K142473 · Sion Biotext Medical , Ltd. · KMJ · Apr 13, 2015 · General Hospital
Device Facts
Record ID
K142473
Device Name
Lubricating Jelly
Applicant
Sion Biotext Medical , Ltd.
Product Code
KMJ · General Hospital
Decision Date
Apr 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6375
Device Class
Class 1
Indications for Use
The Lubricating Jelly is a medical device intended for medical purposes, to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when either sterile or non-sterile fields are required.
Device Story
Lubricating jelly; clear, greaseless, odorless, water-soluble, non-irritating. Used to lubricate body orifices; facilitates entry of diagnostic/therapeutic devices (catheters, endoscopes, surgical instruments). Applied by clinicians in medical settings. Provided in sterile or non-sterile formats; packaged in bacterial barrier tubes or sealed laminated sachets. Benefits patient by reducing friction during medical device insertion, minimizing tissue irritation. No complex processing or software involved.
Clinical Evidence
Bench testing only. Evaluated via physical, chemical, and microbiological performance tests: appearance, weight, pH, viscosity, sterility (ISO 11737-2, USP <71>), preservative effectiveness (USP <51>), dye test (ASTM 1929), and biocompatibility (ISO 10993-1, -5, -10, -11). Sterilization validated per ISO 11137-2 and AAMI TIR 33; packaging integrity per ISO 11607-1. All tests met pre-defined acceptance criteria.
Technological Characteristics
Composition: water, carbomer thickeners, methylparaben, propylparaben, sodium hydroxide, and lubricator. Form factor: tubes or laminated sachets. Sterilization: Gamma radiation (for sterile version). Biocompatibility evaluated per ISO 10993-1:2009.
Indications for Use
Indicated for lubrication of body orifices to facilitate insertion of diagnostic or therapeutic devices (e.g., catheters, endoscopes, surgical instruments) in patients requiring such procedures.
Regulatory Classification
Identification
A patient lubricant is a device intended for medical purposes that is used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community or interconnectedness. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the organization's name and national affiliation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2015
Sion Biotext Medical, Limited c/o Ms. Tali Hazan Talmed Ltd. M.P Upper Galilee Ramont Naftali, Ha Zafon 1383000 Israel
Re: K142473
Trade/Device Name: Lubricating Jelly Regulation Number: 21 CFR 880.6375 Regulation Name: Patient Lubricant Regulatory Class: I Product Code: KMJ Dated: March 5, 2015 Received: March 11, 2015
Dear Ms. Hazan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Tali Hazan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Sion Biotext Medical Ltd. The logo is in blue and teal colors. The address Hagoshrim, 12225, ISRAEL is printed at the bottom left and the website www.sb-medical.com is printed at the bottom right.
Tel: 972-4-6956201, Fax: 972-4-6956202 E-mail:sbmedical@sb-medical.com
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
Indications for Use
| 510(k) Number (if known) | K142473 |
|--------------------------|-------------------|
| Device Name | Lubricating Jelly |
Indications for Use (Describe) The Lubricating Jelly is a medical device intended for medical purposes, to lubricate body onlices to facilitate entry of diagnostic or therapeutic devices when either sterile or non-sterile fields are required.
Image /page/2/Figure/4 description: This image shows a section of the FDA form 3881 (1/14). The section is for FDA use only and includes a digital signature by Mary E. Brooks -S, dated 2015.04.01. The form indicates that it applies to the requirements of the Paperwork Reduction Act of 1995 and provides contact information for the PRA Staff at PRAStaff@fda.hhs.gov.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Sion Biotext Medical Ltd 510(k) Lubricating Jelly
.............................................................................................................................................................................. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------July 30, 2014 4-1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Sion Biotext Medical Ltd. The logo features the company name in bold, dark blue letters, with "Sion Biotext" on the top line and "Medical Ltd." on the second line. Below the company name, there is contact information, including the address "Hagoshrim, 12225, ISRAEL," the website "www.sb-medical.com," the telephone number "972-4-8956201," the fax number "972-4-8956202," and the email address "smedical@sb-medical.com".
# 510(K) SUMMARY FOR SION BIOTEXT MEDICAL'S LUBRICATING JELLY Device
## DATE PREPARED: JULY 25, 2014
#### 510(K) OWNER NAME 1.
Sion Biotext Medical Ltd. Kibbutz Hagoshrim Upper Galilee Zip Code: 12225, Israel Phone: +972-4-6956201, Fax: +972-4-6956202, E-mail: sbmedical@sb-medical.com.
### Contact person:
Ms. Tali Hazan – RA Consultant Phone: +972-(0)50-5292-304 Fax: +972-(0)722448981 Email: tali.hazan@talmed.co.il
#### DEVICE NAME AND CLASSIFICATION 2.
Common/Usual Name: Lubricating Jelly (Sterile or Non-Sterile) Proprietary/Trade name: Lubricating Jelly Classification: Sion Biotext Medical's Lubricating Jelly device has been classified as
Class I device under the following classification names:
| Classification Name | Product Code | 21 CFR Ref. | Review Panel |
|---------------------|--------------|-------------|------------------|
| Lubricant, Patient | KMJ | 880.6375 | General Hospital |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Sion Biotext Medical Ltd. The logo features the company name in a bold, blue font, with "Medical Ltd." in a smaller font size. Below the logo, the address "Hagoshrim, 12225, ISRAEL" is printed, along with the website address "www.sb-medical.com" and contact information including the telephone number 972-4-6956201, fax number 972-4-8956202, and email address sbmedical@sb-medical.com.
### રે
Sion Biotext Medical's Lubricating Jelly device is substantially equivalent to the following Predicate Device:
Dynarex Sterile Lubricating Jelly (by Dynarex Corporation) cleared under 510(k) number K092488 on December 18, 2009.
#### 4. DEVICE DESCRIPTION
Sion's Lubricating Jelly is used to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when either sterile or non-sterile fields are required. Typical uses of the lubricating jelly include (but not limited) insertion of catheters, endoscopes surgical instruments, gynecological, rectal, oral and rubber or plastic etc. The lubricating jelly is clear, greaseless, odorless, water soluble and non-irritating to the skin, tissue and mucous membranes.
Both product types (sterile and non-sterile) are packed in a bacterial barrier tubes or sealed laminated sachets.
Sion's Lubricating Jelly is a body contact material and was evaluated for biocompatibility with accordance to FDA's Memorandum - #G95 1, May 1, 1995 and ISO 10993-1:2009.
#### 5. INTENDED USE
The Lubricating Jelly is a medical device intended for medical purposes, to lubricate body orifices to facilitate entry of diagnostic or therapeutic devices when either sterile or non-sterile fields are required.
#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.
The Lubricating jelly is composed of water, carbomer thickeners, methylparaben, propylparaben, sodium hydroxide and lubricator. These ingredients are the same as those used for the predicate device and formulation is nearly the same. Sion's Lubricating jelly is substantially equivalence to the predicate since it has the very same indication for use; it performs the same and has the same properties and labeling claims. It is intended to be used for the same patients' population.
{5}------------------------------------------------
The only differences between Sion's proposed device and Dynarex cleared device are that Sion provides the device as both 'sterile' and 'non-sterile' while Dynarex provides the product in sterile state only. Minor changes in the formulation were assessed and they do not affect the final device substantial equivalency determination.
#### PERFORMANCE DATA 7.
Sion's Lubricating Jelly has been successfully tested for physical non-clinical performances' tests. These non-clinical testing data performed to evaluate device performances:
| No. | Test Name | Per Standard |
|-----|---------------------------------------------------|-----------------------------|
| 1. | Chemical and Microbiological<br>along Shelf Life: | |
| 1.1 | Appearance | Specification |
| 1.2 | Weight | Specification |
| 1.3 | pH | Specification |
| 1.4 | Viscosity | Specification |
| 1.5 | Sterility and GPT | ISO 11737-2,<br>USP <71> |
| 1.6 | Preservative Effectiveness | USP <51> |
| 1.7 | Dye Test | ASTM 1929 |
| 2. | Biocompatibility: | 10993-1 (General) |
| 2.1 | Cytotoxicity | 10993-5 |
| 2.2 | Sensitization | 10993-10 |
| 2.3 | Irritation - Mucosal (Vaginal) | 10993-10 |
| 2.4 | Acute Systemic Toxicity | 10993-11 |
| 3. | Gammy Sterilization<br>Validation | ISO 11137-2,<br>AAMI TIR 33 |
| 4. | Sterile Packaging Integrity | ISO 11607-1 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Sion Biotext Medical Ltd. The logo is in blue and teal colors. The address Hagoshirm, 12225, ISRAEL, the website www.sb-medical.com, the telephone number 972-4-6956201, the fax number 972-4-6956202, and the email address sbmedical@sb-medical.com are also included in the image.
All tested devices met the tests' requirements and pre-defined acceptance criteria. Therefore it was concluded that under normal use, the device will performed according to its specifications and intended uses.
Performance tests' results supported Sion's labeling claims and the determination of substantial equivalence with predicate device.
#### 8. SUBSTANTIAL EQUIVALENCE
Sion's Lubricating Jelly is substantially equivalent to the predicate device selected in terms of indication for use, technology, performances, physical characterization, sterile product's sterility method of use, target patient population and nature of body contact.
#### 9. CONCLUSIONS
As above described, the evaluation of our device performances demonstrates that it is equivalent and as safe and as effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.