The Ophthalmed Bending Laser Probes (Models A207000, and A307000) are indicated for performing laser endophotocoagulation in the posterior segment of the eye during vitreoretinal surgery at an operating wavelength range from 500 to 1100 nm.
Device Story
Device is a fiberoptic laser probe for vitreoretinal surgery; transmits laser energy from external laser source to surgical site. Consists of 8ft fiberoptic, SMA905 connector, handle with sliding button, and distal stainless steel/PEEK memory tube. Surgeon manipulates sliding button to advance internal straightening tube, adjusting distal tip angle (up to 55 degrees for 20g/23g, 40 degrees for 25g). Used in OR by ophthalmologists. Does not provide illumination or imaging. Benefits include adjustable tip angle for precise laser delivery during posterior segment procedures. Device is sterile, single-use.
Clinical Evidence
Bench testing only. No clinical or animal data provided. Testing included laser transmission (94-95%), spot size, energy distribution, and linearity across 20g, 23g, and 25g models. Abuse testing performed at 1,500mW/1,500 shots showed no degradation in performance compared to predicates.
Indicated for patients undergoing vitreoretinal surgery requiring laser endophotocoagulation in the posterior segment of the eye. Operating wavelength range 500-1100 nm.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human profile in triplicate, suggesting a sense of community and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 15, 2015
Ophthalmed LLC Mr. Jay Mansour Director , QA/RA 1050 Northfield Ct., Suite 280 Roswell, GA 30076
Re: K142830
> Trade/Device Name: Ophthalmed Directional Laser Probes Models A207000. A807000 and A307000 Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQB, HQF Dated: September 22, 2014 Received: September 30, 2014
Dear Mr. Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/10 description: The image shows the name "Deborah L. Falls -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The background is plain, ensuring the text is easily readable. The letters are uniformly sized and spaced, contributing to the overall clarity of the name.
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142830
#### Device Name
Ophthalmed Bending Laser Probes (Models A207000, A807000, and A307000)
Indications for Use (Describe)
The Ophthalmed Bending Laser Probes (Models A207000, and A307000) are indicated for performing laser endophotocoagulation in the posterior segment of the eye during vitreoretinal surgery at an operating wavelength range from 500 to 1100 nm.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------------------------------------|
| | ☐ Temporary hardship (illness, loss of job, etc.) |
| | ☐ Disaster-related hardship (declared disaster area) |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510k Summary
As required by 21 CFR 807.92
#### -SUBMITTER
OPHTHALMED LLC. 1050 NORTHFIELD COURT, SUITE 280. ROSWELL, GA 30076 USA. Tel 770 777 6613 Fax 678 623 3765
Contact person: Jay Mansour Date prepared: June 25, 2015
#### II. DEVICE
Name of device: Ophthalmed bending laser probes (Models A207000, and A307000) Common or usual name: laser probe Classification name: Ophthalmic photocoagulator Regulatory class: II Product code: HQB
#### III. PREDICATE DEVICE
| 510k number | Trade or Proprietary or Model Name | Manufacturer | Primary predicate device |
|-------------|----------------------------------------------------|-------------------|--------------------------|
| K021696 | LF20, LF25 | OPHTHALMED LLC | Yes |
| K050807 | LF100 | OPHTHALMED LLC | No |
| K113857 | Synergetics Directional Endo Ocular<br>Laser Probe | SYNERGETICS | No |
| K121187 | Katalyst laser probes | KATALYST SURGICAL | No |
#### IV. DEVICE DESCRIPTION
This laser probe is made out of 8ft fiberoptic, terminated on one end with Alcon compatible (SMA905) laser connector, and on another side with a handpiece for holding and manipulation during surgery. For protection against damages, a flexible plastic jacket covers the length of the fiber.
The handpiece is terminated by a proximal stainless steel tubing that is 20, 23 or 25 gauge in size, and ending with a distal pre-curved PEEK memory tube that can change angle when activated by the sliding button on the handle side.
When the sliding button on the handle is advanced, an internal straightening tube advances into the distal pre-curved PEEK memory tube, thus causing a reduction of the angle of the pre-curved PEEK memory tube, down to zero degrees at the maximum sliding position,
{4}------------------------------------------------
allowing the selection of the desired angle during surgery. The maximum angle is 55 degrees for 20g and 23g, and 40 degrees for 25g.
This device allows to transmit laser energy from the laser source to the surgical site, facilitated by the aiming beam that is provided by the laser source. This device does not provide illumination and should not be used to illuminate the surgical site.
The user should refer to the instructions for use of the laser source, for information about the increments for power settings, duty cycle, and ranges and increments for timing and interval.
This device does not include any lenses, and there is no focal distance to define. The laser spot size at 4mm away is 1mm.
## v. INDICATIONS FOR USE
The Ophthalmed Bending Laser Probes (Models A207000, and A307000) are indicated for performing laser endophotocoagulation in the posterior segment of the eye during vitreoretinal surgery at an operating wavelength range from 500 to 1100 nm.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICE DEVICE:
Summary of technological characteristics of our device compared to predicate device, in design, material, chemical composition and energy source:
## Design:
All predicate devices have the same main structure and functionality as our device: a fiber optic is terminated on one side via a laser connector that mates with a laser machine, and on the other side terminated by a tip that penetrates the eye so that the laser beam can reach into the surgical site. (K050807, in addition to the above, provides also illumination and aspiration functionalities).
The difference is that only our device and K113857 allow the tip to move, for the surgeon to select the desired angle. The other predicate devices have fixed tip. K113857 has a pre-curved Nitinol tube enclosed within the main (external) stainless steel shaft that projects out of the handle. As the Nitinol tube projects out of the main shaft, the fiber optic follows the curvature of the Nitinol tube. Our device, on the contrary, has a fixed and pre-curved PEEK tip, which changes angle, as a straightening (internal) stainless steel tube penetrates it internally. The fiber optic within the pre-curved tip follows the curvature of the pre-curved tip.
In terms of design, our device uses the same components as K021696 (and K050807, except for illumination and aspiration functions), but with two changes: (a) enhanced
{5}------------------------------------------------
fiber optic specifications, and (b) a movable tip that was tested to be substantially equivalent in its performance to K113857.
#### Material and chemical composition:
All predicate devices have the same component list as our device: glass fiber optic, plastic protective sheath that covers the fiber optic, metallic laser connector, plastic handle and stainless steel tip. In addition to the above, our device and K113857 have other materials, namely, PEEK tube for our device, and Nitinol tube for K113857.
Laser output properties for our device, K113857 and K021696 (and indirectly K050807, having the same fiber optic as K021696) were successfully tested. In other words, the intended functionality of our device was not affected by the change in materials.
Our device is sterile with sterility assurance level of SAL 10-6, as per documented evidence provided within the submission, by running an additional fractional cycle based on currently validated ethylene oxide sterilization cycle.
#### Energy used:
All predicate devices and our device do NOT emit energy, but only transmits laser energy from the connector side to the tip side.
The fiber optic of K021696 and K050807 was cleared originally for 514 to 532 nm, then subsequently for 500 to 1,100 nm.
K121187 was cleared for 500 to 900 nm.
Our device covers 500 to 1,100 nm, as for K021696 and K050807.
Our device, K113857 and K021696 (and indirectly K050807, having the same fiber optic as K021696) have been tested for up to 1,500 mW, with comparable and acceptable laser output characteristics.
#### Biocompatibility testing
The biocompatibility evaluation for Ophthalmed Bending Laser Probes (Models A207000, A807000, and A307000) was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO 10993, as recognized by FDA.
The battery of testing included the following tests:
- Cytotoxicity (L929 Neutral red uptake, 4 concentrations, as per ISO 10993-5:2009, GLP study)
- Irritation (intracutaneous injection, as per ISO 10993-10:2010, GLP study)
- Sensitization (Kligman maximization, as per ISO 10993-10:2010, GLP study)
{6}------------------------------------------------
The device is considered tissue contacting for a duration of less than 24 hours.
## Electrical safety and electromagnetic compatibility (EMC)
Not applicable
## Software verification and validation testing
Not applicable
## Mechanical and acoustic testing
Our device is substantially equivalent to K113857 and indirectly K050807, having the same fiber optic as K021696), as far as laser transmission, spot size, energy distribution and linearity, at baseline and after prolonged conditions.
Three 20g, three 23g and three 25g sterile probes were subjected to various tests, along with predicate devices, as described below, in accordance with 21 CFR 807.92(b)(1):
#### Laser transmission:
- i-Ophthalmed probes as per K021696: 96 to 97%
- ii-Synergetics probes as per K113857: 92 to 97%
- iii-Probes under this submission K142830: 94 to 95%
The above transmission rates were substantially the same prior and after an abuse test based on 1,500 shots at 1,500mW, with pulse duration of 200 msec at 50 msec pulse intervals, with transmission rates measured at 1,000mW.
#### Spot size and energy distribution:
The same laser probes (K021696, K113857 and K142830) were also subjected to additional tests. Projecting the laser beam onto an appropriate surface, photos were taken at 1000mW, indicating no changes before and after the same abuse test of 1500mW, concerning the spot sizes and energy distribution.
The energy distribution was confirmed to be uniform for the three 510k submissions.
Also, the average divergence angle was determined to be 20 degrees for K113857, 14 degrees for K021696 and 16 degrees for K142830.
#### Linearity:
The laser probes of K113857 and K142830 were tested for linearity at 25%, 50%, 75% and 100% of knob sliding. Again, the same abuse test of 1500mW did not affect the results. It was determined that for sizes of 20g and 23g of K142830, the tip bends 20
{7}------------------------------------------------
deg, 37.5 deg, 45 deg and 55 deg, at 25%, 50%, 75% and 100% sliding, respectively, but 15 deg, 27.5 deg, 37.5 deg and 42.5 deg for 25g. In comparison, K113857 reflected 22.5 deg, 42.5 deg, 60 deg and 75 deg for 20g probe, but 20 deg, 30 deg, 40 deg, and 45 to 60 deg for 25g probe, respectively as well (at 25%, 50%, 75% and 100% of sliding).
## Animal study
Not applicable.
## Clinical studies
Not applicable.
## VII. CONCLUSIONS
The various tests, internal and external, document that our device is substantially equivalent to the predicate device. Our laser output is similar, our mechanism is similar, our biocompatibility is tested, our sterility is documented, and our overall structure / components is similar.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.