EnteraLite Infinity Enteral Pump Delivery Set, 1200 mL Enteral Feeding Delivery Set
K142539 · Zevex, Inc. · PIF · Jan 8, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K142539
Device Name
EnteraLite Infinity Enteral Pump Delivery Set, 1200 mL Enteral Feeding Delivery Set
Applicant
Zevex, Inc.
Product Code
PIF · Gastroenterology, Urology
Decision Date
Jan 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The devices in this product family are used to dispense liquid nutrients (feeding solution) at a preprogrammed pump or user controlled rate. These enteral feeding sets interface with the patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The devices may include a bag to contain the feeding solution and/or a spike to connect to a prefilled container.
Device Story
Enteral feeding sets deliver liquid nutrients from a bag or prefilled container to a patient's feeding tube. Sets interface with enteral pumps or operate via gravity. Key feature is the bonded ENFit conical connector, designed to be incompatible with non-enteral connections per ISO 80369-3. An included adapter provides a traditional stepped connector for non-compliant feeding tubes. Used in clinical or home settings by patients or caregivers. Output is controlled flow of nutritional solution. Benefits include reduced risk of misconnection to non-enteral devices and reliable nutrient delivery.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including bond strength, leak testing, volumetric accuracy, biocompatibility, misconnection assessment, dimensional analysis, FMEA, accelerated aging, and ISO 80369-3 performance testing.
Technological Characteristics
Materials: Non-DEHP PVC tubing, polyethylene inlet port cap, polycarbonate cassette, silicone pump tubing, ABS (HF 380) ENFit connector. Design: Single-use, non-sterile, includes ENFit connector and transitional stepped adapter. Connectivity: Mechanical interface to feeding tubes and pumps. Complies with ISO 10993-1 (biocompatibility) and ISO 80369-3 (small-bore connectors).
Indications for Use
Indicated for patients requiring enteral nutrition delivery via feeding tube; compatible with gravity or pump-assisted administration of liquid nutrients.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Zevex, Inc., Enteral Feeding Sets for Gravity and Pump Use (K012147)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them, possibly representing water or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 8, 2015
Zevex, Inc. Dianne Batch Director, RA/QA 4314 Zevex Park Lane Salt Lake City, UT 84123
Re: K142539
> Trade/Device Name: Enteralite Infinity Enteral Pump Delivery Set, 1200 ML Enteral Feeding Delivery Set Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: Class II Product Code: PIF, PIO Dated: November 28, 2014 Received: November 28, 2014
Dear Dianne Batch,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -A" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the overall impression is clean and professional.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142539
Device Name
Enteralite Infinity Enteral Pump Delivery Set, 1200 ML Enteral Feeding Delivery Set
#### Indications for Use (Describe)
The devices in this product family are used to dispense liquid nutrients (feeding solution) at a preprogrammed pump or user controlled rate. These enteral feeding sets interface with the patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The devices may include a bag to contain the feeding solution and/or a spike to connect to a prefilled container.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# for
# Enteral Feeding Sets
## 1. Submission Sponsor
Zevex, Inc. 4314 Zevex Park Lane Salt Lake City, Utah 84123 USA Phone: 801.201.6928 Fax: 801.264.1051 Contact: Dianne Batch, Director, RA/QA
### 2. Date Prepared
September 5, 2014
### 3. Device Identification
| Trade/Proprietary Name: | EnteraLite Infinity Enteral Pump Delivery Sets | |
|---------------------------------|------------------------------------------------|--|
| | 1200 mL Enteral Feeding Delivery Set | |
| Common/Usual Name: | Enteral Feeding Sets | |
| Classification Name: | Gastrointestinal Tubes and Accessories | |
| Classification Regulation: | 876.5980 | |
| Feeding Set Product Code: PIF | | |
| Transition Connector | | |
| Product Code: | PIO | |
| Device Class: | Class II | |
| Classification Panel: | Gastroenterology/Urology | |
### 4. Legally Marketed Predicate Device(s)
K012147, Zevex, Inc., Enteral Feeding Sets for Gravity and Pump Use
### 5. Device Description
These enteral feeding sets are designed to be used with or without an enteral feeding pump, (pump is not in scope for this 510k) model specific. Refer to diagrams below for explanation of components. The enteral sets terminate in a bonded conical connector, the ENFit connector; that is designed to be incompatible with ISO 594-1/2 and other non-enteral feeding connections. The ENFit connector is a new connector and it is anticipated that not all patient feeding tubes will be compatible with it, so the administration sets include an adapter that connects to the ENFit connector on one end while providing the traditional "Christmas Tree" shaped transitional stepped connector on the other.
#### 6.
### Indication for Use Statement
The devices in this product family are used to dispense liquid nutrients (feeding solution) at a preprogrammed pump or user controlled rate. These enteral feeding sets interface with
{4}------------------------------------------------
the patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The devices may include a bag to contain the feeding solution and/or a spike to connect to a prefilled container.
## 7. Substantial Equivalence Discussion
The following table compares the Enteral Feeding Sets to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
In the 'Significant Differences' column of the table, list the differences between the device and the predicate and briefly justify why these differences do not raise safety and effectiveness concerns.
| Manufacturer | Moog Medical Devices | Moog Medical Devices | Significant |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Group | Group | Differences |
| Trade Name | Enteral Feeding Sets | Zevex, Inc., Enteral | |
| | | Feeding Sets For Gravity | |
| | | And Pump Use | |
| 510(k) Number | K142539 | K012147 | N/A |
| Product Code | PIF, PIO | KNT | Updated Codes from FDA |
| Regulation | 876.5980 | 876.5980 | Same |
| Number | | | |
| Regulation Name | Gastrointestinal Tubes | Tubes, Gastrointestinal | Same |
| | and Accessories | (and Accessories) | |
| Indications for | The devices in this | The devices in this product | The difference is in the |
| Use | product family are used | family are used to | distal (or patient) end of |
| | to dispense liquid | dispense liquid nutrients | the device. Where the |
| | nutrients (feeding | (feeding solution) at a | predicate terminates |
| | solution) at a | preprogrammed pump or | only in a traditional |
| | preprogrammed pump or | user controlled rate. | stepped connector, the |
| | user controlled rate. | These enteral feeding sets | proposed device |
| | These enteral feeding | interface with the | terminates in a bonded |
| | sets interface with the | patient's feeding tube and | ENFit connector which is |
| | patient's feeding tube | may use gravity or an | designed to be |
| | and may use gravity or an | enteral feeding pump to | incompatible with all |
| | enteral feeding pump to | dispense feeding solution. | other defined connectors |
| | dispense feeding | The devices may include a | available. The ENFit |
| | solution. The devices | bag to contain the feeding | connector is designed to |
| | may include a bag to | solution and/or a spike to | comply with ISO 80369-3. |
| | contain the feeding | connect to a prefilled | GI Tubes that are |
| | solution and/or a spike to | container. | compatible with the ISO |
| | connect to a prefilled | | 80369-3 compliant |
| | container. | | design will be able to |
| Manufacturer | Moog Medical Devices<br>Group | Moog Medical Devices<br>Group | Significant<br>Differences |
| Trade Name | Enteral Feeding Sets | Zevex, Inc., Enteral<br>Feeding Sets For Gravity<br>And Pump Use | |
| | | | connect directly to the<br>ENFit connector. Non<br>ISO 80369-3 compliant GI<br>tubes will be able to<br>connect to the<br>transitional stepped<br>connector which is<br>designed to adapt to the<br>ENFit connector. |
| Material | Non-DEHP PVC tubing,<br>extruded film (bags), inlet<br>port and outlet port;<br>Polyethylene inlet port<br>cap; polycarbonate<br>connector/pump<br>interface (cassette);<br>silicone pump tubing<br>segment; ABS distal<br>(stepped) connector. | Non-DEHP PVC tubing,<br>extruded film (bags), inlet<br>port and outlet port;<br>Polyethylene inlet port<br>cap; polycarbonate<br>connector/pump interface<br>(cassette); silicone pump<br>tubing segment; ABS distal<br>(stepped) connector. | Similar; the differences<br>between them are that<br>the ABS formulation (HF<br>380) and color of the<br>ENFit connector (purple,<br>Remofin Violet<br>PE43076356-ZT)) are<br>different from the ABS<br>formulation (CYCOLAC<br>HMG47MD 6H3C106,<br>pre-colored) used in the<br>current stepped adapter. |
| Sterile | Non-sterile | Non-sterile | Same |
| Overall Design | | | |
| Length | Set length varies, tubing<br>lengths same as<br>predicate; connector is<br>slightly shorter (~ 0.82"<br>vs. ~2.0") than the<br>current stepped adapter,<br>but with the adapter<br>(~1.74") the overall set<br>length will be less than<br>1/4" different. | Set length varies<br>depending on<br>configuration. | Same<br>The EnFit connector has a<br>slightly larger ID, which<br>has shown no negative<br>impact on flow rates. |
| Diameter | EnFit connector ID is<br>0.116". | Distal connector ID is<br>0.10". | |
| Single-Use | Yes | Yes | Same |
| Shelf Life | 1 yr. | 1 yr. (minimum) | Same |
| Manufacturer | Moog Medical Devices | Moog Medical Devices | Significant |
| | Group | Group | Differences |
| Trade Name | Enteral Feeding Sets | Zevex, Inc., Enteral | |
| | | Feeding Sets For Gravity | |
| | | And Pump Use | |
| Latex Free | Yes | Yes | Same |
| Complies with ISO | Yes | Yes | Same |
| 10993-1 | | | |
# Table 5A – Comparison of Characteristics
{5}------------------------------------------------
{6}------------------------------------------------
## 8. Non-Clinical Performance Data
The following testing has been performed to support substantial equivalence:
- Bond Strength - Passed
- Leak Testing - Passed
- Volumetric accuracy testing Passed ●
- Biocompatibility - Passed
- Misconnection assessment - Passed
- Dimensional analysis — Passed
- Enteral Connector Risk Management Report Acceptable ●
- Human Factors Testing – Acceptable
- Failure Modes and Effects Analysis (FMEA) – Acceptable
- Accelerated Aging - Passed
- ISO 80369-3 Testing as follows:
- . Falling drop positive pressure liquid leakage - Passed
- Stress cracking - Passed
- Resistance to separation from axial load – Passed
- Resistance to separation from unscrewing – Passed
- Resistance to overriding - Passed
- Disconnection by unscrewing - Passed
- Falling drop positive pressure liquid leakage after 20 cycles of connection and separation - Passed
As part of demonstrating safety and effectiveness of Enteral Feeding Sets and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Zevex completed a number of tests. The Enteral Feeding Sets meets all the requirements for overall design, functionality, biocompatibility, and confirms that the output meets the design inputs and specifications. The Enteral Feeding Sets passed all testing stated above as shown by the acceptable results obtained.
The Enteral Feeding Sets along with the ENFit connector and adapter complies with the applicable voluntary standards for biocompatibility. The devices passed all the testing in accordance with national and international standards.
### 9. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
{7}------------------------------------------------
# 10. Statement of Substantial Equivalence
It has been shown in this 510(k) submission that the difference between the Enteral Feeding Sets and the predicate devices do not raise any questions regarding its safety and effectiveness. The Enteral Feeding Sets, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.