K142496 · Collagen Matrix, Inc. · NPL · Nov 24, 2014 · Dental
Device Facts
Record ID
K142496
Device Name
Collagen Dental Membrane - Conformable IIBP
Applicant
Collagen Matrix, Inc.
Product Code
NPL · Dental
Decision Date
Nov 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Collagen Dental Membrane - Conformable IIBP is intended for use in oral surgical procedures as a resorbable membrane material for use in augmentation around implants placed in immediate extraction sockets, delayed extraction sockets; localized ridge augmentation for later implantation; alveolar ridge reconstruction for prosthetic treatment; filling of bone defects; guided bone regeneration in dehiscence defects and guided tissue regeneration procedures in periodontal defects.
Device Story
Collagen Dental Membrane - Conformable IIBP is a non-friable, resorbable barrier membrane derived from purified bovine pericardium. Supplied pre-hydrated and sterile for single-use in oral surgery. Device acts as a physical barrier to facilitate guided bone and tissue regeneration by preventing soft tissue ingrowth into bone defects while remaining permeable to nutrients. Used by dental surgeons in clinical settings; placed over bone graft sites or defects to support healing. Benefits include structural support for bone regeneration and conformability to complex defect contours. Device resorbs gradually over time, eliminating need for surgical removal.
Clinical Evidence
No clinical data submitted. Substantial equivalence established via bench testing, in vitro characterization, and in vivo animal studies (rabbit intra-oral and rat subcutaneous models). Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity, and genotoxicity, all yielding passing results.
Technological Characteristics
Resorbable membrane composed of purified bovine pericardium collagen. Thickness 0.3-1.0 mm. Semi-permeable; cross-linked via proprietary method. Supplied pre-hydrated, sterile (SAL 10^-6), non-pyrogenic. Biocompatible per ISO 10993 standards. Non-friable, conformable, and suture-compatible.
Indications for Use
Indicated for patients requiring oral surgical procedures involving bone augmentation around implants (immediate/delayed extraction sockets), localized ridge augmentation, alveolar ridge reconstruction, bone defect filling, guided bone regeneration in dehiscence defects, and guided tissue regeneration in periodontal defects.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, formed by three stylized human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 24,2014
Collagen Matrix, Inc. Ms. Gloria Zuclich Senior Manager of Regulatory Affairs 15 Thornton Road Oakland, New Jersey 07436
Re: K142496
Trade/Device Name: Collagen Dental Membrane - Conformable IIBP Regulation Number: 21 CFR 872.3930 Regulation Name: Bone grafting material Regulatory Class: II Product Code: NPL Dated: September 04, 2014 Received: September 05, 2014
Dear Ms. Zuclich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
K142496 510(k) Number (if known):
Device Name: Collagen Dental Membrane - Conformable IIBP
Indications for Use:
Collagen Dental Membrane - Conformable IIBP is intended for use in oral surgical procedures as a resorbable membrane material for use in augmentation around implants placed in immediate extraction sockets, delayed extraction sockets; localized ridge augmentation for later implantation; alveolar ridge reconstruction for prosthetic treatment; filling of bone defects; guided bone regeneration in dehiscence defects and guided tissue regeneration procedures in periodontal defects.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
# 510(k) Summary
#### 1. Application Information Applicant Name: Collagen Matrix. Inc. Address: 15 Thornton Road Oakland, New Jersey 07436 Telephone: (201) 405-1477 Fax: (201) 405-1355 Contact Person: Gloria Zuclich Senior Manager of Regulatory Affairs Date of Summary: November 24, 2014 2. Name of the Device Device Trade Name: Collagen Dental Membrane - Conformable IIBP Device Common Name: Collagen Dental Membrane Device Classification Name: Barrier, animal source, intraoral 21 CFR 872.3930 Product Code NPL Device Class II 3. Legally Marketed Devices to Which Substantial Equivalence is Claimed Predicate Device(s): Collagen Dental Membrane IV K090216 Tutodent® Dental Membrane CopiOs™ Pericardium Membrane K073097 4. Description of the Device
Collagen Dental Membrane - Conformable IIBP is a non-friable, resorbable membrane matrix consisting of purified intact collagen tissue derived from bovine pericardium. The membrane device is flexible and conforms to the contours of the defect site. Collagen Dental Membrane - Conformable IIBP is supplied pre-hydrated, sterile, non-pyrogenic, and for single use only.
The collagen dental membrane has a thickness of approximately 0.3 to 1.0 mm and is available in the following sizes 15 x 20 mm, 20 x 30 mm, and 30 x 40 mm.
#### 5. Intended Use
Collagen Dental Membrane - Conformable IIBP is intended for use in oral surgical procedures as a resorbable membrane material for use in augmentation around implants placed in immediate extraction sockets, delayed extraction sockets; localized ridge augmentation for later implantation; alveolar ridge reconstruction for prosthetic
{4}------------------------------------------------
treatment; filling of bone defects; guided bone regeneration in dehiscence defects and guided tissue regeneration procedures in periodontal defects.
## 6. Summary/Comparison of Technical Characteristics
Collagen Dental Membrane - Conformable IIBP and its predicates have similar technological characteristics. In particular, Collagen Dental Membrane - Conformable IIBP is substantially equivalent to Collagen Dental Membrane IV with regard to design and function. The bovine pericardium collagenous tissue material is also purified in the same manner as the bovine dermis collagenous tissue of Collagen Dental Membrane IV device. Collagen Dental Membrane - Conformable IIBP and Tutodent® Dental Membrane and CopiOs™ Pericardium Membrane are similar in design and function and utilize the same source of animal tissue, i.e. bovine pericardium.
| Parameter | Collagen Dental Membrane<br>-Conformable IIBP (This<br>submission) | Collagen Dental Membrane<br>IV K090216 | TutoDent® / CopiOs™<br>K073097 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Intended for use in oral<br>surgical procedures as a<br>resorbable membrane<br>material for use in<br>augmentation around<br>implants placed in immediate<br>extraction sockets, delayed<br>extraction sockets; localized<br>ridge augmentation for later<br>implantation; alveolar ridge<br>reconstruction for prosthetic<br>treatment; filling of bone<br>defects; guided bone<br>regeneration in dehiscence<br>defects and guided tissue<br>regeneration procedures in<br>periodontal defects. | Intended for use in oral<br>surgical procedures as a<br>resorbable membrane<br>material for use in<br>augmentation around<br>implants placed in immediate<br>extraction sockets, delayed<br>extraction sockets; localized<br>ridge augmentation for later<br>implantation; alveolar ridge<br>reconstruction for prosthetic<br>treatment; filling of bone<br>defects; guided bone<br>regeneration in dehiscence<br>defects and guided tissue<br>regeneration procedures in<br>periodontal defects. | Intended for use in oral<br>surgical procedures as a<br>resorbable material for<br>augmentation around<br>implants placed in immediate<br>extraction sockets, delayed<br>extraction sockets; localized<br>ridge augmentation for later<br>implantation; alveolar ridge<br>reconstruction for prosthetic<br>treatment; filling of bone<br>defects after root resection,<br>cystectomy, removal of<br>retained teeth; guided bone<br>regeneration in dehiscence<br>defects and guided tissue<br>regeneration procedures in<br>periodontal defects. |
| Material | Intact purified collagen tissue | Intact purified collagen tissue | Intact purified collagen tissue |
| Collagen Source | Bovine pericardium | Bovine dermis | Bovine pericardium |
| Form | Membrane | Membrane | Membrane |
| Color | White to off-white | White to off-white | White to off-white |
| Physical Integrity | Non-friable | Non-friable | Non-friable |
| Sizes | 15 x 20 mm<br>20 x 30 mm<br>30 x 40 mm | 15 x 20 mm<br>20 x 30 mm<br>30 x 40 mm | 15 x 20 mm<br>20 x 30 mm<br>30 x 40 mm |
| Thickness | Approx. 0.5 mm | Approx. 0.5 mm | Not tested |
| Conformability | Conformable to defect site | Conformable to defect site | Conformable to defect site |
| Suture Strength | Can be sutured | Can be sutured | Can be sutured |
| Porosity | Semi-permeable; permeable<br>to nutrients and<br>macromolecules | Semi-permeable; permeable<br>to nutrients and<br>macromolecules | Semi-permeable; permeable<br>to nutrients and<br>macromolecules |
| Cross-linked | Yes | Yes | No |
| Method of<br>Crosslinking | Proprietary | Proprietary | Not applicable |
A comparison between the subject and predicate devices is shown in the following table.
{5}------------------------------------------------
| Parameter | Collagen Dental Membrane<br>-Conformable IIBP (This<br>submission) | Collagen Dental Membrane<br>IV K090216 | TutoDent® / CopiOs™<br>K073097 |
|-----------------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Resorption<br>Time/In Vivo<br>Stability | Gradual resorption, resorbed<br>in approximately 16 weeks<br>as demonstrated in animal<br>testing | Gradual resorption, 6-9<br>months target | Gradual resorption, resorbed<br>in approximately 8 to 16<br>weeks as demonstrated in<br>animal testing |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible |
| How Supplied | Pre-hydrated | Dry, with hydration prior to<br>implantation | Dry, with hydration prior to<br>implantation |
| Sterility | Sterile, SAL 10-6 | Sterile, SAL 10-6 | Sterile, SAL 10-6 |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic |
| Single Use/Reuse | Single use only | Single use only | Single use only |
| Packaging | Double peel package | Double peel package | Double peel package |
# Nonclinical Tests Submitted
The substantial equivalence of Collagen Dental Membrane - Conformable IIBP and its predicates was demonstrated based on in vitro characterization studies, biocompatibility studies, in vivo animal studies, and clinical history of the predicate devices.
In vitro product characterization testing was performed to demonstrate substantial equivalence of the subject device to its predicate devices. A series of bench tests were conducted which included an evaluation of physical properties such as membrane thickness, conformability, suture strength, and an evaluation of physicochemical properties such as product permeability and hydrothermal transition temperature.
A series of in vitro and in vivo biocompatibility testing was performed to assess safety of the Collagen Dental Membrane-Conformable IIBP as an implantable material. The device passed all applicable FDA Blue Book Memorandum G95-1and ISO 10993-1 testing for the biological evaluation of medical devices.
| Test | Test Method/ Model | Results |
|------------------------------|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity | ISO Agarose Overlay<br>Method, ISO 10993-5 | Non-cytotoxic. No evidence of causing any cell lysis<br>or toxicity. |
| Sensitization | ISO Guinea Pig<br>Maximization Sensitization<br>Test, ISO 10993-10 | No evidence of sensitization was observed. The<br>test article was not considered a sensitizer. The test<br>article extract met the requirements of the test. |
| Intracutaneous<br>Reactivity | Acute Intracutaneous<br>Reactivity in Rabbit, ISO<br>10993-10 | No erythema and no edema from the test extract<br>injected intracutaneously into the rabbits. The test<br>article extract met the requirements of the test. |
| Acute Systemic<br>Toxicity | ISO Systemic Toxicity in<br>Mice, ISO 10993-11 | No mortality or evidence of systemic toxicity. The<br>test article extract met the requirements of the test. |
| Pyrogenicity | USP Pyrogen Study –<br>Material Mediated, ISO<br>10993-11 | Non-pyrogenic. The test article extract met the<br>requirements of the test. |
| Genotoxicity | In Vitro Mouse Lymphoma<br>Assay, ISO 10993-3 | None of the test article treatments induced<br>substantial increases in the number of revertant<br>colonies. Based on the criteria and conditions of the<br>study protocol, the test article is considered non-<br>mutagenic. The test article extract met the<br>requirements of the test. |
{6}------------------------------------------------
| Test | Test Method/ Model | Results |
|---------------------------------|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Genotoxicity | Bacterial Mutagenicity<br>Test - Ames Assay, ISO<br>10993-3 | Non-mutagenic to Salmonella typhimurium strains<br>TA97a, TA98, TA100, and TA1535 and to<br>Escherichia coli strain WP2-uvrA. The test article<br>extract met the requirements of the test. |
| Implantation | Subcutaneous<br>Implantation in Rats | Minimum tissue reaction up to 4 weeks of<br>implantation and no adverse tissue reaction to the<br>host. |
| Subchronic/<br>Chronic Toxicity | Subcutaneous<br>Implantation in Rats | Minimum tissue reaction up to 24 weeks of<br>implantation and no adverse tissue reaction to the<br>host. |
In addition to the in vitro characterization tests, animal studies using a rabbit intra-oral model, as well as a rat subcutaneous model were conducted to evaluate the in vivo stability and local tissue response to the subject device.
Viral inactivation studies were performed to ensure the viral safety of the product.
#### Clinical Test Submitted
Given the similarities between Collagen Dental membrane - Conformable IIBP and the predicate devices, it was determined that a clinical study would not be necessary to demonstrate substantial equivalences.
#### 7. Conclusion of Non-clinical Studies
The results of the in vitro product characterization and biocompatibility testing, as well as the animal study show that Collagen Dental Membrane-Conformable IIBP is as safe and substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.