0.9% Sodium Chloride Injection, USP - 2.5 mL in 3 mL Syringe, 0.9% Sodium Chloride Injection, USP-3mL in 10 mL Syringe, 0.9% Sodium Chloride Injection, USP - 5 mL in 10 mL Syringe, 0.9% Sodium Chloride Injection, USP - 10 mL in 10 mL Syringe
K142620 · Excelsior Medical Corporation · NGT · Dec 12, 2014 · General Hospital
Device Facts
Record ID
K142620
Device Name
0.9% Sodium Chloride Injection, USP - 2.5 mL in 3 mL Syringe, 0.9% Sodium Chloride Injection, USP-3mL in 10 mL Syringe, 0.9% Sodium Chloride Injection, USP - 5 mL in 10 mL Syringe, 0.9% Sodium Chloride Injection, USP - 10 mL in 10 mL Syringe
Applicant
Excelsior Medical Corporation
Product Code
NGT · General Hospital
Decision Date
Dec 12, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
0.9% Sodium Chloride Injection, USP prefilled syringes are intended for flushing of IV catheters and IV tubing only.
Device Story
0.9% Sodium Chloride Injection, USP prefilled syringes are single-use, terminally sterilized devices. The device consists of a polypropylene barrel, plunger, latex-free isoprene rubber piston, and tip cap. It is used by clinicians to flush IV catheters and tubing before and after medication administration, blood sampling, total parenteral nutrition, or hemodialysis. The saline solution is delivered via a standard luer lock connection to maintain catheter patency through hydraulic displacement. The device is provided pre-filled and pre-packaged, supporting clinical workflows by ensuring sterile, ready-to-use saline for vascular access maintenance.
Clinical Evidence
Bench testing only. Evidence includes raw material qualification, stability studies, sterilization validation, and extractables/leachables testing. Biocompatibility testing was not required as materials are identical to predicates, with the exception of the sodium chloride raw material source.
Technological Characteristics
Materials: medical grade polypropylene (barrel/plunger), latex-free isoprene rubber (piston), TPE plus colorant (tip cap). Energy: none (manual). Dimensions: 3 mL and 10 mL syringe form factors. Connectivity: none. Sterilization: steam (10^-6 SAL).
Indications for Use
Indicated for flushing of IV catheters and IV tubing in patients requiring maintenance of catheter patency via hydraulic displacement.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Sterile Field Saline Flush Syringe(s), Excelsior Medical Corporation (K082837)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
December 12, 2014
Excelsior Medical Corporation Mr. John Linfante Vice President RA/QA 1933 Heck Avenue Neptune, NJ 07753
Re: K142620
Trade/Device Name: 0.9% Sodium Chloride Injection, USP Regulation Number: 21 CFR 880.5200 Regulation Name: Saline, Vascular Access Flush Regulatory Class: II Product Code: NGT Dated: September 15, 2014 Received: September 16, 2014
Dear Mr. Linfante
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142620
Device Name
0.9% Sodium Chloride Injection, USP
(2.5 mL in 3 mL Syringe, 3 mL in 10 mL Syringe, 5 mL in 10 mL Syringe, 10 mL in 10 mL Syringe)
Indications for Use (Describe)
0.9% Sodium Chloride Injection, USP prefilled syringes are intended for flushing of IV catheters and IV tubing only.
Type of Use (Select one or both, as applicable)
| <span> <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|------------------------------------------------------------------------------------------------------------------|
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Manufacturer Name: | Excelsior Medical Corporation |
|--------------------|---------------------------------------|
| Address: | 1933 Heck Avenue<br>Neptune, NJ 07753 |
| Contact Name: | John Linfante |
| Title: | Vice President RA/QA |
| Phone Number: | 732-643-6088 |
| Fax Number: | 732-776-7600 |
| Date Prepared: | September 15, 2014 |
# 510(k) Summary
| Device Proprietary Name: | 0.9% Sodium Chloride Injection, USP |
|------------------------------|-------------------------------------|
| Device Common or Usual Name: | Saline, Vascular Access Flush |
| Classification Name: | Saline, Vascular Access Flush |
| Classification Code: | NGT |
| Regulation Number: | 21 CFR Part 880.5200 |
| Device Classification | II |
# Predicate Devices:
Substantial equivalence is claimed to the following devices as related to intended use, design, and material characteristics:
- Excelsior Disposable Syringe W/Normal Saline (0.9% Sodium Chloride), Excelsior ● Medical Corp., K962938
- . Sterile Field Saline Flush Syringe(s), Excelsior Medical Corporation, K082837
# Description of the Device
Excelsior Medical Corporation's 0.9% Sodium Chloride Injection. USP prefilled syringe products are provided as terminally sterilized, single-use, pre-filled, pre-packaged products. The saline solution is delivered through the luer lock of a venous access device to maintain catheter patency via hydraulic displacement. Typically, the venous access device is flushed with normal saline before and after the administration of intermittent medication therapy, blood sampling. total parenteral nutrition, or hemodialysis.
0.9% Sodium Chloride Injection, USP, prefilled syringes are available as follows:
- 2.5 mL in 3 mL Syringe
- 3 mL in 10 mL Syringe ●
- 5 mL in 10 mL Syringe ●
{4}------------------------------------------------
- 10 mL in 10 mL Syringe ●
The product is intended for use with commercially available valves and catheters fitted with a standard mating luer lock or luer taper.
## Intended Use/Indications for Use
0.9% Sodium Chloride Injection, USP, prefilled syringes are intended for flushing of IV catheters and IV tubing only.
## Technological Characteristics
The 0.9% Sodium Chloride Injection, USP syringes subject to this filing were previously cleared by FDA under K962938. The use of alternate syringes, application of syringe labels on nonprinted barrels, and sterilization via steam were cleared under K082837 for Excelsion's Sterile Field Flush saline syringes.
A comparison of the products is provided in the table below.
| Intended use | Saline Flush Syringe<br>(Subject Device) | Sterile Field Saline<br>Flush Syringe(s)<br>K082837 | Excelsior Disposable<br>Syringe W/ Normal<br>Saline<br>K962938 |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Flushing of IV<br>catheters and IV<br>tubing only | Flushing IV catheters<br>and IV tubing | Flushing IV catheters<br>and IV tubing |
| Components | • Barrel – Medical<br>grade polypropylene<br>and silicone lubricant<br>• Plunger - Medical<br>grade polypropylene<br>• Piston - Latex free<br>isoprene rubber<br>• Tip Cap – Medical<br>grade polypropylene<br>and TPE plus<br>colorant | • Barrel – Medical<br>grade polypropylene<br>and silicone lubricant<br>• Plunger - Medical<br>grade polypropylene<br>• Piston - Latex free<br>isoprene rubber<br>• Tip Cap – Medical<br>grade polypropylene<br>and TPE plus<br>colorant | • Barrel – Medical<br>grade polypropylene<br>and silicone lubricant<br>• Plunger - Medical<br>grade polypropylene<br>• Piston - Latex free<br>isoprene rubber<br>• Tip Cap – Medical<br>grade polypropylene<br>and TPE plus<br>colorant |
| Solution | 0.9% Sodium Chloride<br>Injection, USP | 0.9% Sodium Chloride<br>Injection, USP | 0.9% Sodium Chloride<br>Injection, USP |
| Max. Fill<br>level | 10 mL | 10 mL | 10 mL |
| Sterility | Terminally sterilized by<br>steam 10-6 SAI | Terminally sterilized by<br>steam 10-6 SAL | Terminally sterilized by<br>gamma |
{5}------------------------------------------------
| Shelf Life | Intended 2 years | 2 years | 2 years |
|-------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|----------------------------|
| Syringe<br>Barrel | Not printed<br>Syringe Label contains:<br>• Graduation marks<br>• Lot Number<br>• Expiry Date | Not printed<br>Syringe Label contains:<br>• Graduation marks<br>• Lot Number<br>• Expiry Date | Pre-printed syringe barrel |
There are no differences between the subject and predicate device with respect to intended use or technology. The subject device is the exact same device as the predicate device cleared under K962938 with the following exceptions:
- Introduction of alternate sources of syringes ●
- Introduction of alternate source of saline
- Gradation markings applied via label
- Sterilization method is steam ●
These differences are addressed via material qualification, process validation, and finished product release testing. In addition, Excelsior Medical has undertaken performance testing to support that the use of an alternate source of saline does not alter the safety and effectiveness of the product.
### Non-Clinical Testing
The following studies have been performed or are being relied upon to support the safety and effectiveness of the product:
- Raw material qualification
- Stability Studies ●
- Sterilization validation
- Extractables/Leachables
Biocompatibility testing on the subject device is not required as the products are exactly the same as the predicate devices with the exception of the source of the sodium chloride raw material.
### Conclusion
Based on the analysis presented above and the results of the performance testing, the subject device is as safe and effective as the predicate device. No new or different questions of safety and effectiveness than the predicate device are raised by the changes.
Therefore, it is concluded that the product is substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.