K142769 · Icare Newlife Technologies, Inc. · HEK · Mar 20, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K142769
Device Name
Fetal Doppler
Applicant
Icare Newlife Technologies, Inc.
Product Code
HEK · Obstetrics/Gynecology
Decision Date
Mar 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Indications for Use
The iCareNewlife Fetal Doppler is indicated for monitoring fetal heart rate during the antepartum period as a general indication of fetal well-being. The device is for heart sounds through the speaker only. The device is intended to be used by health care professionals including registered nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices.
Device Story
Hand-held ultrasonic fetal heart rate detector; uses piezoelectric ceramic transducers to transmit/receive ultrasound waves; detects Doppler frequency shifts caused by fetal cardiac valve vibration; converts echoes into voltage signals; demodulates Doppler signal to produce audible rhythmical heartbeat sounds via built-in speaker. Operated by healthcare professionals in hospitals, clinics, or private offices. Powered by 9V alkaline battery. Provides real-time auditory feedback of fetal heart activity to assist clinicians in assessing fetal well-being.
Clinical Evidence
Bench testing only. Acoustic output levels measured and verified against FDA guidance. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1. Electrical safety and EMC testing performed per IEC 60601-1, IEC 60601-2-37, and IEC 60601-1-2.
Indicated for monitoring fetal heart rate during the antepartum period (typically >16 weeks gestation) as a general indication of fetal well-being. Intended for use by healthcare professionals (nurses, midwives, ultrasound technicians, physician assistants) in clinical settings.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three interconnected human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 20, 2015
iCare Newlife Technologies, Inc. % Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Consulting Co., Ltd 7th Floor, Jingui Business Building, No. 982 Congyun Rd. Baiyun District, Guangdong 510240 China
Re: K142769
Trade/Device Name: Fetal Doppler Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: HEK Dated: February 17, 2015 Received: February 18, 2015
Dear Mike Gu.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K142769
Device Name Fetal Doppler
Indications for Use (Describe)
The iCareNewlife Fetal Doppler is indicated for monitoring fetal heart rate during the antepartum period as a general indication of fetal well-being. The device is for heart sounds through the speaker only. The device is intended to be used by health care professionals including registered nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Diagnostic Ultrasound Indications for Use
# Fetal Doppler (K142769)
Intended Use: Detect fetal heart rate as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | N | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
*Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{5}------------------------------------------------
In accordance with 21 CFR 807.92 the following summary of information is provided:
| I. SUBMITTER | iCare Newlife Technologies, Inc. |
|---------------------------|-------------------------------------------------------------------------------|
| | Wangjing Science Pioneer Park, Suite E-510A, Chaoyang, Beijing, P.R.C. 100102 |
| Phone: | +86-10-8414-8058 |
| Fax: | +86-10-8414-8058 |
| Primary Contact Person: | Mike Gu |
| | Regulatory Affairs Manager |
| | OSMUNDA Medical Device Consulting Co., Ltd |
| | Tel: (+86) 20-6232 1333 |
| | Fax: (+86) 20-8633 0253 |
| Secondary Contact Person: | Mr. Bo Xiao |
| | Vice President |
| | iCare Newlife Technologies, Inc. |
| Date Prepared: | 17 March, 2015 |
| II. DEVICE | |
| Name of Device: | iCareNewlife Fetal Doppler |
| Common/Usual Name: | Fetal Monitor |
| Classification Names: | Fetal ultrasonic monitor and accessories (21 CFR 884.2660) |
| Regulation Class: | II |
| Product Code: | HEK |
| III. PREDICATE DEVICE | |
Fetal Doppler JPD-100S, K110124.
This predicate has not been subject to a design-related recall.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Newlife iCare. The logo is white text on a pink background. The word "Newlife" is on the top right of the logo, and the word "iCare" is in a larger, cursive font on the bottom left.
#### IV. INDICATION FOR USE
The iCareNewlife Fetal Doppler is indicated for monitoring fetal heart rate during the antepartum period as a general indication of fetal well-being. The device is for hearing fetal heart sounds through the speaker only. The device is intended to be used by health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physician ssistants, by prescription from licensed physicians in hospitals, clinics and private offices.
The Indications for Use statement for the iCareNewlife Fetal Doppler is not identical to the predicate device; however, the differences do not alter the intended diagnostic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for the detecting of fetal heart beat.
#### V. DEVICE DESCRIPTION
The device is a hand-held ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate (FHR). The built-in speaker of the device allows for listening of the fetal heartbeat. The device is generally applied to above 16 weeks gestation.
Ultrasonic wave is transmitted from one piezoelectric ceramic at the front of the uterus of the pregnant women. Echo is received by the other piezoelectric ceramic at the front of the probe when ultrasonic wave reaches the fetal heart. Then it is converted into voltage. This Doppler signal is detected and demodulated from the received signal. And the Doppler frequency is consistent with the rhythm of the fetal systole and diastole. Once cardiac valves vibrate and a Doppler frequency excursion is formed. lt is transmitted an output signal of cardiac valves vibrating, and it is sent to the loudspeaker for getting a rhythmical sound with the fetal heartbeat.
The device is powered by a 9V alkaline battery.
#### VI. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Bench testing
The device was evaluated for levels of acoustic output.
### Biocompatibility testing:
The biocompatibility evaluation for the iCare Newlife Fetal Doppler was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Newlife iCare. The logo is white text on a pink background. The word "Newlife" is above the word "iCare".
## Traditional 510(k) Submission Fetal Doppler
Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- . Cytotoxicity
- Sensitization
- . Irritation
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the iCare Newlife Fetal Doppler. The device complies with the IEC 60601-1, IEC 60601-2-37 standards for safety and the IEC 60601-1-2 standard for EMC.
| Descriptor | Subject device | Fetal Doppler JPD-100S (K110124) |
|------------------------------------------------------|---------------------------------------------|----------------------------------|
| Intended use | detect FHR | detect FHR |
| Sensitivity | 16 weeks gestation | 12 weeks gestation |
| Mode of operation | Continuous wave | Continuous wave |
| Ultrasound<br>frequency | 2.5 MHz | 3.0 MHz |
| Ultrasonic power | $\leq$ 22 mW | 13.3 mW |
| Area corresponding<br>to entrance beam<br>dimensions | 1.57 cm² | 2.65 cm² |
| Acoustic output | 14.01 mW/cm² (actual max<br>value measured) | 5.02 mW/cm² |
| FHR measuring<br>range | 50~210 bpm | 50~210 bpm |
| FHR resolution | 1 bpm | 1 bpm |
| Accuracy | $\pm$ 2 bpm | $\pm$ 2 bpm |
| FHR readout | Speaker sound | Speaker sound |
| Power supply | 9V alkaline battery | 9V alkaline battery |
| Transducer housing<br>material | ABS plastic | ABS plastic |
#### VII. Predicate Device Comparison
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Newlife iCare. The logo is white text on a pink background. The word "Newlife" is on the top right of the logo, and the word "iCare" is in a larger font on the bottom left.
### Substantial Equivalence Discussion
- Intended use The subject and predicate devices have the same intended use to detect the ● FHR. The predicate device can detect the FHR from 12 weeks gestation whereas the subject device can detect the FHR from 16 weeks gestation. The difference does not raise any effectiveness concerns, because the subject device is labeled for detecting the FHR during the antepartum period.
- Technological characteristics - Similarity in design
The subject and predicate devices have the same fundamental design. They have the following same or comparable technological characteristics:
- ★ Mode of operation
- * Ultrasound frequency
- * FHR measuring range
- * FHR resolution
- * Accuracy
- ★ FHR readout
- * Power supply
- Technological characteristics Difference in design ●
The subject device has higher ultrasonic power and smaller area corresponding to entrance beam dimensions, resulting in higher acoustic output. The differences do not raise safety concerns, because the statistical maximum acoustic output of the subject device meets the requirement (≤20 mW/cm² for unfocused CW FHR devices) in the FDA guidance document, "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" (issued on September 09, 2008).
- . Material – The subject and predicate device use the same type of material – ABS plastic. To ensure safety, the subject device has been evaluated for biocompatibility, and the results demonstrated the subject device is safe.
In conclusion, the subject device is substantial equivalent to the predicate device in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.