K142357 · Contract Medical International, GmbH · DYB · Jun 25, 2015 · Cardiovascular
Device Facts
Record ID
K142357
Device Name
DuraSheath Introducer Sheath System
Applicant
Contract Medical International, GmbH
Product Code
DYB · Cardiovascular
Decision Date
Jun 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The DuraSheath Introducer Sheath System is indicated to be used for introduction of interventional and diagnostic devices into the human vasculature, including but not limited to femoral access via a contralateral approach to access the popliteal and infrapopliteal arteries.
Device Story
DuraSheath Introducer Sheath System is a sterile, single-use device for vascular access. Components include a coil-reinforced shaft with an atraumatic distal tip, a hemostasis valve with side port and 3-way stopcock, and a tapered dilator. The sheath features a radiopaque marker and lubricious coating. Used in hospitals/healthcare facilities by clinicians to facilitate the introduction of diagnostic or interventional devices into the vasculature, specifically for femoral access to popliteal and infrapopliteal arteries. The device remains in the patient for up to 24 hours. The stainless steel coil reinforcement provides kink resistance during navigation through tortuous anatomy. The device is manually operated; it does not involve electronic processing or software.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical performance testing, including mechanical testing (kink resistance, coating adhesion, particulate testing), biocompatibility testing, sterilization validation, and accelerated aging studies.
Indicated for patients requiring introduction of interventional and diagnostic devices into the human vasculature, including femoral access for popliteal and infrapopliteal artery procedures.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is simple and professional, reflecting the department's role in public health and human services.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 24, 2015
Contract Medical International GmbH Jan Kloboucnik Director, Regulatory and Quality Affairs Lauensteiner Str, 37 Dresden, 01277 DE
Re: K142357
Trade/Device Name: DuraSheath Introducer Sheath System Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB. DRE Dated: May 21, 2015 Received: May 26, 2015
Dear Jan Kloboucnik,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K142357
Device Name DuraSheath Introducer Sheath System
## Indications for Use (Describe)
The DuraSheath Introducer Sheath System is indicated to be used for introduction of interventional and diagnostic devices into the human vasculature, including but not limited to femoral access via a contralateral approach to access the popliteal and infrapopliteal arteries.
| | Type of Use (Select one or both, as applicable) | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--|--|
| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over=The-Counter Use (21 CFR 801 Subpart C) | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | |
| | Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | Department of Health and Human Services | | |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K142357 - 510(k) Summary
#### 1.1 Submitter
| Submitter: | Contract Medical International GmbH<br>Lauensteiner Straße 37<br>01277 Dresden<br>Germany |
|-----------------|-------------------------------------------------------------------------------------------|
| Contact Person: | Jan Kloboucnik, Director, RA/QA |
| Phone: | +420 494 949 564 |
| Fax: | +420 494 949 565 |
| Email: | jkloboucnik@contract-medical.com |
| Date Prepared: | 24 June 2015 |
#### 1.2 Device
| Device Trade Name: | DuraSheath Introducer Sheath System |
|----------------------|--------------------------------------------------------------------------------|
| Device Common Name: | Introducer Sheath |
| Classification Name: | Introducer, Catheter; and<br>Dilator, Vessel, For Percutaneous Catheterization |
| Regulatory Class: | II |
| Product Code: | DYB; and<br>DRE |
#### Predicate Device 1.3
| Primary Predicate Device: | Pinnacle Destination Guiding Sheath<br>(K091329) |
|---------------------------|--------------------------------------------------|
| Second Predicate Device: | Fortress Introducer Sheath System<br>(K100799) |
#### 1.4 Device Description
The DuraSheath Introducer Sheath System is a sterile, disposable device consisting of (a) a coil reinforced shaft with an atraumatic tip at the distal end; (b) a hemostasis valve with a side port and color coded 3-way stopcock; and (c) a tapered tip dilator with snap-fit hub at the proximal end. The distal end of the sheath contains a radiopaque marker. Lubricous coating is applied to the distal end of the sheath. The system is packaged inside a sealed Tyvek pouch.
The DuraSheath Introducer Sheath System is a prescription medical device that is used only in healthcare facilities or hospitals. The device is placed in patients for up to 24 hours.
{4}------------------------------------------------
#### 1.5 Indications for Use
The DuraSheath Introducer Sheath System is indicated to be used for introduction of interventional and diagnostic devices into the human vasculature, including but not limited to femoral access via a contralateral approach to access the popliteal and infrapopliteal arteries.
The Indications for Use statement for the DuraSheath Introducer Sheath System is not identical to that of the two predicate devices; however, the differences do not alter the intended use of the device nor do they affect the safety and effectiveness of the device relative to the predicate devices. The DuraSheath Introducer Sheath System and the two predicate devices each have the same intended use, namely to support the introduction of interventional and diagnostic devices.
#### 1.6 Comparison of Technological Characteristics with the Predicate Device
The DuraSheath Introducer Sheath System is a manually operated, sterile, single patient use sheath system made predominantly of thermoplastic polymers. The sheath is reinforced with a stainless steel coil in order to provide kink resistance when passed through tortuous paths. With regard to the design, device features, method of sterilization, and mode of operation, the DuraSheath Introducer Sheath System does not differ from the predicate devices. Materials used for manufacture of the DuraSheath Introducer Sheath System are the same or very similar to those contained in the predicate devices. Technological characteristics of the subject device differ from those of the predicates only with respect to materials for selected components and the choice of lubricious coating on the sheath. Both the polyurethane used for the outer layer and the hydrophobic coating applied to the distal end of the sheath of the DuraSheath Introducer Sheath System are commonly used in medical devices, including introducer sheaths and catheters. Differences in technological characteristics between the subject device and the predicates do not raise and concerns of safety and effectiveness, as demonstrated by the performance data collected.
#### 1.7 Performance Data
Nonclinical and clinical performance data demonstrate that the DuraSheath Introducer Sheath System is safe and effective and performs substantially equivalent to the predicates. The following performance data from non-clinical tests are being provided in support of the substantial equivalence determination:
- Mechanical testing, including tests required under relevant international standards, coating adhesion and particulate testing, and transportation integrity testing, performed to verify and validate the design and to demonstrate process capability.
- Full range of Biocompatibility testing to demonstrate biocompatibility. ●
- Sterilization Validation Adoption to confirm sterility of the device upon exposure to the selected sterilization cycle.
- . Accelerated age testing to confirm product performance and end of shelf life.
{5}------------------------------------------------
#### 1.8 Conclusions
Results demonstrate that the DuraSheath Introducer Sheath System performs comparably with predicates and other legally marketed devices. The DuraSheath Introducer Sheath System is substantially equivalent to the two predicates devices in terms of intended use, design and materials, technological characteristics, and principle of operation. Any differences between the subject device and the predicates do not raise any issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.