K142833 · Jeisys Medical, Inc. · OUH · Jun 4, 2015 · General, Plastic Surgery
Device Facts
Record ID
K142833
Device Name
INTRAGen
Applicant
Jeisys Medical, Inc.
Product Code
OUH · General, Plastic Surgery
Decision Date
Jun 4, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The INTRAgen is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Device Story
INTRAgen is a monopolar electrosurgical unit used for electrocoagulation and hemostasis in dermatologic and general surgical procedures. The device delivers high-frequency current to target tissue via single-use active electrodes (KT-07 or KT-15) and a patient return plate. Heat generated by load/contact resistance causes tissue coagulation. The system includes a main frame, handpiece, foot switch, and power cable. It is operated by a clinician in a clinical setting. The device features an impedance monitor to manage energy delivery. Unlike some predicates, it does not use skin cooling or cryogens. The clinician uses the device to achieve hemostasis or coagulate tissue, directly impacting surgical outcomes by controlling bleeding and tissue effects. Benefits include effective coagulation for dermatologic and surgical applications.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing (dimensional accuracy, impedance measurement, output accuracy, waveform analysis, skin temperature safety) and a pre-clinical porcine study. The porcine study evaluated skin structure and collagen effects histologically and via RT-PCR up to 10 weeks post-treatment, demonstrating neocollagenesis and neoelastogenesis.
Technological Characteristics
Monopolar electrosurgical unit; 6 MHz output frequency. Patient-contacting materials: PCB and ABS (ISO 10993-1 compliant). Active electrodes (KT-07, KT-15) sterilized via 20% EtO/80% CO2 (ISO 11135-1). Features impedance monitoring. Power output: 5 W - 136 W. Dimensions: 365.0 x 454.8 x 795.5 mm; Weight: 28 kg. Complies with IEC 60601-1, IEC 60601-2-2, and IEC 60601-1-2.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electrocoagulation and hemostasis. No specific patient population age or gender restrictions provided.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring a stylized image of three human profiles facing right, with a design that resembles a bird or flowing fabric above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 4, 2015
Jeisys Medical Incorporated % Ms. Priscilla Chung LK Consulting Group USA Incorporated 2651 East Chapman Avenue, Suite 110 Fullerton, California 92831
Re: K142833
Trade/Device Name: INTRAgen Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: OUH GEI Dated: April 28, 2015 Received: April 30, 2015
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142833
Device Name
INTRAgen
Indications for Use (Describe)
INTRAgen is intended for dermatologic and general surgical procedures for electro coagulation and hemostasis.
Type of Use (Select one or both, as applicable)
| > Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CRF 807.92] (K142833)
#### 1. Date Prepared
June 3, 2015
#### 2. Submitter's Information
- Name of 510K Applicant / Manufacturer:
Jeisys Medical Inc.
307, Daeryung Techno Town 8th, 481-11, Gasan-dong, Geumcheon-gu, Seoul, 153-775, Korea,
- Contact Name: KyungWon Hwang (Mr.) / RA Manager .
- -Telephone No. : +82 2 2603 6417
- -Fax No. :
-
●
- +82 2 2603 6447 Email Address : lhs1436@jeisys.com
- Registration Number: 3006985163 .
#### Trade Name, Common Name, and Classification 3.
- . Trade Name: INTRAgen
- Common Name: Electrosurgical Unit .
- Classification Name: Electrosurgical cutting and . coagulation device and accessories.
-
- Classification Panel: General & Plastic Surgery
- Classification Regulation: 21 CFR 878.4400 •
- Primary Product Code: ойн • Secondary Product Code: GEI
{4}------------------------------------------------
- . Device Class: ll
### 4. Identification of Predicate Device(s)
The identified predicate devices within this submission are shown as follow;
- 510(k) Number: K023365 .
- Applicant: ●
- Common Name: .
Thermage Inc. Electrosurgical Unit
- Device Name:
- Thermage ThermaCool System
### 5. Description of the Device
The INTRAgen uses the principle that coagulation of cellular tissue is caused using the heat generated by the load or contact resistance by carrying high frequency current to the body.
Active electrodes of INTRAgen have 2 types according to effective area: model KT-07 (7 mm x 7 mm) and model KT-15 (15 mm x 15 mm). Electrical current flows through the tip of the electrode, through the target tissue and to the patient plate. The active electrodes are for single use. Patient contacting materials are PCB (Polychlorinated Biphenyl) and ABS (Acrylonitrile-Butadiene-Styrene Copolymer).
The INTRAgen consists of the following components:
- INTRAgen (Main Frame) .
- . Active Accessory : Active Electrode(KT-07 (7 mm x 7 mm), KT-15 (15 mm x 15 mm)), Handpiece Hanger and Hand-piece
- . Miscellaneous Accessories
: Foot Switch and Power Cable
#### 6. Intended Use
The INTRAgen is indicated for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.
{5}------------------------------------------------
#### 7. Technological Characteristics
Based on a technical feature comparison, the subject device was found to have the same output frequency as the predicate devices. The design and the components of the INTRAgen including the RF generator and the handpiece applicator are similar to the predicate.
However, the subject device has a different power output, tip energy and electrode size from the predicate device. The power output of the subject device is 5 W - 136 W while that of the predicate device is 330 W. The tip energy of the subject device is 10 J – 136 J while that of the predicate device is 52 J - 220 J. The subject device has two types of tip size (7 mm x 7 mm and 15 mm x 15 mm) whereas the predicate device has three types of tip size (25 mm x 25 mm, 15 mm x 15 mm and 30 mm x 30 mm).The differences are insignificant and do not influence the safety or efficacy of the device as shown in the performance tests provided in section 9.
| Description | INTRAgen | |
|-------------------------|--------------------------------------------------------------------------------------|----------------------|
| 510(K) Number | K142833 | |
| Indications for Use | Dermatologic and general surgical procedures for electro coagulation and hemostasis. | |
| Electrosurgical Unit | Major Functions | 1. Monopolar |
| | | 2. Impedance monitor |
| | Performance Specifications | Output Frequency |
| | | Power Output |
| | | |
| | | Tip Energy |
| | | |
| | | Fluence |
| | | |
| Physical Specifications | 1. Dimensions: 365.0 * 454.8 * 795.5 ( mm) | |
| | 2. Weight: 28 kg | |
The detailed information on INTRAgen's specifications is shown in the following table:
{6}------------------------------------------------
| Active Accessory | Monopolar or Bipolar | Monopolar | | | | | | | | | | | | | | | | | | |
|----------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| | Physical<br>Dimensions and<br>Design | KT-07:<br>31 * 30.4 * 30.3 (mm)<br>(width * length * height)<br>KT-15:<br>31 * 30.4 * 29.4 (mm)<br>(width * length *height) | | | | | | | | | | | | | | | | | | |
| | Effective Area | KT-07: 7 * 7 (mm) | | | | | | | | | | | | | | | | | | |
| | | KT-15: 15 * 15 (mm) | | | | | | | | | | | | | | | | | | |
| | Materials(Patient contact) | PCB<br>(polychlorinated biphenyl)<br>ABS (Acrylonitrile-Butadiene-<br>Styrene Copolymer) | | | | | | | | | | | | | | | | | | |
| | Foot Pedal | Physical Specification | 1) Dimensions: 130 * 150 * 134.6 (mm) (width * length *height)<br>2) Cable length: 2 (m)<br>3) Weight: 1 kg | | | | | | | | | | | | | | | | | |
| Power adjustment increment | Level Joules (J) T 5 1 45 2 60 3 75 4 91 5 105 6 134 7 136 | | | | | | | | | | | | | | | | | | | |
| On/Off Time | KT-07 On: 2 sec<br>Off: 0 sec KT-15 On: 1 sec<br>Off: 0 sec | | | | | | | | | | | | | | | | | | | |
| | Use or Non-use of Coupling gels | | Use | | | | | | | | | | | | | | | | | |
| Skin Cooling Methods | | Not use | | | | | | | | | | | | | | | | | | |
| Ointment and/or Drug Product Use | | Not use | | | | | | | | | | | | | | | | | | |
{7}------------------------------------------------
## 8. Substantial Equivalence
Substantial equivalence to the following predicate devices is claimed:
| Description | | Proposed<br>[INTRAgen] | Predicate<br>[Thermage<br>ThermaCool<br>System] | Remark | |
|----------------------------------|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|----------------------------------------------------------------------------------|
| 510(K) Number | | K142833 | K053365 | - | |
| Indications for Use | | The INTRAgen is<br>indicated for use in<br>dermatologic and<br>general surgical<br>procedures for<br>electrocoagulation<br>and hemostasis. | The Thermage<br>ThermaCool System<br>is indicated for use in<br>Dermatologic and<br>General Surgical<br>procedures for<br>electrocoagulation<br>and hemostasis". | Same | |
| Electr<br>osurgi<br>cal<br>Unit | Major Functions | 1. Monopolar<br>2. Impedance<br>monitor | 1. Monopolar<br>2. Impedance<br>monitor<br>3. Temperature<br>monitor | Performance<br>Testing | |
| | Performance<br>Specifications | Output<br>Frequency | 6 MHz | 6 MHz | Same |
| | | Power<br>Output | KT-07:<br>5 W - 35 W<br>KT-15:<br>5 W - 136 W | 330 W | Electrical<br>Safety<br>Testing<br>(IEC<br>60601-2-<br>2) and<br>Animal<br>Study |
| | | Tip Energy | KT-07:<br>10 J - 70 J<br>KT-15:<br>20 J - 136 J | 52 J - 220 J | Performa-<br>nce<br>Testing<br>and |
| | | | | | Animal<br>Study |
| | | Fluence | KT-07:<br>20 – 144 J/cm²<br>KT-15:<br>2 – 60 J/cm² | 225 J/cm² | Performa-<br>nce<br>Testing<br>and<br>Animal<br>Study |
| | Physical Specifications | | 1. Dimensions:<br>365.0 * 454.8 *<br>795.5 (mm)<br>2. Weight: 28 kg | Not known | - |
| Active<br>Accessory | Monopolar or Bipolar | | Monopolar | Monopolar | Same |
| | Physical<br>Dimensions<br>and Design | Dimensions | KT-07:<br>31 * 30.4 * 30.3<br>(mm)<br>(width * length *<br>height)<br>KT-15:<br>31 * 30.4 * 29.4<br>(mm)<br>(width * length *<br>height) | 2.5 * 2.5 (mm),<br>15 * 15 (mm),<br>30 * 30 (mm) | Performa-<br>nce<br>Testing |
| | | Effective<br>Area | KT-07: 7 * 7 (mm)<br>KT-15: 15 * 15<br>(mm) | 15 * 15 (mm),<br>2.5 * 2.5 (mm),<br>30 * 30 (mm) | |
| | | Materials(Patient contact) | PCB<br>(polychlorinated<br>biphenyl)<br>ABS (Acrylonitrile-<br>Butadiene-Styrene<br>Copolymer) | Polymid film | Biocompatibility<br>Testing |
| Foot<br>Pedal | Physical Specification | | 1) Dimensions:<br>130 * 150 *<br>134.6 (mm)<br>(width * length *<br>height)<br>2) Cable length: 2<br>(m)<br>3) Weight: 1 kg | Not known | - |
| Power adjustment increment | Level | Joules (J) | Level | Joules (J) | Performa- |
| | T | 5 | 10.5 | 52 | nce<br>Testing<br>and<br>Animal<br>Study |
| | 1 | 45 | 11.0 | 58 | |
| | 2 | 60 | 11.5 | 65 | |
| | 3 | 75 | 12.0 | 72 | |
| | 4 | 91 | 12.5 | 79 | |
| | 5 | 105 | 13.0 | 87 | |
| | 6 | 134 | 13.5 | 95 | |
| | 7 | 136 | 14.0 | 103 | |
| | | | 14.5 | 112 | |
| | | | 15.0 | 121 | |
| | | | 15.5 | 131 | |
| | | | 16.0 | 140 | |
| | | | 16.5 | 151 | |
| | | | 17.0 | 161 | |
| | | | 17.5 | 172 | |
| | | | 18.0 | 184 | |
| | | | 18.5 | 195 | |
| | | | 19.0 | 207 | |
| | | | 19.5 | 220 | |
| On/Off Time | KT-07 | On: 2 sec<br>Off: 0 sec | Not Known | | Performa-<br>nce<br>Testing<br>and<br>Animal<br>Study |
| | KT-15 | On: 1 sec<br>Off: 0 sec | | | |
| Use or Non-use of Coupling gels | Use | | Use | | Same |
| Skin Cooling Methods | Not use | | Uses a bio-safe<br>cryogen<br>Tetrafluoroethane | | Performa-<br>nce<br>Testing |
| Ointment and/or Drug Product Use | Not use | | Not known | | |
{8}------------------------------------------------
{9}------------------------------------------------
The indications for use and technological characteristics of the INTRAgen device are substantially equivalent to theThermageThermaCool System. The key technological characteristics of the INTRAgen, such as energy type and operating principle, are equivalent to the predicate device.The design and components in the INTRAgen device, including the RF generator and the hand piece applicator are also similar to the design and components found in the predicate ThermageThermaCool System.
The major differences between the subject device and the predicate device are power output, tip energy and electrode size from the predicate device. However, the preclinical study using the porcine model demonstrated the subject device would perform as well as the predicate devices in the market. The INTRAgen device also underwent performance testing, including software validation testing (provided in Section 4) and electrical and mechanical safety testing according to IEC 60601-1 and
{10}------------------------------------------------
electromagnetic compatibility testing according to IEC 60601-1-2 (provided in Section 10) and bench tests (provided in Section 9). These performance tests demonstrated that the differences in the device specifications meet the system requirements and do not raise new safety or effectiveness concerns.
Consequently, it can be concluded that the INTRAgen is substantially equivalent to the predicate ThermageThermaCool System, cleared under 510(k) K053365.
#### 9. Summary of Non-Clinical Performance Data
- . Electrical Safety and Electromagnetic Compatibility
The tests were performed on the INTRAgen to confirm electrical safety and electromagnetic compatibility in accordance with IEC 60601-1, IEC 60601-2-2, and IEC 60601-1-2. The results were acceptable.
- Biocompatibility .
The patient contacting part of the INTRAgen is the active electrode and it is made of PCB (Polychlorinated Biphenyl) and ABS (Acrylonitrile-Butadiene-Stvrene Copolymer). Biocompatibility test has been conducted according to the ISO 10993-1 standards, and the test results support that it is biocompatible.
- . Bench Data
Bench tests were performed to evaluate the dimensional accuracy, the accuracy of measurement for impedance, the accuracy of the active electrode output and output waveform. The test results were acceptable. An additional bench test was performed to demonstrate that the temperature of the skin surface does not go up to the point where it can cause damage to the skin tissue even when the cooling system is not used. The results demonstrated that the temperature of the skin surface did not go up to the point where it could cause damage to the skin tissue.
- . Pre-Clinical
A pre-clinical trial was performed to evaluate performance of the subject device. The test was performed on a porcine tissue model using different joule energy levels and active electrodes.
- Study purpose: The study was designed to evaluate the effects of the novel grid fractional RF treatment on skin structure and collagen in vivo.
{11}------------------------------------------------
- Follow up periods/days: The wound healing response was evaluated histologically and by RT-PCR up to 10 weeks post-RF treatment. - Animal type: Porcine
- Treatment applied: The INTRAgen RF device was used to deliver RF energy to skin.
- Histology evaluation: The study results showed that vigorous wound healing response was initiated post-treatment with progressive increase in inflammatory cell infiltration from day 2 through 10 weeks. Furthermore, through both immune histochemical and PCR studies, it was found that profound neocollagenesis and neoelastogenesis following RF treatment of micropigskin occurred.
In conclusion, the test results of the pre-clinical study demonstrated that the subject device is substantially equivalent to the predicate device.
- · Sterilization
The active electrodes (KT-07 and KT-15) are sterilized in its final packaging using 20% Ethylene Oxide (EtO) and 80% CO2. The EtO sterilization process has been validated in accordance with ISO 11135-1:2007.
· INTRAgen complies with the following recognized standards. The following table lists the non-clinical testing performed.
| Testing Type | Test Description | Test Result |
|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety and<br>Electromagnetic<br>Compatibility Testing | • IEC 60601-1:2012<br>• IEC 60601-2-2:2009<br>• IEC 60601-1-2:2007 | The INTRAgen met all<br>The acceptance<br>criteria in accordance<br>with IEC 60601<br>1:2009, IEC 60601-2-<br>2:2009 and IEC<br>60601-1-2:2007. |
| Biocompatibility<br>Testing | • Cytotoxicity<br>• Sensitization<br>• Intracutaneous Reactivity<br>• Acute Systemic Toxicity<br>• Pyrogen | The active electrodes<br>are biocompatible. |
| Performance Testing | • Dimensional Inspection<br>• Accuracy of Measurement<br>• Energy Output<br>• Current Density<br>• On/off time<br>• Output Waveform<br>• Skin Temperature | The INTRAgen met all<br>the acceptance<br>criteria. |
{12}------------------------------------------------
| Pre-clinical Trial | • Tissue Analysis<br>• Total RNA isolation and<br>real-time RT-PCR<br>· Statistical analysis | The test results<br>demonstrated that the<br>subject device is<br>substantially equivalent<br>to the predicate device. |
|--------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Sterilization | • ISO 11135-1:2007 | The INTRAgen met all<br>the acceptance<br>criteria in accordance<br>with ISO 11135-<br>1:2007. |
#### 10. Summary of Clinical Test
No clinical studies were considered necessary and performed.
#### 11. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR part 807, FDA's "Premarket Notification [510(k)] Submissions for Electrosurgical Devices for General Surgery" and based on the information provided in this premarket notification, Jeisys Medical Inc. concludes that INTRAgen is safe, effective and substantially equivalent to predicate devices for dermatologic and general surgical procedures for electrocoagulation and hemostasis.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.