K142584 · Siemens Medical Solutions USA, Inc. · JAK · Jan 16, 2015 · Radiology
Device Facts
Record ID
K142584
Device Name
iMAR
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Jan 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
iMAR is Siemens' iterative reconstruction software designed to reduce metal artifacts in CT images. It is a user-selected option, to be used with conventional reconstruction methods (WFBP or iterative reconstruction) on Siemens. The amount of metal artifact reduction and corresponding improvement in image quality depends on a number of factors including: composition and size of the metal object, patient size, anatomical location and clinical practice. If is recommended to perform reconstructions with iMAR enabled, in addition to conventional reconstruction without iMAR.
Device Story
iMAR is an iterative reconstruction software option for Siemens CT systems. It processes CT projection and image data to mitigate artifacts caused by large or dense metal objects (e.g., implants). The software corrects for beam hardening, scatter, undersampling, and photon starvation. It is operated by radiology staff in a clinical setting. The clinician selects iMAR as an option during reconstruction, typically generating both iMAR-enabled and conventional images for comparison. By reducing metal-induced artifacts, the device improves image quality, potentially aiding in more accurate diagnostic assessment of anatomy near metal objects.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed software specifications met acceptance criteria. Clinical images were evaluated and judged acceptable by a certified radiologist. No prospective clinical trials were required.
Technological Characteristics
Software-based iterative reconstruction for CT systems. Operates on the Somaris platform. Corrects artifacts in both projection and image domains. Corrects for beam hardening, scatter, undersampling, and photon starvation. Conforms to standards: IEC 60601-1 (2005), IEC 60601-2-44 (2009), IEC 60601-1-6 (2006), ISO 14971 (2010), IEC 62304 (2006), and DICOM (PS 3.1-3.18).
Indications for Use
Indicated for patients undergoing CT imaging who have large or dense metal objects causing artifacts. Used as a user-selected option with conventional reconstruction methods to reduce metal artifacts.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SOMATOM Definition AS Open using MARIS (VA46) (K130901)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2015
Siemens Medical Solutions USA, Inc. % Ms. Eve Davis Regulatory Affairs Specialist 51 Valley Stream Parkway MALVERN PA 19355
Re: K142584
Trade/Device Name: iMAR (iterative Metal Artifact Reduction) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: December 19, 2014 Received: December 22, 2014
Dear Ms. Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### 510(k) Number (if known) K142584
Device Name iMAR (iterative Metal Artifact Reduction)
#### Indications for Use (Describe)
iMAR is Siemens' iterative reconstruction software designed to reduce metal artifacts in CT images. It is a user-selected option, to be used with conventional reconstruction methods (WFBP or iterative reconstruction) on Siemens.
The amount of metal artifact reduction and corresponding improvement in image quality depends on a number of factors including: composition and size of the metal object, patient size, anatomical location and clinical practice.
If is recommended to perform reconstructions with iMAR enabled, in addition to conventional reconstruction without iMAR.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. There is a black line underneath the word.
# 510(k) Summary: iMAR (iterative Metal Artifact Reconstruction) Software
| Company: | Siemens Medical Systems USA, Inc. |
|----------|--------------------------------------------|
| | 51 Valley Stream Parkway Malvern, PA 19355 |
Date Prepared: December 11, 2014
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
1. General Information: Importer / Distributor: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Establishment Registration Number: 2240869
Location of Manufacturing Site SIEMENS AG Healthcare Siemensstrasse 1 D-91301 Forchheim, Germany Establishment Registration Number: 3004977335
## 2. Contact Person:
Eve Davis Regulatory Affairs Specialist Siemens Medical Solutions USA. Inc. 51 Valley Stream Parkway, D-02 Malvern, PA 19355 Phone: (610) 219-7133 Fax: (610) 448-1787 Email: eve.davis@siemens.com
# 3. Device Name and Classification:
| Trade Name: | iMAR (iterative Metal Artifact Reconstruction) SW |
|----------------------------|---------------------------------------------------|
| Classification Name: | Computed Tomography X-Ray System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
## 4. Legally Marketed Predicate Devices:
| Trade Name: | SOMATOM Definition AS Open using MARIS (VA46) |
|-----------------------|-------------------------------------------------------------------------------|
| 510(k) #: | K130901 |
| Clearance Date: | January 2, 2014 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | This predicate device has not been the subject of any design related recalls. |
{4}------------------------------------------------
# SIEMENS
# 5. Device Description:
iMAR (Metal Artifact Reconstruction) SW is a new software option, for Siemens Computed Tomography X-ray systems with the Somaris platform operating system. The iMAR software is designed to reduce metal artifacts caused by large and/or dense metal objects in computed tomography images. This submission includes the following device modifications:
- 1) New optional iMAR Software
- 2) A new specific Indication for Use Statement
- 3) Inclusion of a product Claims List
The iMAR software is based on a modification to the predicate device, previously cleared SOMATOM Definition AS Open using "MARIS" (Metal Implant Reduction in Image Space) VA46 (K130901) computed tomography system cleared under Premarket Notification K130901 on 01/02/2014.
The modified subject device the iMAR Software is within the same classification requlation with a specific indication for use from the primary predicate device: however, the intended use and the general Indication for Use Statement for Siemens' Computed Tomography X-ray system remains the same.
We believe these modifications are eligible for the Traditional 510(k) process since it has the same fundamental scientific technology. Intended use and general Indication for Use as the predicate device system. The new specific Indication for Use statement is supported with non-clinical testing. Documentation is provided to support a claim of substantial equivalence to Siemens' primary predicate device the SOMATOM Definition AS Open using MARIS VA46 (K130901).
# 6. Indication for Use:
iMAR is Siemens' iterative reconstruction software designed to reduce metal artifacts in CT images. It is a user-selected option, to be used with conventional reconstruction methods (WFBP or iterative reconstruction) on Siemens CT systems.
The amount of metal artifact reduction and corresponding improvement in image quality depends on a number of factors including: composition and size of the metal object, patient size, anatomical location and clinical practice.
It is recommended to perform reconstructions with iMAR enabled in addition to conventional reconstruction without iMAR.
# 7. Summary of Technological Characteristics of the Subject Device as Compared to the Predicate Device:
The iMAR Software is designed to reduce metal artifacts caused by large and/or dense metal objects in computed tomography images. The primary predicate device, the SOMATOM Definition AS Open using MARIS (VA46), was 510(k) cleared with the MARIS software which also reduces metal artifacts in scanned images. The subject device (iMAR) is comparable in general Computed tomography indication for use, same intended use, design, material, functionality,
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. There is a black line underneath the word.
technology, energy source and is considered substantially equivalent to the commercially available Siemens' SOMATOM Definition AS Open using MARIS (VA46).
The components of the subject device have many of the same technological characteristics as the components from the predicate device. There are several technological characteristics that differ slightly, as show below. The iMAR Software is designed to reduce metal artifacts caused by large and/or dense metal objects in computed tomography images. The predicate device has the MARIS feature which reduces metal artifact in images.
| Property | Subject Device: iMAR | Predicate Device:<br>SOMATOM Definition AS<br>Open using MARIS (VA46) |
|--------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Physical effects<br>corrected | beam hardening, scatter,<br>undersampling and photon<br>starvation | beam hardening |
| Corrections<br>applied in | projection domain and image<br>domain | projection domain and image<br>domain |
| User- controllable<br>settings | pre-defined parameter sets<br>optimized for a range of<br>different clinical situations | 5 strength levels |
## Subject and Predicate Device Compared Technological Characteristics
Testing and validation is complete. Test results show that the subject device, the iMAR Software, is comparable to the predicate device and therefore is substantially equivalent.
The modifications of the subject device, the iMAR Software, do not alter the intended use or fundamental scientific technology from the 510(k) cleared predicate device, the SOMATOM Definition AS Open using MARIS (VA46).
## 8. Performance Testing:
The modifications described in this premarket notification are supported with software verification/validation. Verification and Validation testing of the device was found acceptable to support claims of substantial equivalence. Non-clinical tests (integration and functional), including bench testing were conducted for the iMAR Software during product development. The Risk Analysis was completed and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria.
Siemens claims conformance to a signed statement of conformance to the following six (6) performance standards: 60601-1; 2005; 60601-2-44; 2009; 60601-1-6: 2006: 14971:2010: 62304 Ed. 1.0 2006. and PS 3.1 - PS3.18.
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software.
{6}------------------------------------------------
# SIEMENS
Evaluation of clinical images was conducted and judged acceptable by a certified radiologist.
### 9. General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a hazard analysis which is used to identify potential hazards. These potential hazards are controlled during software development, and verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
## 10. Conclusion as to Substantial Equivalence:
The predicate device was cleared based on non-clinical supportive information. The subject device bench testing, Verification and validation testing, and the evaluation of clinical images demonstrate that the iMAR Software performs as intended. The non-clinical test data and evaluation of clinical images demonstrate that the iMAR device performance is comparable to the predicate device that is currently marketed for the same intended use.
In summary Siemens is of the opinion that iMAR does not introduce any new potential safety risk and is substantially equivalent to the MARIS functionality of the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.