K142393 · Lake Region Medical · DQX · Nov 25, 2014 · Cardiovascular
Device Facts
Record ID
K142393
Device Name
PREDICATE III GUIDEWIRE
Applicant
Lake Region Medical
Product Code
DQX · Cardiovascular
Decision Date
Nov 25, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
PREDICATE™ III GUIDEWIRES are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.
Device Story
Guidewire used in angiographic procedures to facilitate placement of catheters and interventional devices within coronary and peripheral vasculature. Device consists of Nitinol or Stainless Steel core wire with Stainless Steel coil; available in various lengths (30cm-500cm) and diameters (0.018"-0.038"). Operated by clinicians in clinical settings. Device provides mechanical support and guidance for interventional tools; aids in navigating complex vascular anatomy. Benefits include improved access to target sites during minimally invasive procedures.
Clinical Evidence
Bench testing only. Testing included dimensional, visual/tactile, and mechanical attributes (pull test, camber resistance, wire fracture, corrosion, wire flex, ISO strength of union, particulate residue). Biocompatibility testing included cytotoxicity, hemolysis, and chemical characterization. Results demonstrate compliance with design input summary and equivalence to the predicate device.
Technological Characteristics
Core: Nitinol or Stainless Steel. Coil: Stainless Steel. Joining: Weld. Length: 30cm-500cm. Diameter: 0.018"-0.038". Sterilization: ETO. Standards: ISO 10993, ISO 11070, ISO 11135, ISO 11138.
Indications for Use
Indicated for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature in patients requiring such procedures.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2014
Lake Region Medical Mathew Pexa Regulatory Specialist II 340 Lake Hazeltine Drive Chaska, MN 55318
Re: K142393
Trade/Device Name: Predicate III Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DOX Dated: October 28, 2014 Received: October 30, 2014
Dear Mr. Pexa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Melissa A. Torres -S
For Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for Lake Region Healthcare. The logo consists of a red abstract symbol on the left, followed by the words "Lake Region" in a bold, serif font. Below "Region" is the word "Medical" in a smaller font size.
## INDICATIONS FOR USE STATEMENT
#### 510(k) NUMBER (IF KNOWN): K142393
### device name: PREDICATE™ III GUIDEWIRE
### INDICATIONS FOR USE:
PREDICATE™ III GUIDEWIRE are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.
| PRESCRIPTION | |
|--------------|--|
| USE | |
×
AND/OR
OVER-THE-COUNTER USE
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
PAGE 1 of ।
{3}------------------------------------------------
e Region
PRODUCT: PREDICATE™ III GUIDEWIRE – Vascular Use JBMISSION DATE: August 260, 2014 SUBMISSION TYPE: SPECIAL 510(k)
# 510(k) SUMMARY
November 25, 2014
This summary is being included in the submission in lieu of a statement of availability.
| 2.1 | MANUFACTURER / REGISTRATION INFORMATION | | |
|-----|-----------------------------------------|-----------------|-------------------|
| | Lake Region Medical | Telephone: | 952-448-5111 |
| | 340 Lake Hazeltine Dr. | Fax: | 952-448-3441 |
| | Chaska, Mn 55318 | Contact Person: | Sze Yuen Tan |
| | FDA REGISTRATION NUMBER: 2126666 | Title: | Regulatory Intern |
| 2.2 | DEVICE TRADE NAME / PROPRIETARY NAME | | |
PREDICATE™ III GUIDEWIRE
2.3 DEVICE COMMON NAMES / USUAL NAMES / CLASSIFICATION NAMES CATHETER GUIDEWIRE (DQX)
2.4 CLASS OF DEVICE
These devices are commonly known as guides, guidewires, or spring guidewires. The current classification name and product code is Catheter Guidewire (DQX) and is considered a Class II device per 21 CFR Part 870.1330.
#### 2.5 IDENTIFICATION OF PREDICATE DEVICE(s)
K140485 Mandrel Guidewires (Lake Region Medical)
#### 2.6 DEVICE DESCRIPTION
The guidewire is made of an uncoated Nitinol or Stainless Steel core wire that is tapered at the distal tip where a coil is secured over the core. The wire can have a distal coil or a full length coil that is made of Stainless Steel. The guidewire may contain proximal markers. The Guidewire family is bound by the following parameters:
| Core Material | Stainless Steel or Nitinol |
|----------------------|----------------------------|
| Coil Material | Stainless Steel |
| Coil Length | Full length |
| Joining Agents | Weld |
| Tip Type | Straight, Shaped |
| Tip Flexibilities | Various |
| Overall Lengths | 30cm to 500cm |
| Outside Diameters | 0.018" to 0.038" |
| Sterilization Method | ETO |
2.7 COMPLIANCE WITH APPLICABLE STANDARDS
LRM has determined that no mandatory standards, performance standards, or special controls have been established for these devices under Section 514 of the Medical Device Amendments to Federal Food, Drug, and Cosmetic or by any subsequent regulatory action. However, the following standards are referenced within this filing: ISO 10993, ISO 11070, ISO 11135, ISO 11138.
#### 2.8 INTENDED USE STATEMENT
PREDICATE™ III GUIDEWIRES are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.
NOTE: This modification does not alter its intended use.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Lake Region Healthcare. The logo consists of a red abstract symbol on the left, followed by the words "Lake Region" in a bold, sans-serif font. Below "Region" is the word "Medical" in a smaller font. The logo is clean and professional, suggesting a healthcare organization.
2.11
PRODUCT: PREDICATE™ III GUIDEWIRE – Vascular Use BMISSION DATE: August 260, 2014 SUBMISSION TYPE: SPECIAL 510(k)
#### 2.9 CONTRAINDICATIONS
There are no contraindications listed.
#### 2.10 TECHNOLOGICAL CHARACTERISTICS
The design specifications are substantially equivalent to the existing Mandrel Guidewires.
### QUALITY SYSTEM CONTROL
DESIGN CONTROLS
LRM is in conformance with the design control procedure requirements as specified in 21 CFR Part 820.30. Risk analysis was completed by means of a Failure Mode and Effects Analysis and all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria were met.
### MATERIALS / SUPPLIER / PRODUCT / PROCESS CONTROLS
LRM has formal quality systems in placed to assure that each product manufactured remains equivalent to the predicate products, and that the changes will not have an adverse effect on safety or effective use of the product. The quality systems include Engineering Change Order Review, Material Qualification, Supplier Qualification, Product Qualification, and Process Qualification. These controls are applied to each product size / group.
#### 2.12 QUALIFICATION TESTING
The conclusions drawn from bench testing and biocompatibility testing demonstrate compliance with the design input summary which shows the device is at least as safe and effective as the current legally marketed device.
### BENCH TESTING
In order to demonstrate equivalence of the guidewire, Lake Region Medical performed bench testing to establish requirements. Test devices were manufactured and inspected according to established requirements for visual/tactile, dimensional and mechanical attributes. The devices were then subjected to the following test methods to show the devices comply with the design input summary:
- Dimensional
- Guidewire Pull Test
- Camber Resistance
- Visual Inspection
- Wire Fracture Test ●
- · Corrosion
- Wire Flex Test
- ISO Strength of Union
- Particulate Residue Test
- BIOCOMPATIBILITY TESTING
Biocompatibility testing per the design input summary requirements show the addition of full length coil to the device does not affect the biocompatibility of the device and the device is still in compliance with pre-defined acceptance criteria outlined in the product Design Input Summary. A biocompatibility risk assessment determined the following biocompatibility tests are required:
- Cytotoxicity
- Hemolysis
- Chemical Characterization
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Lake Region Healthcare. The logo consists of a red abstract symbol on the left, followed by the words "Lake Region" in a bold, sans-serif font. Below "Region" is the word "Medical" in a smaller font size.
#### 2.13 CHANGES FROM PREDICATE DEVICE
- Include a Stainless Steel Coil only option as opposed to Stainless Steel, Platinum/Tungsten, -Palladium, and Tungsten
- -Weld as a joining option as opposed to options for Braze, Solder, Epoxy and Weld
- -Full length coil as opposed to 2cm to 30cm distal tip coil
- -Min. Diameter goes to 0.018" as opposed to 0.014"
- -Max. Diameter goes to 0.035" as opposed to 0.038"
- -No coating options as opposed to PTFE coating
#### 2.14 SUBSTANTIAL EQUIVALENCE DATA
The changes included in this 510(k) for the Mandrel guidewire family does not change the indications for use of the Mandrel guidewires and is not a change to the fundamental scientific technology. The information summarized above shows the device will perform as well as the previously marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.