The Vest® Airway Clearance Systems is intended to provide airway clearance therapy when external manipulation of the thorax is the physician's choice or treatment. Indications for this form of therapy are described by the American Association of Respiratory Care (AARC) in the Clinical Practices Guidelines for Postural Drainage Therapy. According to AARC guidelines, specific indications for external manipulation of the thorax include evidence or a suggestion of retained secretions, evidence that the patient is having difficulty with the secretion clearance, or presence of atelectasis caused by mucus plugging. In addition, the Vest® Airway Clearance System is also indicated for external manipulation of the thorax to promote airway clearance or improve bronchial drainage for the purposes of collecting mucus for diagnostic evaluation.
Device Story
The Vest® Airway Clearance System (Model 105) provides High Frequency Chest Wall Oscillation (HFCWO) therapy. The system comprises an Air Pulse Generator, a connecting hose, and an inflatable garment. The generator rapidly inflates and deflates the garment, applying gentle compression and release to the chest wall to create airflow within the lungs, mobilizing mucus toward large airways for clearance via coughing or suctioning. Used in clinical settings (critical care, acute care, skilled nursing, pulmonary labs) and home care, the device is operated by clinicians or patients. The system includes programmable modes (Cough Pause®, Ramp) and Bluetooth™ connectivity. It is intended to assist patients with secretion clearance, potentially reducing mucus plugging and improving bronchial drainage.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included comparative pressure mapping of garments, environmental and mechanical testing, and compliance with electrical safety and electromagnetic compatibility standards (ANSI/AAMI ES60601-1, IEC 60601-1, FCC Part 15B, ETSI EN 300 328).
Technological Characteristics
System utilizes HFCWO via an Air Pulse Generator (100–230 VAC) and inflatable garments. Connectivity includes Bluetooth™. Safety standards met: UL/EN/IEC 60601-1, CAN/CSA C22.2 No. 601.1, and various ETSI standards for wireless/EMC. Materials are identical to the predicate and do not contact patient skin.
Indications for Use
Indicated for pediatric to geriatric patients requiring airway clearance therapy due to retained secretions, difficulty with secretion clearance, or atelectasis caused by mucus plugging, or for bronchial drainage to collect mucus for diagnostic evaluation.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of a human figure with three faces in profile, stacked on top of each other. The symbol is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the border of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 7, 2015
Hill-Rom Private Limited % Mr. Paul Dryden President ProMedic Inc. 24301 Woodsage Dr. Bonita Springs, FL 34134
Re: K142482
Trade/Device Name: The Vest® Airway Clearance System Regulation Number: 21 CFR 868.5665 Regulation Name: Percussor, Powered-Electric Regulatory Class: II Product Code: BYI Dated: April 3, 2015 Received: April 6, 2015
## Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known)
# K142482
Device Name
The Vest® Airway Clearance Systems (Model 105)
# Indications for Use (Describe)
The Vest® Airway Clearance Systems is intended to provide airway clearance therapy when external manipulation of the thorax is the physician's choice or treatment. Indications for this form of therapy are described by the American Association of Respiratory Care (AARC) in the Clinical Practices Guidelines for Postural Drainage Therapy. According to AARC guidelines, specific indications for external manipulation of the thorax include evidence or a suggestion of retained secretions, evidence that the patient is having difficulty with the secretion clearance, or presence of atelectasis caused by mucus plugging. In addition, the Vest® Airway Clearance System is also indicated for external manipulation of the thorax to promote airway clearance or improve bronchial drainage for the purposes of collecting mucus for diagnostic evaluation.
Type of Use (*Select one or both, as applicable*)
| | <b>XX</b> Prescription Use (Part 21 CFR 801 Subpart D) |
|--|--------------------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the Hill-Rom logo. The logo is set against a blue, rounded rectangular background. The text "Hill-Rom" is written in a bold, sans-serif font and is white.
#### 510(k) Summary Page 1 of 5
| Date Prepared | 4-May-15 | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--|
| Hill-Rom Services Pte Ltd<br>1 Yishun Ave 7<br>Singapore, 768923, Singapore | Tel - 011 - 65 65945215<br>Fax - 011 - 65 65945201 | |
| Official Contact: | Christina M. Miracle<br>1069 State Route 46 East<br>Batesville, IN 47006<br>Tel – 812-931-3590<br>Fax - 812-934-1675 | |
| Proprietary or Trade Name: | The Vest® Airway Clearance Systems (Model 105 | |
| Common/Usual Name: | Percussor, Powered-Electric | |
| Classification Code/Name: | BYI – Powered Percussor<br>21 CFR 868.5665<br>Class 2 | |
| Predicate Devices: | K024309 - The Vest® Airway Clearance System | |
#### Device Description:
The Vest® Airway Clearance System consists of an inflatable garment attached to an Air Pulse Generator that rapidly inflates and deflates the inflatable garment. This causes the chest wall to be gently compressed and released, which creates airflow within the lungs. This process moves the mucus toward the large airways where it can be cleared by coughing or suctioning. This type of Airway Clearance Therapy is referred to as High Frequency Chest Wall Oscillation (HFCWO).
The system consists of an Air Pulse Generator, connecting hose, and multiple styles and sizes of garments. The garments are single patient, multi-use.
#### Indications for Use:
The Vest® Airway Clearance Systems is intended to provide airway clearance therapy when external manipulation of the thorax is the physician's choice or treatment. Indications for this form of therapy are described by the American Association of Respiratory Care (AARC) in the Clinical Practices Guidelines for Postural Drainage Therapy. According to AARC guidelines, specific indications for external manipulation of the thorax include evidence or a suggestion of retained secretions, evidence that the patient is having difficulty with the secretion clearance, or presence of atelectasis caused by mucus plugging. In addition, the Vest® Airway Clearance System is also indicated for external manipulation of the thorax to promote airway clearance or improve bronchial drainage for the purposes of collecting mucus for diagnostic evaluation.
#### Patient Population:
The Vest® Airway Clearance System is applicable for a wide variety of settings as recommended by the AARC Guidelines, including: Critical care, in-patient acute care, extended care and skilled
### 510(k) Summary
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image features the logo for Hill-Rom, a company specializing in medical technologies. The logo is displayed in a rounded, rectangular shape with a light blue background. The text "Hill-Rom." is written in bold, white letters, creating a clean and professional look.
# Page 2 of 5 4-May-15
nursing facility care, home care, outpatient / ambulatory care, pulmonary diagnostic (bronchoscopy) laboratory.
The Vest® Airway Clearance System can be used for the pediatric to geriatric population. The system accommodates use across the targeted population by providing a selection of garment sizes which is used along with the Air Pulse Generator.
### Predicate Device Comparison:
The Vest® Airway Clearance Systems (Model 105) is substantially equivalent to the predicate device, Vest® Airway Clearance Systems (K024309). We present a summary of the comparison in Table 1.
#### Table 1 - Substantial Equivalence Comparative Table
| | Proposed Vest® Airway Clearance<br>System | Predicate<br>Vest® Airway Clearance System<br>(K024309) |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CFR<br>Classification | 868.5665 | 868.5665 |
| Classification | BYI | BYI |
| Classification name | Percussor, Powered-Electric | Percussor, Powered-Electric |
| Indications for Use | The Vest® Airway Clearance Systems<br>is intended to provide airway clearance<br>therapy when external manipulation of<br>the thorax is the physician's choice or<br>treatment. Indications for this form of<br>therapy are described by the American<br>Association of Respiratory Care<br>(AARC) in the Clinical Practices<br>Guidelines for Postural Drainage<br>Therapy. According to AARC<br>guidelines, specific indications for<br>external manipulation of the thorax<br>include evidence or a suggestion of<br>retained secretions, evidence that the<br>patient is having difficulty with the<br>secretion clearance, or presence of<br>atelectasis caused by mucus plugging.<br>In addition, the Vest® Airway<br>Clearance System is also indicated for<br>external manipulation of the thorax to<br>promote airway clearance or improve<br>bronchial drainage for the purposes of<br>collecting mucus for diagnostic<br>evaluation. | The Vest® Airway Clearance Systems<br>is intended to provide airway clearance<br>therapy when external manipulation of<br>the thorax is the physician's choice or<br>treatment. Indications for this form of<br>therapy are described by the American<br>Association of Respiratory Care<br>(AARC) in the Clinical Practices<br>Guidelines for Postural Drainage<br>Therapy. According to AARC<br>guidelines, specific indications for<br>external manipulation of the thorax<br>include evidence or a suggestion of<br>retained secretions, evidence that the<br>patient is having difficulty with the<br>secretion clearance, or presence of<br>atelectasis caused by mucus plugging.<br>In addition, the Vest® Airway<br>Clearance System is also indicated for<br>external manipulation of the thorax to<br>promote airway clearance or improve<br>bronchial drainage for the purposes of<br>collecting mucus for diagnostic<br>evaluation. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Hill-Rom logo. The logo is in a blue rounded rectangle. The text "Hill-Rom" is in white and is centered in the rectangle.
### 510(k) Summary Page 3 of 5 4-May-15
| | Proposed Vest® Airway Clearance<br>System | Predicate<br>Vest® Airway Clearance System<br>(K024309) |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Environments of Use | Critical care, in-patient acute care,<br>extended care and skilled nursing<br>facility care, home care, outpatient /<br>ambulatory care, pulmonary<br>diagnostic (bronchoscopy)<br>laboratory. | Critical care, in-patient acute care,<br>extended care and skilled nursing<br>facility care, home care, outpatient /<br>ambulatory care, pulmonary<br>diagnostic (bronchoscopy)<br>laboratory. |
| Patient Population | Pediatric to geriatric population | Pediatric to geriatric population |
| Systems consists of | Air Pulse Generator<br>Connecting hose (60") locking / non-<br>locking<br>Garments of various styles and sizes<br>Chest (permanent)<br>Full Vest (Permanent / SPU / C3)<br>Wrap Vest (Permanent / SPU) | Air Pulse Generator<br>Connecting hose (60-80") non-locking<br>Garments of various styles and sizes<br>Chest (permanent)<br>Full Vest (Permanent / SPU) |
| Software Modes | Manual (Normal)<br>Bluetooth™ - Yes<br>Operational modes:<br>Programmable<br>Cough Pause®<br>Ramp | Manual (Normal)<br>Operational modes:<br>Manual mode |
| Electrical<br>Specifications | 100 – 230 VAC, 50 / 60 Hz<br>3.4 A @ 100 VAC / 2.0 A @ 230 VAC<br>Power plug – NEMA 1-15P (2 pin) | 100 – 230 VAC, 50 / 60 Hz<br>5.0 A max<br>Power plug – NEMA 5-15P (3 pin) |
| Mechanical safety | UL/EN/IEC 60601-1 and CAN/CSA<br>C22.2 No. 601.1 | UL/EN/IEC 60601-1 and CAN/CSA<br>C22.2 No. 601.1 |
### Indications -
The proposed indications for use of providing airway clearance therapy when external manipulation of the thorax is the physician's choice or treatment and to promote airway clearance or improve bronchial drainage for the purposes of collecting mucus for diagnostic evaluation.are identical to the predicate. Discussion: The indications for use are identical to the predicate Vest® Airway Clearance System (K024309).
#### Technology -
The design of the Vest® Airway Clearance System utilizing an Air Pulse Generator, connecting hose, and inflatable / deflatable garment is unchanged from the predicate.
Discussion: The design is identical with minor modifications to the connecting hose and addition of garment styles, but in all substantially equivalent to the predicate Vest® Airway Clearance System (K024309).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Hill-Rom. The logo is in a rounded blue rectangle. The text "Hill-Rom." is in white.
#### 510(k) Summary Page 4 of 5 4-May-15
# Environment of Use -
The proposed environments of use are Critical care, in-patient acute care, extended care and skilled nursing facility care, home care, outpatient / ambulatory care, pulmonary diagnostic (bronchoscopy) laboratory which are identical to the predicate.
Discussion: The intended environments of use are identical to the predicate Vest® Airway Clearance System (K024309).
#### Patient Population -
The Vest® Airway Clearance System (K024309) is intended for pediatric to geriatric population. Discussion: The patient population is identical to the predicate Vest® Airway Clearance System (K024309).
### Discussion of Differences -
The basic differences between the proposed device and the predicate are an update in software, addition of styles of garments and addition of a locking style connecting hose. Testing has demonstrated that these differences do not raise any new safety or risks and thus can be found to be substantially equivalent.
### Non-clinical Testing Summary -
We performed comparative non-clinical bench testing to demonstrate that the Vest® Airway Clearance System and its modifications and additional garment styles and sizes perform equivalent to the predicate Vest® Airway Clearance System (K024309). These tests included:
- Comparative Pressure mapping of garments with predicate and proposed Air Pulse ● Generator
- Environmental and Mechanical testing ●
- ANSI/AAMI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and A2:2010/(R)2012 ●
- 47 CFR FCC Part 15B:2012 (CLASS B)
- IEC 60601-1:2005+CORR,1 (2006) + CORR. 2 (2007) ●
- EN 60601-1:2006 +CORR: 2010 ●
- CAN/CSA-C22.2 No. 6060101 (2008)
- . EN 60601-102: 2007
- ETSI EN 300 328 V1.7.1:2006
- ETSI EN 301 489-1 V1.9.2:2011 ●
- ETSI EN 301 489-17 V2.2.1:2012 ●
### Biocompatibility of Materials -
The materials of the garments are not in direct or indirect contact with the patient skin or tissue. The materials are identical to the predicate as well.
Discussion: Biocompatibility testing is not needed.
#### 510(k) Summary Page 5 of 5
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Hill-Rom. The logo is in a blue rounded rectangle. The text "Hill-Rom." is in white and is centered in the rectangle.
# 4-May-15
# Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.