K142643 · D3 Dc, LLC /Stabilynx, Inc. · MBI · Dec 15, 2014 · Orthopedic
Device Facts
Record ID
K142643
Device Name
Stabilynx PEEK Suture Anchor
Applicant
D3 Dc, LLC /Stabilynx, Inc.
Product Code
MBI · Orthopedic
Decision Date
Dec 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stabilynx Suture Anchor is a soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as: • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Capsular Shift/Capsulolabral Reconstruction, Biceps Tendonesis, Deltoid Repair • Knee: Extra Capsular Repairs (Medial Collateral Ligament, Lateral Collateral Ligament, Posterior Oblique Ligament), Illiotibial Band Tendonesis, Patellar Tendon Repair • Elbow, Wrist, Hand: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Biceps Tendon Reattachment • Foot and Ankle: Medial Instability Repair/Reconstruction, Lateral Instability Repair/Reconstruction, Achilles Tendon Repair/Reconstruction, Midfoot Reconstruction, Hallux Valgus Reconstruction
Device Story
Stabilynx PEEK Suture Anchor is an implantable bone screw system for soft tissue-to-bone fixation. Device consists of PEEK bone screw, UHMWPE sutures, and disposable insertion tool (ABS handle, stainless steel shaft). Surgeon inserts anchor into bone; unique design allows creation of contiguous suture links between adjacent anchors; pre-loaded sutures can be exchanged between anchors. Used in orthopedic reconstructive procedures (shoulder, knee, elbow, wrist, hand, foot, ankle). Benefits include enhanced surgical flexibility for complex soft tissue repairs. Operated by orthopedic surgeons in clinical/OR settings.
Clinical Evidence
Bench testing only. Verification included dimensional/visual inspections, suture loop exchange force, fixation strength (with/without cycle loading), inserter/anchor torque strength, suture knot-pull tensile strength, and repair suture bifurcated loop strength. Biocompatibility testing per ISO 10993-1 (cytotoxicity, systemic toxicity). Sterilization validated per ISO 11135-1 and ISO 10993-7. Packaging/shipping validated per ASTM F88, ASTM 2096-11, and ASTM D4169-09. All tests met acceptance criteria.
Technological Characteristics
Materials: PEEK (bone screw), UHMWPE (suture), ABS (handle), stainless steel (shaft). Principle: Mechanical bone fixation fastener. Connectivity: None. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 10993-1, ISO 11135-1, ISO 10993-7, ASTM F88, ASTM 2096-11, ASTM D4169-09.
Indications for Use
Indicated for patients requiring soft tissue-to-bone fixation during reconstructive surgery of the shoulder, knee, elbow, wrist, hand, foot, and ankle.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with flowing lines extending from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 15, 2014
Stabilynx, Incorporated Mr. Mike Rosenthal Vice President, Engineering Operations 3475-0 Edison Wav Menlo Park, California 94025
Re: K142643
Trade/Device Name: Stabilynx PEEK Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: September 15, 2014 Received: September 17, 2014
Dear Mr. Rosenthal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Department of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K142643
Device Name Stabilynx PEEK Suture Anchor
#### Indications for Use (Describe)
The Stabilynx Suture Anchor is a soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as:
• Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Capsular Shift/Capsulolabral Reconstruction, Biceps Tendonesis, Deltoid Repair
· Knee: Extra Capsular Repairs (Medial Collateral Ligament, Lateral Ligament, Posterior Oblique Ligament), Illiotibial Band Tendonesis, Patellar Tendon Repair
· Elbow, Wrist, Hand: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Biceps Tendon Reattachment
· Foot and Ankle: Medial Instability Repair/Reconstruction, Lateral Instability Repair/Reconstruction, Achilles Tendon Repair/Reconstruction, Midfoot Reconstruction, Hallux Valgus Reconstruction
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 6: 510(k) Summary (21 CFR § 807.92(c))
# I: SUBMITTER INFORMATION
| Submitter: | Stabilynx, Inc.<br>3475-0 Edison Way<br>Menlo Park, CA 94025 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mike Rosenthal<br>Vice President, Engineering Operations, Stabilynx<br>Phone: 650.503.3329<br>Fax: 650.618.1440<br>Email: mrosenthal@d3dc.com |
| Date Summary Prepared: | 15 September 2014 |
#### II: SUBJECT DEVICE INFORMATION
| Device Trade Name: | Stabilynx PEEK Suture Anchor |
|----------------------|--------------------------------------------------------------------------|
| Common Name: | Fastener, Fixation, Nondegradable, Soft Tissue |
| Classification Name: | Smooth or threaded metallic bone fixation fastener<br>(21 CFR §888.3040) |
| Product Code: | MBI |
# III: PREDICATE DEVICE INFORMATION
Stryker PEEK Zip Suture Anchor (K070758) Predicate Device:
No recalls, market withdrawals or safety alerts were identified in FDA's database for the above referenced predicate device.
No reference devices were used in this submission.
#### IV: DEVICE DESCRIPTION:
The Stabilynx PEEK Suture Anchor consists of an ethylene oxide sterilized, implantable, poly-ether-etherketone (PEEK) bone screw and ultra high molecular weight polyethylene (UHMWPE) suture anchor is supplied pre-mounted on a disposable Insertion Tool, which consists of a non-patient contact, acrylonitrile butadiene styrene (ABS) handle molded over an externally-communicating, stainless steel shaft. The anchor comes pre-loaded with two (2) high-strength, polyethylene, non-absorbable sutures. The Stabilynx PEEK Suture Anchor design allows a surgeon to create at least one contiguous suture link between one or more adjacent suture anchors after the suture anchors have been inserted into bone. The suture anchor is designed such that one or both of the pre-loaded sutures may be exchanged with one or more sutures connected to an adjacent suture anchor(s).
{4}------------------------------------------------
# V. INDICATIONS FOR USE:
The Stabilynx Suture Anchor is a soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as:
- . Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Capsular Shift/Capsulolabral Reconstruction, Biceps Tendonesis, Deltoid Repair
- . Knee: Extra Capsular Repairs (Medial Collateral Ligament, Lateral Collateral Ligament, Posterior Oblique Ligament), Illiotibial Band Tendonesis, Patellar Tendon Repair
- . Elbow, Wrist, Hand: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Biceps Tendon Reattachment
- . Foot and Ankle: Medial Instability Repair/Reconstruction, Lateral Instability Repair/Reconstruction, Achilles Tendon Repair/Reconstruction, Midfoot Reconstruction, Hallux Valgus Reconstruction
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
The primary technological principle for the subject and predicate devices is to anchor suture to bone so that the suture may be used to secure soft tissue to bone. The subject and predicate devices accomplish this function by similar means. At a high level, the subject and predicate devices are based on the following same technological elements:
- . PEEK implantable bone screw and non-absorbable sutures
- Ability to secure soft tissue to bone during reconstructive procedures
The following technological differences exist between the subject and predicate devices:
- Sutures in the subject device contain suture loops at either end
- Suture loops of the subject device enable contiguous linking of adjacent suture anchors
# VII. PERFORMANCE DATA:
The following performance data were provided in support of the substantial equivalence determination.
# Bench Testing
Verification tests included: a) Dimensional and Visual Inspections; b) Suture Loop Exchange Force; c) Fixation Strength with and without Cycle Loading; d) Inserter and Anchor Torque Strength; e) Suture Knot-Pull Tensile Strength; and, f) Repair Suture Bifurcated Loop Strength. These tests were conducted on ethylene oxide sterilized units at baseline and on devices subjected to accelerated aging conditions equivalent to three (3) months. Passing results were obtained for all design verification tests.
# Biocompatibility Testing
The biocompatibility evaluation for the Stabilynx PEEK Suture Anchor was conducted in accordance with the requirements defined in ISO 10993-1 "Biological Evaluation of Medical Devices" and FDA's guidance ("Use of International Standard ISO 10993", draft document issued on 24 April 2013 and BlueBook Memorandum #G95-1 dated 01 May 1995). Testing included the following: 1) Cytotoxicity; and, 2) Systemic Toxicity. The device is comprised of: 1) an inserter shaft that is considered a tissue/bone
{5}------------------------------------------------
contacting component for a duration of < 24 hours; and, 2) a bone anchor and suture that are considered permanent implants.
### Sterilization Data
The sterilization parameters for the Stabilynx PEEK Suture Anchor device comply with the requirements prescribed in the applicable standards for ethylene oxide sterilization (ISO 11135-1:2007 "Sterilization of health care products -- Ethylene oxide -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices" and ISO 10993-7:2008 "Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization residuals"). The sterilization cycle ensures a SAL of 10 °. The EO and ECH residuals were below the limits specified in the standard.
#### Packaging and Shipping Validation
Packaging and shipping validation studies were successfully conducted on sterilized Stabilynx PEEK Suture Anchors pursuant to the applicable ASTM guidelines. These tests included seal peel and bubble emission tests at baseline and T= 3 months. Tests were conducted pursuant to ASTM F88 "Standard Test Method for Seal Strength of Flexible Barrier Materials"; and ASTM 2096-11 "Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)". Additionally, the ASTM D4169-09 standard was used to evaluate package performance under simulated transport conditions. All tests results met the acceptance criteria.
#### VIII CONCLUSIONS:
Based on the technological characteristics, indications for use, and bench performance data provided in this premarket notification, the Stabilynx PEEK Suture Anchor Device has a safety and effectiveness profile that is substantially equivalent to the predicate device. The information included in this 510(k) submission demonstrates the same intended use, similar indications for use and technological characteristics of the Stabilynx PEEK Suture Anchor Device as compared to the predicate device. The differences between the subject and predicate devices do not raise different types of safety or effectiveness questions and the performance data demonstrate substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.