MAGNIFY and MAGNIFY-S Spacers

K142498 · Globus Medical, Inc. · OVD · Feb 26, 2015 · Orthopedic

Device Facts

Record IDK142498
Device NameMAGNIFY and MAGNIFY-S Spacers
ApplicantGlobus Medical, Inc.
Product CodeOVD · Orthopedic
Decision DateFeb 26, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MAGNIFY™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™ Spacer is to be filled with autogenous bone graft material, and is to be used with supplemental fixation, such as the CREO®, REVERE® or REVOLVE® Stabilization Systems. The MAGNIFY™-S Spacer is a stand-alone interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™-S Spacer is to be filled with autogenous bone graft material, and is to be used with three titanium alloy screws that accompany each implant.

Device Story

Expandable anterior lumbar interbody fusion device; provides structural stability post-discectomy. Device features height expansion ranges and geometric options; superior/inferior surface protrusions grip vertebral endplates to resist expulsion. Implant filled with autogenous bone graft. MAGNIFY™ requires supplemental fixation (e.g., CREO®, REVERE®, REVOLVE® systems); MAGNIFY™-S functions as stand-alone device using three integrated titanium alloy screws. Used in surgical setting by orthopedic/neurosurgical spine surgeons. Provides mechanical support to facilitate spinal fusion; benefits patient by restoring disc height and stabilizing affected vertebral levels.

Clinical Evidence

Bench testing only. Mechanical testing included static and dynamic compression, static and dynamic compression-shear, subsidence, and expulsion conducted per ASTM F2077 and F2267 and FDA guidance.

Technological Characteristics

Expandable interbody fusion device. Materials: Titanium alloy (ASTM F136, F1295), PEEK polymer (ASTM F2026), and optional hydroxyapatite coating (ASTM F1185). Features protrusions for endplate grip. MAGNIFY™-S includes three titanium alloy screws. Mechanical performance validated via ASTM F2077 and F2267.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous lumbosacral levels (L2-S1) with discogenic back pain, including up to Grade 1 spondylolisthesis or retrolisthesis, who have failed at least 6 months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping to create a sense of depth and connection. The profiles are rendered in a simple, line-art style. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 26, 2015 Globus Medical, Incorporated Kelly Baker, Ph.D. Senior Vice President, Regulatory and Clinical Affairs 2560 General Armistead Avenue Audubon, Pennsylvania 19403 Re: K142498 Trade/Device Name: MAGNIFY™ and MAGNIFY™-S Spacers Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD, MAX Dated: January 29, 2015 Received: January 30, 2015 Dear Dr. Baker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) #### K142498 Device Name MAGNIFY™ and MAGNIFYTM-S Spacers ### Indications for Use (Describe) The MAGNIFY™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™ Spacer is to be filled with autogenous bone graft material, and is to be used with supplemental fixation, such as the CREO®, REVERE® or REVOLVE® Stabilization Systems. The MAGNIFY™-S Spacer is a stand-alone interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™-S Spacer is to be filled with autogenous bone graft material, and is to be used with three titanium alloy screws that accompany each implant. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ # 510(k) Summary: MAGNIFY™ Spacers | Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800 | |-----------------|------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Kelly J. Baker, Ph.D.<br>Senior Vice President, Regulatory and Clinical Affairs | | Date Prepared: | September 4, 2014 | | Device Name: | MAGNIFYTM and MAGNIFYTM.S Spacers | | Classification: | Per 21 CFR as follows:<br>§888.3080 Intervertebral Body Fusion Device<br>Product Codes: OVD, MAX<br>Regulatory Class: II, Panel Code: 87 | Primary Predicate: CALIBER® Spacer (K102293) | Additional | INDEPENDENCE® Spacer (K082252 & K120101) | |-------------|-------------------------------------------| | Predicates: | CALIBER® Spacer (K123231) | | | PATRIOT® Continental® (K072970 & K122097) | # Purpose: The purpose of this submission is to request clearance for the MAGNIFY™ Spacers. # Device Description: MAGNIFY™ Spacers are expandable anterior lumbar interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. The devices are available in various height expansion ranges and geometric options to fit the anatomical needs of a wide variety of patients. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to aid in expulsion resistance. These devices are to be filled with autogenous bone graft material. The MAGNIFY™ Spacer is to be used with supplemental fixation. The MAGNIFY™-S Spacer is to be used with three titanium alloy screws that accompany the implant. MAGNIFY™ Spacers are manufactured from titanium alloy, as specified in ASTM F136, and include an internal component manufactured from radiolucent PEEK polymer, as specified in ASTM F2026. The screws used with MAGNIFY™-S are {4}------------------------------------------------ manufactured from titanium alloy, as specified in ASTM F136 and F1295, and are available with hydroxyapatite (HA) coating, as specified in ASTM F1185. # Indications for Use: The MAGNIFY™ Spacer is an interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™ Spacer is to be filled with autogenous bone graft material, and is to be used with supplemental fixation, such as the CREO®, REVERE® or REVOLVE® Stabilization Systems. The MAGNIFY™-S Spacer is a stand-alone interbody fusion device intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The MAGNIFY™-S Spacer is to be filled with autogenous bone graft material, and is to be used with three titanium alloy screws that accompany each implant. # Performance Data: Mechanical testing (static and dynamic compression, static and dynamic compression-shear, subsidence, and expulsion) was conducted in accordance with the "Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Intervertebral Fusion Device," June 12, 2007, ASTM F2077, and ASTM F2267 to demonstrate substantial equivalence to the predicate devices. # Basis of Substantial Equivalence: MAGNIFY™ Spacers have been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices. MAGNIFY™ Spacers are as safe, as effective, and perform as well as or better than the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...