology Station PACS; enables specialized diagnostic viewing of digital mammography and Digital Breast Tomosynthesis (DBT) images. Inputs: DICOM-compliant medical images/video from modalities (X-ray, CT, MRI, Ultrasound, Nuclear Medicine). Operation: software provides image manipulation tools (zoom, CAD SR interpretation, historical image comparison, DBT scrolling/cine, 2D/3D toggling, slab reconstruction). Output: processed images displayed on cleared mammography-grade monitors. Used in clinical settings by radiologists/physicians for diagnostic interpretation. Benefits: facilitates specialized mammography workflows, improves visualization of breast imaging, and integrates BI-RADS assessment data to support clinical decision-making.
Clinical Evidence
Bench testing only. Verification and validation testing performed per FDA software guidance. Compliance verified against ISO 13485:2003, IEC 62304:2006, ISO 14971:2007/EN 14971:2012, NEMA XR 22/23-2006, IEC/ISO 10918-1:1994, NEMA PS 3.1-3.20 (DICOM), AAPM TG18, SMPTE RP 133-1991, and MQSA standards. No clinical study data provided.
Technological Characteristics
Software-based PACS accessory; runs on off-the-shelf hardware. Features image display/manipulation tools for 2D/3D mammography. Connectivity: DICOM-compliant, web-enabled. Standards: ISO 13485, IEC 62304, ISO 14971, NEMA PS 3.1-3.20, AAPM TG18, SMPTE RP 133, MQSA. Software-only device.
Indications for Use
Indicated for qualified healthcare professionals (radiologists, non-radiology specialists, physicians, technologists) to view and process digital medical images, including digital mammography and Digital Breast Tomosynthesis (DBT), for diagnostic interpretation. Contraindicated for primary interpretation of lossy compressed mammographic images and digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Hologic, Inc. SecurView DX Diagnostic Workstation (K103385)
{0}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with stylized wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 12, 2015
McKesson Medical Imaging Company % Mr. Paul Sumner Director, Regulatory Affairs McKesson Technologies, Inc. 5995 Windward Parkway ALPHARETTA GA 30005
Re: K142850
Trade/Device Name: McKesson Radiology Mammography Plus™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 15, 2014 Received: December 17, 2014
Dear Mr. Sumner:
This letter corrects our substantially equivalent letter of January 6, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2-Mr. Sumner
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert A Ocks
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K142850
Device Name
McKesson Radiology Mammography Plus™
#### Indications for Use (Describe)
McKesson Radiology Mammography Plus™ ("Mammography Plus"), is an accessory to McKesson Radiology Station™, a component of McKesson Radiology™. McKesson Radiology is medical image and information management software that is intended to receive, transmit, store, archive, manage, display, print and process digital medical images, digital medical video and associated patient and medical information. McKesson Radiology™ includes a suite of standalone, web-enabled software components, and is intended for installation and use with off-the-shelf hardware that meets or exceeds minimum specifications.
McKesson Radiology Station™ is the primary software component used for processing and presentation of medical images on display devices with network access to McKesson Radiology. McKesson Radiology Station is intended to process and display lossless and non-lossed medical images provided from DICOM conformant modalities such as X-Ray Radiography (including digital mammography), X-Ray Computed Tomography, Magnetic Resonance Imaging. Ultrasound, and Nuclear Medical images obtained from other DICOM-compliant modalities.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations.
Mammographic images may only be interpreted using cleared monitors intended for mammography display.
McKesson Radiology is indicated for use by qualified healthcare professionals including, but not restricted to, radiologists. non-radiology specialists, physicians and technologists.
Type of Use (*Select one or both, as applicable*)
| <b>☑</b> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------|-----------------------------------------------|
|-------------------------------------------------------|-----------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Empowenniq Healthcare
### 510(k) SUMMARY McKesson Medical Imaging Company's McKesson Radiology Mammography Plus Device
McKesson Medical Imaging Company 130 - 10711 Cambie Road Richmond, B.C. Canada, V6X 3G5
| 510(k) Owner and Contact: | Gilbert Wong Vice President, Regulatory Operations McKesson Medical Imaging Company 130 - 10711 Cambie Road Richmond, BC Canada V6X 3G5 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| | Phone: 604-279-5422 Fax: 604-279-5468 |
| | E-mail: Gilbert.Wong@McKesson.com |
| Date Prepared: | December 15, 2014 |
| Proprietary Name: | McKesson Radiology Mammography Plus™ |
| Common/Usual Name: | PACS |
| Classification Name: | Picture Archiving Communications System |
| Classification Regulation: | 21 CFR 892.2050 |
| Classification Product Code: | LLZ |
| Device Class: | Class II |
| Classification Panel: | Radiology Devices |
| Predicate Devices: | Hologic, Inc. SecurView DX Diagnostic Workstation (K103385) |
### Intended Use / Indications for Use
McKesson Radiology Mammography Plus™ ("Mammography Plus"), is an accessory to McKesson Radiology Station™, a component of McKesson Radiology™. McKesson Radiology is medical image and information management software that is intended to receive, transmit, store, archive, retrieve, manage, display, print and process digital medical images, digital medical video and associated patient and medical information. McKesson Radiology™ includes a suite of standalone, web-enabled software components, and is intended for installation and use with off-the-shelf hardware that meets or exceeds minimum specifications.
{4}------------------------------------------------
Empowering Healthcare
McKesson Radiology Station is the primary software component used for processing and presentation of medical images on display devices with network access to McKesson Radiology. McKesson Radiology Station is intended to process and display lossless and non-lossless compressed medical images provided from DICOM conformant modalities such as X-Ray Radiography (including digital mammography), X-Ray Computed Tomography, Magnetic Resonance Imaging, Ultrasound, and Nuclear Medicine, as well as medical images obtained from other DICOM-compliant modalities.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations.
Mammographic images may only be interpreted using cleared monitors intended for mammography display.
McKesson Radiology is indicated for use by qualified healthcare professionals including, but not restricted to, radiologists, non-radiology specialists, physicians and technologists.
### Technological Characteristics
Mammography Plus, an accessory to the McKesson Radiology Station, the radiology viewing component of McKesson Radiology, consists of image display and manipulation tools that help radiologists zoom in onto an image, interpret Computer Aided Detection (CAD) findings, navigate through and visualize digital mammography and Digital Breast Tomosynthesis (DBT) images, compare historically similar images, and configure their reading environment.
Mammography Plus uses the underlying capabilities of McKesson Radiology's diagnostic viewer, McKesson Radiology Station, to integrate specialized functionality into the diagnostic reading workflows. The McKesson Radiology Station viewer interfaces with the McKesson Radiology Platform to access patient and study information.
### Performance Data
Verification and validation testing were performed on Mammography Plus to ensure it met all design specifications and documentation was provided as recommended by FDA's guidance for industry and FDA staff entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" which included usability and image quality assessments compared to the predicate(s). Performance testing was conducted to verify compliance with specified design requirements in accordance with ISO 13485:2003, IEC 62304:2006 and ISO 14971:2007/ EN 14971:2012. Additionally, performance testing, specifically bench testing, included testing to conform to consensus standards such as NEMA XR 22-2006, NEMA XR 23-
{5}------------------------------------------------
Emponening Healthcare
2006, IEC/ ISO 10918-1:1994, NEMA PS 3.1 - 3.20 (2014) Digital Imaging and Communications in Medicine (DICOM), American Association of Physicists in Medicine (AAPM) TG18 Test Patterns used in conjunction with Assessment of Display Performance for Medical Imaging Systems (AAPM On-line Report No. 03), Society of Motion Picture and Television Engineers (SMPTE) Test Pattern Recommended Practice: Specifications for Medical Diagnostic Imaging Test Pattern for Television Monitors and Hard-Copy Recording Cameras, RP 133-1991 and the Mammography Ouality Standards Act (MQSA) (AS AMENDED BY MQSRA of 1998 and 2004).
In all instances, Mammography Plus functioned as intended and the observed results demonstrated substantial equivalence with the predicate devices.
### Substantial Equivalence
Mammography Plus is substantially equivalent to the identified predicate devices which include Hologic SecurView DX Diagnostic Workstation (K103385), as well as other similar, if not identical, legally commercially available devices in U.S. Interstate commerce. Specifically, Mammography Plus has the same general intended use and similar indications for use, technological characteristics and principles of operation compared to these previously cleared predicate devices.
As with the predicate device(s), Mammography Plus provides end users with the ability for specialized mammography features such as support for interpretation of Computer Aided Detection Structured Reports (CAD SR), Quadrant Zoom, Historical Similar Images (HSI), Zoom to Fit Anatomy, Zoom to Match Region of Interest (ROI), Unviewed Images Tool. Back-to-Back Placement of Images, and Background Air Suppression. Mammography Plus like the predicate device also provides enhanced features which include processing and displaying Digital Breast Tomosynthesis (DBT) images, and DBT-specific tools such as Display Protocols, Scrolling Mode, Cine Speed, Orientation Indications, Toggling between 2D and 3D Views, and reconstruction of thinslab visualizations from the DBT slices (also commonly referred to as "slabbing").
Unlike the predicate device(s), however, similar to other mammography image viewing and reporting devices legally marketed in U.S. interstate commerce, including McKesson Radiology (K140909), Mammography Plus also provides the following functionalities:
- The display of BI-RADS® (Breast Imaging-Reporting and Data System) . Assessment Categories obtained from prior patient mammography and other breast imaging reports,
- . The ability to automatically dim auxiliary monitors, and
- Web-enabled communication functionality for imaging study and patient records ● access.
{6}------------------------------------------------
Empowering Healthcare
None of the additional functionalities raise any new issues of safety and/ or effectiveness.
Thus, Mammography Plus is substantially equivalent to previously-cleared predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.