← Product Code [LLZ](/productcode/LLZ) · K142521

# Synapse 3D Liver and Kidney Analysis (K142521)

_Fujifilm Meidcal Systems U.S.A.,Inc. · LLZ · Dec 1, 2014 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142521

## Device Facts

- **Applicant:** Fujifilm Meidcal Systems U.S.A.,Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Dec 1, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Synapse 3D Liver and Kidney Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Liver and Kidney Analysis accepts DICOM compliant medical images. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Addition to Synapse 3D Base Tools, Synapse 3D Liver and Kidney Analysis uses contrast enhanced images of the body and provides custom workflows and UI, and reporting functions for liver and kidney analysis including, liver and peripheral organ segmentation, and tumor segmentation of intrahepatic and peripheral vessels as well as the approximation of vascular territories is provided using contrast enhanced computed tomographic images.

## Device Story

Software module for Synapse 3D Base Tools; processes DICOM-compliant CT and MR images. Provides tools for liver and kidney analysis, including organ and vessel segmentation, tumor segmentation, and vascular territory approximation. Used in clinical settings by radiologists and clinicians for treatment planning and reporting. Operates on Windows-compatible hardware; integrates with PACS and cardiovascular systems. Transforms raw image data into 3D visualizations and quantitative measurements via custom workflows and UI. Assists clinicians in interpreting complex anatomy, facilitating surgical or interventional planning. Benefits patients through improved visualization and analysis of abdominal organs and vasculature.

## Clinical Evidence

Bench testing only. Verification and validation activities included system-level functionality, segmentation accuracy, measurement accuracy, interfacing, usability, and serviceability testing. Performance testing utilized actual clinical images to confirm accuracy against requirements.

## Technological Characteristics

Windows-based application software; runs on general-purpose hardware. Connectivity via DICOM standard to CT, MR, CR, US, NM, PT, XA, and PACS. Provides 3D visualization, segmentation, and measurement tools. Software module architecture; integrates with Synapse 3D Base Tools. Moderate level of concern per FDA software guidance.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- VitreaAdvanced ([K121213](/device/K121213.md))
- IQQA-Liver Multimodality software ([K131498](/device/K131498.md))

## Reference Devices

- Synapse 3D Base Tools ([K120361](/device/K120361.md))
- Synapse 3D Liver Analysis ([K110186](/device/K110186.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

FUJIFILM Medical Systems U.S.A., Inc. % Mr. Jyh-Shyan Lin Senior Manager, Regulatory, Quality and Clinical Affairs 419 West Avenue STAMFORD CT 06902

December 1, 2014

Re: K142521

Trade/Device Name: Synapse 3D Liver and Kidney Analysis Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 10, 2014 Received: November 12, 2014

Dear Mr. Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K142521

Device Name SYNAPSE 3D Liver and Kidney Analysis

#### Indications for Use (Describe)

Synapse 3D Liver and Kidney Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Liver and Kidney Analysis accepts DICOM compliant medical images. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Addition to Synapse 3D Base Tools, Synapse 3D Liver and Kidney Analysis uses contrast enhanced images of the body and provides custom workflows and UI, and reporting functions for liver and kidney analysis including, liver and peripheral organ segmentation, and tumor segmentation of intrahepatic and peripheral vessels as well as the approximation of vascular territories is provided using contrast enhanced computed tomographic images.

| Type of Use (Select one or both, as applicable)                                |
|--------------------------------------------------------------------------------|
| <div> <span>     Prescription Use (Part 21 CFR 801 Subpart D)   </span> </div> |
| <div> <span>     Over-The-Counter Use (21 CFR 801 Subpart C)   </span> </div>  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## FUJ}}FILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse 3D Liver and Kidney Analysis

## 510(k) Summary

#### Date Prepared:

September 5, 2014

### Submitter's Information:

FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (301) 251-1092 Fax: (203) 602-3785 Contact: Jyh-Shyan Lin

### Device Trade Name:

Synapse 3D Liver and Kidney Analysis

#### Device Common Name:

Medical Image Processing and Analysis Software

#### Regulation Number:

21 CFR 892.2050

#### Device Classification:

Class II

## Device Classification Name:

Picture Archiving and Communications System (PACS)

#### Panel:

Radiology

#### Product Code:

LLZ

{4}------------------------------------------------

## FUJ}{FILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse 3D Liver and Kidney Analysis

### Description of the Device

Synapse 3D Liver and Kidney Analysis (V4.0) (this submission) is updated software of previously-cleared Synapse 3D Liver Analysis (cleared by CDRH via K110186 on 04/07, 2011) with expanded IFU and revised device name. The new feature in Synapse 3D Liver and Kidney Analysis (V4.0) are Kidney Analysis (CT) function and Liver Analysis (MR).

Synapse 3D Liver and Kidney Analysis is medical application software running on Windows standalone and server/client configuration installed on a commercial general-purpose Windowscompatible computer. It offers software tools which can be used by trained professionals, such as radiologists, clinicians or general practitioners to interpret medical images obtained from various medical devices, to create reports, or to develop treatment plans.

Synapse 3D Liver and Kidney Analysis is an optional software module that works with Synapse 3D Base Tools (cleared by CDRH via K120361 on 04/06/2012) which is connected through DICOM standard to medical devices such as CT, MR. CR, US, NM, PT, XA, etc. and to a PACS system storing data generated by these medial devices, and it retrieves image data via network communication based on the DICOM standard. The retrieved image data are stored on the local disk managed by Synapse 3D Base Tools, and the associated information of the image data is registered in the database and is used for display, image processing, analysis, etc. Synapse 3D Liver and Kidney Analysis can handle images of CT and MR. The software can display the images on a display monitor, or printed them on a hardcopy using a DICOM printer or a Windows printer.

The liver and kidney analysis tools both segment the organs, peripheral organs and vessels using similar body part recognition algorithms already available in the FDA-cleared Base Tools (K120361) and the Liver Analysis (CT) (K110186). The technical characteristics and principles of operations are described in details in the Device Description section, which includes the Liver Analysis (CT and MR) and Kidney Analysis (CT). Based on the cleared functions, Synapse 3D Liver and Kidney Analysis enhances the custom workflows and UI to improve the usability.

Synapse 3D Liver and Kidney Analysis with Synapse 3D Base Tools can be integrated with Fujifilm's Synapse PACS, and can be used as a part of a Synapse system. Synapse 3D Liver and Kidney Analysis also can be integrated with Fujifilm's Synapse Cardiovascular for cardiology purposes. In summary, this 510(k) submission introduces the Synapse 3D Liver and Kidney Analysis with the added capability of performing Liver (MR) and Kidney (CT) Analysis.

### Indication for Use

Synapse 3D Liver and Kidney Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, and treatment planning. Synapse 3D Liver and Kidney Analysis accepts DICOM compliant medical images. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

{5}------------------------------------------------

## FUJ}{FILM MEDICAL SYSTEMS U.S.A., Inc.

#### Synapse 3D Liver and Kidney Analysis

Addition to Synapse 3D Base Tools, Synapse 3D Liver and Kidney Analysis uses contrast enhanced images of the body and provides custom workflows and UI, and reporting for liver analysis including, liver and peripheral organ segmentation. Also segmentation. Also segmentation of intrahepatic vessels as well as the approximation of vascular territories is provided using contrst enhanced computed tomographic images.

### Predicate Devices:

- . VitreaAdvanced (K121213), Vital Images, Inc.
- . IQQA-Liver Multimodality software (K131498), EDDA Technology

### Technological Characteristics

The proposed Synapse 3D Liver and Kidney Analysis and the predicate devices, i.e., VitreaAdvanced (K121213) and IQQA-Liver Multimodality software (K131498), are medical application software running on Windows operating system installed on commercial general-purpose Windows-compatible computers. These devices are connected to CT and MR with DICOM standard and retrieve image data via network communications. These devices provide 3D image visualization and manipulation tools for medical images with various user interfaces and measurement tools for analysis of rendered images. The Synapse 3D Liver and Kidney Analysis and the predicate devices support the workflows, UI, and reporting functions for analyzing the images of the liver (MR and CT) and kidney (CT), which are abdomen organs.

## Safety Information

Synapse 3D Liver and Kidney Analysis introduces no new safety or efficacy issues other than those already indentified with the predicate devices. The Risk Management and the results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices." The Synapse 3D Liver and Kidney Analysis labeling contains instructions for use and necessary cautions, warnings and notes to provide the safe and effective use of the device.

## Testing and Performance Information

Synapse 3D Liver and Kidney Analysis is tested successfully with reference to its Software Requirements Specification, as well as design verification and validation documents and Traceability Matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Synapse 3D Liver and Kidney Analysis software, which is found to be safe and effective and substantially equivalent to the predicate devices.

{6}------------------------------------------------

# FUJ}}FILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse 3D Liver and Kidney Analysis

Testing involved system level functionality test, segmentation accuracy test, measurement accuracy test, interfacing test, usability test, serviceability test, labeling test, as well as the test for risk mitigation method analyzed and implemented in the risk management process. In addition, we conducted the bench performance testing using actual clinical images to help demonstrate that the proposed device achieved the expected accuracy performance.

Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests successfully passed.

### Conclusion

This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K142521](https://fda-staging.innolitics.com/device/K142521)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
