Visi Mobile Monitoring System, Visi Mobile Chest Sensor
Applicant
Sotera Wireless, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jul 20, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire), respiration rate, noninvasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, basic arrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole, Atrial Flutter) and alarm in hospital-based facilities; including general medical-surgical floors, intermediate care floors, and emergency departments. Continuous non-invasive blood pressure (cNIBP) measurements have not been evaluated on patients during ambulation. The basic arrhythmia analysis feature is intended for use on patients 18 years of age and older. It has not been evaluated on pediativ patients or neonates. The arrhythmia analysis feature is intended for use by healthcation and treatification and treatment of arrhythmia events. Automated arrhythmia analysis is an adjunct to clinician review of the analysis should precede any therapeutic intervention. The ViSi Mobile Monitoring System may be used as standalone devices or networked to a ViSi Mobile Remote Viewer through wireless 802.11 communication.
Device Story
Body-worn vital signs monitor; features high-resolution touch screen; provides visual/audible alarms. Inputs: ECG, SpO2, respiration, NIBP/cNIBP, pulse rate, temperature. Transforms inputs via arrhythmia analysis algorithm; outputs real-time vital signs and rhythm alerts. Used in hospital settings (general/intermediate/emergency) by clinicians/medically qualified personnel. Operates as standalone or networked via 802.11 to ViSi Mobile Remote Viewer. Output assists clinicians in patient assessment and arrhythmia identification; serves as adjunct to clinical review. Benefits: continuous monitoring of adult patients; enables timely intervention for detected arrhythmia events.
Clinical Evidence
Bench testing only. Performance validated against standard databases (AHA, MIT-BIH, CU, NST) per ANSI/AAMI EC57:2012. QRS detection sensitivity comparable to predicate. Electrical safety (AAMI/ANSI ES60601-1), EMC (IEC 60601-1-2), alarm system (AAMI/ANSI/IEC 60601-1-8), and ECG performance (AAMI/ANSI/IEC 60601-2-27) testing confirmed compliance.
Technological Characteristics
Body-worn monitor with touch screen display. Wireless 802.11 connectivity. Arrhythmia analysis algorithm adapted from Monebo Technologies library. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, AAMI/ANSI/IEC 60601-1-8, and AAMI/ANSI/IEC 60601-2-27 standards.
Indications for Use
Indicated for adult patients (18+) in hospital settings (general medical-surgical, intermediate care, emergency departments) for monitoring ECG, respiration, NIBP, cNIBP, SpO2, pulse rate, and basic arrhythmia (VT, VF, Asystole, Atrial Flutter). Not for pediatric/neonatal use. cNIBP not evaluated during ambulation.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Automated ECG Analysis and Interpretation Analysis Software Library (K062282)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird or abstract human figures, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 20, 2015
Sotera Wireless, Inc. Eben Gordon Regulatory Consultant 10020 Huennekens Street San Diego, California 92121
Re: K142827
Trade/Device Name: ViSi Mobile Monitoring System, ViSi Mobile Chest Sensor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including St-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DSL, DRT, DXN, DQA, FLL Dated: June 10, 2015 Received: June 11, 2015
Dear Eben Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Arthel Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K142827
Device Name ViSi Mobile Monitoring System
Indications for Use (Describe)
The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire), respiration rate, noninvasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, basic arrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole, Atrial Flutter) and alarm in hospital-based facilities; including general medical-surgical floors, intermediate care floors, and emergency departments.
Continuous non-invasive blood pressure (cNIBP) measurements have not been evaluated on patients during ambulation.
The basic arrhythmia analysis feature is intended for use on patients 18 years of age and older. It has not been evaluated on pediativ patients or neonates.
The arrhythmia analysis feature is intended for use by healthcation and treatification and treatment of arrhythmia events. Automated arrhythmia analysis is an adjunct to clinician review of the analysis should precede any therapeutic intervention.
The ViSi Mobile Monitoring System may be used as standalone devices or networked to a ViSi Mobile Remote Viewer through wireless 802.11 communication.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for Sotera Wireless. The logo features the word "Sotera" in a bold, black font. Above the word "Sotera" is a blue heart with three curved lines emanating from it. Below the word "Sotera" is the word "WIRELESS" in a smaller, blue font.
# 510(K) SUMMARY
| Date prepared | July 7, 2015 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Sotera Wireless, Inc.<br>10020 Huennekens Street<br>San Diego, CA 92121<br>T. 858.427.4620; F. 858.999.2487 |
| Trade name | ViSi Mobile Monitoring System |
| Common name | Vital signs monitor |
| Regulation Name | Arrhythmia detector and alarm (including ST-segment measurement and alarm) |
| Classification number | 21 CFR 870.1025 |
| Product code | MHX, DSI, DRT, DXN, DQA, FLL |
| Regulatory class | II |
| Predicate devices | Automated ECG Analysis and Interpretation Analysis Software Library, K062282<br>(Clearance: 03/22/2007)<br>Acuity Central Monitoring Station, K052160 (Clearance: 12/16/2005)<br>ViSi Mobile Monitoring System; K122036 (Clearance: 8/15/2012)<br>ViSi Mobile Monitoring System; K130709 (Clearance: 10/7/2013) |
| Description | The ViSi Mobile Monitoring System is a lightweight, body-worn vital signs<br>monitor featuring a high resolution, full color touch screen display, with visual<br>and audible alarms and alerts. The ViSi Mobile Monitor is designed to<br>continuously non-invasively measure ECG, basic arrhythmias [including<br>ventricular tachycardia, ventricular fibrillation, asystole, and atrial<br>fibrillation/atrial flutter], heart rate, SpO2, blood pressure, pulse rate, respiration<br>rate, and temperature. The ECG, Sp02, and Respiration waveforms are viewable<br>on demand. The ViSi Mobile Monitoring System is capable of one-time and<br>continuous NIBP measurements. |
| 510(k) Number | K142827 |
| Indications for use | The ViSi Mobile Monitoring System is intended for use by clinicians and<br>medically qualified personnel for single or multi-parameter vital signs monitoring<br>of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire),<br>respiration rate, heart rate, noninvasive blood pressure (NIBP), continuous non-<br>invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen<br>saturation of arterial hemoglobin (SpO2), pulse rate, skin temperature, basic<br>arrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole,<br>Atrial Fibrillation/Atrial Flutter) and alarm in hospital-based facilities; including<br>general medical surgical floors, intermediate care floors, and emergency |
departments.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Sotera Wireless. The logo consists of the word "Sotera" in a bold, black font, with the word "WIRELESS" in a smaller, light blue font underneath. Above the word "Sotera" is a light blue graphic that resembles a heart with radio waves emanating from it.
Continuous non-invasive blood pressure (cNIBP) measurements have not been evaluated on patients during ambulation.
The basic arrhythmia analysis feature is intended for use on patients 18 years of age and older. It has not been evaluated on pediatric patients or neonates.
The arrhythmia analysis feature is intended for use by healthcare professionals trained in the identification and treatment of arrhythmia events. Automated arrhythmia analysis is an adjunct to clinical assessment; clinician review of the analysis should precede any therapeutic intervention.
The ViSi Mobile Monitoring System may be used as standalone devices or networked to a ViSi Mobile Remote Viewer through wireless 802.11 communication.
Summary of The basic arrhythmia analysis is based on the same technology as the currently substantial equivalence cleared Monebo Technologies, Inc Arrhythmia library (K062282), which has been adapted for real time analysis in the ViSi System. The ViSi Monitoring System's basic arrhythmia analysis has been validated by comparison to the AHA, MIT-BIH, CU, and NST databases as prescribed in ANSI/AAMI EC57: 2012. The QRS detection sensitivity of the ViSi System was comparable to the predicate device.
> The safety and effectiveness of the design elements implemented into the ViSi System have been confirmed by compliance to Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm (October 28, 2003).
> The ViSi System's basic arrhythmia analysis and alarms incorporates two additional rhythm classifications, artifact and unclassified, to minimize false alarms. Artifact detection prevents noisy signals from being detected as beats, thus excluding those signals from rhythm detection. In this way, false positive ventricular fibrillation and asystole events are minimized. If the artifact condition persists for >30 seconds, an equipment technical alarm is triggered. If the algorithm cannot classify the current rhythm, the rhythm is labeled as unclassified. If the condition persists for >30 seconds, an equipment technical alarm is triggered. These differences do not raise new questions of either safety or effectiveness.
## Non-Clinical Testing
The ViSi System has successfully undergone database and functional testing to demonstrate equivalence to the predicate devices. Verification and validation testing was identified to support the safety and effective of the ViSi System and were relied upon in the determination of substantial equivalence. These nonclinical tests included:
- . Performance against databases for arrhythmia analysis as specified in ANSI/AAMI EC57
- Electrical safety testing per AAMI/ANSI ES60601-1:2005 ●
- . Electromagnetic compatibility testing per IEC 60601-1-2 3rd Edition including Radiated Emissions, Electrostatic Discharge, Radio Frequency
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Sotera Wireless. The logo features the word "Sotera" in a bold, sans-serif font, with a light blue line underneath. Above the word "Sotera" is a light blue heart-shaped icon with three curved lines emanating from it, resembling radio waves. Below the line and the word "Sotera" is the word "WIRELESS" in a smaller, sans-serif font.
Electromagnetic Field Amplitude Modulated, Radio Frequency Common Mode, and Power Frequency Magnetic Field
- Alarm system testing per AAMI/ANSI/IEC 60601-1-8:2006
- . ECG performance according to AAMI/ANSI/IEC 60601-2-27:2011
The ViSi System has been tested and found to comply with the Special Controls Guidance, meet ANSI/AAMI EC57 performance requirements, and comply with recognized consensus standards for medical devices providing adequate support of substantial equivalence to the predicates devices. The results of all the testing demonstrate that the ViSi System is as safe, as effective, and performs as well as or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.