K142759 · Gyrus Acmi, Inc. · GEI · Jan 13, 2015 · General, Plastic Surgery
Device Facts
Record ID
K142759
Device Name
PK Cutting Forceps for ESG-400 Workstation
Applicant
Gyrus Acmi, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PK Cutting Forceps are indicated for electrosurgical coagulation, mechanical cutting, and grasping of tissue during the performance of laparoscopic and open general surgical procedures when used in conjunction with the general purpose ESG-400 workstation.
Device Story
Single-use, sterile bipolar electrosurgical accessory for use with ESG-400 workstation; intended for laparoscopic and open general surgery. Device inputs: manual activation via handpiece switch or footswitch. Device transforms electrical energy from workstation to distal jaws for tissue coagulation, mechanical cutting, and grasping. Modifications from predicate include redesigned handle for improved ergonomics and jaw coatings/internal mechanical bridge for performance. Operated by surgeons in OR settings. Output is physical tissue effect (coagulation/cutting) observed directly by surgeon. Benefits include effective tissue management during surgery. Device is a modification of Gyrus HALO PKS Cutting Forceps.
Clinical Evidence
No clinical testing conducted. Evidence based on non-clinical bench testing (electrical, mechanical, functional, stability) and preclinical ex-vivo evaluations using bovine and porcine tissue. Bench testing confirmed thermal margins and coagulation performance are equivalent or superior to predicate. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, pyrogen) passed.
Technological Characteristics
Single-use, sterile bipolar electrosurgical forceps. Materials: patient-contacting parts compliant with ISO 10993-1. Energy: bipolar electrosurgical energy from ESG-400 generator. Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18, IEC 60601-1-2 (EMC). Sterilization: Ethylene Oxide (ISO 11135). Connectivity: wired connection to ESG-400 workstation.
Indications for Use
Indicated for electrosurgical coagulation, mechanical cutting, and grasping of tissue during laparoscopic and open general surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Gyrus ACMI Inc. HALO PKS Cutting Forceps (K100896)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2015
Gyrus ACMI Incorporated Mr. Dolan Mills Sr. Specialist, Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772
Re: K142759
Trade/Device Name: PK Cutting Forceps for ESG-400 workstation Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 4, 2014 Received: December 8, 2014
Dear Mr. Mills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
Device Name PK Cutting Forceps for ESG-400 workstation K142759
Indications for Use (Describe)
The PK Cutting Forceps are indicated for electrosurgical coagulation, mechanical cutting, and grasping of tissue during the performance of laparoscopic and open general surgical procedures when used in conjunction with the general purpose ESG-400 workstation.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Gyrus ACMI, Inc. PK Cutting Forceps for ESG-400 Workstation
#### General Information
| Manufacturer: | Gyrus ACMI, Inc., an Olympus company<br>136 Turnpike Road<br>Southborough, MA 01772-2104<br>Phone: 1-800-262-3540<br>Fax: 1-901-373-0260 |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 3003790304 |
| Contact Person: | Dolan Mills<br>Senior Specialist, Regulatory Affairs |
| Date Prepared: | September 24, 2014 |
Device Description
| Classification Name: | Electrosurgical Cutting & Coagulation<br>Device and Accessories |
|----------------------|-------------------------------------------------------------------|
| Regulatory Class | Class 2 |
| Regulation Number | 21 CFR 878.4400 |
| Review Panel | General and Restorative Surgery Panel |
| Product Code | GEI |
| Trade Name(s): | PK Cutting Forceps, ESG PK Cutting<br>Forceps, ESG PK CF |
| Generic/Common Name: | Electrosurgical cutting and coagulation<br>device and accessories |
#### Predicate Device
Gyrus ACMI Inc. HALO PKS Cutting Forceps: K100896 This predicate has not been subject to a design-related recall.
#### Product Description
The PK Cutting Forceps are indicated for electrosurgical coagulation, mechanical cutting, and grasping of tissue during the performance of laparoscopic and open general surgical procedures when used in conjunction with the general purpose ESG-400 workstation. The device is a single use, sterile accessory, to be used in
{4}------------------------------------------------
conjunction with the bipolar outputs of the Olympus workstation with PK software (ESG-400).
The proposed device is a modification of the Gyrus HALO PKS Cutting Forceps and includes an improved handle and improved jaw performance. The device handle has been redesigned with internal changes to improve ease of use. The jaws are identical as those found on the predicate Gyrus HALO PKS Cutting Forceps except that coatings have been added, and a mechanical bridge has been added inside the shaft.
#### Technological Characteristics
The Gyrus ACMI PK Cutting Forceps is a single use, sterile accessory for use with the ESG-400 Generator with updated PK software only, and is intended to provide the same performance as the predicate Gyrus HALO PKS Cutting Forceps. The modified device and the currently marketed device share intended use and the same operating principle, similar patient contacting materials, and the same processes of packaging and sterilization. Other than the coating, the distal jaws are identical as those found on the predicate Gyrus HALO PKS Cutting Forceps. The device may be activated via the handpiece coagulation switch or footswitch.
# Material
The device uses similar patient-contacting materials that are utilized in the predicate devices, as well as other legally marketed devices manufactured by Gyrus ACMI.
The patient contacting items are classified in accordance with ISO 10993-1, as an external communicating, tissue/bone/dentin device for limited exposure (<24hrs.). ISO10993-1 and FDA Blue Book memo #G95-1 guidelines recommend that these direct patient contact parts have supporting data for cytotoxicity, sensitization and irritation. Full biocompatibility testing to ISO10993-1 for the device category was completed and is available for patient contacting materials.
The battery of testing included the following tests:
- Cytotoxicity
- Sensitization ●
- Irritation ●
- Pyrogen Testing ●
Based on the material assessment, patient contacting materials were tested in accordance with ISO 10993-1, and results are considered passing.
{5}------------------------------------------------
### Intended Use / Indications for Use
The PK Cutting Forceps are indicated for electrosurgical coagulation, mechanical cutting, and grasping of tissue during the performance of laparoscopic and open general surgical procedures when used in conjunction with the general purpose ESG-400 workstation.
## Compliance to Standards
The design of the device complies with the following standards:
For pro-code GEI - FDA Recognized Consensus Standards: IEC 60601-2-2
Additional standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18 ISO 10993-1, 5, 7, 10, Biological Evaluation of Medical Devices ISO 14971, Risk Analysis ISO 15223-1:2012, Medical Devices - Symbols to be used ISO 11135:2014, Sterilization of Health Care Products, EO Validation ISO 11138: 2006, Sterilization of health care products: Biological Indicators ISO 11607-1:2006, Packaging for Terminally Sterilized Medical Devices ISO 11607-2:2006, Packaging for Terminally Sterilized Medical Devices ISO 11737-1:2006, Sterilization of Medical Devices - Microbiological Methods ISO 11737-2:2009, Sterilization of Medical Devices - Microbiological Methods
Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971. The design verification tests were identified and performed as a result of risk analysis assessment.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the device complies with the applicable clauses of IEC 60601-1, IEC 60601-2-2, and IEC 60601-2-18 standards for safety, and the IEC 60601-1-2 standard for EMC.
#### Summary of Sterilization and Shelf Life Discussion
The device is provided sterile for single-use. They are sterilized using Ethylene Oxide, using a cycle validated in accordance with ISO 11135 to provide a sterility assurance level of 10°.
{6}------------------------------------------------
The Shelf Life period for the device was determined through an analysis of the shelflife stability of the materials used in the design of the device, as well as an analysis of the packaging materials and processes used with other Gyrus ACMI devices. Accelerated shelf-life studies were conducted to support an initial one year shelf life, with real time testing in process to confirm an initial one vear expiration date. Additional studies (accelerated and real time) are planned to support a three-year expiration date in the future.
# Summary of Performance Testing
Performance tests were executed to ensure that the device functions as intended and meets design specifications.
Evidence of safety and effectiveness was obtained from two primary areas:
- 1) non-clinical (electrical, mechanical, functional, stability) performance testing
- 2) preclinical (bench, ex-vivo tissue) evaluations and testing
Non-clinical: Basic safety and performance testing was performed in accordance with IEC 60601-1 and IEC 60601-2-2. In addition, verification and comparison bench studies were conducted to evaluate the mechanical and functional performance. Testing included: system testing, torque strength, endurance, force testing, reliability, ship testing, baseline performance testing, age testing, environmental conditioning, durability, dimensional verification, ergonomics, system compatibility, grasping, thermal margin, cutting, rotation, coagulation, and basic functionality.
Stability: Samples were subjected to accelerated aging to confirm that the device maintains functionality and continues to meet specifications over time. The results of the accelerated age testing demonstrate that the device will be stable for the stated shelf-life. In addition, real time age testing will be carried out to confirm the results of the accelerated age testing. Samples were also subjected to environmental conditioning and ship testing.
Preclinical: Evidence obtained from preclinical bench tissue (ex vivo) demonstrates that the device performs substantially equivalent to the predicate devices in relevant aspects associated with usability, cutting, coagulation, and tissue grasping. For bench testing, the selected tissue medium was confirmed to be appropriate for testing based on clinical experience with the predicate.
Bench tissue - evaluated ex vivo using bovine and porcine tissue:
- Thermal margin -
- । Visual comparison of coagulation
Testing demonstrated that the device performs as well as or better than the predicate device. Non-clinical bench testing and preclinical performance testing shows that the
{7}------------------------------------------------
device performance is substantially equivalent to the predicate, and that the modifications raise no new issues of safety and effectiveness.
No clinical testing was conducted. The use of Electrosurgical Cutting and Coagulation Devices has been documented in published literature and indicates safe and effective use for the target indications and expected patient populations.
## Substantial Equivalence and Conclusion
The device utilizes similar bipolar electrosurgical energy to cut and coagulate tissue as that used in the predicate device.
The device uses similar patient-contacting materials in similar quantities that are utilized in the predicate device, as well as other legally marketed devices manufactured by Gyrus ACMI and others.
The performance of the device was compared against the known performance characteristics of the predicate device. Testing demonstrated that the performance requirements were met, and that the PK Cutting Forceps exhibited comparable performance characteristics to the predicates.
In summary, the PK Cutting Forceps is substantially equivalent to the predicate device and presents no new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.