K142383 · Ge Medical Systems, LLC · IZL · Sep 25, 2014 · Radiology
Device Facts
Record ID
K142383
Device Name
Optima XR200amx, Optima XR220amx
Applicant
Ge Medical Systems, LLC
Product Code
IZL · Radiology
Decision Date
Sep 25, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Optima XR200amx and Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR). Optima XR200amx and Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects). The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room. The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position. These devices are not intended for mammographic applications.
Device Story
Self-contained, battery-operated mobile radiographic imaging systems; generate diagnostic x-ray images for detection of disease or injury (e.g., orthopedic damage, tumors, pneumonia, foreign objects). Operated by clinicians in clinical settings (emergency, ICU, OR, etc.) for patients unable to be moved to traditional radiology rooms. Input: x-ray exposure initiated via wired or remote exposure switch. Output: radiographic images (film, CR, or DR via GE Wireless Detector). Digital detector supports patients of all sizes; withstands 352 lbs load. Benefits: enables bedside imaging, reducing patient transport risks; digital detector provides comparable image quality at lower dose than CR.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including module verification, system integration, performance, safety, and simulated use validation.
Technological Characteristics
Mobile, battery-operated radiographic system. Digital radiography (DR) via GE Wireless Detector. Flat panel detector supports 352 lbs load. Connectivity: wired or remote wireless exposure switch. Software: embedded system control. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for adult and pediatric patients for general-purpose diagnostic radiographic examinations of skull, spinal column, chest, abdomen, and extremities. Used in clinical environments including emergency, ICU, NICU, cardiac, and OR for immobile or difficult-to-transport patients. Not for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Brivo XR285amx, Optima XR200amx, and Optima XR220amx (K103476)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2014
GE Medical Systems, LLC % Mr. Christopher Paulik Regulatory Affairs Leader 3000 N. Grandview Blvd. WAUKESHA WI 53188
Re: K142383
Trade/Device Name: Optima XR200amx, Optima XR220amx Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: August 22, 2014 Received: August 26, 2014
Dear Mr. Paulik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Sm. 7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K142383
Device Name Optima XR200amx and Optima XR220amx
#### Indications for Use (Describe)
The Optima XR200amx and Optima XR220amx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR220amx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).
Optima XR200amx and Optima XR220amx are self-contained; battery operated mobile radiographic imaging systems designed to generate diagnostic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).
The Optima XR200amx and Optima XR220amx are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.
The systems are indicated for taking radiographic exposures of the skull, spinal column, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.
These devices are not intended for mammographic applications.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has a swirling pattern around the letters, giving it a dynamic and recognizable appearance. The logo is simple, yet iconic, representing the brand's long history and global presence.
# Attachment A: Section 5 – 510(k) Summary
| Document Name | Document<br>Number | Page Number |
|----------------------------|--------------------|-------------|
| Section 5 - 510(k) Summary | N/A | 5 Pages |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The letters are surrounded by a swirling pattern. The logo is simple and recognizable, and it is often used to represent the company.
Section 5: 510(k) Summary
Optima XR200amx and Optima XR220amx
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular frame. The color of the logo is a light blue.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | August 22, 2014 |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems, LLC<br>3000 N. Grandview Blvd<br>Waukesha, WI 53188, USA |
| Primary Contact Person: | Chris Paulik<br>Regulatory Affairs Leader<br>GE Healthcare<br>262-548-2010<br>Christopher.Paulik@med.ge.com |
| Secondary Contact Person: | John L. Schmidt<br>Regulatory Affairs Manager<br>GE Healthcare<br>262-548-4964<br>John.L.Schmidt@med.ge.com |
| Device Trade Name: | Optima XR200amx and Optima XR220amx |
| Common/Usual Name: | Optima XR200amx and Optima XR220amx |
| Classification Names:<br>Product Code: | Class II, System, X-ray, Mobile, 21 CFR 892.1720<br>IZL |
| Predicate Device(s): | Brivo XR285amx, Optima XR200amx, and Optima XR220amx<br>K103476 |
| Device Description: | The Optima XR200amx and Optima XR220amx are intended to take<br>exposures, using a wired or remote exposure switch, utilizing film or<br>computed radiography (CR), however the Optima XR220amx utilizes the GE<br>Wireless Detector, which is intended to replace radiographic film screen<br>systems in all general purpose diagnostic procedures, for digital<br>radiography (DR). |
| operated mobile radiographic imaging systems designed to generate<br>diagnostic radiographic images (medical x-rays) that may increase the<br>ability to detect disease or injury early enough for a medical problem to be<br>managed, treated, or cured. Medical x-rays are used in many types of<br>examinations and procedures, some examples include: x-ray radiography<br>(to find orthopedic damage, tumors, pneumonias, foreign objects).<br><br>The Optima XR200amx and Optima XR220amx systems are indicated for<br>use on adult and pediatric patients for general-purpose diagnostic<br>radiographic examinations and procedures. Its mobility enables general-<br>purpose radiographic procedures throughout the clinical environment, or<br>as needed within the emergency, intensive care, premature birth ward,<br>cardiac and operating departments, for patients that may not be able to<br>to be moved or in cases where it is unsafe or impractical to move them to a<br>traditional RAD room.<br><br>The flat panel detector provides increased functionality to enable images<br>of patients of all sizes, and can produce comparable quality images with<br>as little as half the dose of traditional computer radiography (CR), cassettes<br>and other flat panel detectors with lower DQE. The digital detector is<br>designed to withstand a distributed load of 352Lbs to accommodate<br>certain larger patients.<br><br>The systems are indicated for taking radiographic exposures of the skull,<br>spinal column, chest, abdomen, extremities, and other body parts with the<br>patient sitting, standing, or lying in the prone or supine position.<br><br>These devices are not intended for mammographic applications. | |
| <b>Intended Use:</b> The Optima XR200amx and Optima XR220amx are intended to take<br>exposures utilizing film or computed radiography (CR), however the Optima<br>XR220amx utilizes the GE Wireless Detector, which is intended to replace<br>radiographic film screen systems in all general purpose diagnostic<br>procedures, for digital radiography (DR).<br><br>Optima XR200amx and Optima XR220amx are self-contained; battery<br>operated mobile radiographic imaging systems designed to generate<br>diagnostic radiographic images (medical x-rays) that may increase the<br>ability to detect disease or injury early enough for a medical problem to be<br>managed, treated, or cured. Medical x-rays are used in many types of<br>examinations and procedures, some examples include: x-ray radiography<br>(to find orthopedic damage, tumors, pneumonias, foreign objects).<br><br>The series are indicated for use on adult and pediatric patients for general-<br>purpose diagnostic radiographic examinations and procedures. Its<br>mobility enables general-purpose radiographic procedures throughout the<br>clinical environment, or as needed within the emergency, intensive care,<br>premature birth ward, cardiac and operating departments, for patients<br>that may not be able to be moved or in cases where it is unsafe or<br>impractical to move them to a traditional RAD room.<br><br>The systems are indicated for taking radiographic exposures of the skull | |
| | spinal column, chest, abdomen, extremities, and other body parts with the<br>patient sitting, standing, or lying in the prone or supine position.<br>These devices are not intended for mammographic applications. |
| Technology: | The Optima XR200amx and Optima XR220amx employ the same<br>fundamental scientific technology as the predicate devices. The only<br>difference is that exposures can now be initiated using either a wired or<br>remote exposure switch. |
| Determination<br>of Substantial<br>Equivalence: | Summary of Non-Clinical Tests:<br>The Optima XR200amx and Optima XR220amx and their applications<br>comply with voluntary standards. The following quality assurance<br>measures were applied to the development of the system:<br>Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) New risk were identified for potential user error or misuse attributed with<br>the addition of the remote hand switch. These risks were reviewed and<br>mitigated with design controls and labeling. The mitigations were verified<br>and validated as a part of the design verification and validation testing<br>that has been executed with acceptable results.<br>Summary of Clinical Tests:<br>The subject of this premarket submission, Optima XR200amx and Optima<br>XR220amx, did not require clinical studies to support substantial<br>equivalence for the option of initiating x-ray exposures with the use of the<br>remote exposure switch.<br>Design verification and validation testing was performed to confirm that<br>the safety and effectiveness of the devices has not been affected. The test<br>plans and results have been executed with acceptable results. |
| Conclusion: | The Optima XR200amx and Optima XR220amx devices incorporate a<br>wired as well as a remote hand switch for the user to initiate x-ray<br>exposures. The addition of the optional wireless hand switch to these<br>systems do not result in any new potential safety risks, they have the same<br>technological characteristics, and perform as well as the devices currently<br>on the market. |
| After analyzing design verification and validation testing on the bench it is<br>the conclusion of GE Healthcare that the Optima XR200amx and Optima<br>XR220amx to be as safe, as effective, and performance is substantially | |
| equivalent to the predicate devices. | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The frame is made up of curved lines that resemble stylized leaves or flourishes. The logo is colored in a bright, solid blue.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular frame. The color of the logo is a light blue.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular shape. The circle is a light blue color.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.