UNIEVER Disposable Epidural Anesthesia Needle is intended to be used for injection into the epidural space/or placing the epidural catheter into the epidural space. UNIEVER Disposable Nerve Blockade Needle is intended to be used for injection of local anesthetic agent near the nerve for temporary pain control.
Device Story
Disposable stainless steel needles used for epidural anesthesia or nerve blockade; available in various gauges (14G-26G) and lengths (25mm-260mm); feature removable stylets and various tip designs (Huber, Hustead, Crawford, Back-Cut, K-3). Used by clinicians in clinical settings to inject anesthetic agents or place catheters. Provides physical access to epidural space or nerve sites; facilitates pain management. Benefits include temporary pain control and targeted anesthetic delivery.
Clinical Evidence
Bench testing only. Verification and validation testing performed to internal functional specifications. Compliance with ISO 594-1, ISO 7864, and ISO 9626 confirmed. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1. Pyrogenicity, endotoxin, and particulate testing for CSF contact performed.
Technological Characteristics
Single-use, terminally sterilized (EO, SAL 10^-6) needles. Materials comply with ISO 7864 and ISO 9626. Available in gauges 14G-26G and lengths 25mm-260mm. Tip designs include Huber, Hustead, Crawford, Back-Cut, and K-3. Includes removable stylet.
Indications for Use
Indicated for patients requiring epidural anesthesia or catheter placement, and patients requiring local anesthetic injection near a nerve for temporary pain control.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
Life-Tech ProLong Continuous Nerve Block Set (K081864)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
January 14, 2015
Unisis Corp. % Ms. Diane Rutherford Submissions Manager Ken Block Consulting, Inc. 1201 Richardson Drive Suite 280 Richardson, Texas 75080-4403
Re: K142553
> Trade/Device Name: Uniever Disposable Epidural Anesthesia Needle, Uniever Disposable Nerve Blocked Needle Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: December 12, 2014 Received: December 15, 2014
Dear Ms. Rutherford,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin -. Keith -S
Erin Keith Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known) K142553
Device Name
Uniever Disposable Epidural Anesthesia Needle, Uniever Disposable Nerve Blocked Needle
#### Indications for Use (Describe)
The Uniever Disposable Epidural Anesthesia Needle is intended to injection into the epidural space / or placing the epidural catheter into the epidural space.
Uniever Disposable Nerve Blockade Needle is intended to be used for injection of local anesthetic agent near the nerve for temporary pain control.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5. 510(k) SUMMARY
# K142553
| Submitter: | UNISIS Corp<br>2675-1, NISHIKATA<br>Koshigaya-shi, Saitama 343-0822 Japan | |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Contact Person: | Unisis Corp<br>Mr. Tadashi Yasuda<br>Medical Information Department<br>2623-1, NISHIKATA<br>Koshigaya-shi, Saitama 343-0822 Japan<br>TEL: 048-990-8250<br>FAX: 048-990-8251<br>m-info@unisis.co.jp | |
| Date Prepared: | August 28, 2014 | |
| Trade Name: | UNIEVER Disposable Epidural Anesthesia Needle<br>UNIEVER Disposable Nerve Blockade Needle | |
| Common Name: | Epidural Anesthesia Needle<br>Nerve Block Needle | |
| Classification Name: | NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) | |
| Product Code: | BSP | Class 2 868.5150 |
| Predicate Devices: | * K131006 LCCS AN-E Epidural Needle<br>* K040965 Pajunk Tuohy needle, Quincke needle, Chiba needle and<br>K113207 Crawford needle [Sono]<br>* K041843 Epimed Blunt Nerve Block Needle<br>* K081864 Life-Tech ProLong Continuous Nerve Block Set | |
| Device Description: | The UNIEVER Disposable Epidural Anesthesia Needles and UNIEVER Disposable<br>Nerve Blockade Needles are available in an array of sizes, lengths, and bevel and<br>tip designs. Both have a tightly fitting removable stylet.<br><br>The UNIEVER Disposable Epidural Anesthesia Needles are available with Huber,<br>Hustead, or Crawford tips. Available sizes are 14G to 25G (Huber), 16G to 19G<br>(Hustead), and 16G to 22G (Crawford) for lengths of 30-160mm.<br><br>The UNIEVER Disposable Nerve Blockade Needles are available with Back-Cut,<br>K-3, or Huber tips. Available sizes are 18G, 20G to 26G for lengths of 25-260mm<br>(Back Cut and K-3), and 20G, 22G and 25G for lengths of 30-160mm (Huber). | |
| Statement of<br>Intended Use: | UNIEVER Disposable Epidural Anesthesia Needle is intended to be used for<br>injection into the epidural space/or placing the epidural catheter into the epidural<br>space.<br><br>UNIEVER Disposable Nerve Blockade Needle is intended to be used for injection<br>of local anesthetic agent near the nerve for temporary pain control. | |
{4}------------------------------------------------
Summary of Technological Characteristics: As with the predicates, the UNIEVER Disposable Epidural Anesthesia and UNIEVER Disposable Nerve Blockade Needles are single use, terminally sterilized devices available in various gauge/length combinations including the overall combined ranges of 14G - 26G and 25mm to 260mm. The proposed devices share technological characteristics with the predicate devices. The proposed devices also have some differences in technological characteristics from those of the predicate devices. The differences in the technological characteristics are minor and reflect market strategy and do not impact the safety, effectiveness, or substantial equivalence of the device.
UNIEVER Disposable Epidural Anesthesia Needle offers various needle types as do the predicate devices. The UNIEVER Disposable Epidural Anesthesia Needle and both predicates identify the identical biocompatibility category, contact, and duration. UNIEVER Disposable Epidural Anesthesia Needles as well as both predicates are sterilized using ethylene oxide (EO) with Unisis specifying an SAL of 10-°.
Minor differences do exist between the UNIEVER Disposable Epidural Anesthesia Needle and the predicates. For example, the UNIEVER Disposable Epidural Anesthesia Needles offer a 14G needle while the two predicates have 15G [K110194] or 16G [K112515] as the largest gauge offered. The UNIEVER Disposable Epidural Anesthesia Needles also differ in the minimum and maximum needle lengths offered with the UNIEVER Epidural needles ranging from 30mm -160mm with the predicates offering 50mm - 150mm [K131006] and 20mm-180mm [K040965, K113207]. All the lengths offered for the UNIEVER Disposable Epidural Anesthesia Needles fall within the ranges offered by the identified predicates.
The UNIEVER Disposable Nerve Blockade Needle various needle types as do the predicate devices. Both the UNIEVER Blockade and the K041843 predicate identify the identical biocompatibility category, contact, and duration. UNIEVER Disposable Nerve Blockade Needles are sterilized using ethylene oxide (EO) to an SAL of 10°.
Minor differences do exist between the UNIEVER Disposable Nerve Blockade Needle and the predicates. For example, the UNIEVER Disposable Nerve Blockade Needles offer a 26G needle while the two predicates have 25G [K041843] or 20G [K081864] as the smallest gauge offered. The UNIEVER Disposable Nerve Blockade Needles also differ in the minimum and maximum needle lengths offered with the UNIEVER Blockade needles ranging from 25mm - 260mm with the predicates offering 38mm - 203.2mm [K041843] and 25mm - 150 mm [K081864].
As all the needles offered are tested for compliance to the same international standards (ISO 7864 and ISO 9626) this difference does not impact the safety, effectiveness, or substantial equivalence of the device.
{5}------------------------------------------------
- Tests were performed on the UNIEVER Disposable Spinal Epidural Anesthesia Summary of Performance Testing: Needles and the UNIEVER Disposable Nerve Blockade Needles including verification/validation testing to internal functional specifications which demonstrated that the devices are safe and effective. Testing confirmed that the UNIEVER Disposable Spinal Epidural Anesthesia Needles and the UNIEVER Disposable Nerve Blockade Needles comply with relevant voluntary safety standards, specifically ISO standards 594-1, 7864, and 9626. In addition, evaluations and validations have been performed to demonstrate compliance to the applicable standards for biocompatibility (ISO 10993-1) and sterilization including additional endotoxin and particulate testing for CSF contact. Biocompatibility testing performed includes cytotoxicity, sensitization and irritation. Pyrogenicity was also tested.
- Conclusion: Unisis Corp considers the UNIEVER Disposable Epidural Anesthesia Needle and the UNIEVER Disposable Nerve Blockade Needle to be substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.