K142575 · C.R. Bard, Inc. · EZD · Oct 31, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K142575
Device Name
Bard RiteCath Intermittent Urinary Catheter
Applicant
C.R. Bard, Inc.
Product Code
EZD · Gastroenterology, Urology
Decision Date
Oct 31, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Bard RiteCath Intermittent Urinary Catheter is intended for use by adult and pediatric, male and female patients for draining urine from the bladder. Pediatric patients include neonates, infants, children and adolescents.
Device Story
Bard RiteCath Intermittent Urinary Catheter; biocompatible PVC catheter; used for bladder drainage. Device components: funnel, shaft with two staggered eyelets, tip. Tip configurations: straight or coude. Sizes: 6–18 Fr; lengths: 6", 10". Operation: tip inserted through urethra into bladder; urine drains through eyelets, shaft, and exits funnel. Single-use device. Used in clinical or home settings; operated by patients or healthcare providers. Provides non-invasive method for urine evacuation; benefits patients with urinary retention or bladder emptying dysfunction.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing included cytotoxicity, sensitization, and vaginal mucosal irritation per ISO 10993. Toxicological risk assessment performed for DINP. Functional performance testing conducted per BS EN 1616:1997 + A1:1999.
Technological Characteristics
Material: Polyvinyl chloride (PVC). Configuration: Funnel, shaft with two staggered eyelets, straight or coude tip. Sizes: 6–18 Fr. Lengths: 6", 10". Sterilization: Ethylene oxide per ANSI/AAMI/ISO 11135-1:2007. Single-use. No software or electronic components.
Indications for Use
Indicated for adult and pediatric (neonates, infants, children, adolescents) male and female patients requiring bladder drainage.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three human profiles facing right, layered on top of each other, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 31, 2014
C.R. Bard, Inc. Michele Davis Regulatory Affairs Project Manager 8195 Industrial Blvd Covington, GA 30014
> K142575 Trade/Device Name: Bard RiteCath Intermittent Urinary Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: Class II Product Code: EZD, EZC Dated: September 11, 2014 Received: September 12, 2014
Dear Michele Davis,
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P.
Lerner -S
for
Benjamin Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
### K142575
Device Name
Bard RiteCath Intermittent Urinary Catheter
#### Indications for Use (Describe)
The Bard RiteCath Intermittent Urinary Catheter is intended for use by adult and pediatric, male and female patients for draining urine from the bladder. Pediatric patients include neonates, infants, children and adolescents.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Bard Medical Division C.R. Bard, Inc. 8195 Industrial Blvd. Covington, GA 30014
Image /page/3/Picture/3 description: The image contains the word "BARD" in a stylized, green font. To the right of "BARD" is a vertical line, followed by the word "MEDICAL", also in green. The text appears to be a logo or brand name.
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Bard RiteCath Intermittent Urinary Catheter 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Submitter: | BARD Medical Division<br>C. R. BARD, Inc.<br>8195 Industrial Blvd.<br>Covington, GA 30014<br>Establishment Registration Number: 1018233 |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Michele Davis, RAC<br>Regulatory Affairs Project Manager<br>Bard Medical Division<br>Tel: 770-784-6274<br>Fax: 770-385-4706 |
| Date: | September 11, 2014 |
| Subject Device: | Trade Name: Bard® RiteCath <sup>TM</sup> Intermittent Urinary Catheter<br>Common Name: Urological Catheter<br>Classification Name: Urological catheter and accessories<br>Regulation: 21 CFR 876.5130<br>Regulatory Class: II<br>Primary Product Code: EZD<br>Secondary Product Codes: EZC |
| Predicate Device: | Legally marketed device to which substantial equivalence is claimed<br>• Bard RiteCath Intermittent Urinary Catheter, K133470<br>• Coloplast A/S Self Cath Catheter, K100878 |
- Coloplast A/S Self Cath Catheter, K100878
#### Device Description
The Bard RiteCath Intermittent Urinary Catheter is a biocompatible, polyvinyl chloride (PVC) catheter used to drain urine from the bladder. The catheter consists of a funnel, shaft with two staggered eyelets and a tip. The tip is available in a straight or coude configuration. The tip of the catheter passes through the urethra into the bladder to allow urine to drain into the eyelets and then through the catheter shaft, exiting through the funnel. The catheter will be offered in multiple French sizes (6 – 18 Fr.), lengths (6", 10" and two tip designs (straight and coude). The product is ethylene oxide sterilized (per ANSI/AAMI/ISO 11135-1:2007, Sterilization
{4}------------------------------------------------
of health care products – ethylene oxide – Part 1: Requirements for development, validation and routine control of a sterilized process for medical devices). The catheter is for single use.
#### Intended Use
The Bard RiteCath Intermittent Urinary Catheter is intended for use by adult and pediatric, male and female patients for draining urine from the bladder. Pediatric patients include neonates, infants, children and adolescents.
The indications for use statement for the Bard RiteCath Intermittent Urinary Catheter is not identical to the predicate device; however, the differences do no alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have intended use of urine drainage from the bladder (bladder drainage).
### Comparison of Technological Characteristics with the Predicate Devices
The Bard RiteCath Intermittent Urinary Catheter has the same technological characteristics as the predicate devices. The subject and predicate devices are based on the following same technological elements:
- Catheter material
- Catheter length ●
- Tip configuration
- Drainage eye position
- Condition of use
- Sterilization
#### Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility
The biocompatibility evaluation of the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO 10993 'Biological Evaluation of Medical Devices Part 1: Evaluation of Testing''' May 1, 1995, and ISO 10993-1:2009 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process." The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- Vaginal Mucosal Irritation
In addition a toxicological risk assessment for diisononyl phthalate (DINP) was performed per ISO 10993-17:2002 "Biological evaluation of medical devices – Part 17: Establishment of allowable limits for leachable substances."
The subject device is considered mucosal contacting with limited exposure.
{5}------------------------------------------------
#### Nonclinical functional performance testing
Nonclinical functional performance testing of the subject device was performed in accordance with BS EN 1616: 1997 + A1:1999, Sterile urethral catheters for single use.
#### Conclusions
The Bard RiteCath Intermittent Urinary Catheter has the similar technological characteristics and is intended for the same use of draining urine from the bladder as the predicate devices. The subject device is substantially equivalent to the legally, marketed predicate devices and nonclinical test data demonstrates that the subject device is safe and effective.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.