← Product Code [MAX](/productcode/MAX) · K142634

# Biomet Spine Fusion System (K142634)

_Biomet Spine, LLC · MAX · Mar 12, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142634

## Device Facts

- **Applicant:** Biomet Spine, LLC
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Mar 12, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

When used as a lumbar intervertebral body fusion device, the Biomet Spine Intervertebral Body Fusion System ("Biomet Spine Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Biomet Spine Fusion System is to be implanted via an anterior or posterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Spinal Fixation System. The expandable interbody fusion devices are not indicated for vertebral body replacement.

## Device Story

Biomet Spine Fusion System consists of rectangular or semi-rectangular intervertebral body fusion implants; hollowed central area for autogenous bone graft; transverse grooves on surfaces for stability. Available in various sizes/configurations; expandable versions include PEEK spacer components and titanium alloy endplates with predetermined lordotic angles (0, 8, 12 degrees). Implanted by surgeons via anterior or posterior approach; requires supplemental fixation (e.g., Biomet Spinal Fixation System). Device maintains disc space height and facilitates spinal fusion in DDD patients; provides structural support during bone healing. Not indicated for vertebral body replacement.

## Clinical Evidence

Bench testing only. Performance testing conducted per FDA guidance for spinal systems and intervertebral body fusion devices. Testing included static/dynamic axial compression, shear compression, static torsion (ASTM F2077), subsidence (ASTM F2267), and particulate wear analysis (ASTM F1877). Results demonstrate substantial equivalence to predicates.

## Technological Characteristics

Materials: PEEK (ASTM F2026), Ti-6Al-4V ELI (ASTM F136), CP Titanium (ASTM F1580), Tantalum markers (ASTM F560). Design: Rectangular/semi-rectangular, hollow center, transverse surface grooves. Expandable models: PEEK spacers, titanium endplates, 0/8/12 degree lordotic angles. Dimensions: 7-14mm height, 11mm width, 25-35mm length. Sterile.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Spinal Elements Lucent ([K081968](/device/K081968.md))
- Biomet Spine Fusion System ([K083815](/device/K083815.md))
- Medtronic/AMT Wave ([K083626](/device/K083626.md))
- Globus Medical Caliber ([K102293](/device/K102293.md))

## Reference Devices

- Spinal Elements Lucent ([K122967](/device/K122967.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping to create a sense of depth and connection. The profiles are rendered in a simple, line-art style.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 12, 2015

Biomet Spine, LLC Mr. Mike Medina Director, Regulatory Affairs 310 Interlocken Parkway, Suite 120 Broomfield, Colorado 80021

Re: K142634

Trade/Device Name: Biomet Spine Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: February 6, 2015 Received: February 9, 2015

Dear Mr. Medina:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Mr. Mike Medina

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K142634

Device Name

Biomet Spine Fusion System

#### Indications for Use (Describe)

When used as a lumbar intervertebral body fusion device, the Biomet Spine Fusion System ("Biomet Spine Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis at the involved level(s). The Biomet Spine Fusion System is to be implanted via an anterior or posterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Spine Spinal Fixation System. The expandable interbody fusion devices are not indicated for vertebral body replacement

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in large, bold, sans-serif font. Below it, in a smaller font, is the word "SPINE". The logo is simple and modern, with a focus on the company name.

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with
the requirements of 21 CFR § 807.92.

| Preparation Date:                   | February 4, 2015                                                                               |
|-------------------------------------|------------------------------------------------------------------------------------------------|
| Applicant/Sponsor:                  | Biomet Spine LLC.                                                                              |
|                                     | 310 Interlocken Parkway, Suite 120                                                             |
|                                     | Broomfield, CO 80021                                                                           |
| Contact Person:                     | Mike Medina                                                                                    |
|                                     | Director, Regulatory Affairs                                                                   |
|                                     | Phone: 303-501-8548                                                                            |
|                                     | Fax: 303-501-8444                                                                              |
| Trade Name:                         | Biomet Spine Fusion System                                                                     |
| Common Name:                        | Intervertebral Body Fusion Device, Lumbar                                                      |
| Device Class:                       | Class II                                                                                       |
| Classification Name (Product Code): | Intervertebral body fusion device MAX                                                          |
| Device Panel – Regulation No.:      | Orthopedic – 21 CFR 888.3080                                                                   |
| Primary Predicate:                  | Spinal Elements Lucent K081968                                                                 |
| Additional Predicates:              | Biomet Spine Fusion System K083815, Medtronic/AMT Wave K083626, Globus Medical Caliber K102293 |

{4}------------------------------------------------

#### Device Description & Technological Characteristics:

The Biomet Spine Fusion System implants are intervertebral body fusion/vertebral body replacement devices consisting of a rectangular or semi-rectangular shape and various heights and footprints. The devices have a hollowed out central area to accommodate bone graft. The upper and lower surfaces of the devices have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The implants are available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The implants are made of PEEK (OPTIMA LT1®) per ASTM F2026, titanium alloy (Ti-6A1-4V ELI) per ASTM F136, commercially pure titanium per ASTM F1580, and may include tantalum markers per ASTM F560. The expandable Biomet Spine Fusion System implants incorporate PEEK (OPTIMA LT1® per ASTM F2026) spacer components and titanium alloy (Ti-6Al-4V ELI per ASTM F136) endplate components that orient into predetermined lordotic angles of 0, 8 and 12 degrees. The expandable implants are available in 7 to 14mm heights, an 11mm width, and 25, 30 and 35mm lengths. The Biomet Spine Fusion System is intended to be sterile.

## Intended Use / Indications for Use:

When used as a lumbar intervertebral body fusion device, the Biomet Spine Intervertebral Body Fusion System ("Biomet Spine Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Biomet Spine Fusion System is to be implanted via an anterior or posterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Spinal Fixation System. The expandable interbody fusion devices are not indicated for vertebral body replacement.

{5}------------------------------------------------

#### Summary of Technologies:

The technological characteristics of the subject Biomet Spine Fusion System components remain the same as, or similar to, the predicate devices in regards to intended use, indications for use, design, materials, manufacturing methods, fundamental technology, and operational principles.

## Performance Data:

Mechanical testing was conducted in accordance with FDA's Guidance for Industry and FDA Staff - Spinal System 510(k)s dated May 3, 2004 and Class II Special Controls Guidance Document: Intervertebral Body Fusion Device dated June 12, 2007. Performance testing was performed to demonstrate substantial equivalence to the predicate device(s). Performance testing included tests per ASTM F2077 (static and dynamic axial compression, static and dynamic shear compression, and static torsion), ASTM F2267 (subsidence) and particulate wear analysis per ASTM F1877. In all instances, the modified device met or exceeded predicate device performance, functioned as intended and therefore demonstrated substantial equivalence to the predicate device(s).

## Substantial Equivalence:

The modified Biomet Spine Fusion System implants included in the product line extension have the same intended use, indications, technological characteristics, and principles of operation as the previously cleared Biomet Spine Fusion System (K083815). Spinal Elements Lucent (K081968, K122967), AMT, Inc./Medtronic Wave (K083626) and Globus Medical Caliber (K102293). The minor differences in the new components do not raise any new issues of safety or effectiveness. Performance data presented also demonstrated comparable properties to the previously cleared devices.

## Conclusion:

The Biomet Spine Fusion System is substantially equivalent to the predicate systems as a spinal fusion device in regards to intended use, indications for use, fundamental technology including design, materials, manufacturing methods, sterility, and operational principles.

{6}------------------------------------------------

Furthermore, mechanical testing and other supporting information sufficiently demonstrate the substantial equivalence of the subject components to the Biomet Spine Fusion System to the predicate system. Based on this information, the subject modifications do not raise any new issues regarding the safety or efficacy when compared to its predicates.

---

**Source:** [https://fda-staging.innolitics.com/device/K142634](https://fda-staging.innolitics.com/device/K142634)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).