K142692 · Embolx, Inc. · MJN · May 31, 2015 · Cardiovascular
Device Facts
Record ID
K142692
Device Name
Occlusion Balloon Catheter
Applicant
Embolx, Inc.
Product Code
MJN · Cardiovascular
Decision Date
May 31, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Embolx Occlusion Balloon Catheter is intended for use in the peripheral vasculature where temporary occlusion is desired and offers a vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Embolx Occlusion Catheter is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.
Device Story
Coaxial dual-lumen catheter; distal occlusion balloon; two embedded radiopaque bands for fluoroscopic positioning. Proximal hub features two ports: one for guidewire/fluid delivery, one for balloon inflation/deflation via handheld syringe. Inserted through femoral artery via guiding catheter; used in peripheral vasculature. Physician operates device under fluoroscopic guidance to select target vessels; balloon conforms to vessel wall to stop/control blood flow. Enables targeted delivery of contrast media or therapeutic agents. Benefits patient by allowing localized treatment or flow control in peripheral vessels. Single-use; EtO sterilized.
Clinical Evidence
Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity), sterilization/packaging validation, and mechanical performance (dimensional verification, balloon fatigue/compliance/burst, inflation/deflation time, bond/tip pull strength, flexibility, torque, radiopacity, coating integrity, and flow rates).
Technological Characteristics
Coaxial dual-lumen catheter; 2.9F proximal/2.2F distal diameter; 110cm length. Compliant balloon (up to 5mm diameter, 11mm length). Radiopaque markers. Infusion pressure up to 900 psi. Compatible with guidewires ≤ 0.016". EtO sterilization. No software/electronics.
Indications for Use
Indicated for temporary occlusion of peripheral vasculature vessels (up to 5mm diameter) to control blood flow and to assist in delivery of diagnostic/therapeutic agents.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black. The logo is simple and recognizable, and it is often used on official documents and websites related to the Department of Health & Human Services.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 31, 2015
Embolx, Inc. % Craig Coombs Coombs Medical Device Consulting, Inc. 1193 Sherman Street Alameda, California 94501
Re: K142692
> Trade/Device Name: Embolx Occlusion Balloon Catheter Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: MJN Dated: May 28, 2015 Received: May 29, 2015
Dear Craig Coombs,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
Page 2 - Craig Coombs
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142692
Device Name Embolx Occlusion Balloon Catheter
### Indications for Use (Describe)
The Embolx Occlusion Balloon Catheter is intended for use in the peripheral vasculature where temporary occlusion is desired and offers a vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Embolx Occlusion Catheter is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "EMBOLX" in a bold, sans-serif font. To the left of the word is a circular design made up of many different sized circles. The circles are arranged in a spiral pattern, with the larger circles on the outside and the smaller circles on the inside.
# Section 5: 510(k) Summary
## A. Device Information:
| Category | Comments |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Embolx, Inc.<br>5760 Arboretum Dr., Los Altos, CA 94024<br>Tel: 408.888.7792<br>Fax: 650.968.4069<br>Contact: Michael Allen, CEO & President<br>Email: mallen@embolx.com |
| Correspondent Contact<br>Information: | Craig Coombs<br>Coombs Medical Device Consulting<br>1193 Sherman Street<br>Alameda, CA 94501<br>Tel: 510-337-0140<br>Fax: 510-337-0416<br>Email: CraigJCoombs@Gmail.com |
| Device Common Name: | Catheter, Intravascular Occluding, Temporary |
| Device Classification Number<br>& Name: | 21 CFR 870.4450<br>Vascular Clamp |
| Device Classification &<br>Product Code: | Class II,<br>MJN |
| Device Proprietary Name: | Embolx Occlusion Balloon Catheter |
### Predicate Device Information:
| Predicate Device: | Micrus™ Ascent™ Balloon Catheter |
|----------------------------------------------------|--------------------------------------------------------------------|
| Predicate Device Manufacturer: | Codman Neurovascular (a division of Johnson & Johnson Medical Ltd) |
| Predicate Device Common Name: | Catheter, Intravascular Occluding, Temporary |
| Predicate Device Premarket Notification # | K103780 |
| Predicate Device Classification: | 21 CFR 870.4450 |
| Predicate Device Classification &<br>Product Code: | Class II,<br>MJN |
## B. Date Summary Prepared
29 May 2015
## C. Description of Device
The Embolx Occlusion Balloon Catheter is a coaxial dual lumen device that consists of an occlusion balloon at the distal end with two embedded radiopaque bands to allow for visualization and positioning of the device under fluoroscopic guidance. The proximal hub consists of two ports: one port for use by the guidewire and delivery of fluids and the second port for inflation and deflation of the balloon. The low profile balloon is manufactured of a compliant material that allows ease of insertion and withdrawal from the vasculature and conforms to the vessel wall. The balloon is inflated and deflated with a hand held syringe. The device is supplied sterile by EtO and is intended for single use.
The occlusion catheter has an outside diameter of 2.9F proximally and 2.2F distally. The
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "EMBOLX" in a bold, sans-serif font. To the left of the word is a design made up of many circles of different sizes. The circles are arranged in a spiral pattern, with the larger circles on the outside and the smaller circles on the inside.
occlusion balloon on the distal end can be inflated up to 5mm in diameter and 11mm in length. The usable length of the device is 110cm. The device can withstand an infusion pressure up to 900 psi.
# D. Indications for Use
The Embolx Occlusion Balloon Catheter is intended for use in the blood vessels of the peripheral vasculature where temporary occlusion is desired and offers a vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Embolx Occlusion Balloon Catheter is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.
## E. Comparison to Predicate Device
The Embolx Occlusion Balloon Catheter is substantially equivalent in intended use, indications for use, technology, design, performance and materials to the predicate device, the Micrus™ Ascent™ Occlusion Balloon Catheter (K103780).
Both devices provide the ability to select specific blood vessels in the peripheral vasculature (of up to 5 mm in diameter) in which temporary control or complete occlusion of blood flow is desired, as well as provide the ability to deliver fluids to the target vessels such as diagnostic and therapeutic agents. The predicate device also has 5mm diameter balloons in its catalogue.
The predicate device is indicated for peripheral and neurovascular use, whereas the application device is only indicated for peripheral use. Since the indication for use of the application device is a subset of the predicate device, the differences do not raise new questions of safety or efficacy.
Both devices are coaxial dual lumen catheters with distal end balloons that are inserted through a guiding catheter via the femoral artery; both devices use guidewires for target vessel selection and have radiopaque markers for visualization under fluoroscopy.
The Embolx Occlusion Balloon Catheter has a working length of 110 cm, whereas the predicate device has a working length of 150 cm due to its indications for use in neurovascular applications as well as peripheral vascular applications. Both devices are single use only and ethylene oxide sterilized.
Both devices are nearly the same outside diameter. The predicate catheter is 2.9Fr in outside diameter and the application device ranges from 2.9Fr on the proximal end to 2.2Fr on the distal end. The predicate device can accommodate guidewires that are 0.014" in diameter or smaller, whereas the application device can accommodate guidewires that are 0.016" or smaller in diameter.
Other than working length of the catheter, all other significant aspects of the Embolx Occlusion Balloon Catheter and the predicate device are comparable. The testing described below demonstrates that the minor differences in the devices do not raise any unresolved issues of safety or efficacy.
Embolx concludes that the application and predicate devices are substantially equivalent.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "EMBOLX" in a bold, sans-serif font. To the left of the word is a circular design made up of many small circles. The circles are arranged in a spiral pattern, with the smallest circles in the center and the largest circles on the outside.
### F. Summary of Supporting Data
Where appropriate, all testing was conducted on sterilized and aged test articles. Substantial equivalence was demonstrated through the following non-clinical testing:
- Cytotoxicity .
- Sensitization
- . Irritation
- Acute Systemic Toxicity
- Complement Activation (C3a and Sc5b-9)
- Hemocompatibility ●
- ASTM Hemolysis (direct and indirect) o
- In-vivo Thrombogenicity (canine) o
- Partial Thromboplastin Time (PTT) O
- o Prothrombin Time Assay (PT)
- Pyrogenicity (LAL)
- . Material-mediated Pyrogencity
- Appropriate Sterilization and Packaging Testing .
- Sterilization validation O
- Packaging seal integrity O
- o Dye penetration
- o Transit testing
- . Dimensional Verification
- Balloon Prep, Deployment, & Retraction (simulated use)
- Balloon Rated Burst Volume ●
- Balloon Fatigue
- . Balloon Compliance
- . Balloon Inflation/Deflation Time
- . Balloon Position Test
- . Catheter Bond Strength (tensile testing)
- . Tip Pull Strength (tensile testing)
- . Flexibility & Kink Test
- . Torque Strength
- . Radiopacity
- Coating Integrity
- . Catheter Body Burst Test
- Catheter body Leakage test .
- Contrast Media Flow Rate
- . Corrosion Resistance
Embolx concludes that these nonclinical tests demonstrate that the Embolx Occlusion Balloon Catheter is as safe, as effective, and performs as well as or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.