# Fixtemp Cement (K142446)

_Exacta Dental Direct, Inc. · EMA · Jun 4, 2015 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K142446

## Device Facts

- **Applicant:** Exacta Dental Direct, Inc.
- **Product Code:** [EMA](/productcode/EMA.md)
- **Decision Date:** Jun 4, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The intended use of the Fixtemp Cement is as a dental luting agent for use in interim dental restorations to retain temporary restorations.

## Device Story

FixTemp Cement is a zinc oxide-based, eugenol-free dental luting agent used for temporary cementation of provisional crowns, bridges, and semi-permanent implants. It functions via chemical reaction to create an adhesive bond. The device is supplied in double cartridges or syringes for 1:1 ratio dispensing. It is applied by dental professionals directly onto the temporary restoration and into the cavity. By providing a secure yet temporary attachment, it allows for interim restoration retention without the potential sensitivity issues associated with eugenol-containing cements. It is intended for use in a clinical dental setting.

## Clinical Evidence

Bench testing only. Performance verified against ISO 3107:2004 standards. Testing confirmed working time, film thickness (<20 µm), and compressive strength (>8 MPa) are equivalent to the predicate device.

## Technological Characteristics

Zinc oxide-based, eugenol-free dental luting cement. Formulated as a two-paste system (1:1 ratio) in double cartridges or syringes. Complies with ISO 3107:2004 standards. Storage at room temperature (15-25°C).

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- TempoCem ([K110759](/device/K110759.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles facing to the right. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 4, 2015

Exacta Dental Direct Inc. c/o Mr. Richard Keen Compliance Consultants 1151 Hope Street Stamford, CT 60803

Re: K142446

Trade/Device Name: FixTemp Cement Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: February 28, 2015 Received: March 6, 2015

#### Dear Mr. Keen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## 4. Indications for Use Statement

no 510(K) number assigned_ 510(K) Number (If k K142446 Name: FixTemp™ Cement

#### INDICATIONS FOR USE

FixTemp ™ Cement is a dental luting agent indicated for:

- temporary cementation of provisional crowns and bridges, ●
- cementing of semi-permanent implants. ●

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use | XXX |
|------------------|-----|
|------------------|-----|

or

| Over - The - Counter Use |                      |
|--------------------------|----------------------|
|                          | (Per 21 CFR 801.109) |

(Optional Format 1-2-96)

#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

| Page 14 | Fixtemp Cement |
|---------|----------------|
|---------|----------------|

{3}------------------------------------------------

## 5. 510(k) Summary

(per 21 CFR 807.92)

#### 28 May 2015

Sponsor Mr. John Pankuch President

Exacta Dental Direct, Inc.

44780 Macomb Industrial Dr. Clinton Township, MI 48036 (800) 474-7665 416-253-6060 416-253-1911 Fax john@exactadental.com

Consultant

Mr. Richard Keen Compliance Consultants 1151 Hope Street Stamford, CT 06907-1659 203 329 2700 F 203 329 2345 rkeen@fda-complianceconsultants.com

Proprietary Name: Fixtemp Cement Common Name Fixtemp Cement Device Classification Name Dental Cement Classification Number: 21 CFR 872.3275 Product Code EMA Reviewing Group Dental Devices Panel Device Classification Class II Establishment registration No. Owner/Operator 9043538/ Registration # 1836392 Predicate Device K110759/ TempoCem / DMG USA, Inc. Trademark Notice: All Trademarks used other than those of Exacta Dental Direct, Inc. Exacta

Dental Direct, Inc. are registered to their respective owners.

#### Device Description

The Fixtemp Cement is designed as a luting cement for temporary dental applications. This dental cement having the active ingredients of zinc oxide does not include Eugenol. This lack of Eugenol does not reduce structural properties but allows a choice among healthcare professionals where Eugenol can cause objectionable results.

#### Indications For Use

FixTemp "10 Cement is a dental luting agent indicated for:

- temporary cementation of provisional crowns and bridges, ●
- cementing of semi-permanent implants.

#### Intended Use

The intended use of the Fixtemp Cement is as a dental luting agent for use in interim dental restorations to retain temporary restorations.

The scientific concept on which this device is based is the principle that a chemical reaction with zinc oxide forms an adhesive bond to act as a dental luting agent. This device functions by creating adhesion by chemical action.

Fixtemp Cement

{4}------------------------------------------------

# 5. 510(k) Summary

(per 21 CFR 807.92)

#### Substantial Equivalence

Exacta Dental Direct, Inc. has determined that the Fixtemp Cement is substantially equivalent to the performance of a predicate Device. Fixtemp Cement is equal to this predicate. The differences between these devices are incidental and not significant. Both devices use similar technological characteristics and principles.

#### Testing

The Fixtemp Cement has benefited from design, development, testing and production procedures that conform to ISO 13485 Quality Systems.

Testing has confirmed this device meets its product specification. A series of factory tests are conducted to verify the intended signals are accurate and can maintain performance over its useful life. Testing has established this device is equal to predicate medical device in terms of working time, film thickness, and compressive strength (see comparison table).

#### Conclusion

There are no substantial differences between the Fixtemp Cement defined in this 510(k) submission and the stated predicate device. This device is equal to predicates currently used in other similar medical devices.

Exacta Dental Direct, Inc. continues to search all appropriate sources for information relating to safety and effectiveness and maintains an in-house reporting device to identify adverse safety and effectiveness information and as such, applicable data is recorded for this product.

{5}------------------------------------------------

### Table 12-1 Substantially Equivalent Comparison to Predicate.doc comparison of the FixTemp ™ Cement to the predicate: TempoCem K110759

| Description/Feature           | Predicate Device TempoCem K110759                                                                                               | FixTemp Cement                                                                 | Similarity | Difference | Comment                                                                 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------|------------|-------------------------------------------------------------------------|
| regulatory requirements       |                                                                                                                                 |                                                                                |            |            |                                                                         |
| Description                   | temporary luting cement containing Eugenol that is based on zinc oxide in double cartridges                                     | Eugenol free, temporary luting cement based on zinc oxide in double cartridges |            | Different  | The absence of Eugenol simply eliminates the depression of sensitivity. |
| Classification Name:          | Dental cement                                                                                                                   | Dental cement                                                                  | Same       |            | Both devices are dental cements                                         |
| Classification                | II                                                                                                                              | II                                                                             | Same       |            | Both devices are subject to same FDA regulation.                        |
| Regulation Number             | 21 CFR 872.3275 (a)                                                                                                             | 21 CFR 872.3275 (b)                                                            | Same       |            | Both are subject to same FDA regulations                                |
| Product Code                  | EMB                                                                                                                             | EMA                                                                            |            | Different  | Product codes reflect presence or absence of Eugenol                    |
| Difference related to Eugenol | pleasant clove taste oil                                                                                                        | Eugenol may inhibit future dental procedures                                   |            | Different  | Some practitioners prefer Eugenol and some avoid it.                    |
| Indications for Use           | Indicated for temporary cementation of crowns and bridges or provisional cementation of crowns and bridges on implant abutments | Indicated for luting of temporary dental prosthesis (crowns, inlays, onlays)   | Same       |            |                                                                         |
| Intended Use                  | Temporary luting and cementation of temporary crowns and bridges                                                                | Temporary luting and cementation of temporary crowns and bridges               | Same       |            |                                                                         |
| What condition does it treat? | Applied directly on the temporary restoration and into the cavity.                                                              | Applied directly on the temporary restoration and into the cavity.             | Same       |            |                                                                         |

{6}------------------------------------------------

### Table 12-1 Substantially Equivalent Comparison to Predicate.doc comparison of the FixTemp ™ Cement to the predicate: TempoCem K110759

| Description/Feature                                                   | Predicate Device                                                                                                                                  | FixTemp tm Cement                                                                                                | Similarity | Difference | Comment |  |  |  |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------|------------|---------|--|--|--|
|                                                                       | TempoCem K110759                                                                                                                                  |                                                                                                                  |            |            |         |  |  |  |
| characteristics                                                       |                                                                                                                                                   |                                                                                                                  |            |            |         |  |  |  |
| Compliance to standard                                                | ISO3107                                                                                                                                           | ISO3107                                                                                                          | same       |            |         |  |  |  |
| container                                                             | double cartridge or syringe                                                                                                                       | double cartridge or syringe                                                                                      | same       |            |         |  |  |  |
| Dispenser                                                             | injector / syringe                                                                                                                                | injector / syringe                                                                                               | same       |            |         |  |  |  |
| Paste Ratio                                                           | 1:1 ?                                                                                                                                             | 1:1                                                                                                              | same       |            |         |  |  |  |
| Appearance                                                            | Homogeneous and<br>smoothly consistent                                                                                                            | Homogeneous and<br>smoothly consistent                                                                           | same       |            |         |  |  |  |
| Film Thickness (ISO<br>3107)                                          | Less than 20 µm                                                                                                                                   | Less than 20 µm                                                                                                  | same       |            |         |  |  |  |
| Setting Time (ISO 3107)                                               | 4 min                                                                                                                                             | Approx. 4 - 7 min                                                                                                | same       |            |         |  |  |  |
| Compressive strength (ISO<br>3107)                                    | 8 MPa                                                                                                                                             | More than 8MPa                                                                                                   | same       |            |         |  |  |  |
| temporary cement                                                      | yes                                                                                                                                               | yes                                                                                                              | same       |            |         |  |  |  |
| temporary cementation of<br>provisional crowns & bridges              | yes                                                                                                                                               | yes                                                                                                              | same       |            |         |  |  |  |
| cementing of semi-permanent<br>crown or bridges on top of<br>implants | yes                                                                                                                                               | yes                                                                                                              | same       |            |         |  |  |  |
| Meets ISO 3107:2004                                                   | yes                                                                                                                                               | yes                                                                                                              | same       |            |         |  |  |  |
| Working time                                                          | 1 minute                                                                                                                                          | More than 1 min                                                                                                  | same       |            |         |  |  |  |
| Storage                                                               | Room temperature, dry<br>(15-25°C)                                                                                                                | Room temperature, dry<br>(15-25°C)                                                                               | Same       |            |         |  |  |  |
| applications                                                          |                                                                                                                                                   |                                                                                                                  |            |            |         |  |  |  |
| What is the treatment<br>environment?                                 | Applied directly on the<br>temporary restoration and<br>into the cavity.                                                                          | Applied directly on the<br>temporary restoration and<br>into the cavity.                                         | Same       |            |         |  |  |  |
| Is device a treatment<br>device?                                      | Yes                                                                                                                                               | Yes                                                                                                              | Same       |            |         |  |  |  |
| What patient population<br>does this apply?                           | for temporary tooth filling.<br>to affix dental devices<br>such as crowns or bridges,<br>or to be applied to a tooth<br>to protect the tooth pulp | for temporary luting<br>cements to affix dental<br>prosthesis such as crowns<br>or bridges, inlays and<br>onlays | Same       |            |         |  |  |  |
| Anatomical Sites                                                      | all human teeth                                                                                                                                   | all human teeth, except<br>allergies                                                                             | Same       | Same       |         |  |  |  |

Page 129 to 133

---

**Source:** [https://fda-staging.innolitics.com/device/K142446](https://fda-staging.innolitics.com/device/K142446)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).