The OnControl Bone Marrow Biopsy System by Vidacare
Applicant
Vidacare, LLC
Product Code
KNW · Gastroenterology, Urology
Decision Date
Dec 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K142377 · Dec 17, 2014
The OnControl Bone Marrow Biopsy System by Vidacare
Vidacare, LLC
Published clinical literature (Reid et al., 1996; 1999) reporting on routine clinical practice outcomes
The sponsor uses retrospective clinical data from published literature to characterize the failure rates of manual bone marrow biopsy procedures in pediatric patients, establishing the clinical need for the OnControl system.
Pediatric bone marrow biopsy; Standard of care failure rates; Literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Reid et al. (1996/1999) clinical cohort; Retrospective analysis of clinical practice
Pediatric patients undergoing trephine biopsies; Sample Size: 605 to 822 patients; Number of Sites: 13 institutions
Not applicable for this study
Adequacy of bone marrow trephine biopsy specimens
Indications for Use
The OnControl™ Bone Marrow Biopsy System is intended for bone marrow aspiration and biopsy in adult and pediatric patients age 2 and older.
Device Story
The OnControl Bone Marrow Biopsy System consists of a reusable power rotary driver and a disposable sterile needle set. The needle set includes a 304 stainless steel cannula with a beveled cutting tip, stylet, and internal core-capturing thread. The clinician uses the power driver to insert the needle through the bone cortex; the stylet is then removed for aspiration. The driver is reattached to the cannula to advance the needle to the desired depth for biopsy specimen collection. The device is used in clinical settings by trained physicians. The powered rotary mechanism provides tactile feedback, assisting in precise needle placement. Compared to manual biopsy methods, the system is designed to improve specimen quality and size, reduce procedure time, and potentially decrease patient pain and the number of attempts required, thereby reducing anesthesia time.
Clinical Evidence
Evidence includes a randomized controlled trial in pediatric patients (ages 2-18) and multiple adult studies. Pediatric data indicates the system is safe, effective, and provides good quality biopsies in less time compared to manual methods. Adult studies (e.g., Swords et al., Berenson et al.) demonstrate superior core specimen size/quality, reduced procedure time, and decreased pain. A tactile feedback study in simulated bone showed a 97% insertion success rate for the power driver versus 48.5% for manual needles.
Indicated for bone marrow aspiration and biopsy in adult and pediatric patients age 2 and older.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
The OnControl™ Bone Marrow System by Vidacare® (K072045)
Powered EZ-IO® Pediatric Bone Marrow Aspiration System (OnControl Bone Marrow Aspiration System) (K070759)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent. The logo is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2014
Vidacare, LLC Ms. Diana F. Montez, BSN, RN Research, Clinical and Regulatory Assistant 4350 Lockhill Selma Road Shavano Park, Texas 78023
Re: K142377
Trade/Device Name: The OnControl™ Bone Marrow Biopsy System by Vidacare® Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW, FCG Dated: November 14, 2014 Received: November 17, 2014
Dear Ms. Montez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Binita S. Ashar -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K142377
Device Name
The OnControl Bone Marrow Biopsy System by Vidacare
Indications for Use (Describe)
The OnControl Bone Marrow Biopsy System is intended for bone marrow aspiration and biopsy in adult and pediatric patients age 2 and older.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
008_ 510(k) Summary [807.92(c)]
Image /page/3/Picture/3 description: The image shows the word "vidacare" in red, with a stylized font. The letters are connected and have rounded edges. There is a registered trademark symbol after the "e" at the end of the word.
50 I ockhill Selr Shavano Park. TX 78249+2095 210-375-8500 www.vidacare.com
#### SUMMARY
Submitter's name [807.92(a)(1)]: Vidacare LLC
Address:
4350 Lockhill Selma Road Shavano Park, TX 78249-2095
210-375-8500
Phone:
Fax number: 210-375-8537
Name of contact person: Diana Montez, BSN, RN diana.montez@teleflex.com
Date Summary was prepared: July 10, 2014
[807.92(a)(2)]:
Trade Name of the device: The OnControl™ Bone Marrow Biopsy System by Vidacare®
Common or usual name: Bone Marrow Biopsy Needle (21 CFR 876.1075, KNW, FCG) Classification name: Gastroenterology-Urology Biopsy Instrument
The legally marketed devices to which we are claiming equivalence [807.92(a)(3)]:
| 510(k) number | Trade or Proprietary or<br>Model Name | Manufacturer |
|---------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------|
| K072045 | The OnControl™ Bone<br>Marrow System by Vidacare® | Vidacare Corporation |
| K070759 | Powered EZ-IO® Pediatric<br>Bone Marrow Aspiration<br>System<br>(new trade name, OnControl<br>Bone Marrow Aspiration<br>System). | Vidacare Corporation |
{4}------------------------------------------------
# Device Description [807.92(a)(4)]:
The OnControl Bone Marrow Biopsy System by Vidacare® consists of a reusable Power Driver and a disposable sterile needle set in a sealed procedure tray. The procedure tray contains 1 driver connector with sterile sleeve, 1 biopsy needle set with depth guide, an ejector rod and alignment quide. The OnControl Bone Marrow Biopsy System Needle Set consists of a cannula made of 304 stainless steel, with a beveled cutting tip and stylet. The needle sets are 11 gauge, 102 mm and 152 mm with an internal core capturing thread at the distal tip of the cannula. The proposed device will utilize identical needle sets (gauge, length and materials) as currently utilized in the predicate device K072045 The OnControl Bone Marrow Biopsy System by Vidacare®. The proposed device will also utilize the power rotary driver technology as used for the predicates K072045, The OnControl Bone Marrow Biopsy System by Vidacare® and K070759 Powered EZ-10® Pediatric Bone Marrow Aspiration System (new trade name, OnControl Bone Marrow Aspiration System).
Upon activation, the power rotary driver assists the clinician in inserting the needle set through the cortex of the bone. The driver is then separated from the hub of the needle set by retracting the OnControl Connector release mechanism. The inner stylet is used only to penetrate the cortex and is then removed. The standard Luer lock hub on the cannula permits attachment of a syringe for aspiration. The depth guide which slides up and down the cannula is moved to the desired depth marking. The power rotary driver is then reattached to the biopsy cannula and the driver is activated and advanced the desired depth to obtain the biopsy specimen and then withdrawn. The driver is then separated from the needle assembly for specimen removal.
This submission requests the addition of the pediatric population, age 2 and older, as an expanded indication to previously cleared K072045. The OnControl Bone Marrow Biopsy System with the same Indication For Use, for bone marrow aspiration and biopsy utilizing the same insertion technique and devices as the predicate.
### (Proposed) Indications for Use:
The OnControl™ Bone Marrow Biopsy System is intended for bone marrow aspiration and biopsy in adult and pediatric patients age 2 and older.
{5}------------------------------------------------
## Proposed Expanded Indication Discussion:
Reason for requested expanded indication: This device expanded indication will enhance the diagnostic abilities of physicians trained in the retrieval of bone/bone marrow biopsy samples from pediatric patients with addition of an option to traditional manual devices (with their associated poor rates of adequate specimen retrieval). 12 This could possibly improve patient outcomes by offering an option when clinicians believe the control and rotary driver will obtain better bone marrow biopsy samples with fewer attempts (allowing for less time under anesthesia with associated potential risks) for diagnostic accuracy, and faster access to treatment when necessary.
Overview: Bone marrow evaluation is essential for the diagnosis of hematological malignancies and non-malignant disease in children; and is also useful for staging certain diseases and response to therapy.3 Core biopsy length has been found to be critical in diagnoses, relapse prediction or identifying residual disease postchemotherapy. The larger the amount of marrow obtained the greater chance of finding a focal lesion.4
The current practice of obtaining trephine biopsies and bone marrow aspirates in children via the manual method has a poor success rate for obtaining adequate specimens. In a study by the European Neuroblastoma Study Group, 139 of 822 (17%) biopsy specimens were inadequate; containing less than 0.5 cm of well-preserved bone marrow as assessed by two central reviewers. In 13 institutions submitting at least 20 cores, failure rates ranged from 2.6 to 50%. " Reid, et al. reported 25% of 605 pediatric trephine biopsies were inadequate.2
Safety and Efficacy Considerations (includes non-clinical test summary): The results of a recent randomized controlled trial with pediatric patients (ranging from 2-18 years old) requiring a bone marrow biopsy demonstrate that the OnControl™ Bone Marrow Biopsy System by Vidacare® product is safe and effective for bone marrow biopsy in pediatric patients. Researchers concluded that the OnControl™ Bone Marrow Biopsy System obtained biopsies safely, in less time and of good quality when compared with traditional manual biopsy retrieval methods. 9
The powered rotary technology for obtaining aspirations in pediatric patients previously cleared with predicate K070759, Powered EZ-10® Pediatric Bone Marrow Aspiration System (new trade name, OnControl Bone Marrow Aspiration System) has been used safely without any reported complications since clearance in 2007. The powered technology and rotary driver design of Vidacare's vascular access devices have proven
{6}------------------------------------------------
safe for the indication of vascular access in pediatric bones, including the FDA cleared sites of proximal humerus, and the proximal and distal tibia and distal femur.
The OnControl Bone Marrow Biopsy System, K072045, is currently cleared for use in adults only. Multiple studies have evaluated the use of the OnControl system in adults and found decreased time of procedure, decreased pain, and improved core biopsy specimens when compared with traditional manual biopsy methods.67,8,9,10 Swords et al demonstrated the OnControl system yielded core specimens that were of superior size and quality when compared to the manual technique in a randomized controlled trial in adults. There have been no reported serious complications for either of the Vidacare predicates reported from the literature or the FDA MDR system since clearance in 2007 for K070759 in pediatric patients or K072045 in adults.
A tactile feedback study was designed to address concerns that the power rotary driver design, a component of the OnControl Bone Marrow and Biopsy System, might limit procedural control and the ability of clinicians to discern catheter tip location when accessing bone. This study compared the ability of clinicians to attain precision of needle tip placement by tactile feedback only when using three devices (two manual devices and the power rotary device) in simulated bone. The Vidacare power rotary driver had an insertion success rate of 97% compared with 48.5% for the traditional manual biopsy needle. 11
## [807.92(b)(3)]:
### Conclusion:
In consideration of published studies, the cited clinical study, the safety record of both the predicate devices, and also the experience with the rotary powered technique of the EZ-IO vascular access devices we conclude that adding the pediatric population to expand the indications for use of K072045 is safe and effective, using the same technique and equipment. The only change to the proposed Directions for Use labeling would be the addition of the pediatric population age 2 and older to the indications.
### [807.92(a)(5)]:
### (Proposed) Indications for Use:
The OnControl™ Bone Marrow Biopsy System is intended for bone marrow aspiration and biopsy in adult and pediatric patients age 2 and older.
{7}------------------------------------------------
### [807.92(a)(6)]:
# Summary of the technological characteristics of this device compared to the predicate devices:
The expanded indication requires no new technology to facilitate the safe application of the product. There have been no changes to the design or components of the devices cleared under K072045 and therefore, the comparison of technological characteristics listed below are identical:
- ·Driver Design Features
- •Firmware
- •Needle Design
- ·Technique
- ·Anatomical sites
- ·Sterility
- •Where Device is Used
- ·Biocompatibility
- ·Ergonomics of the patient-user interface
{8}------------------------------------------------
References (available upon request):
- 1. Reid MM, Roald B. Adequacy of bone marrow trephine biopsy specimens in children. J. Clin. Pathol. 1996;49:226-229
- Reid MM, Roald B. Deterioration in performance in obtaining bone marrow trephine biopsy つ cores from children. J. Clin. Pathol. 1999;52:851-852.
- 3. Abla O. Friedman J. and Doyle J. Performing bone marrow aspiration and biopsy in children: Recommended guidelines. Paediatr Child Health 2008; 13:499-501.
- 4. Islam A and Henderson ES. Value of long-core biopsy in the detection of discrete bone marrow lesions. Histopathology 1988;12:641-648.
- 5. Falcon-Cantrill M. Thomas P. Saldivar V. Assanasen C. Comparison of a rotary powered bone marrow aspiration and biopsy device to the traditional manual device in children. Pediatr Blood Cancer 2013;60(S2);S8. Doi:10.1002/pbc.24509 (Full manuscript pending publication)
- 6. Cohen SC. Gore JM. Evaluation of a powered intraosseous device for bone marrow sampling. Anticancer Research 2008;28:3843-3846
- 7. Swords RT, Anguita J, Higgins RA, et al. A prospective randomised study of a rotary powered device (OnControl) for bone marrow aspiration and biopsy. J Clin Pathol 2011:64:809-813.
- 8. Miller LJ, Philbeck TE , Montez DF, et al. Powered bone marrow biopsy procedures produce larger core specimens, with less pain, in less time than with standard manual devices. Hematology Reports 2011;3:e8.
- 9. Berenson JR, Yellin O, Blumenstein B, et al. Using a powered bone marrow biopsy system results in shorter procedures, causes less residual pain to adult patients, and yields larger specimens. Diagnostic Pathology 2011:6:23.
- 10. Reed LJ, Raghupathy R, Strakhan M et al. The OnControl bone marrow biopsy technique is superior to the standard manual technique for hematologists-in-training: a prospective, randomized comparison. Hematology Reports 2011;3(e21). doi:10.4081/hr.2011.e21
- 11. Garcia G, Miller LJ, Philbeck, T, Bolleter S, Montez D. Tactile feedback allows accurate insertion of a powered bone access device for vertebroplasty and bone marrow sampling procedures.J Vasc and Interv Radiol 2011;22(3):S86
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.