K142500 · Terumo Cardiovascular Systems Corporation · MJJ · Jan 29, 2015 · Cardiovascular
Device Facts
Record ID
K142500
Device Name
One Way Soft Valve
Applicant
Terumo Cardiovascular Systems Corporation
Product Code
MJJ · Cardiovascular
Decision Date
Jan 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Indications for Use
The One Way Soft Valve is intended to be used for the prevention of retrograde flow in ¼/' or smaller lines subject to arterial pressure during cardiopulmonary bypass procedures for use up to six hours.
Device Story
One Way Soft Valve is a single-use, disposable, mechanical check valve; prevents retrograde flow in cardiopulmonary bypass (CPB) circuits. Device features transparent PVC housing with sleeve-like valve mechanism; includes directional arrow for orientation. Used in clinical settings during CPB procedures; operated by perfusionists or surgical staff. Input is blood flow within 1/4" arterial lines; device mechanically restricts flow in opposing direction. Output is unidirectional flow; ensures circuit integrity during bypass. Benefits include smaller prime volume (approx. 9 mL) compared to predicate, facilitating use in cardiovascular procedure convenience kits and smaller diameter lines.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via in-vitro bench testing: tubing connection bond strength, mechanical/structural integrity/burst strength, opening pressure, negative pressure/retrograde flow performance, and circulation/hemolysis evaluation.
Technological Characteristics
Materials: Flexible PVC housing and valve. Principle: Mechanical one-way sleeve-style check valve. Dimensions: 1/4" ports. Prime volume: ~9 mL. Sterilization: Ethylene oxide (SAL 10^-6). Biocompatibility: ISO 10993 compliant for external communicating, circulating blood, limited exposure (≤ 24 hours).
Indications for Use
Indicated for prevention of retrograde flow in 1/4" or smaller lines subject to arterial pressure during cardiopulmonary bypass procedures for up to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Quest Medical Retroguard Arterial Safety Valve (K922356)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, ribbon-like element beneath them. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 29, 2015
Terumo Cardiovascular Systems Corporation Garry A. Courtney Director, Regulatory Affairs 28 Howe Street Ashland, Massachusetts 01721
Re: K142500
Trade/Device Name: One Way Soft Valve Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II Product Code: MJJ Dated: January 2, 2015 Received: January 5, 2015
Dear Garry A. Courtney,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Terumo Medical Corporation logo. The logo consists of a green circle with a stylized "T" inside, followed by the word "TERUMO" in green, sans-serif font. A registered trademark symbol appears to the right of the word "TERUMO".
## SECTION 4 – Indications for Use
One Way Soft Valve
510(k) Number (if known): K 142500
Device Name: One Way Soft Valve
# Indications For Use:
The One Way Soft Valve is intended to be used for the prevention of retrograde flow in ¼/' or smaller lines subject to arterial pressure during cardiopulmonary bypass procedures for use up to six hours.
| Prescription Use <span style="text-decoration: overline;">XX</span><br>(Part 21 CFR 801 Subpart D) | OR | Over-The-Counter Use ____________________<br>(Part 21 CFR 807 Subpart C) |
|----------------------------------------------------------------------------------------------------|----|--------------------------------------------------------------------------|
| (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF<br>NEEDED) | | |
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Terumo. The logo consists of a circle with the letter T inside of it, followed by the word "TERUMO" in all capital letters. The logo is green.
## SECTION 5 – 510(k) Summary One Way Soft Valve
| Submitter Information | 5-2 |
|-------------------------------------------------|-----|
| Device Names/Classification | 5-2 |
| Identification of Predicate Device | 5-2 |
| Device Description | 5-2 |
| Comparison of Indication | 5-2 |
| Principles of Operation, Technology, and Design | 5-3 |
| Performance Evaluations | 5-4 |
| Substantial Equivalence Statement | 5-4 |
| Additional Safety Information | 5-4 |
| Conclusion for 510(k) Summary | 5-4 |
{4}------------------------------------------------
### Submitter Information:
This submission was prepared in August 2014 by: Suzanne Grenier, RAC. COA Regulatory Affairs Senior Specialist Terumo Cardiovascular Systems Corporation 125 Blue Ball Road Elkton, MD 21921 Telephone: 443-745-8443
This submission was prepared for:
Terumo Cardiovascular Systems Corporation 28 Howe Street Ashland, MA 01721 Facility Registration No. 1212122
#### Device Names/Classifications:
| Proprietary Name | Classification Name | Common Name |
|--------------------|--------------------------------|--------------------------|
| One Way Soft Valve | Cardiopulmonary Bypass | CPB Check Valve, |
| | Blood Reservoir (ProCode: MJJ) | Retrograde Flow, In-Line |
#### Identification of Predicate Device:
The device submitted in this 510(k) maintains characteristics that are substantially equivalent in indications for use, design, technology including features, materials, and principles of operation to the Quest Medical Retroguard Arterial Safety Valve - K922356.
#### Device Description:
The One Way Soft Valve is a transparent one-way valve with ¼" inlet and outlet ports. The One Way Soft Valve prevents retrograde flow. A directional arrow is printed on the valve to provide the user with a visual indication of the allowable flow direction within the valve is constructed entirely of PVC and is to be used in lines subject to arterial pressures.
#### Comparison of Indication
The One Way Soft Valve product and the predicate, Quest Medical Retroguard Arterial Safety Valve product, are intended to be used to prevent retrograde flow during cardiopulmonary bypass procedures for up to six hours.
Subject Device Indication: The One Way Soft Valve is intended to be used for the prevention of retrograde flow in ¼" or smaller lines subject to arterial pressure during cardiopulmonary bypass procedures for use up to six hours.
Predicate Device Indication: Per the IFU, the Quest Medical Retroguard Arterial Safety Valve is "for prevention of retrograde flow when used with centrifugal pump."
The differences between the intended use for the subject device and the predicate device do not constitute a new intended use. The wording differences are details of device specifications (e.g. dimensions), specified circuit placement (e.g. lines subject to arterial pressures, line size), and
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "TERUMO" in green font. To the left of the word is a green circle with a "T" inside. The font is sans-serif and the word is in all capital letters.
circuit component choices (e.g. pump choice), which do not significantly impact the overall intended use of the valves in preventing retrograde flow during cardiopulmonary bypass procedures. These differences are not critical to the overall intended use of the device preventing retrograde flow during cardiopulmonary bypass procedures. Additionally, the differences do not affect the safety and effectiveness of the device when used as labeled. Performance testing was conducted with substantially equivalent results and is presented in this submission.
#### Principles of Operation, Technology, and Design:
The One Way Soft Valve is a single-use, disposable device that prevents retrograde flow in lines subject to arterial pressure by utilizing a closed path, mechanical, one-way valve. The design allows flow through the valve in one-direction and limits the flow in the opposing direction. The operating principles of the Ouest Medical Retroguard Arterial Safety Valve are identical to that of the device subject to this submission, the One Way Soft Valve.
The primary differences in design between the One Way Soft Valve and the predicate device are found in the material selection, the type of check-valve used, port dimensions, and overall prime volume.
- The first of these differences is the material choice. The One Way Soft Valve is o constructed of a flexible PVC housing and a flexible PVC valve, while the predicate valve is constructed of a rigid ABS housing and flexible silicone valve. Flexible Polyvinylchloride (PVC) is a common material used in Cardiopulmonary Bypass Circuits. Performance testing results are presented within this submission. The subject devices were tested to clinically relevant pressures to ensure that they could withstand both positive and negative pressures and still maintain desired performance.
- The second difference is the type of check-valve used. The One Way Soft Valve uses a 0 sleeve-like valve, while the predicate uses a duckbill valve design. Both designs allow flow through the valve in one-direction and limit flow in the opposing direction.
- The third difference is the adapter (or port) dimensions. The One Way Soft Valve has a o smaller port size (1/4" compared to 3/8"), which allows the One Way Soft Valve to be used in smaller diameter lines.
- Lastly, the One Way Soft Valve has a smaller prime volume than the predicate device. 0 The One Way Soft Valve has a prime volume of approximately 9 mL. The predicate device has a prime volume of approximately 20 mL. Submission of this 510(k) was primarily motivated by the clinical need for a smaller, more versatile valve for use in Cardiovascular Procedure Convenience Kits.
Testing was completed that demonstrates the One Way Soft Valve is as safe and effective as the predicate. Any differences in design between the One Way Soft Valve and the predicate do not raise different questions of safety and effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Terumo. The logo consists of a green circle with a white "T" inside of it. To the right of the circle is the word "TERUMO" in green, block letters.
#### Performance Evaluations:
Terumo Cardiovascular Systems has conducted in-vitro performance evaluations for the purpose of demonstrating that the One Way Soft Valve device is substantially equivalent to the predicate Quest Medical Retroguard Arterial Safety Valve.
Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. Substantial equivalence is demonstrated with the following in-vitro performance evaluations:
- One Way Soft Valve Tubing Connection (Bond) Strength Evaluation 0
- 0 One Way Soft Valve Mechanical/Structural Integrity/Burst Strength Evaluation
- o One Way Soft Valve Opening Pressure Evaluation
- 0 One Way Soft Valve Negative Pressures and Retrograde Flow Evaluation
- o One Way Soft Valve Circulation and Hemolysis Evaluation
The Terumo Cardiovascular Systems Corporation One Way Soft Valve exhibits substantially equivalent performance to the predicate, the Quest Medical Retroguard Arterial Safety Valve-K922356. Performance testing was conducted with substantially equivalent results and is presented in this submission.
#### Substantial Equivalence Statement:
The One Way Soft Valve device and the predicate device. the Quest Medical Retroguard Arterial Safety Valve - K922356, are substantially equivalent in intended use, principles of operation and technology, design and materials, and performance. Any noted differences between the subject device and the predicate devices do not raise new issues of safety and effectiveness.
#### Additional Safetv Information:
- Sterilization conditions have been validated to provide a Sterility Assurance Level (SAL) of . 106. Terumo further asserts that the ethylene oxide residues will not exceed the maximum residue limits at the time of product distribution.
- Terumo maintains biocompatibility studies as recommended in the FDA General Program o Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing." [External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration]. The blood contacting materials were found to be biocompatible.
#### Conclusion:
In summary, Terumo deems the Terumo Cardiovascular Systems Corporation One Way Soft Valve device is substantially equivalent to the predicate device, the Quest Medical Retroguard Arterial Safety Valve - K922356. with respect to intended use, duration of use, design, materials. principles of operation, and performance. It is further noted that any recognized differences do not raise any new issues of patient/user safety or product effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.