K142379 · Biotronik, Inc. · LIT · Dec 5, 2014 · Cardiovascular
Device Facts
Record ID
K142379
Device Name
Passeo-35 PTA Catheter
Applicant
Biotronik, Inc.
Product Code
LIT · Cardiovascular
Decision Date
Dec 5, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Passeo-35 Peripheral Dilatation Catheter is intended for dilatation of stenotic segments in peripheral vessels and arteriovenous dialysis fistulae.
Device Story
The Passeo-35 is an over-the-wire, dual-lumen percutaneous transluminal angioplasty (PTA) catheter used by physicians to dilate stenotic peripheral vessels and arteriovenous dialysis fistulae. The device features a balloon with radiopaque markers at each end for fluoroscopic positioning and a tapered soft tip for lesion advancement. It is compatible with 0.035" guide wires. The operator connects an inflation device to a proximal Luer port to inflate the balloon to a specific diameter and pressure, as defined by the compliance chart. The catheter shaft is coated with a hydrophobic silicone to improve trackability. The device is delivered via standard percutaneous access over a guide wire. By inflating the balloon within a stenotic lesion, the device mechanically dilates the vessel, potentially improving blood flow and resolving obstructive lesions. The device is provided sterile and is intended for single use.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing only, including biocompatibility (ISO 10993), mechanical performance (tensile strength, torque, kink resistance, fatigue, burst pressure), and simulated use testing in anatomical models.
Technological Characteristics
Over-the-wire 2-lumen balloon catheter. Materials: Pebax (shaft/tip), Poly-Amide (balloon), Polycarbonate (manifold), Pt/Ir (markers). Hydrophobic silicone coating. Sterilization: EO gas. Compatible with 0.035" guide wires. Dimensions: 5F shaft, 3-10mm balloon diameters, lengths up to 200mm. Standards: ISO 10993, ISO 25539-2, ASTM F2081, ISO 10555.
Indications for Use
Indicated for dilation of stenosis in renal, iliac, femoral, popliteal, and infrapopliteal arteries and treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is simple and recognizable, representing the department's role in promoting health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 5, 2014
BIOTRONIK, Inc. Mr. Jon Brumbaugh Vice President, Regulatory Affairs and Compliance 6024 Jean Road Lake Oswego, Oregon 97035
Re: K142379
> Trade/Device Name: Passeo-35 Peripheral Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: November 4, 2014 Received: November 5, 2014
Dear Mr. Brumbaugh,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
K142379 510(k) Number (if known):
Device Name: Passeo-35 Peripheral Dilatation Catheter
Indications for Use:
The Passeo-35 Peripheral Dilatation Catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
# Prescription Use _____________________________________________________________________________________________________________________________________________________________
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of the letters "BIO" inside of a circle on the left, followed by the word "BIOTRONIK" in large, bold, dark blue letters. Below the word "BIOTRONIK" is the phrase "excellence for life" in a smaller, lighter blue font.
# Passeo-35 Peripheral Dilatation Catheter Special 510(k) Premarket Notification 510(k) Summary
| Name and Address of Sponsor: | BIOTRONIK, Inc.<br>6024 Jean Road<br>Lake Oswego, OR 97035 |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact Person and Phone Number: | Jon Brumbaugh<br>Vice President, Regulatory Affairs and Compliance<br>BIOTRONIK, Inc.<br>6024 Jean Road<br>Lake Oswego, OR 97035<br>Phone: (888) 345-0374<br>Fax: (503) 635-9936<br>jon.brumbaugh@biotronik.com |
| Date Prepared: | December 5, 2014 |
| Device Name: | |
| Proprietary Name: | Passeo-35 Peripheral Dilatation Catheter |
| Common Name: | Percutaneous Transluminal Angioplasty (PTA)<br>Catheter |
| Classification: | Class II (21 CFR 870.1250) |
| Classification Name: | Catheter, angioplasty, peripheral, transluminal |
| Product Code: | LIT |
#### Predicate Device:
| | 510(k) # | Device Name | Manufacturer | Date of Clearance |
|------------|----------|-------------|--------------|-------------------|
| Predicate: | K082933 | Passeo-35 | BIOTRONIK | 3-Nov-2008 |
#### General Description:
The Passec-35 Peripheral Dilatation Catheter (Passeo-35) is intended for dilatation of stenotic segments in peripheral vessels. One radiopaque marker is located at either end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. The balloon catheter includes a tapered soft tip to facilitate advancement of the catheter.
The balloon catheter shaft has two Luer ports at the proximal end. One port) serves for connecting an inflation device to inflate/deflate the balloon. The other port) enables insertion of the guide wire. The balloon catheter is a dual lumens contained within one tube. The smaller lumen is the balloon inflation/deflation lumen permits the use of guide wires with a maximum diameter of 0.035" to facilitate advancement of the Passeo-35 catheter towards and through the lesion(s) to be dilated. The balloon catheter is compatible with introducer) sizes according to the recommendations on the label. The balloon catheter has a silicone coating (hydrophobic) to improve the trackability characteristics.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of the BIOTRONIK name in bold, dark blue letters, with a light blue circle containing the letters "BIO" to the left. Below the name is the phrase "excellence for life" in a smaller, light blue font.
#### Indication for Use:
The Passeo-35 Peripheral Dilatation Catheter is indicated to dilate stenosis in the renal, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
#### Purpose of Submission:
BIOTRONIK submits this 510(k) for clearance of additional device size configurations for the Passeo-35 including additional balloon lengths up to 200mm. In addition, device and packaging materials changes are documented and labeling is updated for clarity as well as to reflect the new sizes and RBPs.
| Substantial Equivalence of Passeo-35 to existing (predicate) Passeo-35 product line | | | |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameters | Passeo-35<br>K082933 | Passeo-35<br>Subject Device | Rationale for Substantial<br>Equivalence |
| Proprietary name | Passeo-35 Peripheral<br>Dilatation Catheter | Passeo-35 Peripheral<br>Dilatation Catheter | Same |
| Common name | PTA catheter | PTA catheter | Same |
| Classification | Class II (21 CFR 870.1250) | Class II (21 CFR 870.1250) | Same |
| Classification<br>name | Catheter, angioplasty,<br>peripheral, transluminal | Catheter, angioplasty,<br>peripheral, transluminal | Same |
| Product code | LIT | LIT | Same |
| Intended use / Indications for Use | | | |
| Intended use | The Passeo-35 Peripheral<br>Dilatation Catheter is intended<br>for dilatation of stenotic<br>segments in peripheral<br>vessels and arteriovenous<br>dialysis fistulae. | The Passeo-35 Peripheral<br>Dilatation Catheter is intended<br>for dilatation of stenotic<br>segments in peripheral<br>vessels and arteriovenous<br>dialysis fistulae. | Same |
| Indications for<br>Use | The Passeo-35 Peripheral<br>Dilatation Catheter is<br>indicated to dilate stenosis in<br>the renal, iliac, femoral,<br>popliteal and infrapopliteal<br>arteries and for the treatment<br>of obstructive lesions of native<br>or synthetic arteriovenous<br>dialysis fistulae. | The Passeo-35 Peripheral<br>Dilatation Catheter is<br>indicated to dilate stenosis in<br>the renal, iliac, femoral,<br>popliteal and infrapopliteal<br>arteries and for the treatment<br>of obstructive lesions of native<br>or synthetic arteriovenous<br>dialysis fistulae. | Same |
| Substantial Equivalence of Passeo-35 to existing (predicate) Passeo-35 product line | | | |
| Parameters | Passeo-35<br>K082933 | Passeo-35<br>Subject Device | Rationale for Substantial<br>Equivalence |
| Contraindications | All general contraindications<br>for percutaneous transluminal<br>angioplasty (PTA) are<br>contraindications for this<br>device. Contraindications for<br>this device and peripheral<br>dilatation catheters in general<br>are:<br>• Lesions that cannot be<br>reached or treated with the<br>system<br>• Large amounts of acute or<br>subacute thrombus at the<br>target lesion<br>• Perforated vessels<br>• Lesion that lies within or<br>adjacent to an aneurysm<br>• Uncorrected bleeding<br>disorders<br>• Renal insufficiency or an<br>allergy to contrast media<br>Furthermore, all procedure-<br>related contraindications as<br>described in the national and<br>international guidelines of the<br>respective medical<br>associations apply. | All general contraindications<br>for percutaneous transluminal<br>angioplasty (PTA) are<br>contraindications for this<br>device. Contraindications for<br>this device and peripheral<br>dilatation catheters in general<br>are:<br>• Lesions that cannot be<br>reached or treated with the<br>system<br>• Large amounts of acute or<br>subacute thrombus at the<br>target lesion<br>• Perforated vessels<br>• Lesion that lies within or<br>adjacent to an aneurysm<br>• Uncorrected bleeding<br>disorders<br> | Same |
| Intended user | Physicians competent in PTA<br>procedures | Physicians competent in PTA<br>procedures | Same |
| Method of<br>placement | Standard percutaneous<br>access to site over a guide<br>wire, with fluoroscopic<br>visualization | Standard percutaneous<br>access to site over a guide<br>wire, with fluoroscopic<br>visualization | Same |
| Sterilization / Shelf life / Packaging | | | |
| Sterilization | EO gas | EO gas | Same |
| Sterilization<br>System | Sauter EO Sterilizer | Sauter EO Sterilizer<br>or<br>Sterichem EO Sterilizer | Does not alter intended use.<br>Validation testing according to<br>design controls support<br>equivalence. BIOTRONIK<br>uses either sterilizer in<br>production. |
| SAL | 10-6 | 10-6 | Same |
| Shelf life | 3 years | 3 years | Same |
| Substantial Equivalence of Passeo-35 to existing (predicate) Passeo-35 product line | | | |
| Parameters | Passeo-35<br>K082933 | Passeo-35<br>Subject Device | Rationale for Substantial<br>Equivalence |
| Protective sheath | Balloon has a protective<br>sheath. Spiral dispenser<br>sealed in a Tyvek® and<br>PET/PE pouch.<br>Product is packed in an outer<br>cardboard carton. | Balloon has a protective<br>sheath. Spiral dispenser<br>sealed in a Tyvek® and<br>PET/PE pouch.<br>Product is packaged in an<br>outer cardboard carton. | Same |
| Shelf life | 3 years | 3 years | Same |
| Instructions for<br>Use and labeling | As provided in K082933. | Minor wording & symbol<br>changes | Changes do not alter intended<br>use, technological<br>characteristics, or raise<br>different questions of safety<br>and effectiveness. |
| Device Design | | | |
| Device<br>description | Over the wire 2-lumen balloon<br>catheter | Over the wire 2-lumen balloon<br>catheter | Same |
| Radiopaque<br>markers | 2 – one at each end of the<br>balloon<br>Material: 90% Pt / 10% Ir<br>Length: 1.5 mm | 2 – one at each end of the<br>balloon<br>Material: 90% Pt / 10% Ir<br>Length: 1.5 mm | Same |
| Usable length<br>[cm] | 80 and 130 | 80, 90 and 130 | Does not alter intended use.<br>Performance testing<br>according to design controls<br>support equivalence. |
| Introducer<br>sheath<br>compatibility | 5F (Balloon Ø: 3 – 7 mm)<br>6F (Balloon Ø: 8 – 10 mm) | 5F (Balloon Ø: 3 – 7 mm)<br>6F (Balloon Ø: 8 – 10 mm) | Same |
| Crossing profile | Ø: 3-7mm: max. 0.074 inches<br>Ø: 8-10mm: max. 0.083<br>inches | Ø: 3-7mm: max. 0.074 inches<br>Ø: 8-10mm: max. 0.083<br>inches | Same |
| Guide wire<br>compatibility | 0.035" | 0.035" | Same |
| Shaft outer<br>diameter [F] | 5 | 5 | Same |
| Shaft inner<br>diameter [mm] | 0.96 | 0.98 | The larger inner diameter of<br>Passeo-35 LE does not alter<br>the intended use or raise<br>different safety and<br>effectiveness questions.<br>Performance testing data<br>showed substantial<br>equivalence to the predicate. |
| Balloon diameter<br>[mm] | 3.0, 4.0, 5.0, 6.0, 7.0, 8.0, 9.0,<br>10.0 | 3.0, 4.0, 5.0, 6.0, 7.0, 8.0, 9.0,<br>10.0 | Same |
| Substantial Equivalence of Passeo-35 to existing (predicate) Passeo-35 product line | | | |
| | Passeo-35 | Passeo-35 | Rationale for Substantial |
| Parameters | K082933 | Subject Device | Equivalence |
| Balloon length<br>[mm] | 20, 40, 60, 80, 100 | 20, 40, 60*, 80*, 100*, 120,<br>150, 170, 200<br>(* new balloon lengths for<br>some balloon diameter sizes,) | Same intended use and does<br>not raise different safety and<br>effectiveness questions.<br>Performance testing data<br>demonstrate substantial<br>equivalence. |
| Balloon wrapping | 5 folds | 5 folds | Same |
| Balloon Nominal<br>pressure [atm] | 7 | 7 | Same |
| Balloon RBP<br>[atm] | 20 (Balloon Ø: 3 - 4 mm)<br>16 (Balloon Ø: 5 - 6 mm)<br>14 (Balloon Ø: 7 - 8 mm)<br>12 (Balloon Ø: 9 - 10 mm) | 20 (Balloon Ø: 3mm)<br>18 (Balloon Ø: 4mm)<br>16 (Balloon Ø: 5 - 6 mm)<br>14 (Balloon Ø: 7 - 8 mm)<br>12 (Balloon Ø: 9mm)<br>11 (Balloon Ø: 10 mm) | Same intended use and does<br>not raise different safety and<br>effectiveness questions.<br>Performance testing data<br>demonstrate substantial<br>equivalence. |
| Materials of Construction (Direct and indirect patient contact) | | | |
| Tip | Pebax 5533 SA01 | Pebax 5533 SA01 | Same |
| Balloon…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.