LiNA PowerBlade Plus HC

K142757 · Lina Medical Aps · GEI · Oct 23, 2014 · General, Plastic Surgery

Device Facts

Record IDK142757
Device NameLiNA PowerBlade Plus HC
ApplicantLina Medical Aps
Product CodeGEI · General, Plastic Surgery
Decision DateOct 23, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The LiNA PowerBlade Plus HC is intended for use in open and laparoscopic surgery where grasping, coagulating and transecting is indicated.

Device Story

LiNA PowerBlade Plus HC is a sterile, single-use 5mm bipolar forceps instrument for open and laparoscopic surgery. Device features 330mm length and 10mm jaw opening. Operation involves grasping tissue, applying bipolar electrical current for coagulation, and using a mechanical sharp blade for transection. Surgeon operates device via handle-integrated rotation wheel for jaw positioning. Coagulation activation occurs via hand-piece button (when using 3-pin connector with ValleyLab/LigaSure generators) or foot pedal (when using 2-pin connector with standard electrosurgical generators). Device provides surgeons with precise tissue manipulation and hemostasis control, potentially reducing procedure time and improving surgical outcomes.

Clinical Evidence

No clinical testing performed. Substantial equivalence supported by descriptive information and bench verification testing confirming functionality and compliance with recognized standards (IEC 60601-1, IEC 60601-2-2).

Technological Characteristics

Bipolar electrosurgical forceps; 5mm diameter, 330mm length, 10mm jaw opening. Single-use, ethylene oxide sterilized. Compatible with standard electrosurgical generators (bipolar outlet) via 2-pin connector or ValleyLab generators (LigaSure outlet) via 3-pin connector. Features mechanical sharp blade for transection and rotation wheel for jaw positioning. Complies with IEC 60601-1 and IEC 60601-2-2.

Indications for Use

Indicated for use in open and laparoscopic surgery for grasping, coagulating, and transecting tissue.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is presented in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 23, 2014 LINA Medical ApS % Ms. Christine E. Nichols RAC Boston Biomedical Associates 100 Crowley Drive, Suite 216 Marlborough, Massachusetts 01752 Re: K142757 Trade/Device Name: LiNA PowerBlade Plus™ HC Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 24, 2014 Received: September 25, 2014 Dear Ms. Nichols: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K142757 Device Name LiNA PowerBlade Plus™ HC Indications for Use (Describe) The LiNA PowerBlade Plus™ HC is intended for use in open and laparoscopic surgery where grasping, coagulating and transecting is indicated Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 5 510(k) Summary #### 1. Submission Sponsor LINA Medical ApS Formervangen 5 2600 Glostrup Denmark Phone: +45 43 29 66 66 Fax: -45 43 29 66 99 Contact: Louisa Memborg, Regulatory Affairs Officer ### Submission Correspondent Christine E. Nichols RAC Boston Biomedical Associates 100 Crowley Dr., Suite 216 Marlborough, MA 01752 ### Date Prepared September 24, 2014 ### 2. Device Identification Trade/Proprietary Name: Common/Usual Name: Classification Name: Classification Regulation: Product Code: Device Class: Classification Panel: ### 3. Predicate Device LiNA PowerBlade Plus™ K132837 ### 4. Device Description The PowerBlade Plus HC is a sterile 5 mm single use instrument available in one 330mm length version with a 10mm jaw opening. The proposed device is a bipolar forceps device that grasps and coagulates utilizing electrical current. Transecting is done with a mechanical sharp LiNA PowerBlade Plus™ HC electrosurgical, cutting & coagulation & accessories Electrosurgical cutting and coagulation device and accessories 21 CFR 878.4400 GEI Class II General & Plastic Surgery {4}------------------------------------------------ blade. The LiNA PowerBlade Plus HC includes a rotation wheel on the handle that rotates the grasper jaws at the tip to improve positioning and ergonomics during the surgical procedure. The device is single use ethylene oxide sterilized and is compatible with most standard electrosurgical generators that provide a bipolar outlet. Two types of connectors are provided: the two pin connector is for use with most standard electrosurgical generators with a bipolar outlet, and a 3-pin connector that is attached to the cable of the LiNA PowerBlade Plus HC for use with ValleyLab™ generators with a LigaSure™ outlet. When the 3-pin connector is used, the control button on the hand piece can be used to activate the coagulation. When using the 2pin connector the coagulation function of the PowerBlade Plus HC is foot pedal controlled. #### 5. Intended Use The LiNA PowerBlade Plus HC is intended for use in open and laparoscopic surgery where grasping, coagulating and transecting is indicated. #### 6. Comparison of Technological Characteristics This special 510(k) is a modification to the LiNA PowerBlade Plus previously cleared by the FDA via 510(k) K132837. No changes were made to the intended use, indications for use, energy type, materials, sterilization method or fundamental scientific technology. The LiNA PowerBlade Plus HC has the following differences from the predicate: Added hand control activation button for coagulation when device is used with the ValleyLab Generators with a LigaSure outlet, and the addition of a 3 pin connector adapter to allow connection with the ValleyLab Generators that have the LigaSure outlet. The three pin connector adapter can be easily removed from the cable so that the cable is compatible with most ESUs. #### Non-Clinical Performance Data Verification testing was performed as part of design controls to verify functionality of the proposed device and compliance with the same recognized standards as the predicate (IEC60601-1, and IEC60601-2-2). #### Clinical Testing There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The descriptive information detailed in this submission supports the substantial equivalence of the device. ### 7. Conclusion Statement of Substantial Equivalence The differences between the LiNA PowerBlade Plus HC and the predicate LiNA PowerBlade Plus do not raise any new questions regarding its safety and effectiveness. Verification testing and compliance with voluntary recognized standards, demonstrate that the LiNA PowerBlade Plus is substantially equivalent to the predicate device LiNA PowerBlade Plus in terms of design, components, principals of operation, sterilization, performance characteristics, and intended use. The LiNA PowerBlade Plus, as designed is determined to be substantially equivalent to the referenced predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...