K142834 · Neurowave Systems, Inc. · OLT · Jun 23, 2015 · Neurology
Device Facts
Record ID
K142834
Device Name
DiscoverEEG System, Model DE-401
Applicant
Neurowave Systems, Inc.
Product Code
OLT · Neurology
Decision Date
Jun 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The DiscoverEEG System, Model DE-401 is intended to be used for measuring and recording the electrical activity of a subject's brain, obtained by placing non-invasive electrodes on the head. The DiscoverEEG, DE-401 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and at home and for clinical research. The medical use of data acquired by the DiscoverEEG is to be performed under the direction and interpretation of a licensed medical professional. The DiscoverEEG, Model DE-401 does not provide any diagnostic conclusion about the subject's condition.
Device Story
Wearable, medical-grade EEG monitor; acquires/stores up to 4 channels of EEG signals via non-invasive disposable electrodes. Components: Acquisition Module (signal digitization), Memory Module (processing, storage, power), Disposable Electrode Array, and PC-based Data Viewer Software. Operates by digitizing analog EEG signals; performs impedance checks, lead-off detection, and artifact removal. Produces raw EEG waveforms and processed spectral variables (power spectrum, MEF, SEF, frequency bands). Used in OR, ICU, ER, clinical, and home settings by clinicians or researchers. Data transferred to PC for review; supports conversion to standard formats (EDF+, tab-delimited). Provides signal quality metrics (impedance, artifact flags) to assist clinicians in data interpretation. Benefits include portable, continuous EEG monitoring for clinical research and patient assessment without providing automated diagnostic conclusions.
Clinical Evidence
No clinical data. Bench testing only; device performance evaluated against design specifications and safety standards (IEC 60601-1, 60601-1-11, 60601-2-26, 60601-1-2).
Technological Characteristics
Wearable EEG monitor; 4-channel acquisition; silver-silver chloride disposable electrodes. Battery-powered; standalone or wireless data transfer. PC-based software for data retrieval/review. Complies with IEC 60601-1, 60601-1-11, 60601-2-26, and 60601-1-2. Amplifier: ≥110 dB CMRR, ≥100 Meg Ohm input impedance.
Indications for Use
Indicated for acquiring EEG signals in OR, ICU, ER, clinical settings, and home environments for clinical research; intended for use under the direction and interpretation of a licensed medical professional; no diagnostic conclusions provided.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
NeuroFAST Monitoring System, Model NF-201 (K092477)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2015
NeuroWave Systems Inc. Tatjana Zikov President 2490 Lee Blvd., Suite 300 Cleveland Heights, OH 44118
Re: K142834
Trade/Device Name: DiscoverEEG System, Model DE-401 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLT, OMC Dated: May 10, 2015 Received: May 11, 2015
Dear Tatjana Zikov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aguel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142834
Device Name DiscoverEEG System, DE-401
### Indications for Use (Describe)
The DiscoverEEG System, Model DE-401 is intended to be used for measuring and recording the electrical activity of a subject's brain, obtained by placing non-invasive electrodes on the head. The DiscoverEEG, DE-401 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and at home and for clinical research. The medical use of data acquired by the DiscoverEEG is to be performed under the direction and interpretation of a licensed medical professional. The DiscoverEEG, Model DE-401 does not provide any diagnostic conclusion about the subject's condition.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| 区 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Contact Details 1.
| Applicant Name: | NeuroWave Systems Inc.<br>2490 Lee Blvd., Suite 300<br>Cleveland Heights, OH 44118<br>(216) 361-1591 (telephone)<br>(216) 361-1554 (fax) |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Tatjana Zikov,<br>President<br>tzikov@neurowavesystems.com |
| Date Prepared: | September 29, 2014 |
#### 2. Device Name
| Trade Name: | DiscoverEEG System, Model DE-401 |
|-----------------------|----------------------------------|
| Common/Usual Name: | EEG Monitor |
| Classification Name: | Electroencephalograph |
| Product Code: | OMC |
| Regulation Number: | 21 CFR 882.1400 |
| Device Class: | Class II |
| Classification Panel: | Neurology |
#### Legally Marketed Predicate Device 3.
| 510(k)<br>Number | Product<br>Code | Trade Name | Manufacturer |
|------------------|-----------------|----------------------------------------------|--------------------------------------------|
| K092477 | OMC | NeuroFAST Monitoring<br>System, Model NF-201 | NeuroWave Systems, Inc. |
| K101830 | OLV,<br>OMC | Zmachine | Consolidated Research of<br>Richmond, Inc. |
#### 4. Description of Device
The DiscoverEEG System, Model DE-401, is a wearable, medical-grade EEG device that acquires and stores up to four electroencephalograms (EEGs) obtained from noninvasive electrodes placed on a subject's head. The acquired EEG waveforms, as well as, processed EEG spectral variables are continuously stored by the system for later retrieval. The data can be transferred from the DiscoverEEG hardware to a computer for review.
The DiscoverEEG System, Model DE-401 has four main components:
{4}------------------------------------------------
- . Acquisition Module: contains the circuitry for acquisition of the EEG signals through connections to the electrodes placed on the subject's head. The acquired signals are digitized and then passed to the Memory Module, for signal processing and storage.
- . Memory Module: provides power to the system and contains memory and EEG processing circuitry (impedance check, lead off check, pre-processing incl. artifact removal, processed spectral variables). Includes LED indicators for assurance of proper operation and other user interface elements. May contain a wireless option for data transfer to Data Viewer Software in realtime.
- . Disposable Electrode Array: contains noninvasive, disposable, single patient electrodes (5 electrodes, 4 EEG channels) and skin preparation materials. The packaging will guide the user on how to prepare the subject's skin and how to place the electrodes.
- . Data Viewer Software: PC based software used for retrieving and reviewing acquired raw EEG waveforms (as well as processed data) from the DiscoverEEG hardware. This software will also be capable of converting the retrieved EEG data files to standard formats (i.e. EDF+, tab-delimited). Such converted data files can then be reviewed using off-the-shelf bio-signal review software that accepts the converted format.
#### 5. Intended Use
The intended use of DiscoverEEG System, Model DE-401 is consistent with the classification 21 CFR 882.1400, Electroencephalograph:
"An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head."
The DiscoverEEG System, Model DE-401 is intended to be used for measuring and recording the electrical activity of a subject's brain, obtained by placing non-invasive electrodes on the head. The DiscoverEEG, DE-401 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and at home and for clinical research. The medical use of data acquired by the DiscoverEEG is to be performed under the direction and interpretation of a licensed medical professional. The DiscoverEEG, Model DE-401 does not provide any diagnostic conclusion about the subject's condition.
#### 6. Substantial Equivalence Comparison
The technology of the DiscoverEEG DE-401 device is equivalent to other EEG monitor devices. The EEG signal is acquired in analog format, digitized, processed and saved for later user interpretation. An Electrode (Sensor) Array is provided with the system for the convenience of the user in applying the acquisition electrodes. The Electrode Array is made of the materials that have been shown to pass cytotoxicity, primary skin irritation and sensitization as required (refer to section 15 for addition details).
{5}------------------------------------------------
| Feature | DiscoverEEG System, Model<br>DE-401 | NeuroFAST Monitoring<br>System,<br>Model NF-201<br>(092477) | Zmachine<br>(101830) |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The DiscoverEEG System,<br>Model DE-401 is intended to<br>be used for measuring and<br>recording the electrical<br>activity of a subject's brain,<br>obtained by placing non-<br>invasive electrodes on the<br>head. The DiscoverEEG, DE-<br>401 is indicated for use in<br>acquiring<br>electroencephalographic<br>(EEG) signals in the OR, ICU,<br>ER, clinical settings and at<br>home and for clinical research.<br>The medical use of data<br>acquired by the DiscoverEEG<br>is to be performed under the<br>direction and interpretation of<br>a licensed medical<br>professional. The<br>DiscoverEEG, Model DE-401<br>does not provide any<br>diagnostic conclusion about<br>the subject's condition. | The NeuroFAST Monitor NF-<br>201 is intended to be used for<br>measuring and recording the<br>electrical activity of a<br>patient's brain, obtained by<br>placing electrodes on the<br>head. The NeuroFAST NF-<br>201 is indicated for use in<br>acquiring<br>electroencephalographic<br>(EEG) signals in the OR, ICU,<br>ER, clinical settings and for<br>clinical research. The<br>NeuroFAST Monitor is to be<br>used under the direction and<br>interpretation of a licensed<br>medical professional. The<br>NeuroFAST Monitor NF-201<br>does not provide any<br>diagnostic conclusion about<br>the patient's condition. | The CR1 Zrnachine is a<br>single-channel EEG<br>acquisition and analysis<br>system, designed for use in<br>the home or clinical<br>environments. This device<br>is intended to be used by<br>qualified healthcare<br>practitioners to monitor the<br>wake and sleep states of<br>adult patients and as an<br>adjunct to their diagnosis<br>of sleep disorders. |
| Modalities | EEG | EEG | EEG |
| Environment<br>of Use | Operating room, intensive care<br>unit, emergency room, clinical<br>settings and At Home where<br>EEG monitoring is used. | Operating room, intensive<br>care unit, emergency room,<br>and clinical settings where<br>EEG monitoring is used. | In the home or clinical<br>environment |
| Power Source | Battery | 120 Volt 60Hz AC power | Li Ion Battery |
| System<br>Components | Electrode (Sensor) Array,<br>Acquisition and Memory<br>Modules, Data Viewer<br>Software | Patient Module, Patient Cable,<br>Display Module, Data Cable<br>and EasyPrep Electrode Kit | Zmachine device, EEG<br>sensors, Sensor cable, wall<br>charger |
| Sensing<br>Electrodes | Silver-silver chloride<br>disposable EEG electrodes | Silver-silver chloride<br>disposable EEG electrodes | Silver-silver chloride<br>disposable electrodes |
| Feature | DiscoverEEG System, Model<br>DE-401 | NeuroFAST Monitoring<br>System,<br>Model NF-201<br>(092477) | Zmachine<br>(101830) |
| Screen<br>Display<br>Details | Data Viewer Software:<br>1) Raw EEG Waveforms<br>2) Signal Quality<br>information: Channel<br>Connection flags,<br>Impedance values,<br>Artifact Detection flags,<br>Noise Level<br>3) Spectral Parameters: EEG<br>power spectrum and<br>frequency bands, 50%<br>50% Median Edge<br>Frequency (MEF), and<br>95% Spectral Edge<br>Frequency (SEF) | Displays:<br>1) Raw EEG Waveforms<br>2) Signal Quality<br>information: Channel<br>Connection flags,<br>Impedance values,<br>Artifact Detection flags,<br>Noise Level<br>3) Spectral Parameters: EEG<br>power spectrum and<br>frequency bands, 50%<br>50% Median Edge<br>Frequency (MEF), and<br>95% Spectral Edge<br>Frequency (SEF) | Displays:<br>Sleep/Wake Statistics |
| Stored EEG<br>data available | Yes - SD Card | Yes – through removable<br>storage | Yes - microSD |
| EEG<br>Channels/<br>Montage | Up to 4 channels (2 fronto-<br>temporal channels, and 2<br>temporal channels) | 2 bilateral fronto-temporal<br>channels viewed concurrently | 1 channel, differential<br>mastoid |
| Real Time<br>EEG Display | Yes, if wireless option is<br>present | Yes | No |
| Processed<br>EEG<br>Bandwidth | User Selectable in Data<br>Viewer software<br>Low Filter: 0.125 or 0.5 Hz<br>High Filter: 30 or 70 Hz | User Selectable<br>Low Filter: 0.125 or 0.5 Hz<br>High Filter: 30 or 70 Hz | 0.5-380 Hz |
| Automatic<br>Artifacting | Yes | Yes | Unknown |
| Amplifier<br>Common<br>Mode<br>Rejection<br>Ration<br>(CMRR) | ≥110 dB | ≥100 dB | Unknown |
| Amplifier<br>Input<br>Impedance | ≥100 Meg Ohm | ≥50 Meg Ohm | 10 Gig Ohms |
| Electrode<br>Impedance<br>Test | Yes | Yes<br>Continuous and on demand by<br>the user | Yes |
| Feature | DiscoverEEG System, Model<br>DE-401 | NeuroFAST Monitoring<br>System,<br>Model NF-201<br>(092477) | Zmachine<br>(101830) |
| Derived /<br>Processed<br>EEG Measures | Power spectrum parameters<br>derived from FFT: | Power spectrum parameters<br>derived from FFT: | Wake/Sleep Determination |
| | 1) Power spectrum displayed<br>as Density Spectral Array<br>(DSA) | 1) Power spectrum<br>displayed as Density<br>Spectral Array (DSA) | |
| | 2) 95% Spectral Edge<br>Frequency (SEF) | 2) 95% Spectral Edge<br>Frequency (SEF) | |
| | 3) Median Edge Frequency<br>(MEF) | 3) Median Edge Frequency<br>(MEF) | |
| | 4) Spectral Powers in EEG<br>frequency bands<br>traditionally used to<br>quantify EEG signals (α,<br>β1, β2, δ, θ and γ) | 4) Spectral Powers in EEG<br>frequency bands<br>traditionally used to<br>quantify EEG signals (α,<br>β1, β2, δ, θ and γ) | |
| | Signal Quality information<br>(Channel connection flags,<br>Impedance values, Artifact<br>Detection flags, Noise Level) | Signal Quality information<br>(Channel connection flags,<br>Impedance values, Artifact<br>Detection flags, Noise Level) | |
| Contains<br>patient<br>isolation | Battery powered, thus no<br>connection between patient<br>and mains | Yes - Patient module<br>(analog-to digital converter)<br>provides 6kV electrical<br>isolation of the patient from<br>the monitor | Battery powered, thus no<br>connection between patient<br>and mains |
| Display<br>Interface | Yes, LEDs to ensure proper<br>system connection and<br>operation. | Yes - High-resolution, color,<br>graphical user interface and<br>touch screen | Yes, To ensure proper<br>system connection and<br>operation. |
| Event Markers | Yes, artifact detection and<br>signal quality events<br>User selectable events<br>included in Data Viewer<br>software | Yes, user selectable | Yes, Sleep/Wake Events |
{6}------------------------------------------------
{7}------------------------------------------------
#### Non-clinical Testing 7.
Laboratory testing, performed on identical hardware to the DiscoverEEG subject of this submission, demonstrated that the DiscoverEEG, Model DE-401 meets its design and functional requirements. Actual device functions and features were evaluated against the device specifications and in all instances the DiscoverEEG, Model DE-401 performed as expected and no unexpected behavior was observed.
Prior to commercial introduction, the device will meet the requirements of UL medical electrical equipment standards for safety and the IEC particular standard for electroencephalographs.
{8}------------------------------------------------
To ensure safety, the DiscoverEEG, Model DE-401 will comply with all applicable requirements of IEC60601-1 and IEC60601-1-11, and will also comply with the particular requirements for the safety of electroencephalographs established in IEC60601-2-26.
To ensure electromagnetic compatibility, the DiscoverEEG, Model DE-401 will comply with all applicable requirements of IEC60601-1-2, and will also comply with the particular requirements for the safety of electroencephalographs established in IEC60601-2-26.
#### 8. Clinical Testing
The DiscoverEEG, Model DE-401 device is an electroencephalographic device comprised of hardware that has been bench tested to assess safety and effectiveness and to establish substantial equivalence with the predicate devices. We believe further clinical data is not required to demonstrate performance for the DiscoverEEG, Model DE-401 for the indication for use subject to this submission.
#### 9. Conclusion
The DiscoverEEG, Model DE-401, when compared to its predicate devices, has the similar intended use and technological characteristics and principles of operation. The nonclinical tests demonstrate that the device is safe, as effective, and performs at least as safely and effectively as the legally marketed devices. Thus, in indications, intended use and technology, the DiscoverEEG, Model DE-401 is substantially equivalent to the predicated devices. Minor technological difference as documented in the Substantial Equivalence Comparison table above, raise no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.