The INTEGRA TITAN™ Modular Total Shoulder System is a Total Shoulder Arthroplasty which is indicated for: Severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicated that alternative methods of treatment are unsatisfactory. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. - revision of a failed primary component) Shoulder Hemiathroplasty is also indicated for: Ununited humeral head factures, Avascular necrosis of the humeral head, Rotator cuff arthropathy, Deformity and/or limited motion. The humeral component is intended for cemented or uncemented use. The glenoid component is intended for cemented use only.
Device Story
Modular total shoulder system; includes metaphyseal bodies, humeral stems, humeral heads, and all-polyethylene glenoid components. Used for total shoulder arthroplasty or hemiarthroplasty. Humeral components allow cemented or uncemented fixation; glenoid components require cement. Surgeon selects components to match patient humeral geometry; provides proximal fixation and stable bone loading. Humeral heads available in concentric and eccentric configurations to articulate against natural glenoid or polyethylene glenoid component. Used in clinical/surgical settings by orthopedic surgeons to restore joint function, reduce pain, and address complex humeral fractures or degenerative conditions.
Clinical Evidence
Bench testing only. Performance verification included taper axial disassembly, fatigue testing, maximum static load testing, impact assembly testing, and suture verification. All tests met acceptance criteria.
Technological Characteristics
Modular shoulder prosthesis; metal/polymer construction. Humeral components (cemented/uncemented) and all-polyethylene glenoid components (cemented). Features include metaphyseal bodies, humeral stems, and concentric/eccentric humeral heads. Keeled or pegged (3-peg) glenoid options.
Indications for Use
Indicated for patients with severely painful/disabled shoulder joints due to osteoarthritis, rheumatoid arthritis, or traumatic arthritis; proximal humerus fracture-dislocations; or other conditions requiring shoulder arthroplasty (e.g., revision of failed primary component). Hemiarthroplasty indicated for ununited humeral head fractures, avascular necrosis, rotator cuff arthropathy, or deformity/limited motion.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Ascension® TITAN™ Modular Total Shoulder System (K112438)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the seal is a stylized image of three human profiles facing to the right, with a stylized eagle above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 14, 2014
Ascension Orthopedics % Mr. Frederic Testa Regulatory Affairs Director Integra LifeSciences Corporation 311 Enterprise Drive Plainsboro, New Jersey 08536
Re: K142413
Trade/Device Name: INTEGRA® TITAN™ Modular Total Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS, HSD Dated: September 15, 2014 Received: September 16, 2014
Dear Mr. Testa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K142413
#### Device Name
INTEGRA® TITAN™ Modular Total Shoulder System
#### Indications for Use (Describe)
The INTEGRA TITAN™ Modular Total Shoulder System is a Total Shoulder Arthroplasty which is indicated for: Severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicated that alternative methods of treatment are unsatisfactory. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. - revision of a failed primary component) Shoulder Hemiathroplasty is also indicated for: Ununited humeral head factures, Avascular necrosis of the humeral head, Rotator cuff arthropathy, Deformity and/or limited motion. The humeral component is intended for cemented or uncemented use. The glenoid component is intended for cemented use only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
| 807.92(a)(1) – Submitter Information | |
|--------------------------------------|----------------------------------------------------------------------|
| Name | Ascension Orthopedics, Inc. |
| Address | 8700 Cameron Road, Suite 100<br>Austin, TX 78754 |
| Name of Contact Person | Frederic Testa |
| Phone Number | (609) 936-3630 |
| Fax Number | (609) 275-9445 |
| Establishment Registration Number | 1651501 |
| Date Prepared | November 11th, 2014 |
| 807.92(a)(2) – Name of device | |
| Trade or Propriety Name | INTEGRA® TITAN™ Modular Total Shoulder System |
| Common or Usual Name | Shoulder Joint Metal/Polymer Semi-Constrained Cemented<br>Prosthesis |
| Classification Name | Prostheses, Shoulder, Semi-Constrained, Metal/Polymer<br>Cemented |
| Classification Panel | Orthopedic |
| Regulation | Class II (under 21CFR 888.3660) |
| Product Code | KWS, HSD |
### 510(k) Summary
Ascension® TITAN™ Modular Total Shoulder System (K112438)
### 807.92(a)(4) - Device description
The TITAN™ Modular Total Shoulder System consists of a line of metaphyseal bodies, humeral stems, humeral heads and all polyethylene glenoid components. The body, stem and humeral head may be used by themselves, as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with the glenoid, as a total replacement. The metaphyseal bodies and humeral stems are shaped to provide proximal fixation and optimal fixation area. Their variable length and proximally-filling shape are designed to accommodate the natural humeral geometry and provide stable fixation. proximal bone loading and proper head placement. The humeral heads are offered with both concentric and eccentric articulating surfaces. The humeral head may articulate against the natural glenoid bone, if it is of sufficient quality, or against the all polyethylene cemented glenoid. The glenoid has multiple options: keeled or standard pegged (3 pegs). All glenoid options are designed to function with both the concentric and eccentric heads.
The humeral components are intended for cemented or uncemented use, while the glenoid component is for use with cement only.
| 807.92(a)(5) – Intended Use of the device | |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The INTEGRA® TITAN™ Modular Total Shoulder System is indicated for use as a hemi or total shoulder replacement for: |
{4}------------------------------------------------
| | Severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicated that alternative methods of treatment are unsatisfactory. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. – revision of a failed primary component) Shoulder Hemiarthroplasty is also indicated for: Ununited humeral head factures Avascular necrosis of the humeral head Rotator cuff arthropathy Deformity and/or limited motion. The humeral component is intended for cemented or uncemented use. The glenoid component is intended for cemented use only. |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## 807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate
The INTEGRA® TITAN™ Modular Total Shoulder System is similar in design and materials to the predicate device, Ascension® TITAN™ Modular Total Shoulder System (K112438). The INTEGRA® TITAN™ Modular Total Shoulder System has similar indications for use, intended use and fundamental scientific technology as its predicate. The differences between the predicated and proposed device do not raise any new issues regarding safety and effectiveness; therefore, the INTEGRA® TITAN™ Modular Total Shoulder System is considered substantially equivalent to the predicate device.
### 807.92(b)(1-2) - Nonclinical Tests Submitted
Testing to verify the performance of the INTEGRA® TITAN™ Modular Total Shoulder System included the following:
- Taper Axial Disassembly Test
- . Fatigue Test
- Maximum Static Load Test ●
- Impact Assembly Test
- Suture Verification Report ●
The results of these performance tests met their respective acceptance criteria and demonstrate that the INTEGRA® TITAN™ Modular Total Shoulder System is safe for the intended use, and is substantially equivalent to the predicate device identified.
### 807.92(b)(3) - Conclusions drawn from non-clinical data
The design features, materials, intended use, and overall fundamental scientific technology of the INTEGRA® TITAN™ Modular Total Shoulder System are substantially equivalent to the predicate device. The safety and effectiveness of the INTEGRA® TITAN™ Modular Total Shoulder System is adequately supported by the substantial equivalence information, materials information, and performance data provided within this Premarket Notification submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.