Mini and Micro Fragments Reconstruction System-NEOFIX
Applicant
Neoortho Productos Orthopedicos S/A
Product Code
HRS · Orthopedic
Decision Date
Dec 16, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Mini and Micro Fragments Reconstruction System – NEOFIX is intended for small bone fracture fixation, arthrodesis, reconstruction, and osteotomy fixation. Examples include small bones of the hand, wrist, foot and ankle. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone.
Device Story
System consists of titanium plates and screws for orthopedic fixation; plates available in various geometric configurations and thicknesses (0.7-2.7 mm); screws include self-tapping cortical, locking, and non-locking designs. Used by orthopedic surgeons in clinical/surgical settings for trauma and reconstructive procedures. Surgeon selects appropriate plate/screw configuration based on patient anatomy and fracture type; implants provide mechanical stabilization of bone segments to facilitate healing. Benefits include restoration of bone alignment and stability in small bone fractures, including those with osteopenic bone.
Clinical Evidence
Bench testing only. Evidence includes detailed dimensional analysis, engineering analysis, and mechanical testing of subject designs to demonstrate performance equivalence to predicate devices. No clinical data provided.
Technological Characteristics
Materials: Commercially pure titanium (ASTM F67) and Ti-6Al-4V ELI alloy (ASTM F136). Components: Plates (0.7-2.7 mm thickness) and screws (1.5-2.7 mm diameter). Features: Locking and non-locking designs, self-tapping threads, hexalobular and cruciform instrument interfaces. Mechanical fixation principle. Sterilization: Not specified (standard medical device processes).
Indications for Use
Indicated for small bone fracture fixation, arthrodesis, reconstruction, and osteotomy fixation in patients requiring stabilization of small bones, including hand, wrist, foot, and ankle. Locking systems are indicated for fractures in osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes (USA) 1.5 mm Mini Fragment LCP System (K090047)
Synthes (USA) Modular Mini Fragment LCP System (K063049)
Synthes (USA) Locking Distal Radius Plating System (K012114)
Mini and Micro Fragments Reconstruction System - NEOFACE (K102641)
Synthes (USA) Distal Radius Plate System (K982732)
Small and Large Fragments Osteosynthesis System NEOFIX (K113733)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles within the eagle's body. The eagle is facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Neoortho Productos Orthopedicos S/A % Mr. Kevin Thomas Regulatory Affairs PaxMed International, LLC 12264 El Camino Real Suite 400 San Diego, California 92130
December 16, 2014
Re: K142419
Trade/Device Name: Mini and Micro Fragments Reconstruction System-Neofix Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: October 15, 2014 Received: October 16, 2014
Dear Mr. Thomas,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K142419
Device Name
Mini and Micro Fragments Reconstruction System - NEOFIX
Indications for Use (Describe)
Mini and Micro Fragments Reconstruction System - NEOFIX is intended for small bone fracture fixation, arthrodess. reconstruction, and osteotomy fixation. Examples include small bones of the hand, wrist, foot and ankle. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
## 510(k) Summary NEOORTHO Produtos Ortopédicos S/A Mini and Micro Fragments Reconstruction System – NEOFIX K142419
December 8, 2014
## ADMINISTRATIVE INFORMATION
| Manufacturer Name | NEOORTHO Produtos Ortopédicos S/A | |
|---------------------------|----------------------------------------------|------------------|
| | Rua Ângelo Domingos Durigan, 607, Cascatinha | |
| | Curitiba, Paraná 82025-100, Brazil | |
| | Telephone: | +55 41 3535-1000 |
| | Fax: | +55 41 3535-1018 |
| Official Contact | Mariana de Oliveira Quinzani | |
| | Regulatory Affairs | |
| Representative/Consultant | Kevin A. Thomas, PhD | |
| | Floyd G. Larson | |
| | PaxMed International, LLC | |
| | 12264 El Camino Real, Suite 400 | |
| | San Diego, CA 92130 | |
| | | Telephone: |
| | | Fax: |
| | | Email: |
| | | |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Mini and Micro Fragments Reconstruction System – NEOFIX |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Plate, Fixation, Bone<br>Screw, Fixation, Bone |
| Classification Name | Single/multiple component metallic bone fixation appliances and<br>accessories<br>Smooth or threaded metallic bone fixation fastener |
| Classification Regulations<br>Product Code | 21 CFR 888.3030, 21 CFR 888.3040, Class II<br>HRS, HWC |
| Classification Panel<br>Reviewing Branch | Orthopedic Products Panel<br>Joint Fixation Devices Branch Two (JFDB2) |
{4}------------------------------------------------
#### INTENDED USE
Mini and Micro Fragments Reconstruction System – NEOFIX is intended for small bone fracture fixation, arthrodesis, reconstruction, and osteotomy fixation. Examples include small bones of the hand, wrist, foot and ankle. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone.
#### DEVICE DESCRIPTION
Mini and Micro Fragments Reconstruction System – NEOFIX consists of plates and screws in a variety of designs and sizes. Straight plates are provided in various lengths and may be contoured to adapt to patient specific anatomy. Straight plates are available with screw holes to accommodate non-locking and locking screw designs. Plates also are provided in various geometric configurations that are commonly used in trauma and reconstructive surgery. The plates range in thickness from 0.7 mm to 2.7 mm, and are manufactured from commercially pure titanium conforming to ASTM F67 Standard Specification for Unalloyed Titanium for Surgical Implant Applications or Ti-6Al-4V ELI alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications.
Self-tapping cortical screws are provided in locking and non-locking designs. The locking screws are provided with a thread diameter of 2.4 mm with a hexalobular instrument interface (Star Head) and a tapered thread tip to facilitate insertion. Locking smooth pins also are provided with a diameter of 2.4 mm in various lengths. The smooth pins are used with the Distal Radius Volar Plates. Non-locking cortical self-tapping screws are provided with thread diameters of 1.5, 2.0, 2.3, 2.4 and 2.7 mm, and in various lengths from 5 mm to 32 mm. The non-locking screws with thread diameters of 1.5, 2.0 and 2.3 mm have a cruciform instrument interface; the nonlocking screws with diameters of 2.4 and 2.7 mm have a hexalobular instrument interface. All screws are manufactured from Ti-6AI-4V ELI alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications.
## EQUIVALENCE TO MARKETED DEVICE
NEOORTHO Produtos Ortopédicos S/A submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
K090047, Synthes (USA), Synthes (USA) 1.5 mm Mini Fragment LCP System K063049, Synthes (USA), Synthes (USA) Modular Mini Fragment LCP System
{5}------------------------------------------------
K102694, Synthes (USA), Synthes (USA) 2.4 mm Variable Angle LCP Dorsal Distal Radius Plates
K071184, Synthes (USA), Synthes Variable Angle-locking Compression Plate (VA-LCP) Distal Radius System
K083694, Synthes (USA), Synthes 2.4 mm VA-LCP Two-Column Volar Distal Radius Plates
K091644, Synthes (USA), Synthes (USA) 2.4 mm LCP Volar Column Distal Radius Plates
K092556, Synthes (USA) 2.4 mm VA-LCP Two-Column Narrow Volar Distal Radius Plates
K012114, Synthes (USA), Synthes Locking Distal Radius Plating System
K102641, NEOORTHO Produtos Ortopédicos S/A, Mini and Micro Fragments Reconstruction System - NEOFACE
K982732, Synthes (USA), Synthes (USA) Distal Radius Plate System K113733, NEOORTHO Produtos Ortopédicos S/A, Small and Large Fragments Osteosynthesis System NEOFIX
The subject device and the predicate devices have the same intended use and have the same technological characteristics. The subject and predicate devices are all fabricated from the same or similar materials. The subject and predicate devices encompass the same range of physical dimensions, and share similar characteristics including locking and non-locking plate and screw designs. The subject and predicate devices are packaged using the same materials, and are to be sterilized by the same methods. Performance data provided to demonstrate substantial equivalence included detailed dimensional analysis of the subject and predicate device designs, engineering analysis, and mechanical testing of the subject designs. Any differences in the technological characteristics do not raise new issues of safety or efficacy. All of the subject device final finished components are manufactured in the same facilities using identical materials and identical manufacturing processes as used for the previously cleared predicate devices in K102641 and K113733, and therefore are substantially equivalent to the predicates with regard to materials and biocompatibility.
Overall, the subject device has the following similarities to the predicate devices:
- has the same intended use.
- uses the same operating principle, ●
- incorporates the same basic designs, ●
- incorporates the same or very similar materials, and
- has similar packaging and is to be sterilized using the same processes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.