NITRILE EXAMINATION GLOVES WITH ALOE & VITAMIN E, POWDER- FREE
K142685 · Siam Sempermed Corp., Ltd. · LZA · Feb 11, 2015 · General Hospital
Device Facts
Record ID
K142685
Device Name
NITRILE EXAMINATION GLOVES WITH ALOE & VITAMIN E, POWDER- FREE
Applicant
Siam Sempermed Corp., Ltd.
Product Code
LZA · General Hospital
Decision Date
Feb 11, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Non-sterile, Powder free Nitrile Examination Glove with Polymer Coating, Aloe, Vitamin E, White and Pink is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Non-sterile, powder-free nitrile examination glove; synthetic nitrile latex construction; polymer inner coating containing aloe and Vitamin E. Ambidextrous design; available in white or pink. Used by clinicians in medical settings to provide a barrier against contamination between patient and examiner. Device functions as a physical protective garment covering hand and wrist. No electronic or software components.
Clinical Evidence
Bench testing only. Device meets ASTM D6319-10 for physical properties (tensile strength, elongation, dimensions), ASTM D5151-06 for freedom from pinholes (AQL 2.5), and ASTM D6124-06 for residual powder. Biocompatibility testing (ISO 10993-10) confirmed the material is a non-irritant and non-sensitizer.
Technological Characteristics
Synthetic nitrile latex; polymer coating with aloe and Vitamin E. Meets ASTM D6319-10 (physical properties), ASTM D5151-06 (pinholes), and ASTM D6124-06 (powder). Ambidextrous, single-use, non-sterile. Biocompatibility per ISO 10993-10.
Indications for Use
Indicated for use as a protective medical glove worn on the examiner's hand to prevent contamination between patient and examiner. Suitable for general medical examination use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Nitrile Examination Gloves with Aloe & Vitamin E, Powder-free (K121549)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest a sense of community and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 11, 2015
Siam Sempermed Corp., LTD c/o John Calhoun Manager of Regulatory Affairs Sempermed USA, Inc. 13900 49th Street North Clearwater, Florida 33762
Re: K142685
> Trade/Device Name: Non-Sterile, Powder Free Nitrile Examination Gloves with Polymer Coating, Aloe, Vitamin E, White and Pink Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: January 9, 2015 Received: January 12, 2015
Dear Mr. Calhoun,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# ATTACHMENT 6
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K142685
Device Name
Non-sterile, Powder free Nitrile Examination Glove with Polymer Coating, Aloe, Vitamin E. White and Pink
Indications for Use (Describe)
This is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## ATTACHMENT 5
## 510 (k) Summary
#### 1.0 Submitter
| Company name: | Siam Sempermed Corp., Ltd |
|-----------------|-------------------------------------------------------------|
| Address: | 10 Soi 10 Phetkasem Rd. Hatyai<br>Songkhla. Thailand 90110 |
| Phone: | (+66) 74 344 663 |
| Fax: | (+66) 74 344 677 |
| Contact Person: | Mr. Anan Pruksanusak (Managing Director) |
### 2.0 Official Correspondent
| Company name: | Sempermed USA Inc. |
|-----------------------|-------------------------------------------------------|
| Address: | 13900 49th Street North<br>Clearwater, USA , FL 33762 |
| Phone: | 727-787-7250 |
| Fax: | 727-787-7558 |
| Contact person: | John V. Calhoun |
| 3.0 Preparation Date: | September 11, 2014 |
### 4.0 Identification of Legally Marketed Predicate Device
Trade/Proprietary Name: Nitrile Examination Gloves with Aloe & Vitamin E, Powder-free Device Description: Non-sterile, Powder-free Nitrile Examination Glove with Polymer Coating, Aloe and Vitamin E, Blue
510(k): K121549
Device name: Examination glove
{4}------------------------------------------------
Device Classification Name: Patient Examination Gloves (21 CFR 880.6250) Device Class: Class 1 Product Code: LZA Classification Panel: General Hospital (Part 880)
#### 5.0 Device description
The subject device is a patient examination glove made of a synthetic nitrile latex compound. It is non-sterile, powder-free, with a polymer inner coating of aloe and Vitamin E (per 21 CFR 880.6250, Class I). The device can be either white or pink in color. The device is ambidextrous and can be worn on either the left or right hand. The device meets ASTM D6319-10: Standard Specification for Nitrile Examination Gloves for Medical Application.
The subject device is substantially equivalent to legally marketed Nitrile Examination Gloves identified as Product Code LZA.
#### 6.0 Intended Use of Device
Non-sterile, Powder free Nitrile Examination Glove with Polymer Coating, Aloe, Vitamin E, White and Pink is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
#### 7.0 Summary of Technological Characteristics of Device and Comparison to Legally Marketed Predicate Device
Non-sterile, Powder free Nitrile Examination Glove with Polymer Coating, Aloe, Vitamin E, White and Pink is substantially equivalent to the predicate device, which differs only by color. A side-by side comparison of the predicate device with the subject device is presented in Table 1 on the following pages.
{5}------------------------------------------------
### Table 1. Side-by-side Comparison of Predicate Device with Subject Device
| | | Predicate: K121549 | Subject device (K142685) |
|--------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | Acceptance<br>Criteria/Standards | Non-Sterile, Powder-free<br>Nitrile Examination Glove with<br>Polymer Coating. Aloe and<br>Vitamin E, Blue | Non-Sterile, Powder-free<br>Nitrile Examination Glove with<br>Polymer Coating. Aloe and<br>Vitamin E, White and Pink |
| Indications for<br>Use | Medical Glove Guidance<br>Manual | This is a medical device<br>intended for medical purposes<br>that is worn on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner | This is a medical device<br>intended for medical purposes<br>that is worn on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner |
| Device Design | - | A protective garment that<br>covers the hand and wrist with<br>openings for fingers and<br>thumb | A protective garment that<br>covers the hand and wrist with<br>openings for fingers and<br>thumb |
| Material | ASTM D6319-10 | Nitrile Synthetic latex | Nitrile Synthetic latex |
| Single Use | Medical Glove Guidance<br>Manual | Single use | Single use |
| Ambidextrous | Medical Glove Guidance<br>Manual | Ambidextrous | Ambidextrous |
| Sterility | Not applicable | Not applicable | Not applicable |
| Dimensions | ASTM D6319-10 | Meets ASTM D6319-10<br>-Overall Length:<br>220 mm min (XS, S)<br>230 mm min (M, L, XL)<br>-Width (+/- 10 mm)<br>Size XS:70 mm<br>Size S: 80 mm<br>Size M: 95 mm<br>Size L: 110 mm<br>Size XL: 120 mm<br>-Thickness:<br>Finger: 0.05 mm min<br>Palm: 0.05 mm min | Meets ASTM D6319-10<br>-Overall Length:<br>220 mm min (XS, S)<br>230 mm min (M, L, XL)<br>-Width (+/- 10 mm)<br>Size XS:70 mm<br>Size S: 80 mm<br>Size M: 95 mm<br>Size L: 110 mm<br>Size XL: 120 mm<br>-Thickness:<br>Finger: 0.05 mm min<br>Palm: 0.05 mm min |
| Physical<br>Properties | ASTM D6319-10<br>(aging= 70 +/- 2°C for 166 +/-<br>2 hrs; 100 +/- 2° C for 22 +/-<br>0.3 hrs) | Meets ASTM D6319-10<br>(aging= 70 +/- 2°C for 166 +/- 2<br>hrs; 100 +/- 2° C for 22 +/- 0.3<br>hrs)<br>-Tensile Strength:<br>14 MPa min (before aging)<br>14 MPa min (after aging)<br>-Ultimate Elongation:<br>500% min (before aging)<br>400% min (after aging) | Meets ASTM D6319-10<br>(aging= 70 +/- 2°C for 166 +/- 2<br>hrs; 100 +/- 2° C for 22 +/- 0.3<br>hrs)<br>-Tensile Strength:<br>14 MPa min (before aging)<br>14 MPa min (after aging)<br>-Ultimate Elongation:<br>500% min (before aging)<br>400% min (after aging) |
| Freedom from<br>Pinholes | ASTM D5151-06 | Meets AQL 2.5, Inspection<br>Level G-1 | Meets AQL 2.5, Inspection<br>Level G-1 |
| Residual Powder | ASTM D6124-06 | <2.0 mg/pc | <2.0 mg/pc |
| Color | - | Lavender Blue | White and Pink |
| Labeling | Medical Glove Guidance Manual | Includes English, French, and Spanish languages | Draft includes English only and lacks other minor elements pending completion of FINAL design; brand name(s) pending |
| Biocompatibility | Primary Dermal Irritation in Rabbits and Guinea Pig Closed Patch Sensitization Test (ISO 10993-10) | Under the conditions of these studies, the test article was a non-irritant and non-sensitizer | Under the conditions of these studies, the test article was a non-irritant and non-sensitizer |
{6}------------------------------------------------
#### 8.0. Conclusion:
The physical performance of the subject device (Nitrile Examination Gloves with Aloe & Vitamin E, Powder-free) is substantially equivalent to predicate K121549 and will perform according to the glove performance and biocompatibility standards referenced. Both are manufactured from the same nitrile synthetic latex material using the same production process. Only the colors differ. Based on the intended use, physical properties, and technological characteristics, the subject device is safe and effective as a legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.