← Product Code [KWP](/productcode/KWP) · K142558

# Caspian OCT/MESA Mini/DENALI Mini Spinal System (K142558)

_K2m · KWP · Dec 22, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142558

## Device Facts

- **Applicant:** K2m
- **Product Code:** [KWP](/productcode/KWP.md)
- **Decision Date:** Dec 22, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Caspian OCT/MESA Mini Spinal System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) when used with autograft or allograft and is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), spondylolisthesis, fracture/dislocation, revision of previous cervical spine surgery, tumors, atlantoaxial fracture with instability, occipitocervical dislocation. The occipital bone screws are limited to occipital fixation only. The rod and hook components are in the cervical/upper thoracic (C1-T3) spine. The pedicle screws are limited to placement in the T1-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine. The Caspian OCT/MESA Mini/ DENALI Mini Spinal System can also be linked to the RANGE® Spinal System using rod connectors or transitional rods.

## Device Story

Top-loading, posterior spinal fixation system; components include rods, plates, screws, hooks, and wires. Used as adjunct to fusion for stabilization of cervical and upper thoracic spine (occiput-T3). Operated by surgeons in clinical/OR settings. Provides mechanical immobilization of spinal segments. System allows linkage to RANGE Spinal System via connectors. Benefits patients by providing structural support during bone healing/fusion process.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated via engineering analysis and comparison to previous testing conducted in accordance with ASTM F1717 and ASTM F2706.

## Technological Characteristics

Posterior spinal fixation system; top-loading design. Components: rods, plates, screws, hooks, wires. Materials and sizes consistent with predicate systems. Mechanical testing per ASTM F1717 and ASTM F2706.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Caspian OCT/MESA Mini/DENALI Mini Spinal System ([K101084](/device/K101084.md))
- Caspian OCT/MESA Mini/DENALI Mini Spinal System ([K092640](/device/K092640.md))
- Caspian OCT/MESA Mini/DENALI Mini Spinal System ([K081107](/device/K081107.md))
- Caspian OCT/MESA Mini/DENALI Mini Spinal System ([K121808](/device/K121808.md))
- Biomet Altius ([K033961](/device/K033961.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, flowing line style.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 22, 2014

K2M, Incorporated Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive Southeast Leesburg, Virginia 20175

Re: K142558

Trade/Device Name: Caspian OCT/MESA Mini/DENALI Mini Spinal System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: December 5, 2014 Received: December 8, 2014

Dear Ms. Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

#### K142558

Device Name

Caspian OCT/MESA Mini/DENALI Mini Spinal System

#### Indications for Use (Describe)

The Caspian OCT/MESA Mini Spinal System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) when used with autograft or allograft and is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), spondylolisthesis, fracture/dislocation, revision of previous cervical spine surgery, tumors, atlantoaxial fracture with instability, occipitocervical dislocation. The occipital bone screws are limited to occipital fixation only. The rod and hook components are in the cervical/upper thoracic (C1-T3) spine. The pedicle screws are limited to placement in the T1-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine. The Caspian OCT/MESA Mini/ DENALI Mini Spinal System can also be linked to the RANGE® Spinal System using rod connectors or transitional rods.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) SUMMARY Caspian OCT/MESA Mini/ DENALI Mini Spinal System

### Submitter

K2M, Inc. 751 Miller Drive SE Leesburg, VA 20175 Contact Person: Nancy Giezen Telephone: 703-777-3155 Date Prepared: 12/19/2014

### Classification

| Trade Name:       | Caspian OCT/MESA Mini/ DENALI Mini Spinal System |
|-------------------|--------------------------------------------------|
| Common Name:      | Spinal Fixation System                           |
| Regulatory Class: | Class II                                         |

Classification Name(s): Spinal Interlaminal Fixation Orthosis (21 CFR 888.3050, Product Code KWP)

### Predicate Device(s)

Primary Predicate K2M Caspian OCT/MESA Mini/DENALI Mini Spinal System (K101084) Additional Predicates K2M Caspian OCT/MESA Mini/DENALI Mini Spinal System (K092640, K081107, K121808) Biomet Altius (K033961)

# Device Description

The Caspian OCT/MESA Mini/ DENALI Mini Spinal System is a top-loading, multiple component, posterior (occipital-cervical-thoracic) spinal fixation system. The purpose of the current submission is to add modified rods, plates and screws to the system.

Function: The system functions as an adjunct to fusion to provide immobilization of stabilization of the posterior cervical and thoracic spine.

# Intended Use

The CASPIAN OCT/MESA Mini/ DENALI Mini Spinal System is intended to provide stabilization as an adjunct to function of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) when used with autograft or allograft and indicated for the following:

- DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies)
- Spondylolisthesis ●
- Spinal stenosis
- Fracture/dislocation
- Revision of previous cervical spine surgery ●

{4}------------------------------------------------

- Tumors ●
- Atlantoaxial fracture with instability ●
- Occipitocervical dislocation ●

The occipital bone screws are limited to occipital fixation only.

The rod and hook components are intended for use in the cervical/upper thoracic (C1-T3) spine. The pedicle screws are limited to placement in the T1-T3 in treating thoracic conditions only. The pedicle screws are intended to be placed in or treat conditions involving the cervical spine. The wires to be used with Caspian OCT/MESA Mini/ DENALI Mini Spinal System are intended for attachment to the posterior cervical spine.

The CASPIAN OCT/MESA Mini/ DENAIL Mini System can also be linked to the Range Spinal System using rod connectors or transitional rods.

# Technological Comparison to Predicate(s)

The proposed modifications were compared to predicate systems and the design features, materials and sizes were found to be substantially the same as these systems.

### Non-clinical Performance Evaluation

The worst case components of the Caspian OCT/MESA Mini/ DENALI Mini Spinal System were previously tested in accordance with ASTM F1717 and ASTM F2706 and determined to be equivalent to predicate devices. The subject implants were determined not to represent a new worst case using engineering analyses.

# Conclusion

There are no significant differences between the Caspian OCT/MESA Mini/ DENALI Mini Spinal System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K142558](https://fda-staging.innolitics.com/device/K142558)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).