← Product Code [LLZ](/productcode/LLZ) · K142514

# UniWeb 8.0 (K142514)

_Ebm Technologies Incorporated · LLZ · Dec 18, 2014 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142514

## Device Facts

- **Applicant:** Ebm Technologies Incorporated
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Dec 18, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

UniWeb™ 8.0 is a device that receives digital images and data from various sources (including but not limited to Fullfield digital mammography, MR scanners, CR, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other DICOM formatted imaging sources).Images and data can be stored, communicated, processed and displayed within the system and or across a computer network at distributed locations. UniWeb™ 8.0 is utilized only on FDA cleared monitors for interpretation of mammography images in a primary diagnostic setting. Moreover, it can display FFDM(Full-field digital mammography) post-processed images in DICOM for presentation only formats. This device will only use lossless compression when displaying mammography images. For image diagnosis, this device can only be used by trained professionals. However, for image review, the users will not be limited to physicians, nurses and technicians.

## Device Story

UniWeb™ 8.0 is a Picture Archiving and Communication System (PACS) designed to receive, store, communicate, process, and display digital medical images and data. It ingests DICOM-formatted inputs from diverse modalities, including full-field digital mammography (FFDM), MR, CR, ultrasound, and radiographic devices. The system operates across computer networks, enabling distributed access to images. For mammography, the device mandates the use of FDA-cleared monitors and employs lossless compression to maintain image integrity for primary diagnostic interpretation by trained professionals. Beyond diagnostic use, the system facilitates image review by broader clinical staff, including nurses and technicians. By centralizing image management and providing high-fidelity display capabilities, the device supports clinical workflows, enabling efficient review and diagnostic decision-making across healthcare facilities.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and intended use consistent with existing PACS devices.

## Technological Characteristics

PACS software for DICOM image management. Supports lossless compression for mammography. Operates on networked computer systems. Requires FDA-cleared monitors for primary diagnostic interpretation. Compliant with DICOM standards for image communication and processing.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

December 18, 2014

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

EBM Technologies Incorporated % Mr. John Su Quality Manager 5f., No. 516, Sec. 1, Neihu Road., Neihu Dist. Taipei. 114 TAIWAN

Re: K142514 Trade/Device Name: UniWeb™ 8.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 12, 2014 Received: November 19, 2014

Dear Mr. Su:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert A. Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K142514

Device Name UniWeb™ 8.0

UniWeb™ 8.0 is a device that receives digital images and data from various sources (including but not limited to Fullfield digital mammography, MR scanners, CR, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other DICOM formatted imaging sources).Images and data can be stored, communicated, processed and displayed within the system and or across a computer network at distributed locations.

UniWeb™ 8.0 is utilized only on FDA cleared monitors for interpretation of mammography images in a primary diagnostic setting. Moreover, it can display FFDM(Full-field digital mammography) post-processed images in DICOM for presentation only formats. This device will only use lossless compression when displaying mammography images. For image diagnosis, this device can only be used by trained professionals. However, for image review, the users will not be limited to physicians, nurses and technicians.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

---

**Source:** [https://fda-staging.innolitics.com/device/K142514](https://fda-staging.innolitics.com/device/K142514)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).